← Product Code [NKB](/productcode/NKB) · K230116

# Vital™ Spinal Fixation System (K230116)

_Zimvie Spine (Zimmer Biomet Spine, Inc.) · NKB · Oct 11, 2023 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K230116

## Device Facts

- **Applicant:** Zimvie Spine (Zimmer Biomet Spine, Inc.)
- **Product Code:** [NKB](/productcode/NKB.md)
- **Decision Date:** Oct 11, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Vital/Vitality Spinal Fixation System implants are non-cervical spinal fixation devices intended for posterior pedicle screw fixation (T1-S2/ilium), posterior hook fixation (T1-L5), or anterolateral fixation (T8-L5). Pedicle screw fixation is indicated for skeletally mature patients and for adolescent patients. These devices are indicated as an adjunct to fusion for all of the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e. scoliosis, and/or lordosis, Scheuermann's Disease), tumor, stenosis, pseudoarthrosis and/or failed previous fusion. When used as an adjunct to fusion, the Vital/ Vitality Spinal Fixation System is intended to be used with autograft and/or allograft. In addition the Vital/Vitality Spinal Fixation System is intended for treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft, having implants attached to the lumbosacral spine and or ilium with removal of the implant after attainment of a solid fusion. Levels of pedicle screw fixation for these patients are L3-sacrum/ilium. When used for posterior non-cervical pedicle screw fixation in pediatric patients, the Vital/Vitality System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The Vitality System is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach. The use of the Vital/Vitality Spinal Fixation System in skeletally mature patients may include the fixation of the Instinct® Java™ Spinal Fixation System* hooks, APEX Spinal System™* hooks, or fixation of the Universal Clamp® Spinal Fixation System* to the rods of the Vitality Spinal Fixation System. The Vital/Vitality Spinal Fixation System may also be used in skeletally immature patients when connected with the Universal Clamp Spinal Fixation System. In order to achieve additional levels of fixation in skeletally mature patients, the Vital/Vitality Spinal Fixation System may be connected to the Virage® OCT Spinal Fixation System* and the Instinct Java Spinal Fixation System offered by Zimmer Biomet Spine, using rod connectors.

## Device Story

Thoracolumbar and sacroiliac fixation system; provides temporary stabilization during spinal fusion; consists of rods, pedicle screws, hooks, and connectors; customized to patient anatomy; implanted by surgeons in clinical settings; used with autograft/allograft; implants removed after solid fusion; system updated to include MR Conditional labeling for 1.5T and 3.0T environments.

## Clinical Evidence

Bench testing only. MRI evaluation performed to quantify RF-induced heating, static magnetic field interactions, and image artifact generation in 1.5T and 3.0T environments per ASTM F2052, ASTM 2182, and ASTM 2213.

## Technological Characteristics

Materials: medical grade titanium, titanium alloy, and cobalt chromium alloy. Components: rods, pedicle screws, hooks, connectors. MR Conditional per ASTM F2052, ASTM 2182, and ASTM 2213. Single-use implants. Compatible with specified Zimmer Biomet spinal systems.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Vital™ Spinal Fixation System ([K203507](/device/K203507.md))

## Reference Devices

- Instinct® Java™ Spinal Fixation System
- APEX Spinal System™
- Universal Clamp® Spinal Fixation System
- Virage® OCT Spinal Fixation System

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

October 11, 2023

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Zim Vie Spine (Zimmer Biomet Spine, Inc.) Hanna Aucoin Regulatory Affairs Senior Specialist 10225 Westmoor Drive Westminster, Colorado 80021

Re: K230116

Trade/Device Name: Vital™ Spinal Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB Dated: September 7, 2023 Received: September 8, 2023

Dear Hanna Aucoin:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Eileen
Digitally signed
by Eileen Cadel -
S
Cadel -S Date: 2023.10.11
10:48:56 -04'00'
for

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K230116

Device Name VitalTM Spinal Fixation System

#### Indications for Use (Describe)

The Vital/Vitality Spinal Fixation System implants are non-cervical spinal fixation devices intended for posterior pedicle screw fixation (T1-S2/ilium), posterior hook fixation (T1-L5), or anterolateral fixation (T8-L5). Pedicle screw fixation is indicated for skeletally mature patients and for adolescent patients.

These devices are indicated as an adjunct to fusion for all of the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e. scoliosis, and/or lordosis, Scheuermann's Disease), tumor, stenosis, pseudoarthrosis and/or failed previous fusion. When used as an adjunct to fusion, the Vital/ Vitality Spinal Fixation System is intended to be used with autograft and/or allograft.

In addition the Vital/Vitality Spinal Fixation System is intended for treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft, having implants attached to the lumbosacral spine and or ilium with removal of the implant after attainment of a solid fusion. Levels of pedicle screw fixation for these patients are L3-sacrum/ilium.

When used for posterior non-cervical pedicle screw fixation in pediatric patients, the Vital/Vitality System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The Vitality System is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

The use of the Vital/Vitality Spinal Fixation System in skeletally mature patients may include the fixation of the Instinct® Java™ Spinal Fixation System* hooks, APEX Spinal System™* hooks, or fixation of the Universal Clamp® Spinal Fixation System* to the rods of the Vitality Spinal Fixation System. The Vital/Vitality Spinal Fixation System may also be used in skeletally immature patients when connected with the Universal Clamp Spinal Fixation System.

In order to achieve additional levels of fixation in skeletally mature patients, the Vital/Vitality Spinal Fixation System may be connected to the Virage® OCT Spinal Fixation System* and the Instinct Java Spinal Fixation System offered by Zimmer Biomet Spine, using rod connectors.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the ZimVie logo with the text "K230116 Page 1 of 3" in the upper right corner. The logo consists of a stylized geometric shape in light and dark blue, followed by the company name "ZimVie" in a dark blue, bold font. A thin gray line is located underneath the company name. The text in the upper right corner appears to be a document identifier and page number.

#### 10/11/2023

# 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR § 807.92.

| Sponsor Information        |                                                                                                                   |
|----------------------------|-------------------------------------------------------------------------------------------------------------------|
| Name                       | Zimmer Biomet Spine, Inc. (d/b/a ZimVie Spine)                                                                    |
| Address                    | 10225 Westmoor Dr.<br>Westminster, CO 80021                                                                       |
| Establishment Registration | 3012447612                                                                                                        |
| Contact Person(s)          | Hanna Aucoin<br>Regulatory Affairs Senior Specialist<br>Phone : (816) 562.2948<br>Email : Hanna.Aucoin@zimvie.com |
|                            | Alex Pawlowski<br>Regulatory Affairs Director<br>Phone : (303) 264.8255<br>Email : Alex.Pawlowski@zimvie.com      |

| Device Information                  |                                                                  |
|-------------------------------------|------------------------------------------------------------------|
| Proprietary Name                    | Vital™ Spinal Fixation System                                    |
| Common Name                         | Pedicle Screw Spinal Fixation System                             |
| Device Class                        | Class II                                                         |
| Device Panel                        | Orthopedic Panel (87)                                            |
| Regulation Number                   | 21 CFR § 888.3070                                                |
| Classification Name-Product Code(s) | Thoracolumbosacral Pedicle Screw System (NKB)                    |
| Predicate Devices                   | Primary Predicate:<br>Vital™ Spinal Fixation System<br>(K203507) |

## Device Description

The Vital™ Spinal Fixation System is a thoracolumbar and sacroiliac fixation system designed to aid in the surgical correction of several types of spinal conditions. The system consists of a variety of spinal rods, pedicle screws, hooks and connectors intended only to provide temporary stabilization during the development of a solid fusion of the spine with bone graft. The system can be rigidly locked into a variety of configurations, with each construct being customized to the patient's anatomy. All implants are single use only and should not be reused under any circumstances. The implant system is intended to be removed after solid fusion has occurred.

The system also includes instrumentation for insertion, securing and removal of the implants are made from medical grade titanium alloy; select rods are also available in medical grade cobalt chromium alloy. Implants made from medical grade titanium, medical grade titanium alloy, and medical grade cobalt chromium may be used together. Never use titanium alloy, and/or cobalt chromium

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for ZimVie, a medical device company. The logo consists of a stylized "Z" symbol in blue, followed by the text "ZimVie" in a bold, dark blue font. In the upper right corner of the image, there is the text "K230116 Page 2 of 3".

with stainless steel in the same implant construct. The Vital™ Spinal Fixation System is compati components from other cleared spinal fixation systems.

# Intended Use / Indications for Use

The Vital/Vitality Spinal Fixation System implants are non-cervical spinal fixation devices intended for posterior pedicle screw fixation (T1-S2/lium), posterior hook fixation (T1-L5), or anterolateral fixation (T8-L5). Pedicle screw fixation is indicated for skeletally mature patients and for adolescent patients.

These devices are indicated as an adjunct to fusion for all of the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis, Scheuermann's Disease), tumor, stenosis, pseudoarthrosis and/or failed previous fusion. When used as an adjunct to fusion, the Vital/Vitality Spinal Fixation System is intended to be used with autograft and/or allograft.

In addition the VitallVitality Spinal Fixation System is intended for treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft, having implants attached to the lumbosacral spine and or ilium with removal of the implant affer attainment of a solid fusion. Levels of pedicle screw fixation for these patients are L3-sacrum/ilium. When used for posterior non-cervical pedicle screw fixation in pediatric patients, the Vital/Vitality System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The Vitality System is intended to be used with autograft. Pediatric pedicle screw fixation is limited to a posterior approach.

The use of the Vital/Vitality Spinal Fixation System in skeletally mature patients may include the fixation of the Instinct® Java™ Spinal Fixation System* hooks, APEX Spinal System™* hooks, or fixation of the Universal Clamp® Spinal Fixation System* to the Vital/Vitality Spinal Fixation System. The Vital/Vitality Spinal Fixation System may also be used in skeletally immature patients when connected with the Universal Clamp Spinal Fixation System.

In order to achieve additional levels of fixation in skeletally mature patients, the Vital/Vitality Spinal Fixation System may be connected to the Virage® OCT Spinal Fixation System* and the Instinct Java Spinal Fixation System offered by Zimmer Biomet Spine, using rod connectors.

## Substantial Equivalence Assessment

The technological characteristics of the subject Vital™ Spinal Fixation System components remain the same as, or similar to, the predicate Vital™ Spinal Fixation System (K203507) in regard to intended use, indications for use, design, manufacturing methods, fundamental technology, and operational principles. The purpose of this submission is to seek clearance of the MRI conditional information provided in the Instructions for Use for the Vital™ Spinal Fixation System. Additionally, this submission discloses various minor changes to the previously cleared Vital™ Spinal Fixation System that were introduced via Lettersto-File.

Specifically, the subject Vital™ Spinal Fixation System is manufactured using the same materials and processes used to manufacture the previously cleared Vital System. Additionally, the subject devices have the same intended use as the predicate devices. As such, the performance assessments of the subject devices are substantially equivalent to the predicate devices. Furthermore, the subject Vital™ Spinal Fixation System has the same methods and technological characteristics as its cleared predicate Vital™ Spinal Fixation System (K203507).

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for ZimVie, a medical device company. The logo consists of a stylized letter Z in blue and a sans-serif wordmark in dark blue. Above the logo are the alphanumeric characters K230116 and the text "Page 3 of 3".

### Performance Testing

Results from the Magnetic Resonance Imaging evaluation demonstrate that the Vital™ Spinal Fixation System is MR Conditional in accordance with ASTM F2052, ASTM 2182, and ASTM 2213. Data and scientific rationales were used to quantify the RF-induced heating, static magnetic field interactions, and image artifact generation of the Vital™ Spinal Fixation System in the 1.5 Tesla (T) and 3.0 T MRI clinically relevant environments.

### Substantial Equivalence Conclusion

The subject Vital™ Spinal Fixation System's indications for use, intended use, design, materials, and performance assessments are substantially equivalent to the currently marketed Vital™ Spinal Fixation System. Furthermore, the addition of "MR Conditional" information on the labeling does not impact the safety and effectiveness of the system. As such, Zimmer Biomet Spine, Inc. concludes that the subject Vital™ Spinal Fixation System is substantially equivalent to the predicate Vital™ Spinal Fixation System.

---

**Source:** [https://fda-staging.innolitics.com/device/K230116](https://fda-staging.innolitics.com/device/K230116)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).