K230152 · Shanghai United Imaging Healthcare Co., Ltd. · LNH · May 23, 2023 · Radiology
Device Facts
Record ID
K230152
Device Name
uMR Omega
Applicant
Shanghai United Imaging Healthcare Co., Ltd.
Product Code
LNH · Radiology
Decision Date
May 23, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The uMR Omega system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces sagittal, transverse, coronal, and oblique cross sectional images, and spectroscopic images, and that display internal anatomical structure and/or function of the head, body and extremities. These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist the diagnosis. Contrast agents may be used depending on the region of interest of the scan.
Device Story
uMR Omega is a 3.0T superconducting MRI system with a 75cm bore. It acquires MR signals via RF coils and gradient coils, processed by a spectrometer and computer system. Modifications include a new RF power amplifier (uXD2201), additional RF coils (Tx/Rx Knee Coil-24, SuperFlex Body Wide-24), new VSM models, and a dockable patient table. It incorporates a camera-based 'uVision' positioning workflow and a Resoundant acoustic driver for MRE. New software features include 4D Flow Quantification, MRE, SNAP, CEST, T1Rho, and FSP+. Used in clinical settings by trained physicians/technicians. Output consists of anatomical and functional MR images and derived physical parameters. These outputs assist physicians in diagnostic decision-making, potentially improving diagnostic accuracy and patient comfort through advanced imaging sequences and streamlined positioning.
Clinical Evidence
Bench testing only. Performance evaluation reports provided for new features including 4D Flow (gre_fq), MRE (epi_se_mre), CEST (fse_mx), T1rho (gre, fse_mx), mPLD ASL, and silica gel imaging. Compliance with electrical safety (ES 60601-1), EMC (IEC 60601-1-2), and MR safety (IEC 60601-2-33) standards confirmed. Biocompatibility testing (ISO 10993-5, 10993-10) demonstrated no cytotoxicity, irritation, or sensitization.
Technological Characteristics
3.0T superconducting magnet; 75cm bore; 45mT/m gradient amplitude; 200T/m/s slew rate; 2 transmit channels; up to 96 receive channels. Materials tested per ISO 10993. Connectivity via DICOM/NEMA standards. Software includes pulse sequence control and image processing algorithms. Sterilization not applicable (non-invasive).
Indications for Use
Indicated for use as a magnetic resonance diagnostic device (MRDD) to produce cross-sectional images and spectroscopic images of the head, body, and extremities to display internal anatomical structure and/or function. Used by trained physicians to assist in diagnosis. Contrast agents may be used.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 23, 2023
Shanghai United Imaging Healthcare Co., Ltd. % Xin Gao Regulatory Affairs Specialist No.2258 Chengbei Road, Jiading District Shanghai, Shanghai 201807 China
Re: K230152
Trade/Device Name: uMR Omega Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH Dated: April 24, 2023 Received: April 24, 2023
Dear Xin Gao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
D.G.K.
Daniel M. Krainak, Ph.D. Assistant Director Magnetic Resonance and Nuclear Medicine Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230152
Device Name uMR Omega
Indications for Use (Describe)
The uMR Omega system is indicated for use as a magnetic device (MRDD) that produces sagittal. transverse, coronal, and oblique cross sectional images, and that display internal anatomical structure and/or function of the head, body and extremities.
These images and the physical parameters derived from the interpreted by a trained physician yield information that may assist the diagnosis. Contrast agents may be used depending on the region of interest of the scan.
| Type of Use (Select one or both, as applicable) |
|------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><span style="font-size:16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> |
| <div style="display:flex; align-items:center;"><span style="font-size:16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the words is a stylized "U" shape, which is also in bold. The color of the logo is a dark gray or blue.
### 510 (k) SUMMARY
### 1. Date of Prepared
April 24, 2023
#### 2. Sponsor Identification
Shanghai United Imaging Healthcare Co., Ltd. No.2258 Chengbei Rd. Jiading District, 201807, Shanghai, China
Contact Person: Xin GAO Position: Regulatory Affairs Specialist Tel: +86-021-67076888-5386 Fax: +86-021-67076889 Email: xin.gao@united-imaging.com
#### 3. Identification of Proposed Device(s)
Trade Name: uMR Omega Common Name: Magnetic Resonance Diagnostic Device Model: uMR Omega Product Code: LNH Regulation Number: 21 CFR 892.1000 Device Class: II
#### 4. Identification of Predicate Devices(s)
Predicate Device 510(k) Number: K220332 Device Name: uMR Omega Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: II Product Code: LNH
#### 5. Device Description
The uMR Omega is a 3.0T superconducting magnetic resonance diagnostic device with a 75cm size patient bore. It consists of components such as magnet, RF power amplifier, RF coils, gradient power amplifier, gradient coils, patient table, spectrometer, computer, equipment cabinets, power distribution system, internal communication system, and vital signal module etc. The uMR Omega Magnetic Resonance Diagnostic Device is designed to conform to NEMA and DICOM
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" shape, also in a dark color, with a vertical line running down the center, creating a unique design element.
standards.
uMR Omega has been previously cleared by FDA via K220332.
The modification performed on the uMR Omega (K220332) in this submission is due to the following changes that include:
- (1) Addition of the Radio Frequency Power Amplifier model: uXD2201.
- (2) Addition of Radio Frequency Coils: Tx/Rx Knee Coil -24 and SuperFlex Body Wide - 24.
- (3) Addition of VSM models: uVWMERP, uMVRX and mmw100.
- (4) Addition and modification of pulse sequences
- New sequences: gre_snap, gre_quick_4dncemra, gre_trass, epi_se_mre, gre_fq. a)
- b) Added Associated options for certain sequences: T1Rho, CEST, FSP+ (eT1, eGM), silica gel imaging, mPLD, whole heart coronary angiography imaging.
- Name change of pulse sequence: gre (old name: gre sp), gre fine (old name: c) gre_bssfp_fi), gre_quick (gre_quick_wfi merge to gre_quick), gre_quick (gre_bloustracking merge to gre_quick).
- (5) Addition of imaging processing methods: 4D Flow Quantification, Magnetic Resonance Elastography (MRE), SNAP, CEST, T1Rho, FSP+.
- (6) Addition of Dockable Patient Table.
- (7) Add the positioning workflow (uVision) with camera.
- (8) Addition of Resoundant Acoustic Driver System for Magnetic Resonance Elastography (MRE).
These modifications do not affect the intended use or alter the fundamental scientific technology of the device.
#### 6. Intended Use
The uMR Omega system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces sagittal, transverse, coronal, and oblique cross sectional images, and spectroscopic images, and that display internal anatomical structure and/or function of the head, body and extremities.
These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist the diagnosis. Contrast agents may be used depending on the region of interest of the scan.
# 7. Technological Characteristic
The differences from the predicate device are discussed in the comparison table in this submission as below.
| ITEM | Proposed Device<br>uMR Omega | Predicate Device<br>uMR Omega (K220332) | Remark |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| General | | | |
| 510(K) No. | | K220332 | |
| ITEM | Proposed Device<br>uMR Omega | Predicate Device<br>uMR Omega (K220332) | Remark |
| Product Code | LNH | LNH | Same |
| Regulation No. | 21 CFR 892.1000 | 21 CFR 892.1000 | Same |
| Class | II | II | Same |
| Indications For Use | The uMR Omega system is<br>indicated for use as a magnetic<br>resonance diagnostic device<br>(MRDD) that produces sagittal,<br>transverse, coronal, and oblique<br>cross sectional images, and<br>spectroscopic images, and that<br>display internal anatomical<br>structure and/or function of the<br>head, body and extremities.<br>These images and the physical<br>parameters derived from the<br>images when interpreted by a<br>trained physician yield<br>information that may assist the<br>diagnosis. Contrast agents may<br>be used depending on the region<br>of interest of the scan. | The uMR Omega system is<br>indicated for use as a magnetic<br>resonance diagnostic device<br>(MRDD) that produces sagittal,<br>transverse, coronal, and oblique<br>cross sectional images, and<br>spectroscopic images, and that<br>display internal anatomical<br>structure and/or function of the<br>head, body and extremities.<br>These images and the physical<br>parameters derived from the<br>images when interpreted by a<br>trained physician yield<br>information that may assist the<br>diagnosis. Contrast agents may<br>be used depending on the region<br>of interest of the scan. | Same |
| Magnet system | | | |
| Field Strength | 3.0 Tesla | 3.0 Tesla | Same |
| Type of Magnet | Superconducting | Superconducting | Same |
| Patient-accessible<br>bore dimensions | 75cm | 75cm | Same |
| Type of Shielding | Actively shielded, OIS<br>technology | Actively shielded, OIS<br>technology | Same |
| Magnet<br>Homogeneity | 2.30ppm @ 50cm DSV<br>0.80ppm @ 45cm DSV<br>0.38ppm @ 40cm DSV<br>0.08ppm @ 30cm DSV<br>0.02ppm @ 20cm DSV<br>0.002ppm @ 10cm DSV | 2.30ppm @ 50cm DSV<br>0.80ppm @ 45cm DSV<br>0.38ppm @ 40cm DSV<br>0.08ppm @ 30cm DSV<br>0.02ppm @ 20cm DSV<br>0.002ppm @ 10cm DSV | Same |
| Gradient system | | | |
| Max gradient<br>amplitude | 45mT/m | 45mT/m | Same |
| Max slew rate | 200T/m/s | 200T/m/s | Same |
| Shielding | active | active | Same |
| Cooling | water | water | Same |
| RF system | | | |
| ITEM | Proposed Device<br>uMR Omega | Predicate Device<br>uMR Omega (K220332) | Remark |
| Resonant<br>frequencies | 128.23 MHz | 128.23 MHz | Same |
| Number of transmit<br>channels | 2 | 2 | Same |
| Number of receive<br>channels | Up to 96 | Up to 96 | Same |
| Amplifier peak<br>power per channel | 18 kW, 20 kW | 18 kW | Note 1 |
| RF Coils | | | |
| Volume Transmit<br>Coil | Yes | Yes | Same |
| Head & Neck Coil -<br>24 | Yes | Yes | Same |
| Body Array Coil -<br>12 | Yes | Yes | Same |
| Breast Coil - 10 | Yes | Yes | Same |
| Flex Coil Large - 8 | Yes | Yes | Same |
| Flex Coil Small - 8 | Yes | Yes | Same |
| Knee Coil - 12 | Yes | Yes | Same |
| Lower Extremity<br>Coil - 36 | Yes | Yes | Same |
| Shoulder Coil - 12 | Yes | Yes | Same |
| Small Loop Coil | Yes | Yes | Same |
| Spine Coil - 32 | Yes | Yes | Same |
| Wrist Coil - 12 | Yes | Yes | Same |
| Cardiac Coil - 24 | Yes | Yes | Same |
| Temporomandibular<br>Joint Coil - 4 | Yes | Yes | Same |
| Foot & Ankle Coil -<br>24 | Yes | Yes | Same |
| Head Coil - 32 | Yes | Yes | Same |
| Head Coil - 12 | Yes | Yes | Same |
| Carotid Coil - 8 | Yes | Yes | Same |
| Infant Coil - 24 | Yes | Yes | Same |
| Body Array Coil -<br>24 | Yes | Yes | Same |
| Head & Neck Coil -<br>48 | Yes | Yes | Same |
| Spine Coil - 48 | Yes | Yes | Same |
| Head Coil - 64 | Yes | Yes | Same |
| SuperFlex Body -<br>24 | Yes | Yes | Same |
| ITEM | Proposed Device<br>uMR Omega | Predicate Device<br>uMR Omega (K220332) | Remark |
| SuperFlex Large -<br>12 | Yes | Yes | Same |
| SuperFlex Small -<br>12 | Yes | Yes | Same |
| Tx/Rx Knee Coil -<br>24 | Yes | No | Note 2 |
| SuperFlex Body<br>Wide - 24 | Yes | No | Note 3 |
| Dimensions | Patient Table:<br>width 640mm, height 880mm,<br>length 2620mm<br>Dockable Patient Table:<br>width 826mm, height 880mm,<br>length 2578mm | width 640mm, height 880mm,<br>length 2620mm…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.