K230199 · Dentsply Sirona, Inc. · NXC · Oct 18, 2023 · Dental
Device Facts
Record ID
K230199
Device Name
Byte Aligner System
Applicant
Dentsply Sirona, Inc.
Product Code
NXC · Dental
Decision Date
Oct 18, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5470
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Byte Aligner System is indicated for the treatment of tooth malocclusion in patients with permanent dentition (i.e., all second molars). The Byte Aligner System positions teeth by way of continuous gentle force.
Device Story
The Byte Aligner System is a series of clear, thermoformed PETG aligners used for orthodontic tooth movement. Treatment plans are generated using CAD-CAM software; model archforms are created via rapid prototyping (stereolithography). Aligners are thermoformed over these models. The device is prescribed by dental clinicians. Patients wear the aligners 20-22 hours daily, replacing them in a sequential series to apply continuous gentle force to teeth and roots, causing pressure to the periodontium to achieve incremental repositioning. The system supports the use of attachments, cut-outs, and slits to facilitate specific tooth movements. The device is non-sterile and intended for single-patient, multiple-use.
Clinical Evidence
No clinical data included. Substantial equivalence is supported by bench testing, literature review, and comparison to predicate/reference devices.
Technological Characteristics
Materials: PETG (copolyester) plastic sheets. Principle: Orthodontic tooth movement via continuous gentle force from elastic deformation of thermoformed aligners. Form factor: Sequential series of custom-fit trays. Manufacturing: Thermoforming over 3D-printed model archforms. Connectivity: None. Sterilization: Non-sterile.
Indications for Use
Indicated for treatment of tooth malocclusion in patients with permanent dentition (all second molars).
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 18, 2023
Dentsply Sirona Inc. Sobrin Laura Corporate RA Manager 221 West Philadelphia Street, Suite 60W York, Pennsylvania 17401
Re: K230199
Trade/Device Name: Byte Aligner System Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: January 25, 2023 Received: January 25, 2023
Dear Sobrin Laura:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Michael E. Adjodha -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K230199
Device Name
Byte Aligner System
Indications for Use (Describe)
The Byte Aligner System is indicated for the treatment of tooth malocclusion in patients with permanent dentition (i.e., all second molars). The Byte Aligner System positions teeth by way of continuous gentle force.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10pt;"> <span style="font-family:Arial;"> <span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> </span> |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size:10pt;"> <span style="font-family:Arial;"> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(k) SUMMARY For Byte Aligner System (K230199)
- 1. Submitter Information:
Dentsply Sirona 221 West Philadelphia Street Suite 60W York, PA 17404
Contact Person: Laura Sobrin Telephone Number: 717-849-4434 Fax number: 717-849-4343
Date Prepared: October 13, 2023
### 2. Device Name:
- Proprietary Name: Byte Aligner System ●
- Classification Name: Orthodontic plastic bracket ●
- CFR Number: 21 CFR 872.5470 .
- Device Class: Class II ●
- Product Code: NXC, Aligner, Sequential .
### 3. Predicate & Reference Devices:
| Device Name | Device Type | 510(k) | Company Name |
|-------------------------------------------------------------|-------------|---------|--------------------------------------------------------------------------------------------------|
| Byte Aligner System | Predicate | K180346 | Straight Smile, LLC (part of<br>Dentsply Sirona) |
| Sureclear Aligners | Reference | K171860 | OraMetrix (part of Dentsply<br>Sirona), now Dentsply Sirona<br>Orthodontics Inc. |
| Mouthguard and Aligner<br>Materials | Reference | K062828 | Dentsply International (part of<br>Dentsply Sirona), now<br>Dentsply Sirona Orthodontics<br>Inc. |
| Invisalign System with<br>Mandibular<br>Advancement Feature | Reference | K181739 | Align Technology Inc. |
{5}------------------------------------------------
| Device Name | Device Type | 510(k) | Company Name |
|-------------------------------|-------------|---------|---------------------------|
| Ormco Spark Aligner<br>System | Reference | K182826 | Sybron Dental Specialties |
#### 4. Description of Device:
The Byte Aligner System consists of a sequential series of aligners fabricated of clear, thin, lightweight and flexible thermoformed polyethylene terephthalate glycol (PETG) [copolyester] plastics.
The mechanism of action of the proposed Byte Aligner System is mechanical in nature; that is, tooth movement induced by orthodontic force application using continuous gentle forces. The aligners are made from elastic thermoplastic material that applies continuous gentle force to the teeth for treatment of tooth malocclusion. Treatment plans are modeled using CAD-CAM (computer-aideddesign and computer-aided-manufacturing) software. Model archforms are created using rapid prototyping techniques like stereo-lithography. The aligners are thermoformed over the model archforms and are not themselves 3D printed.
With the aligners seated on the patient's upper and/or lower dental arches, the teeth are progressively and gradually straightened over time by mechanical forces that are delivered via minor, incremental changes in tooth positions in each subsequent set of aligner trays that have been customized for the patient, delivered as a sequential series of trays to progressively reposition the teeth. The mechanics of how this happens is based on the forces caused by the elastic deformation of the aligner when worn by the patient. The forces are applied to the teeth, and also to the roots, causing pressure to the periodontium in the direction of the desired movement. Each aligner moves the patient's teeth in small increments from their original state.
- న. Indications for Use:
The Byte Aligner System is indicated for the treatment of tooth malocclusion in patients with permanent dentition (i.e., all second molars). The Byte Aligner System positions teeth by way of continuous gentle force.
#### 6. Comparison of Technological Characteristics:
The purpose of this Traditional 510(k) is to gain U.S. premarket clearance for the modification to the predicate device Byte Aligner System (K180346).
- . The proposed and predicate device have the same intended use, patient population and indications for use.
- The proposed and predicate devices are made by thermoforming PETG (copolyester) plastic sheets.
- The proposed and predicate devices have the same packaging configurations. ●
The key differences between the predicate device (K180346) and the proposed device are:
{6}------------------------------------------------
- The proposed device can be made by thermoforming additionally cleared PETG (copolyester) ● plastic sheets (K062828), and which follows the same manufacturing processes and materials as reference device Sureclear Aligners (K171860).
- . Reference to two additional cleared orthodontic software systems (K213916, K171860) that can be used for treatment planning software prior to the fabrication of the proposed device.
- Revise product label to only include date of manufacture and exclude an expiration date to be . consistent with the reference device Sureclear Aligners (K171860).
- Separate Instructions for Use for patient and doctor. Update patient Instructions for Use to be . more aligned with requirements in FDA guidance document, "Guidance on Medical Device Patient Labeling". Create separate Instructions for Use geared towards the dentist.
- . Allow for the use of cut-outs, slits and attachments similar to the reference devices (K181739 and K182826).
An overview of the similarities and differences between the proposed devices, predicate device, and reference devices is given in Table FS.1.
{7}------------------------------------------------
| Item of<br>Comparison | Proposed device<br>Byte Aligner<br>System<br>(K230199) | Predicate device<br>Byte Aligner<br>System (K180346) | Reference device<br>Sureclear Aligners<br>(K171860) | Reference device<br>Essix plastic<br>sheets (K062828) | Reference device<br>Invisalign System<br>with Mandibular<br>Advancement<br>Feature<br>(K181739) | Reference device<br>Ormco Spark<br>Aligner System<br>(K182826) | Discussion |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Straight Smile,<br>LLC (part of<br>Dentsply Sirona) | Straight Smile,<br>LLC (part of<br>Dentsply Sirona) | OraMetrix (part of<br>Dentsply Sirona),<br>now Dentsply<br>Sirona<br>Orthodontics Inc. | Dentsply<br>International<br>(part of Dentsply<br>Sirona), now<br>Dentsply Sirona<br>Orthodontics Inc. | Align<br>Technology, Inc. | Sybron Dental<br>Specialties | The proposed<br>and predicate<br>devices are<br>Dentsply Sirona<br>devices. Some<br>reference devices<br>are Dentsply<br>Sirona devices or<br>third-party<br>devices. |
| Intended<br>population | Patients with<br>permanent dentition | Patients with<br>permanent<br>dentition | Patients with<br>permanent<br>dentition | | Patients with<br>permanent<br>dentition | Patients with<br>permanent<br>dentition | Same |
| Indications for<br>Use | The Byte Aligner<br>System is indicated<br>for the treatment of<br>tooth malocclusion<br>in patients with<br>permanent dentition<br>(i.e., all second<br>molars). The Byte<br>Aligner System<br>positions teeth by<br>way of continuous<br>gentle force. | The Byte Aligner<br>System is<br>indicated for the<br>treatment of tooth<br>malocclusion in<br>patients with<br>permanent<br>dentition (i.e., all<br>second molars).<br>The Byte Aligner<br>System positions<br>teeth by way of<br>continuous gentle<br>force. | Sureclear aligners<br>are indicated for<br>the treatment of<br>tooth malocclusion<br>in patients with<br>permanent<br>dentition. | Mouthguard and<br>Aligner Materials<br>are indicated for<br>the fabrication of<br>orthodontic and<br>dental appliances<br>such as aligners,<br>bite planes,<br>mouthguards.<br>nightguards,<br>snoring<br>appliances.<br>splints, retainers,<br>repositioners, and<br>temporary<br>bridges. | The Invisalign<br>System is<br>indicated for the<br>orthodontic<br>treatment of<br>malocclusion. | The Ormco™<br>Spark™ Aligner<br>System is<br>indicated for the<br>alignment of<br>teeth during<br>orthodontic<br>treatment of<br>malocclusion in<br>patients with<br>permanent<br>dentition (i.e., all<br>second molars).<br>The Ormco™<br>Spark™ Aligner<br>System positions<br>teeth by way of<br>continuous gentle<br>force. | Same as the<br>predicate device.<br>Similar to the<br>reference<br>devices. |
| Item of | Proposed device | Predicate device | Reference device | Reference device | Reference device | Reference device | Discussion |
| Comparison | Byte Aligner<br>System<br>(K230199) | Byte Aligner<br>System (K180346) | Sureclear Aligners<br>(K171860) | Essix plastic<br>sheets (K062828) | Invisalign System<br>with Mandibular<br>Advancement<br>Feature<br>(K181739) | Ormco Spark<br>Aligner System<br>(K182826) | |
| Mechanism of<br>action | Sequential aligners<br>are made from elastic<br>thermoplastic<br>materials that apply<br>continuous gentle<br>force to the teeth in<br>the form of stored<br>energy imparted in<br>the aligner trays at<br>the time of their<br>fabrication. This<br>energy slowly<br>dissipates over time,<br>and re-positions the<br>teeth by way of<br>continuous gentle<br>force. | Sequential aligners<br>are made from<br>elastic thermoplastic<br>materials that apply<br>continuous gentle<br>force to the teeth in<br>the form of stored<br>energy imparted in<br>the aligner trays at<br>the time of their<br>fabrication. This<br>energy slowly<br>dissipates over time,<br>and re-positions the<br>teeth by way of<br>continuous gentle<br>force. | Sequential aligners<br>are made from<br>elastic thermoplastic<br>materials that apply<br>continuous gentle<br>force to the teeth in<br>the form of stored<br>energy imparted in<br>the aligner trays at<br>the time of their<br>fabrication. This<br>energy slowly<br>dissipates over time,<br>and re-positions the<br>teeth by way of<br>continuous gentle<br>force. | | Sequential<br>aligners apply<br>continuous gentle<br>force to the teeth<br>and position the<br>mandible forward | Orthodontic tooth<br>movement occurs<br>through forces<br>applied by the<br>appliance to the<br>dentition as each<br>tooth follows the<br>programmed<br>displacement<br>based on a dental<br>clinician's<br>prescription. | Same as the<br>predicate device<br>and reference<br>device Sureclear<br>Aligners<br>(K171860).<br>Similar to the<br>reference<br>devices. |
| Aligner Tray<br>Raw material<br>used | PETG (copolyester) | PETG<br>(copolyester) | PETG<br>(copolyester) | Various<br>thermoplastic<br>sheets including<br>PETG<br>(copolyester)<br>plastic sheets | Polyurethane-<br>polyester<br>composite resin | Polyurethane-<br>polyester<br>composite resin | Similar.<br>Proposed device<br>uses the same<br>plastic sheet<br>materials in the<br>predicate device<br>and reference<br>device Sureclear<br>Aligners<br>(K171860). |
| Manufacturing<br>process | Thermoforming<br>plastic sheets over<br>3D printed model | Thermoforming<br>plastic sheets over<br>3D printed model | Thermoforming<br>plastic sheets over<br>3D printed model | Thermoforming<br>plastic sheets<br>over 3D printed<br>model | Thermoforming<br>plastic sheets<br>over 3D printed<br>model | Thermoforming<br>plastic sheets<br>over 3D printed<br>model | Same |
| Item of<br>Comparison | Proposed device | Predicate device | Reference device | Reference device | Reference device | Reference device | Discussion |
| | Byte Aligner<br>System<br>(K230199) | Byte Aligner<br>System (K180346) | Sureclear Aligners<br>(K171860) | Essix plastic<br>sheets (K062828) | Invisalign System<br>with Mandibular<br>Advancement<br>Feature<br>(K181739) | Ormco Spark<br>Aligner System<br>(K182826) | |
| Tooth<br>movements | Tooth movements<br>available include<br>the full range<br>necessary to<br>provide full<br>orthodontic<br>treatment of<br>patients with<br>permanent<br>dentition, consistent<br>with the Indications<br>for Use. | Tooth movements<br>available include<br>the full range<br>necessary to<br>provide full<br>orthodontic<br>treatment of<br>patients with<br>permanent<br>dentition,<br>consistent with the<br>Indications for<br>Use. | Tooth movements<br>available include<br>the full range<br>necessary to<br>provide full<br>orthodontic<br>treatment of<br>patients with<br>permanent<br>dentition,<br>consistent with the<br>Indications for<br>Use. | | Tooth<br>movements<br>consistent with<br>the Indications<br>for Use,<br>including<br>mandibular<br>advancement. | Tooth<br>movements<br>consistent with<br>the Indications<br>for Use. | Same as the<br>predicate and<br>reference device,<br>Sureclear<br>aligners<br>(K171860)<br>Similar to other<br>reference devices<br>(K181739 &<br>K182826) |
| Sterility state | Non-sterile | Non-sterile | Non-sterile | Non-sterile | Non-sterile | Non-sterile | Same |
| Single Patient<br>use | Single patient,<br>multiple use | Single patient,<br>multiple use | Single patient,<br>multiple use | Single patient,<br>multiple use | Single patient,<br>multiple use | Single patient,<br>multiple use | Same |
| Shelf Life | No shelf life | 2 years | No shelf life | No shelf life | No shelf life | No shelf life | Same as the<br>reference devices |
| OTC or Rx | Rx | Rx | Rx | Rx | Rx | Rx | Same |
| Attachments | Available | Available | Available | | Available | Available | Same |
| Cut-outs/Slits | Available | | | | Available | Available | Same as the<br>reference devices<br>(K181739,<br>K182826) |
| Item of<br>Comparison | Proposed device<br>Byte Aligner<br>System<br>(K230199) | Predicate device<br>Byte Aligner<br>System (K180346) | Reference device<br>Sureclear Aligners<br>(K171860) | Reference device<br>Essix plastic<br>sheets (K062828) | Reference device<br>Invisalign System<br>with Mandibular<br>Advancement<br>Feature<br>(K181739) | Reference device<br>Ormco Spark<br>Aligner System<br>(K182826) | Discussion |
| In use<br>duration | Aligners are worn<br>for 7-14 days (1-2<br>weeks) or as<br>directed by the<br>dentist. Aligners are<br>to be worn 20-22<br>hours (24/7) except<br>when eating,<br>drinking, brushing<br>or flossing. | Each month, wear<br>the first set for 1<br>week, the second set<br>for 1 week, and the<br>third set for 2<br>weeks. | Aligners should be<br>worn 20 to 22 hours<br>daily. Each set of<br>aligners should be<br>replaced after<br>approximately 2<br>weeks of use. | | Aligners are<br>worn for<br>approximately 1-<br>2 weeks of 20-22<br>hours of wear per<br>day. This is<br>repeated for<br>duration as<br>prescribed by the<br>dental<br>practitioner. | Aligners are<br>worn every one<br>or two weeks as<br>part of the<br>personalized<br>treatment plan. | Similar<br>Aligners are<br>worn for 7-14<br>days (1-2 weeks)<br>for the proposed<br>device and<br>ranges between 1<br>to 2 weeks for<br>the predicate and<br>reference<br>devices. The<br>proposed labeling<br>standardizes the<br>wear time for the<br>end user<br>regardless of<br>aligner stage and<br>makes it easier<br>for the patient to<br>adhere to their<br>tray change<br>regimen. |
| Software used<br>for ordering<br>workflow | Yes | Yes | Yes | | | | Same |
## Table FS.1: Similarities and Differences between the Proposed device, Predicate device, and Reference devices
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
#### 7. Non-Clinical Performance Data
Testing data that was referenced, or data relied upon to demonstrate substantial equivalence include:
- a) Biocompatibility evaluation in accordance with EN ISO 10993-1:2018 (Biological evaluation of medical devices – Part 1: Evaluation and testing with a risk management process) was performed and it was determined that no additional testing was required to demonstrate biological safety of proposed devices. Testing on file in Sureclear Aligners (K171860) remains valid.
- b) Shelf-life information included in Mouthguard and Aligner Materials (K062828) and Sureclear Aligners (K171860) was reviewed, in addition to literature, and considered to be sufficient evidence to support the addition of alternate PETG (copolyester) materials and modification to the finished device shelf-life.
- c) Literature and material data sheet review was performed on the predicate device (K180346) PETG polyester materials and the proposed PETG polyester materials and confirmed that the materials are very similar in characteristics and performance.
#### 8. Clinical Performance Data
No data from human clinical studies was included to support the substantial equivalence of the proposed Byte Aligner System.
#### 9. Conclusion
The information included in this Traditional 510(k) supports the substantial equivalence of the proposed Byte Aligner System with the predicate device Byte Aligner System (K180346). The proposed device has the same intended use and indications for use as the predicate device, and is an aligner system fabricated by thermoforming PETG (copolyester) plastics that induce tooth movement by orthodontic force application using continuous gentle forces.
The proposed device represents an extension of predicate device Byte Aligner System (K180346) by allowing for the addition of alternate PETG (copolyester) materials to be used for manufacturing of aligners and removing the shelf life of the device based on the nature of the Byte Aligner System and well-known polymer science. The technological differences between the predicate device Byte Aligner System (K180346) and the proposed device are supported with the data included in premarket notifications of reference devices Sureclear aligners (K171860) and Mouthguard and Aligner materials (K062828) and additional data provided in this submission in support of the equivalency between the polyester materials. In addition, this premarket submission incorporates slits and cut-outs into the aligner and allows for the use of attachments which is similar to the additional reference devices (K181739, K182826).
Based on the intended use, indications for use, and technological features comparison, and the review of literature, data on file, and testing by reference, it can be concluded that the proposed Byte Aligner System is substantially equivalent to the predicate device Byte Aligner System (K180346).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.