K230155 · Mick Radio-Nuclear Instruments, Inc. · JAQ · Aug 30, 2023 · Radiology
Device Facts
Record ID
K230155
Device Name
Mick Valencia Applicator Set
Applicant
Mick Radio-Nuclear Instruments, Inc.
Product Code
JAQ · Radiology
Decision Date
Aug 30, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5700
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The intended use of the Mick Valencia Applicator is to facilitate the placement of a radioactive source near the target area on intact skin for remote afterloading brachytherapy treatment.
Device Story
The Mick Valencia Applicator is a passive, reusable accessory for remote afterloading brachytherapy. It consists of a tungsten-shielded applicator with a plastic cap, available in 20mm and 30mm diameters. The device is placed on the patient's intact skin and locked into position; it serves as a physical guide for a radioactive source delivered from an external HDR afterloader via a transfer tube. The applicator provides a defined path for the source to deliver a specific dose distribution to the target volume. It does not control the afterloader or the radiation dose. Used in clinical settings by radiation oncologists, physicists, dosimetrists, and therapists. The system is a closed path, preventing source contact with bodily fluids. Benefits include precise, homogeneous dose delivery to surface lesions while shielding surrounding tissue.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on design, material, and performance comparisons to the predicate device. Bench testing was performed to verify connection integrity to transfer tubes.
Technological Characteristics
Materials: Tungsten, plastic, stainless steel. Principle: Passive guide for HDR brachytherapy source. Form factor: 20mm and 30mm diameter applicators. Connectivity: Mechanical click-fit connection to HDR afterloader transfer tubes. Sterilization: Provided non-sterile; reusable, cleanable, and disinfectable. MR status: MR unsafe.
Indications for Use
Indicated for patients with skin or surface lesions (benign or malignant) requiring HDR brachytherapy treatment, regardless of age, gender, or ethnicity.
Regulatory Classification
Identification
A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.
{0}------------------------------------------------
August 30, 2023
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Mick Radio-Nuclear Instruments, Inc. % James Hurlman Director, Quality Management & Regulatory Affairs 521 Homestead Avenue MOUNT VERNON NY 10550
Re: K230155
Trade/Device Name: Mick Valencia Applicator Set Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radionuclide applicator system Regulatory Class: Class II Product Code: JAQ Dated: August 2, 2023 Received: August 3, 2023
Dear James Hurlman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Locon Werder
Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230155
### Device Name
MICK Valencia Applicator Sets
Indications for Use (Describe)
The intended use of the MICK Valencia Applicator is to facilitate the placement of a radioactive source near the target area on intact skin for remote afterloading brachytherapy treatment.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|------------------------------------------------------------|
| <span></span> | <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
| | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for MICK, a BEBIG Medical company. The word "MICK" is written in large, blue, sans-serif font. Below the word "MICK" is the phrase "A BEBIG Medical company" written in a smaller, orange, sans-serif font.
Mick Radio-Nuclear Instruments, Inc. A BEBIG Medical Company
> 521 Homestead Ave Mount Vernon, NY 10550 USA www.micknuclear.com
Main: +1 914 667 3999 Fax: +1 914 665 8834 Toll Free: +1 877 597 MRNI E-Mail: isaac.strickland@micknuclear.com
### Traditional 510(k) Summary of Safety and Effectiveness Information As required by section 807.92(c) K230155
August 28, 2023
#### 1. General Provisions
| Common Name: | Remote controlled radionuclide brachytherapy source (JAQ |
|-------------------|----------------------------------------------------------|
| Proprietary Name: | Mick Valencia Applicator Set |
| Owner Name | Mick Radio-Nuclear Instruments, Inc. |
| Address | 521 Homestead Avenue<br>Mount Vernon, New York 10550 |
| Telephone | (914) 667-3999 |
| Fax | (914) 665-8834 |
| Contact Name | Isaac Strickland |
| Contact email | isaac.strickland@micknuclear.com |
#### Name of Predicate Device: 2.
| Original<br>510(k) # | Trade/ Device Name | Regulation<br>Number | Regulatory<br>Classification | Product<br>Code | Review Panel |
|----------------------|------------------------------|----------------------|------------------------------|-----------------|--------------|
| K073107 | Valencia Skin Applicator Set | 21 CFR 892.5700 | Class II | JAQ | Radiology |
#### 3. Classification
Class 2, Product Code: JAQ Regulation Number: 21 CFR 892.5700
{4}------------------------------------------------
#### 4. Intended Use
The intended use of the Mick Valencia Applicator is to facilitate the placement of a radioactive source near the target area on intact skin for remote afterloading brachytherapy treatment.
#### Device Description 5.
The Mick Valencia Applicator is used for remote afterloading treatment on intact skin/surface. The applicator with associated plastic cap is placed onto the patient, locked into position, and connected to the remote afterloading equipment (via transfer tube) to facilitate the brachytherapy treatment in accordance with the physician approved treatment parameters. The radioactive source travels from the machine into the applicator and deliver the desired dose distribution to the defined treatment volume. When the treatment is completed, the source returns to the machine. The applicator is disconnected from the treatment unit and removed from the patient.
The applicator does not control the treatment unit, it strictly provides a treatment path for the radioactive source. The Mick Valencia applicator is temporarily placed during the treatments designed as a closed system to prevent the radioactive source from coming in contact with bodily fluids.
The device is the same as the legally marketed device cited, both in design and materials.
Image /page/4/Picture/6 description: The image shows two sets of metal objects. Each set contains two objects that appear to be probes or sensors. The objects have a cylindrical base and a rounded or cylindrical top, with a thin metal rod extending from the side.
MICK Valencia Skin Applicator
Elekta Valencia Skin Applicator
Image /page/4/Figure/9 description: The image shows a technical drawing with several dimensions labeled. The dimensions include Dim. A, Dim. B, Dim. C, Dim. D, and Dim. E. The drawing also labels the source cable and shielding of the device. Additionally, a measurement of 3.75 is shown on the right side of the image.
MICK Valencia Skin Applicator
{5}------------------------------------------------
| Set | Model | Description | Dim A | Dim B | Dim C | Dim D | Dim E |
|------------------|---------|-----------------------------|-------|-------|-------|--------|-------|
| 1503<br>(BEBIG) | 1503-20 | Valencia applicator<br>20mm | 20 | 28 | 95 | 91.25 | 105 |
| | 1503-30 | Valencia applicator<br>30mm | 30 | 36 | 115 | 111.25 | 129 |
| 1504<br>(Varian) | 1504-20 | Valencia applicator<br>20mm | 20 | 28 | 113 | 109.25 | 123 |
| | 1504-30 | Valencia applicator<br>30mm | 30 | 36 | 113 | 109.25 | 127 |
#### 6. Manufacturing Process
This device is manufactured according to Good Manufacturing Practices (GMPs) as defined in 21CFR part 820. The processes used to fabricate these devices are identical to those used for the predicate device described in this 510(k) notification.
#### 7. Non-Clinical Tests
Non-clinical testing is performed on the Mick Valencia Applicator connection to the transfer tube (the transfer tubes are supplied by the Afterloader manufacturer). These are bench top tests, performed on every applicator produced.
#### 8. Afterloader Compatibility
The Mick Valencia Applicator Set is compatible with the following HDR afterloaders, BEBIG Medical GmbH model Saginova®. Varian model GammaMedplus™ and Varian model Bravos afterloaders. The Mick Valencia Applicator sets are not compatible with Elekta afterloaders.
#### 9. Biocompatibility
The materials in the subject device are the same as the previously cleared predicate device and thus, there are no new biocompatibility concerns.
#### 10. Summary of Similarities and Differences
The Mick Valencia Applicator Sets are substantially equivalent in design, manufacture, construction and materials and have the same intended use and performance characteristics to the predicate device. The fundamental scientific technology is unchanged from the original predicate device. The connectors used on the Mick Valencia Applicator sets, to attach to the afterloaders transfer tubes, are substantially equivalent to the predicate device in their fundamental design for a standard click fit type connector.
Some differences from the predicate device are as follows;
1) Our device has more of a dome shape then the predicate device,
{6}------------------------------------------------
- 2) the shaft and tube diameters are slightly thicker, and
- 3) the Mick Valencia Applicator set comes with two marker caps; one of the marker caps is used for applicator placement, while the other cap is placed on the target area for treatment.
- 4) The Mick Valencia Applicator set comes in two sizes, in 20 mm & 30 mm diameters and both sizes are compatible with the BEBIG Medical GmbH model Saginova®, the Varian model GammaMedplus™ and the Varian model Bravos afterloaders.
#### 11. Comparison Table
The MICK Valencia Applicator Set has the same functionality, performance and design as the predicate device the Nucletron/Elekta Valencia Skin Applicator. The equivalence is based on a comparison of clinical, technical and biological aspects of the two devices. The following table identifies the clinical, technical and biological aspects of the predicate device and the MICK Valencia Skin Applicator Set:
| Equivalence Table | Predicate Device<br>Nucletron/Elekta<br>Valencia Skin Applicator<br>K073107 | MICK Valencia Skin<br>Applicator Sets<br>K230155 |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|--------------------------------------------------|
| Clinical Aspects | | |
| Intended Purpose: HDR Brachytherapy of<br>skin/surface lesions | Yes | Yes |
| Suitable for any patient with skin/surface lesions<br>(benign or malignant) regardless of their age,<br>gender, ethnicity or predisposition. | Yes | Yes |
| Can be used for any site on the body surface that<br>has been prescribed HDR Brachytherapy<br>treatment | Yes | Yes |
| Applicator is placed in direct contact with the<br>treatment area/skin surface. | Yes | Yes |
| Long term side effects are limited to a slight mark<br>or discoloration of the skin and/or hair loss in the<br>treatment area and is dependent on the<br>prescription dose | Yes | Yes |
| Patient is under the care of a Radiation<br>Oncologist, physicists, dosimetrists and radiation<br>therapists trained in Radiation Oncology and,<br>specifically, brachytherapy techniques | Yes | Yes |
| Technical Aspects | | |
| Available in 20 mm and 30mm diameter | Yes | Yes |
| Only one source position is used for the<br>treatment | Yes | Yes |
| Source delivery through a single channel (transfer | Yes | Yes |
| | | |
| tube) | | |
| Tungsten shielded applicator with reinforced<br>shielding in the dome to reduce dose from the<br>backside of the applicator | Yes | Yes |
| Includes flattening filter for homogeneous dose<br>distribution | Yes | Yes |
| Plastic caps available: providing a flat surface<br>improved homogeneous dose | Yes | Yes |
| Filtering low dose scatter radiation | Yes | Yes |
| Improved applicator positioning by<br>imprinting skin surface | Yes | Yes |
| 2 types of caps available (with and without<br>a marker) | Yes | Yes |
| The device is reusable | Yes | Yes |
| The applicator can be cleaned and disinfected | Yes | Yes |
| Applicator is MR unsafe | Yes | Yes |
| Provides and unobstructed path for the<br>radioactive source | Yes | Yes |
| Connects securely to remote afterloading<br>brachytherapy equipment via transfer tubes | Yes | Yes |
| Biological Aspects | | |
| The applicators are made from the same<br>materials. (tungsten, plastic caps, stainless steel) | Yes | Yes |
| The applicators have the same manufacturing<br>process | Yes | Yes |
| The applicators come in direct contact with skin<br>surface | Yes | Yes |
| Caps are provided for treatment to provide<br>filtration of scatter radiation. | Yes | Yes |
| The applicators are made from materials that are<br>biocompatible | Yes | Yes |
| Provided Non-Sterile | Yes | Yes |
{7}------------------------------------------------
#### Summary of technological considerations: 12.
The Mick Valencia Applicator Set is substantially equivalent to the cleared predicate device Valencia Skin Applicator Set, 510(k): K073107
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.