K230178 · Ndd Medizintechnik AG · BZG · Oct 19, 2023 · Anesthesiology
Device Facts
Record ID
K230178
Device Name
EasyOne Sky Spirometer
Applicant
Ndd Medizintechnik AG
Product Code
BZG · Anesthesiology
Decision Date
Oct 19, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1840
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The EasyOne Sky spirometer is intended to conduct diagnostic spirometry testing on adults and pediatric patients starting at age 4. The EasyOne Sky spirometer is used by general practitioners, specialists and healthcare professionals. The EasyOne Sky is used in hospitals, clinics, pharmacies, in clinical settings, and in occupational medicine.
Device Story
EasyOne Sky (EOS) is a handheld, battery-powered diagnostic spirometer; utilizes ultrasonic transit-time measurement (TrueFlow Sensor FT) to acquire patient flow data; transmits data via Bluetooth Low Energy to a host mobile device running the EasyOne Mobile (EOMA) application. EOMA triggers measurements, communicates with the sensor, and visualizes data. Essential accessories include the single-patient-use EasyOne FlowTube (EOFT) and optional EasyOne Filter FT. Used by clinicians in hospitals, clinics, and occupational medicine to perform spirometry tests. Output includes flow-volume loops and various respiratory parameters (FVC, FEV1, etc.) displayed on the mobile device. Benefits include fast, accurate, and reliable pulmonary function assessment; supports clinical decision-making for respiratory conditions.
Clinical Evidence
Bench testing only. Performance validated against ATS/ERS/ISO 26782/23747 standards for accuracy, repeatability, and linearity. Biocompatibility evaluated per ISO 10993 and ISO 18562 series. Electrical safety and EMC testing per IEC 60601-1/1-2/1-6.
Technological Characteristics
Handheld, battery-powered (2x AA). Ultrasonic transit-time flow measurement. Bluetooth Low Energy connectivity. Dimensions: 111 x 84 x 33 mm. Weight: 140g. IP22 rating. Single-patient-use breathing tube (EasyOne FlowTube). Software: Firmware (MDSW) for sensor data acquisition; mobile application for visualization. Complies with IEC 60601-1, IEC 62304, and ISO 26782.
Indications for Use
Indicated for diagnostic spirometry testing in adults and pediatric patients aged 4 and older. Used by general practitioners, specialists, and healthcare professionals in hospitals, clinics, pharmacies, and occupational medicine settings.
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 19, 2023
NDD Medizintechnik AG Andreas Senn Director Quality, RA & CA Technoparkstrasse 1 Zurich, 8005 Switzerland
Re: K230178
Trade/Device Name: EasyOne Sky Spirometer Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: Class II Product Code: BZG Dated: September 5, 2023 Received: September 6, 2023
Dear Andreas Senn:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
## Rachana Visaria -S
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices
{2}------------------------------------------------
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230178
Device Name EasyOne Sky spirometer
Indications for Use (Describe)
The EasyOne Sky spirometer is intended to conduct diagnostic spirometry testing on adults and pediatric patients starting at age 4.
The EasyOne Sky spirometer is used by general practitioners, specialists and healthcare professionals.
The EasyOne Sky is used in hospitals, clinical settings, and in occupational medicine.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|----------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
In accordance with 21 CFR 807.92 the following information is provided for EasyOne Sky spirometer.
## ADMINISTRATIVE INFORMATION
| Date prepared | October 18, 2023 |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submission type: | Traditional 510(k) |
| Purpose of 510(k): | Introduction of a new device |
| Submitter | NDD Medizintechnik AG<br>Technoparkstrasse 1<br>CH-8005 Zürich<br>Note: NDD stands for New Diagnostic Design |
| Official Contact | Andreas Senn<br>Director Quality, RA & CA<br>Ndd Medizintechnik AG, Switzerland<br>Phone : +41 44 512 65 41<br>e-mail : ase@ndd.ch |
| Alternative Contact | Roman Eicher<br>Regulatory Affairs Manager & Safety Officer<br>Ndd Medizintechnik AG, Switzerland<br>Phone : +41 44 512 65 24<br>e-mail : rei@ndd.ch |
| US agent | Robert Weisman<br>NDD MEDICAL TECHNOLOGIES, INC.<br>300 Brickstone Sq Ste 604<br>ANDOVER, MA US 01810<br>Phone: 978 4700923 Ext<br>Email: Rweisman@Nddmed.com |
## DEVICE NAME AND CLASSIFICATION
| Trade name: | EasyOne Sky Spirometer |
|----------------------|------------------------|
| Common name: | Diagnostic Spirometer |
| Regulation number: | 21 CFR 868.1840 |
| Classification name: | Diagnostic spirometer |
{5}------------------------------------------------
| Regulatory class: | Class II |
|-------------------|----------|
| Product Code: | BZG |
#### PREDICATE DEVICE
| Primary Predicate Device: | EasyOne Air Spirometer, K161536 |
|---------------------------|---------------------------------|
| Reference Devices: | GoSpiro, K163249, K202837 |
#### INDICATIONS FOR USE
The EasyOne Sky spirometer is intended to conduct diagnostic spirometry testing on adults and pediatric patients starting at age 4. The EasyOne Sky spirometer is used by general practitioners, specialists and healthcare professionals. The EasyOne Sky is used in hospitals, clinics, pharmacies, in clinical settings, and in occupational medicine.
#### DEVICE DESCRITION
The product EasyOne Sky (EOS) is a medical electrical equipment for spirometry testing. It consists of the hand-held TrueFlow Sensor FT (SeNe) comprising an ultrasonic flow sensor that conducts air flow measurements, the application EasyOne Mobile (EOMA), and the breathing mouthpiece EasyOne FlowTube (EOFT). The EOS can be used with the optional accessories EasyOne Filter FT and a nose clip.
Key functions:
- . The key function of the TrueFlow Sensor FT (SeNe) is to acquire patient flow data by measuring transit-times of ultrasonic pulses and to send data wireless (via Bluetooth Low Energy) to the host system. The measurements and data transmission are performed by the firmware of the SeNe defined as MDSW with clinical functions.
- . The key function of the application EasyOne Mobile (EOMA) is to trigger the start of the measurement with the TrueFlow Sensor FT, to communicate with it and to visualize the data received from it.
- . The EasyOne FlowTube (EOFT) is an individually packaged, single-patient-use breathing tube and is intended to canalize patient breath through the flow sensor tube. The EOFT is an essential accessory to EOS.
{6}------------------------------------------------
## The host system, for example a smartphone, where EOMA runs on, is not part of the EOS medical electrical equipment.
Image /page/6/Picture/2 description: The image shows a spirometer and a smartphone displaying an app for tracking spirometry tests. The spirometer, labeled "EasyOne Sky", is held in a hand. The smartphone screen shows a list of activities, including spirometry tests from February 17th and 19th, with the app displaying the times of the tests.
## Figure 1: Visualization of the EOS system including the TrueFlow sensor FT with the inserted EasyOne Flow Tube and the application EasyOne Mobile installed on a mobile device.
## SUMMARY OF TECHNOLOGICAL CHARACTERISTICS AND COMPARISON TO PREDICATE DEVICE
| Criteria | EasyOne Sky<br>Subject Device | EasyOne Air<br>Primary Predicate Device | Predicate comparison<br>results | |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | K230178 | K161536 | - | |
| Manufacturer | NDD Medizintechnik AG | NDD Medizintechnik AG | Same | |
| Regulation number | 21 CFR 868.1840 | 21 CFR 868.1840 | Same | |
| Classification name | Diagnostic Spirometer | Diagnostic Spirometer | Same | |
| Product Code(s) | BZG | BZG | Same | |
| Regulatory Class | | II | Same | |
| Intended use | Diagnostic spirometry<br>testing | Diagnostic spirometry<br>testing | Same | |
| Indications for use | The EasyOne Sky<br>spirometer is intended to<br>conduct diagnostic<br>spirometry testing on adults<br>and pediatric patients<br>starting at age 4. The<br>EasyOne Sky spirometer is<br>used by general<br>practitioners, specialists<br>and healthcare<br>professionals. The EasyOne | The EasyOne Air spirometer<br>is intended for prescription<br>use only to conduct<br>diagnostic spirometry<br>testing of adults and<br>pediatric patients over 4<br>years old. The EasyOne Air<br>spirometer is used by<br>general practitioners,<br>specialists, and health care<br>professionals, in hospitals | Substantially Equivalent | |
| | | | | |
| | Sky is used in hospitals,<br>clinics, pharmacies, in<br>clinical settings, and in<br>occupational medicine. | and clinics, in pharmacies,<br>and in clinical settings in<br>occupational medicine. | | |
| Type of use | Prescription use | Prescription use | Same | |
| Patient population | Adults and pediatric<br>patients over 4 years old | Adults and pediatric<br>patients over 4 years old | Same | |
| Users | General practitioners,<br>specialists, healthcare<br>professionals | General practitioners,<br>specialists, healthcare<br>professionals | Same | |
| Use environment | Hospital and clinical settings<br>Pharmacies<br>Occupational medicine | Hospital and clinical<br>settings<br>Pharmacies<br>Occupational medicine | Same | |
| Single patient use | No.<br>Accessories are for single<br>patient use only | No.<br>Accessories are for single<br>patient use only | Same | |
| Description | The EasyOne Sky<br>spirometer is a hand-held<br>battery powered device,<br>which uses digital ultrasonic<br>flow measurement<br>technology for fast,<br>accurate and reliable<br>operation. The TrueFlow<br>Sensor FT firmware<br>performs the measurement<br>and calculates the<br>parameters in accordance<br>with American Thoracic<br>Society and European<br>Respiratory Society<br>requirements. The device<br>transmits data to the<br>EasyOne Mobile application<br>over a Bluetooth<br>connection to display the<br>test results and further data<br>management. | The EasyOne Air spirometer<br>is a hand-held battery<br>powered device, which uses<br>digital ultrasonic flow<br>measurement technology<br>for fast, accurate and<br>reliable operation. The<br>EasyOne Air firmware<br>performs the measurement<br>and calculates the<br>parameters in accordance<br>with American Thoracic<br>Society and European<br>Respiratory Society<br>requirements.<br>The device has an<br>integrated touchscreen to<br>display the test results. Data<br>can be transmitted to the<br>EasyOne Connect PC<br>Software via cradle or over<br>a Bluetooth connection for<br>further data management. | Similar | |
| System components | Hand-held TrueFlow Sensor<br>FT<br>EasyOne Mobile application | Hand-held TrueFlow sensor<br>with rechargeable battery<br>Docking USB Cradle for<br>charging and data transfer<br>(EasyOne Connect<br>Software)<br>AC/DC power supply<br>USB cables | Similar, both devices consist<br>of a flow sensor and a<br>software application to<br>display and manage data.<br>They differ in their power<br>supply and other<br>accessories. | |
| Accessories for single<br>patient use | Breathing Mouthpiece:<br>EasyOne FlowTube | Breathing Mouthpiece:<br>EasyOne FlowTube | Same | |
| | Optional filter: EasyOne<br>Filter FT | Optional filter: EasyOne<br>Filter FT | | |
| Software<br>Requirements | Android, iOS<br>Hard disk capacity minimum<br>250 MB | Embedded firmware | Similar software<br>functionality but are<br>different in implementation | |
| Spirometry standards | Meets ATS/ERS/ISO<br>waveform testing<br>requirements | Meets ATS/ERS/ISO<br>waveform testing<br>requirements | Same | |
| Forced expiratory vital<br>capacity test (FVC) | FVC, FEV1, FEV0.75 (only<br>shown for children ≤6yr),<br>FEV1/FVC, FEV0.75/FVC<br>(only shown for children<br>≤6yr), FET, PEF | FVC, FEV1, FEV0.75, FEV1/FV<br>C, FEV0.75/FVC, FET, PEF | Same, Parameters for basic<br>spirometry tests are the<br>same as defined in ATS /<br>ERS 2019 Table 9. | |
| Flow volume loop test<br>(FVL) | FVC, FEV1, FEV0.75,<br>FEV1/FVC, FEV0.75/FVC,<br>PEF, FET, FIVC | FVC, FEV1, FEV0.75, FEV1/FV<br>C, FEV0.75/FVC, PEF, FET, FIV<br>C | Same | |
| | Additional tests and<br>parameters | FVC: additional parameter:<br>- FEF25-75, FEF25, FEF50,<br>FEF75<br>- MMEF, MEF75, MEF50,<br>MEF25, (naming)<br>FVL additional parameter:<br>- PIF<br>- FEF25-75, FEF25, FEF50,<br>FEF75<br>- MMEF, MEF75, MEF50,<br>MEF25, (naming) | FVC additional parameter:<br>ATI, BEV, EOTV, FEF10,<br>FEF25, FEF25-75, FEF25-<br>75_6, FEF40, FEF50,<br>FEF50/FVC, FEF50/VCmax,<br>FEF60, FEF75, FEF75-85,<br>FEF80, FET25-75, FEV.25,<br>FEV.5,<br>FEV.5/FVC, FEV.75/FEV6,<br>FEV.75/VCmax,<br>FEV1/FEV6, FEV1/FVC6,<br>FEV1/VC, FEV1/VCmax,<br>FEV3/FVC, FEV3/VCmax,<br>FEV3, FEV6, FVC, MEF20,<br>MEF25, MEF40,<br>MEF50, MEF60, MEF75,<br>MEF90, MMEF, MTC1,<br>MTC2, MTC3, MTCR,<br>PEFT, t₀, VC, VCmax<br>FVL additional parameter:<br>ATI, BEV, CVI, E50/150,<br>EOTV, FEF10, FEF25, FEF25-<br>75, FEF25-75_6,<br>FEF40, FEF50, FEF50/FVC,<br>FEF50/VCmax, FEF60,<br>FEF75, FEF75-85,<br>FEF80, FET25-75, FEV.25,<br>FEV.5, FEV.5/FVC,<br>FEV.75/FEV6,<br>FEV.75/VCmax, FEV1/FEV6,<br>FEV1/FIV1,FEV1/FIVC,<br>FEV1/VC, FEV1/VCmax,<br>FEV3/FVC,FEV3/VCmax,<br>FEV3, FEV6, FIF25, FIF 25-<br>75, FIF50,<br>FIF50/FEF50, FIF75, FIV.25,<br>FIV.5, FIV1, MEF20, MEF25,<br>MEF40, MEF50, MEF60,<br>MEF75, MEF90, MIF25,<br>MIF50, MIF75,<br>MMEF, MMIF, MTC1,<br>MTC2, MTC3, MTCR, PEFT,<br>PIF, t₀, VC,<br>VCmax<br>MVV additional parameter:<br>MVV, MVV6, MVVtime, Rf,<br>VCext, VT<br>SVC additional parameter:<br>ERV, IC, IRV, Rf, VC, VCex,<br>VCin, VCmax, VT | Additional parameters<br>which go beyond basic<br>spirometry testing<br>according ATS / ERS 2019<br>recommendation differ<br>between devices. |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
| Measurement<br>principle | Ultrasonic transit-time<br>measurement (TrueFlow) | Ultrasonic transit-time<br>measurement (TrueFlow) | Same |
|--------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Measuring accuracy<br>• Volume<br>• Flow<br>• Flow PEF | • ±2 % or 0.050L<br>• ±2 % or 0.020L/s<br>(except PEF)<br…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.