K230145 · Ge Medical Systems Information Technologies, Inc. · DXQ · Mar 14, 2023 · Cardiovascular
Device Facts
Record ID
K230145
Device Name
ONE-CUF
Applicant
Ge Medical Systems Information Technologies, Inc.
Product Code
DXQ · Cardiovascular
Decision Date
Mar 14, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1120
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Indirect measurement of blood pressure
Device Story
ONE-CUF is a non-sterile, single-patient use blood pressure cuff accessory for non-invasive blood pressure (NIBP) measurement systems. It consists of an inflatable, non-distensible bladder sized to encircle a patient's limb, with flanges for flexible tubing and connectors (bayonet or DINACLICK) to attach to NIBP hoses. During operation, the cuff is inflated to occlude an artery, facilitating automated NIBP measurement. It is used in clinical settings by healthcare providers to obtain blood pressure readings. The device does not contain serviceable parts.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including environmental conditioning, leak testing, pressure capacity, mechanical strength, tensile testing, and biocompatibility (cytotoxicity, sensitization, irritation per ISO 10993-1:2018). Summative usability testing was performed.
Technological Characteristics
Materials: Polyester/polyvinyl chloride laminate, Zephyron-Ion solvent-based ink. Sensing: Inflatable bladder for arterial occlusion. Connectivity: Single/dual tube with bayonet or DINACLICK connectors. Standards: IEC 80601-2-30:2018, ISO 81060-1:2007. Pressure range: 0-300 mmHg. Leak rate: max 0.6 sccm. Non-sterile, single-patient use.
Indications for Use
Indicated for use as an accessory in non-invasive blood pressure measurement systems. Available in pediatric and adult sizes. Non-sterile, single-patient use. Not designed, sold, or intended for use except as indicated.
Regulatory Classification
Identification
A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
March 14, 2023
GE Medical Systems Information Technologies, Inc. % Michelle Johnson Regulatory Affairs Manager GE Healthcare 9900 Innovation Drive Wauwatosa, Wisconsin 53226
Re: K230145
Trade/Device Name: One-cuf Regulation Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II Product Code: DXO Dated: January 18, 2023 Received: January 18, 2023
Dear Michelle Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Robert T. Kazmierski -S
for
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230145
Device Name ONE-CUF
Indications for Use (Describe)
The blood pressure cuff is an accessory used in con-invasive blood pressure measurement systems. It is non-sterile, single-patient use. It is available in pediatric and adult sizes. The cuff is not designed, sold or intended for use except as indicated.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size:10px"> <svg height="10" width="10"> <rect height="10" style="fill:none;stroke:#000000;stroke-width:2" width="10"></rect> <line style="stroke:#000000;stroke-width:2" x1="0" x2="10" y1="0" y2="10"></line> <line style="stroke:#000000;stroke-width:2" x1="0" x2="10" y1="10" y2="0"></line> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:10px"> <svg height="10" width="10"> <rect height="10" style="fill:none;stroke:#000000;stroke-width:2" width="10"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| In accordance with 21 CFR 807.92 the following summary of information is provided: | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | January 9, 2023 |
| Primary Contact Person: | Michelle Johnson<br>Senior Quality Staff Manager - Regulatory Affairs<br>GE Medical Systems Information Technologies, Inc.<br>Phone: +1 414 429 9263<br>Email: Michelle.Johnson@ge.com |
| Secondary Contact Person: | Jennifer Strauther<br>Regulatory Affairs Leader<br>GE Medical Systems Information Technologies, Inc.<br>9900 Innovation Drive<br>Wauwatosa, WI 53226<br>Phone: +1 262 330 2112<br>Email: Jennifer.Strauther@ge.com |
| Device Trade Name: | ONE-CUF |
| Common/Usual Name: | Blood Pressure Cuff |
| Classification Names: | Blood Pressure Cuff |
| Regulation: | 21 CFR 870.1120 Blood Pressure Cuff |
| Classification: | II |
| Product Code: | DXQ |
| Predicate Device: | SOFT-CUF (K120125) |
| Intended Use: | Indirect measurement of blood pressure |
| Indications for Use: | The blood pressure cuff is an accessory used in conjunction with<br>non-invasive blood pressure measurement systems. It is non-sterile,<br>single-patient use. It is available in pediatric and adult sizes. The<br>cuff is not designed, sold or intended for use except as indicated. |
{4}------------------------------------------------
#### 5.1 Device Description:
ONE-CUF blood pressure cuffs are accessories used in conjunction with noninvasive blood pressure (NIBP) measurement systems to obtain a blood pressure measurement.
Blood pressure cuffs are non-sterile and single-patient use and may not be reprocessed for use on additional patients. They are available in pediatric and adult sizes. The devices are not designed, sold or intended for use except as indicated.
Blood pressure cuffs do not include serviceable parts or components.
Non-Invasive Blood Pressure Cuffs incorporate an inflatable non-distensible bladder, sized to encircle a patient's limb. The cuff includes one or two flanges for attaching flexible tubing. This allows air to flow in and out of the cuff bladder for inflation and deflation. Inflation allows for occlusion of an artery to facilitate the measurement of automated noninvasive blood pressure (NIBP). The cuff tubes are terminated with connectors that allow for attachment to a blood pressure hose. This cuff will be offered with two different connection systems, bayonet and DINACLICK,
#### 5.2 Comparison of Technological Characteristics with the Predicate Device
The Non-Invasive Blood Pressure Cuffs employ the same fundamental scientific technology as its predicate device. The differences between the ONE-CUF and the
| Feature/Function | Predicate Device/System<br>Name and 510 (k)<br>SOFT-CUF Blood Pressure Cuff K120125 | Proposed Device/System Name<br>ONE-CUF Blood Pressure Cuff | Comparison Result |
|-----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| Intended Use | | | |
| • Indications for Use | Indications for Use<br>GE CRITIKON blood pressure cuffs are<br>accessories used in conjunction with<br>noninvasive blood pressure (NIBP)<br>measurement systems. SOFT-CUF and<br>CLASSIC-CUF cuffs and inflation systems<br>are non-sterile and limited reuse (may be<br>single-patient use or optional limited reuse).<br>They are available in neonatal, pediatric and<br>adult sizes. The devices are not designed,<br>sold or intended for use except as indicated. | Indications For Use<br>The blood pressure cuff is an accessory<br>used in conjunction with non-invasive<br>blood pressure measurement systems. It is<br>non-sterile, single-patient use. It is available<br>in pediatric and adult sizes. The cuff is not<br>designed, sold or intended for use except as<br>indicated. | Equivalent. |
| • Use | Non-sterile, single patient use or limited<br>reuse | Non-sterile, Single patient use | Equivalent |
| • Population | Adult, pediatric, neonatal | Adult and Pediatric | Equivalent. |
| • Intended Use | Indirect measurement of blood pressure | Indirect measurement of blood pressure | Identical |
| Environmental Specifications - Operating Conditions | | | |
| • Temperature | 0° C to 46° C | 0° C to 40° C | Equivalent |
#### predicate SOFT-CUT do not raise any new or questions on the safety and effectiveness.
{5}------------------------------------------------
| Feature/Function | Predicate Device/System<br>Name and 510 (k)<br>SOFT-CUF Blood Pressure Cuff K120125 | Proposed Device/System Name<br>ONE-CUF Blood Pressure Cuff | Comparison Result |
|---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| Environmental Specifications - Storage Conditions | | | |
| • Relative Humidity | 15% to 90% humidity, non-condensing | 15% to 90% humidity, non-condensing | Identical |
| • Temperature | -20° C to 55° C | -20° C to 55° C | Identical |
| • Relative Humidity | 0% to 95% humidity, non-condensing | 0% to 95% humidity, non-condensing | Identical |
| Physical Specifications | | | |
| • Limb Circumference<br>(Ranges in cm) | Neonatal #1 (3-6)<br>Neonatal #2 (4-8)<br>Neonatal #3 (6-11)<br>Neonatal #4 (7-13)<br>Neonatal #5 (8-15)<br>Infant (8-13)<br>Child (12-19)<br>Child Long (12-19)<br>Small Adult (17-25)<br>Small Adult Long (17-25)<br>Adult (23-33)<br>Adult Long (23-33)<br>Large Adult (31-40)<br>Large Adult Long (31-40)<br>Thigh (38-50) | Infant (8-13)<br>Child (12-19)<br>Small Adult (17-25)<br>Adult (23-33)<br>Adult Long (23-33)<br>Large Adult (31-40) | Equivalent |
| • Bladder Size | Various bladder sizes for neonate, pediatric<br>and adult populations | Various bladder sizes for pediatric and adult<br>populations | Equivalent |
| Feature/Function | Predicate Device/System<br>Name and 510 (k)<br>SOFT-CUF Blood Pressure Cuff K120125 | Proposed Device/System Name<br>ONE-CUF Blood Pressure Cuff | Comparison Result |
| • Cuff Connector<br>Configurations | Adult and Pediatric 2-Tube configuration:<br>• Screw Connector<br>• Sub-Miniature Connector<br>• Mated Subminiature Connector<br>• Dual CLICK-IT (branded as<br>"DINACLICK") Connector<br>Adult and Pediatric 1-Tube configuration:<br>• Bayonet Connector<br>• Screw<br>• Sub-Miniature Connector<br>Neonatal<br>• 1-Tube and 2-Tube Male Slip Luer Taper<br>• 1-Tube and 2-Tube SNAP-IT (branded as<br>"NeoSnap") Connectors | Adult and Pediatric 2-Tube configuration:<br>• Dual CLICK-IT (branded as<br>"DINACLICK") Connector<br>Adult 1-Tube configuration:<br>• Bayonet Connector | Equivalent. |
| Materials | | | |
| • Cuff / Integrated<br>Bladder | SOFT-CUF: Laminate of polyester /<br>polyvinyl chloride | ONE-CUF: Laminate of polyester /<br>polyvinyl chloride | Equivalent. |
| • Ink | SOFT-CUF: Zephyron-Ion Solvent Based | ONE-CUF: Zephyron-Ion Solvent Based | Identical. |
| Compatibility | | | |
| • Device | Single / dual tubes, adaptors and bulb / valve<br>assemblies for use with manual and<br>automated sphygmomanometers. | ONE-CUF: Single / dual tubes for use with<br>manual and automated<br>sphygmomanometers. | Equivalent. |
| Performance Specifications | | | |
| • Leak Rate | Maximum allowable pressure loss rate is 0.6<br>sccm (standard cubic cm per minute) | Maximum allowable pressure loss rate is<br>0.6 sccm (standard cubic cm per minute) | Identical |
| • Air Leakage | Less than 4mmHg/min | Less than 4mmHg/min | Identical |
| • Pressure Range | 0-300 mmHg | 0-300 mmHg | Identical |
| Standards Compliance (Applicable Sections) | | | |
| • Automated/Manual<br>Sphygmomanomete<br>rs | SP10:2002/A1:2003/A2:2006 | IEC 80601-2-30:2018 (applicable sections)<br>ISO 81060-1:2007 (applicable sections) | Equivalent |
| • Automated/Manual<br>Sphygmomanomete<br>rs | EN 1060-1:1995/A1:2002/A2:2009<br>(applicable sections)<br>EN 1060-2:1995/C1:2002/A1:2009<br>(applicable sections)<br>IEC 80601-2-30:2009 (applicable sections)<br>ISO 81060-1:2007 (applicable sections) | IEC 80601-2-30:2018 (applicable sections)<br>ISO 81060-1:2007 (applicable sections) | Equivalent |
| Feature/Function | Predicate Device/System<br>Name and 510 (k)<br>SOFT-CUF Blood Pressure Cuff K120125 | Proposed Device/System Name<br>ONE-CUF Blood Pressure Cuff | Comparison Result |
| <i>Biocompatibility</i> | | | |
| Biocompatibility<br>Standard Compliance | Cuff fabric, ink, loop evaluated per<br>ISO10993-1:2009<br>IrritationSensitizationCytotoxicity | Cuff fabric and loop evaluated per<br>ISO10993-1:2018<br>IrritationSensitizationCytotoxicity | Equivalent |
{6}------------------------------------------------
{7}------------------------------------------------
## 5.3 Determination of Substantial Equivalence:
#### Summary of Non-Clinical Testing:
The following performance data were provided in support of the substantial equivalence determination.
- Environmental Conditioning ●
- Leak
- Pressure Capacity ●
- Mechanical Strength ●
- Tensile Testing ●
- Standards .
- Biocompatibility ●
#### Biocompatibility testing
Biocompatibility testing was performed in accordance with per ISO 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. In accordance with FDA guidance document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" issued September 4, 2020.
- Chemical characterization ●
- . Cytotoxicity
- Sensitization ●
- Irritation/intracutaneous reactivity
#### Compliance with Voluntary Standards
The ONE-CUF Warmer is designed and tested for compliance with the following performance standards:
{8}------------------------------------------------
- 1. IEC 80601-2-30:2018 Medical electrical equipment Part 2-30: Particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers
- 2. ISO 81060-1:2007 Non-invasive sphygmomanometers Part 1: Requirements and test methods for non-automated measurement type
#### Animal Study
The ONE-CUF did not required animal tests to support substantial equivalence.
#### Clinical Studies
The ONE-CUF did not required clinical tests to support substantial equivalence.
## Human Factors Validation
The ONE-CUF performed summative usability to support substantial equivalence.
#### Conclusion:
Based on the indications for use, technological characteristics, and comparison to predicate devices, the subject ONE-CUF has been shown to be substantially equivalent to legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.