K230159 · Soundbite Medical Solutions, Inc. · PDU · Aug 28, 2023 · Cardiovascular
Device Facts
Record ID
K230159
Device Name
SoundBite® Crossing System XS Peripheral
Applicant
Soundbite Medical Solutions, Inc.
Product Code
PDU · Cardiovascular
Decision Date
Aug 28, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SoundBite® Crossing System XS Peripheral is indicated to facilitate the intraluminal placement of conventional guidewires or treatment devices beyond peripheral artery chronic total occlusions. The SoundBite® Crossing System XS Peripheral is not intended for use in the carotid arteries.
Device Story
System facilitates guidewire/treatment device placement beyond peripheral artery chronic total occlusions (CTOs). Used in catheterization labs by physicians. Consists of AC-powered mobile SoundBite® Console XS, foot switch, and single-use sterile SoundBite® Active Wire XS 14P with accessories. Console generates high-amplitude, short-duration mechanical pulses (shock waves) transmitted to wire; causes distal tip to move axially (reciprocating motion), acting as micro-jackhammer to cross occlusions. Console includes RFID for wire identification, usage tracking, and connection/tip activity monitoring. Physician operates via touchscreen/foot switch; output allows guidewire advancement through CTOs. Benefits include improved recanalization capability in peripheral arteries.
Clinical Evidence
Bench testing only. No clinical data. Testing included electrical/mechanical safety, EMC, biocompatibility (ISO 10993-1), usability (IEC 62366-1), and performance testing (simulated use, durability, steerability, kink resistance, torque strength, tip motion/temperature).
Technological Characteristics
System: AC-powered console, titanium alloy wire (0.36mm OD). Wire features PTFE coating, 10mm radiopaque coil, RFID tag, and thermoplastic elastomer proximal sleeve. Connectivity: RFID-based wire identification/tracking. Sterilization: Ethylene Oxide. Standards: ANSI/AAMI ES60601-1, IEC 60601-1-2, ISO 14971, IEC 62304, ISO 10993-1.
Indications for Use
Indicated for patients requiring intraluminal placement of guidewires or treatment devices beyond peripheral artery chronic total occlusions (CTOs). Contraindicated for use in carotid arteries.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
SoundBite® Crossing System – Peripheral (14P) (K210839)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 28, 2023
Soundbite Medical Solutions, Inc. Diane Marceau Director, Ouality Assurance and Regulatory Affairs 2300 Alfred Nobel, Suite 230 Montreal, Quebec H4S 2A4 Canada
Re: K230159
Trade/Device Name: SoundBite® Crossing System XS Peripheral Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: PDU Dated: July 18, 2023 Received: July 18, 2023
Dear Diane Marceau:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Ariel Ariel G. Ashshakoor -S --- Date: 2
sn
G. Ash-shakoor -S Date: 2023.08.28 11:45:39
For
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K230159
Device Name SoundBite® Crossing System XS Peripheral
Indications for Use (Describe)
The SoundBite® Crossing System XS Peripheral is indicated to facilitate the intraluminal placement of conventional guidewires or treatment devices beyond peripheral artery chronic total occlusions.
The SoundBite® Crossing System XS Peripheral is not intended for use in the carotid arteries.
Type of Use (Select one or both, as applicable)
| <span style="font-size:100%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# SOUNCBITE 510(k) Summary
Image /page/3/Picture/2 description: The image shows the text 'K230159' in a bold, sans-serif font. The text is black against a white background. The characters are evenly spaced and easily readable.
| As required by 21 CFR 807.92 | |
|---------------------------------|---------------------------------------------------------|
| Date Prepared: | August 25, 2023 |
| Trade/Device name: | SoundBite® Crossing System XS Peripheral |
| Common Name: | System for Crossing Total Occlusions |
| Classification Name: | Catheter for Crossing Total Occlusions |
| Classification Product Code: | PDU |
| Regulation: | 21 CFR 870.1250, Percutaneous Catheter |
| Regulatory Class: | Class II |
| Manufacturer: | Soundbite Medical Solutions, Inc. |
| | 2300 Alfred Nobel, Suite 230 |
| | Montreal, QC H4S 2A4, Canada |
| Establishment Registration No.: | 3013290687 |
| Regulatory Contact: | Diane Marceau |
| | Director, Quality Assurance and Regulatory Affairs |
| | Tel: (514) 956-2525 x 3352 |
| | Email: diane.marceau@soundbitemedical.com |
| Predicate Device: | SoundBite® Crossing System – Peripheral (14P) [K210839] |
| Reference Device: | SoundBite® Crossing System - Peripheral [K192211] |
#### Device Description:
The SoundBite® Crossing System XS Peripheral, is a recanalization device designed to help physicians place a conventional guidewire or treatment device in the intraluminal space beyond peripheral chronic total occlusions (CTOs). It is intended for use in a professional healthcare facility, such as a catheterization laboratory, also known as a Cath Lab. The system consists of the AC-powered, reusable, mobile SoundBite® Console XS with a Foot Switch, and a single-use, sterile SoundBite® Active Wire XS 14P packaged together with two accessories, the SoundBite® Curving Tool 14 and the SoundBite® Torquer.
The AC-powered, reusable, mobile SoundBite® Console XS is an iteration of the predicate device, re-engineered for improved usability but with similar design, functional, safety, and performance specifications. It generates controlled, high-amplitude, short-duration mechanical pulses (i.e., shock waves) that are transmitted to the connected SoundBite® Active Wire XS 14P and cause the distal end of the wire to move back and forth (axially), acting like a micro-jackhammer.
The single-use SoundBite® Active Wire XS 14P is a 350 cm long metallic wire with an outer diameter of 0.36 mm (0.014") over its usable length. It is similar in construction to the Active Wire of the predicate and reference devices, but 50 cm longer. The proximal end of the SoundBite® Active Wire XS 14P includes a connector assembly that encapsulates a wire section reducer for easy connection to the SoundBite® Console XS, and a Radio Frequency ldentification (RFID) tag that allows the SoundBite® Console XS to detect and uniquely identify the wire and track its use. A thermoplastic elastomer sleeve was also added to proximal length of the wire during use.
### Intended Use / Indications for Use:
The SoundBite® Crossing System XS Peripheral is indicated to facilitate the intraluminal placement of conventional guidewires or treatment devices beyond peripheral artery chronic total occlusions.
The SoundBite® Crossing System XS Peripheral is not intended for use in the carotid arteries.
{4}------------------------------------------------
# SOUNCBITE 510(k) Summary
### Substantial Equivalence Comparison:
The SoundBite® Crossing System XS Peripheral has the same intended use and indications for use, the same fundamental technology and principles of operation, and substantially equivalent technological characteristics as the predicate device, the SoundBite® Crossing System – Peripheral (14P) (K210839). A side-by-side comparison of key device characteristics is presented in the following table:
| Device Characteristic | SoundBite® Crossing System - Peripheral<br>(14P) / K210839 | SoundBite® Crossing System XS Peripheral |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Mechanism of Action | Mechanical pulses sent along the length of the<br>SoundBite® Active Wire cause the distal tip to<br>accelerate axially in a reciprocating (back-and-<br>forth) motion, acting like a micro-jackhammer. | Mechanical pulses sent along the length of the<br>SoundBite® Active Wire cause the distal tip to<br>accelerate axially in a reciprocating (back-and-<br>forth) motion, acting like a micro-jackhammer. |
| Mobile Console | SoundBite® Console | SoundBite® Console XS |
| Energy Source | AC mains-powered | AC mains-powered |
| Software System | Firmware only | Firmware and Linux-based Software |
| User Interface | 20-character monochrome screen,<br>4 physical knobs, 4 lights | 10-inch color, touchscreen on the operator<br>side; 3-inch color, LCD on the physician side |
| Wire Identification | No | Radiofrequency identification (RFID) |
| Wire Tracking | No | Defect logging and usage time tracking |
| Wire Monitoring | No | Connection quality and tip activity monitoring |
| Wire Activation Means | Foot Switch | Foot Switch |
| Dimensions | 130 cm x 59 cm x 49 cm | 106 cm x 55 cm x 30 cm |
| Weight | 160 kg | 64 kg |
| Sound Level when Active | 70 dB | < 55 dB |
| Electrical Safety and<br>EMC | ANSI/AAMI ES60601-1:2005+A1:2012<br>IEC 60601-1-2:2014 (Ed. 4) | ANSI/AAMI ES60601-1:2005+A1:2012<br>IEC 60601-1-2:2014 (Ed. 4)<br>FCC Part 15 Subpart B |
| Active Wire | SoundBite® Active Wire 14P | SoundBite® Active Wire XS 14P |
| Wire Identification | No | Radiofrequency identification (RFID) |
| Wire Connection to the<br>SoundBite® Console | Titanium alloy section reducer (requires<br>manual alignment and locking) | Titanium alloy section reducer with self-<br>aligning, self-locking connector |
| Proximal Wire Sleeve | No (Off-the-shelf sterile sheath specified) | Thermoplastic elastomer sleeve (100 cm) |
| Wire Length | 300 cm | 350 cm |
| Wire Working Length | 145 cm | 175 cm |
| Radiopaque Marker | 10 mm radiopaque coil starting at 0.75 mm<br>from the distal tip | 10 mm radiopaque coil starting at 0.75 mm<br>from the distal tip |
| Wire Diameter | 0.36 mm (0.014") | 0.36 mm (0.014") |
| Wire Tip Diameter | 0.27 mm (0.0105") | 0.28 mm (0.0109") |
| Wire Material | Titanium alloy | Titanium alloy |
| Wire Coating | PTFE (polytetrafluoroethylene) coating on<br>main body of the wire; no coating on distal end | PTFE (polytetrafluoroethylene) coating on<br>main body of the wire; no coating on distal end |
| Single-use | Yes | Yes |
| Wire Sold Sterile | Yes; Ethylene Oxide Sterilization | Yes; Ethylene Oxide Sterilization |
| Shelf Life | 24 months | 24 months |
| Sterile Accessories | SoundBite® Curving Tool<br>Off-the-shelf Torquer | SoundBite® Curving Tool 14<br>SoundBite® Torquer |
{5}------------------------------------------------
# SOUNCBITE 510(k) Summary
### Summary of Non-Clinical Testing:
Design verification and validation testing was conducted in compliable FDA-recognized consensus standards, FDA guidance, and current industry standards following systematic risk assessment in accordance with ISO 14971:2019.
The SoundBite® Console XS was assessed for substantial equivalence with the SoundBite® Active Wire XS 14P (i.e., electrical and mechanical safety, electromagnetic compatibility, functionality, and serviceability).
The medical device software system has been designed, and verified in compliance with IEC 62304.
Usability was assessed under anticipated use conditions in compliance with IEC 62366-1 and FDA guidance.
The sterility of the SoundBite® Active Wire XS 14P and its accessories to the point of care is assured by validated Ethylene Oxide sterilization processes, shelf-life testing, and sterity testing under foreseeable storage and transport conditions.
The biocompatibility of the SoundBite® Active Wire XS 14P was assessed in compliance with FDA-recognized standard ISO 10993-1 and FDA guidance for Cytotoxicity, Sensitization, Irritation/Intracutaneous Reactivity, Acute Systemic Toxicity, Pyrogenicity, and Hemocompatibility.
The safety and performance of the SoundBite® Active Wire XS 14P were assessed, where appropriate using the supplied accessories, in the following areas:
- > Physical and dimensional specifications
- > Console connection and disconnection force
- > Proximal sleeve function
- Simulated use (crossing and durability) >
- > Wire coating integrity
- Particulates >
- > Corrosion resistance
- Lubricity in tortuous model >
- > Steerability
- > Kink resistance
- > Tensile strength on coil, proximal end assembly, and section reducer
- > Torque strength
- > Torqueability
- > Tip flexibility
- > Tip motion (maximum displacement)
- > Tip motion (vessel damage)
- Tip temperature >
### Conclusion:
The results from risk assessment and non-clinical testing of the SoundBite® Crossing System XS Peripheral do not raise new questions of safety or effectiveness. The data submitted with this 510(k) premarket notification demonstrate that the SoundBite® Crossing System XS Peripheral is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.