K230133 · Gelesis, Inc. · QFQ · Jan 19, 2024 · Gastroenterology, Urology
Device Facts
Record ID
K230133
Device Name
Plenity
Applicant
Gelesis, Inc.
Product Code
QFQ · Gastroenterology, Urology
Decision Date
Jan 19, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5982
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Plenity® is indicated to aid in weight management in adults with excess weight or obesity, with a Body Mass Index (BMI) of 25 to 40 kg/m2, when used in conjunction with diet and exercise.
Device Story
Plenity is a non-systemic, space-occupying weight management device consisting of porcine gelatin capsules containing absorbent hydrogel particles (cellulose and citric acid). Patients ingest three capsules with water 20-30 minutes before lunch and dinner. In the stomach, capsules disintegrate, releasing particles that hydrate up to 100 times their weight, occupying approximately 25% of stomach volume. The resulting hydrogel matrix increases volume, elasticity, and viscosity of stomach contents to create a sensation of fullness. The gel passes through the GI tract, maintaining structure until reaching the colon, where water is released and reabsorbed; particles are eliminated via bowel movements. Used by patients for self-managed weight loss; no physician intervention required for OTC use. Benefits include non-pharmacological, non-systemic weight management aid.
Clinical Evidence
No new clinical studies conducted for OTC switch. Original De Novo clearance supported by GLOW and GLOW-EX studies. GLOW pivotal study (n=not specified) showed 58.6% of patients lost ≥5% total body weight at 6 months (95% CI 51.8-65.4%); 27% lost ≥10%. TEAE incidence (71.3% Plenity vs 70.6% Placebo) and device-related AEs (39.5% Plenity vs 30.3% Placebo) showed no significant difference. No unanticipated adverse device effects or deaths reported.
Technological Characteristics
Porcine gelatin capsules containing cellulose and citric acid hydrogel particles. Capsule disintegration time: NMT 15 min. Media uptake: NLT 70 g/g. Elastic modulus: NLT 500 Pa. Biocompatibility per ISO 10993-1:2018. Non-sterile. Shelf-life: 30 months. Shipping/environmental testing per ASTM D4169:2016 and ASTM D999-08:2015.
Indications for Use
Indicated for weight management in adults with excess weight or obesity (BMI 25-40 kg/m2) as an adjunct to diet and exercise.
Regulatory Classification
Identification
This device is an ingested material that transiently occupies space in the stomach. The device passes from the body via the natural gastrointestinal tract.
Special Controls
In combination with the general controls of the FD&C Act, the ingested, transient, space occupying device for weight management and/or weight loss is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible for its intended use.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions for use, as follows:
(i) Performance bench testing in a simulated use model must evaluate device disintegration and device hydration state throughout the gastrointestinal tract;
(ii) Bioburden and moisture content assessments must evaluate device infection risk throughout the labeled shelf life; and
(iii) Performance data must support the shelf life of the device by demonstrating continued package integrity and device functionality over the labeled shelf life.
(3) Clinical performance testing must demonstrate the device performs as intended and evaluate the following:
(i) Weight change;
(ii) All adverse events, including obstruction, dilation, diarrhea, constipation, and dehydration; and
(iii) Interaction with representative medications.
(4) Physician and patient device labeling must state:
(i) The clinical benefit of the device as assessed by using percent total body weight loss;
(ii) Treatment must be offered in combination with diet and exercise;
(iii) Instructions on how to use the device as intended including how to avoid interaction with medication; and
(iv) The shelf life of the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 19, 2024
Gelesis Inc. Giorgio Sorcini 501 Boyston St., Ste. 6102 Boston, Massachusetts 02116
Re: K230133 Trade/Device Name: Plenity Regulation Number: 21 CFR 876.5982 Regulation Name: Ingested, Transient, Space Occupying Device For Weight Management and/or Weight Loss Regulatory Class: II Product Code: QFQ Dated: December 21, 2023 Received: December 22, 2023
Dear Giorgio Sorcini:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Anthony Lee -S 2024.01.19 15:03:16 -05'00'
Anthony Lee, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230133
Device Name Plenity®
Indications for Use (Describe)
Plenity® is indicated to aid in weight management in adults with excess weight or obesity, with a Body Mass Index (BMI) of 25 to 40 kg/m2, when used in conjunction with diet and exercise.
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Gelesis. The logo consists of the word "GELESIS" in a dark green sans-serif font. Above the word is a graphic of orange dots arranged in a semi-circular pattern, resembling a sun or a burst of energy.
K230133 page 1 of 5
#### 510(k) Summary
# Regulatory Information
Classification Name: Ingested, transient, space occupying device for weight management and/or weight loss.
Common Name: Weight management device
Regulation Number: 21 CFR 876.5982
Classification: Class II
Product Code: QFQ
Device Trade Name: Plenity®
Indications for Use: Plenity® is indicated to aid in weight management in adults with excess weight or obesity, with a Body Mass Index (BMI) of 25 to 40 kg/m², when used in conjunction with diet and exercise.
| Sponsor Information: | Gelesis Srl<br>Via Europa, 187<br>73021 Calimera (LE)<br>Italy |
|----------------------|-------------------------------------------------------------------------------------------------------------|
| Contact Person: | Giorgio Sorcini<br>Sr. Regulatory Affairs Director<br>Phone: +39 347 5047105<br>email: gsorcini@gelesis.com |
Date Prepared: January 16th, 2024
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Gelesis. The logo consists of the word "GELESIS" in a simple, sans-serif font. Above the word is a stylized graphic of orange dots arranged in a semi-circular pattern, resembling a sun or a burst of energy. The overall design is clean and modern.
K230133 page 2 of 5
# Device description
Plenity is a porcine gelatin capsule that contains thousands of absorbent hydrogel particles (0.75 grams [g] per capsule); each particle is approximately the size of a grain of salt. Plenity OTC is non-systemic and works directly in the gastrointestinal (GI) tract. Plenity OTC hydrogel is made from two naturally derived ingredients, cellulose and citric acid, that form a three-dimensional matrix designed to occupy volume in the stomach and small intestine, to create a sensation of fullness. Plenity OTC is provided non-sterile.
Image /page/4/Picture/4 description: The image shows a pile of white powder next to several empty white capsules. The capsules are arranged in a semi-circle around the powder. The powder is granular and appears to be a fine powder. The capsules are cylindrical and have a rounded end.
#### Plenity OTC capsules with hydrogel exposed
The capsules disintegrate in the stomach and release the Plenity OTC particles, which can hydrate up to 100 times their original weight. When fully hydrated, the individual non-clustering Plenity OTC particles from the 2.25 g/dose occupy about a quarter of the average stomach volume. The gel particles mix with ingested foods, creating a larger volume with higher elasticity and viscosity in the stomach and small intestine.
Plenity OTC passes through the digestive system, maintaining its three-dimensional structure in the stomach and small intestine before breaking down in the colon. The water is then released and reabsorbed by the body. Plenity particles are eliminated through normal bowel movements; they are not absorbed by the body.
Patients consume three (3) capsules (2.25 g/dose) with water before both lunch and dinner. Plenity OTC is supplied in peelable pods that provide the two doses patients take daily. Each individual pod holds a single dose of three (3) capsules. Plenity OTC pods are supplied in a monthly kit.
## Predicate Device
The Plenity OTC device is identical to the currently cleared device Plenity (DEN180060) in terms of its technological characteristics and mechanism of action. Although this application proposes to switch from prescription required to over-the-counter use, Plenity OTC has the same intended use as its predicate, Plenity (DEN180060).
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Gelesis. The logo consists of the word "GELESIS" in a simple, sans-serif font. Above the word is a graphic of orange dots arranged in a semi-circular pattern, resembling a sun or a burst of energy.
# Technological Characteristics
As noted above, Plenity OTC remains technologically identical to its predicate device. See the table below for the device specifications.
# Table: Plenity OTC Specifications
| | Plenity / Plenity OTC |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| General Specifications | |
| Capsule disintegration<br>time | NMT 15 minutes |
| Media uptake after<br>disintegration | NLT 70 g/g |
| Capsule content<br>uniformity | 760 mg ± 10% |
| Media uptake | NLT 70 g/g |
| Elastic modulus | NLT 500 Pa |
| Loss on drying | NMT 10% |
| Dosing | Plenity OTC should be taken with water twice a day, 20 to 30<br>minutes before lunch and 20 to 30 minutes before dinner. Each<br>dose includes 3 capsules of Plenity (provided in a single pod) |
## Summary of Nonclinical/Bench Studies
No non-clinical/bench studies were specifically conducted in support of the switch to over-thecounter use as the device is the same of the predicate. The submitted device has an upgraded labeling interface to support self-selection ad use without the intervention of a prescriber.
## Biocompatibility
Plenity and Plenity OTC devices underwent biocompatibility assessment according to ISO 10993-1:2018 requirements for a device having permanent (> 30 days) surface contact with mucosal membranes. All the biological endpoints have been addressed by biocompatibility studies and/or rationales based on scientific literature, as the device components have long and documented story of use with the same contact. Endpoints considered are:
- . Cytotoxicity, tested in compliance with ISO 10993-5 (elution and Agar mediated assays).
- Sensitization, tested in compliance with ISO 10993-10 (Guinea Pig Maximization Test). .
- Irritation: tested in compliance with ISO 10993-10 (Oral Mucosal Irritation Study in . Hamsters - Collar Method - 14 Days).
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Gelesis. The logo consists of the word "GELESIS" in a dark green sans-serif font. To the right of the word is a graphic of orange dots arranged in a semi-circular pattern, resembling a sun or a burst of energy.
- Acute Systemic Toxicity: tested in compliance with ISO 10993-11 (Oral Toxicity Study in . Mice, 7 Days).
- Genotoxicity: tested in compliance with ISO 10993-3 ( Bacterial Reverse Mutation Study, . Genotoxicity - Mouse Lymphoma Assay and Mouse Peripheral Blood Micronucleus Study)
- Sub-chronic Toxicity: tested in compliance with ISO 10993-11 (Safety and Tolerability 90 Days Study in Rats with Recovery)
- Chronic Systemic Toxicity: assessed based on chemical characterization according to ISO 10993-18 and literature data.
- Reproductive/Developmental Toxicity: assessed by literature on known experience with the constituent materials.
- . Biodegradation and toxicokinetic: assessed based on degradation products studies according to ISO 10993-13 compared with literature data.
- . Implantation: assessed by literature data on constituent substances, but not required for devices administered orally.
# Bench models
Studies were performed on a bench model to evaluate swelling properties of the hydrogel at each relevant point in a simulated gastro-intestinal transit.
A specific assay was performed on bench to assess, by an internally developed method (RD-0518.01), correct hydration with no lumps of the hydrogel in simulated clinical conditions.
# Stability and shipping studies
Stability studies in compliance with ICH guidances have been performed with Plenity pods (primary packaging level) and are fully applicable to Plenity OTC as it is the same device apart from secondary package labeling. Ongoing stability studies confirmed on bench a shelf-life of 30 months for the most recent version of Plenity / Plenity OTC.
Similarly, ship testing was performed to verify that the packaging configurations for Plenity met the expected shipping and storage conditions during commercial use. The shipping/environmental testing was performed per ASTM D4169:2016 and ASTM D999-08:2015. The test environment included exposure to vibration, stacking, compression and impact simulations.
# Usability studies
Usability studies in compliance with IEC 62366-1:2015/AMD1:2020 have been performed to support transition of Plenity to OTC distribution. As the graphical interface of the Plenity OTC box was provided of a BMI calculation and selection table and other information relevant to let the patient self-identify as intended or unintended user, a first summative usability study, conducted on 27 subjects, identified some weakness in communication of requirements for safe selection and use of the device. Based on these results, device labeling has been improved towards a more effective and clear communication. The upgraded labeling design was validated by a further
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Gelesis. The logo consists of the word "GELESIS" in a dark green sans-serif font. To the right of the word is a stylized sunburst design made up of small orange circles. The circles are arranged in a semi-circular pattern, with the largest circle at the top and the smallest circles at the bottom.
summative comprehension study, involving 28 subjects, which demonstrated an effective communication of requirements for a safe device self-selection and use.
# Summary of Clinical Information
No clinical studies were specifically conducted in support of the switch to over-the-counter use. The Plenity was originally granted by De Novo submission based on results of GLOW and GLOW-EX studies. It demonstrated robust efficacy in GLOW pivotal study, with 58.6% of patients losing ≥5% total body weight in 6 months (95% CI 51.8 to 65.4%), and 27% of patient loosing 10% or more. The incidence of treatment emergent adverse events (TEAEs) in the GLOW study (71.3% in Plenity arm versus 70.6% Placebo) and the incidence of AEs related to Plenity (39.5% in Plenity arm versus 30.3% in Placebo) were no different in the device group compared to Placebo. There were no unanticipated adverse device effects or deaths in the study. There were no SAEs associated with the use of Plenity. There was one SAE in the Placebo treatment group. This exemplary safety profile was maintained for the 21 subjects exposed to Plenity for one year in the GLOE-EX study.
## Conclusion
Gelesis has demonstrated that the proposed Plenity OTC device is substantially equivalent to its predicate device. The intended use and mechanism of action are identical. The device characteristics and specifications. which includes materials. design and performance are unchanged. Plenity OTC meets the same special controls regarding biocompatibility and nonclinical performance as the predicate device.
Plenity OTC was tested and shown to perform according to its specifications that are the same as those of the Plenity (DEN180060). The validation of Plenity OTC labeling by Human Factors studies has provided confidence that the Plenity OTC interface has been properly designed to support effective self-selection, based on device Indications for Use, and recognition of any needs to consult a doctor in specific situations.
After review of the Plenity OTC risk analysis, it was concluded that there were no new unacceptable safety and effectiveness risks identified as a result of switching from prescription to over-the-counter use that would increase risk to patients and/or present unacceptable residual risks.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.