Denture Base Resin is used for the fabrication of removable dentures.
Device Story
Denture Base Resin is a PMMA-based material supplied as resin blocks (discs or horseshoe shapes) for use in dental laboratories. The device is processed via CAD/CAM milling to fabricate removable dentures. It is intended for use by dental professionals to create denture bases that are combined with prosthetic teeth. The device serves as a permanent restorative solution for patients requiring removable prosthetics. Clinical benefit is derived from the restoration of oral function and aesthetics through custom-milled, biocompatible denture bases. The device is used in a clinical or laboratory setting; the healthcare provider uses the milled output to fit the patient.
Clinical Evidence
No clinical data. Bench testing only. Performance verified against ISO 20795-1, ISO 4049, and ISO 7491. Evaluated for surface characteristics, color stability, porosity, flexural strength (87 MPa), flexural modulus (2340 MPa), residual monomer (1.29%), sorption (23.64 µg/mm3), solubility (0.25 µg/mm3), and fracture toughness.
Technological Characteristics
Materials: PMMA, methyl methacrylate, ethylene glycol dimethacrylate, titanium dioxide, ferric oxide, iron oxide yellow, ferric tetroxide. Form factor: Milled resin blocks (discs/horseshoes). Principle: CAD/CAM subtractive manufacturing. Standards: ISO 20795-1, ISO 4049, ISO 7491. Biocompatibility: ISO 10993-1 (permanent mucosal contact).
Indications for Use
Indicated for the fabrication of removable dentures in patients requiring prosthetic dental restoration.
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
Predicate Devices
IvoBase CAD for Zenotec, IvoBase CAD Bond, and Modelling Liquid (K151142)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 17, 2023
Aidite (Qinhuangdao) Technology Co., Ltd % Julie Chen Consultant ICAS Group 155 Pingbei Rd, Minghang Shanghai, Shanghai 201100 CHINA
Re: K230115
Trade/Device Name: Denture Base Resin Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, Or Rebasing Resin Regulatory Class: Class II Product Code: EBI Dated: January 13, 2023 Received: January 17, 2023
Dear Julie Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael E. Adjodha -S
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K230115
### Device Name
Denture Base Resin
Indications for Use (Describe)
Denture Base Resin is used for the fabrication of removable dentures.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
{3}------------------------------------------------
# 510(K) Summary K230115
#### I. SUBMITTER:
Aidite (Oinhuangdao) Technology Co., Ltd No.9 Dushan Road, Economic and Technological Development Zone,Qinhuangdao City China Contact Person: Chen Yingying Title: Registered Engineer Tel: 15033560085 Email: chenyingying@aidite.com
Submission Correspondent: Julie Chen Email: cl.julie@hotmail.com Tel:+86 139 1804 5781
Summary prepared: 03/16/2023
#### II. DEVICE
Name of Device: Denture Base Resin Trade Name: Denture Base Resin Common Name: Denture Relining, Repairing, or Rebasing Resin Regulation Number: 21 CFR PART 872.3690 Classification Name: Denture Relining, Repairing, or Rebasing Resin Regulatory Class: II Product Code: EBI
#### III. PREDICATE DEVICE
Primary predicate device: K151142(IvoBase CAD for Zenotec, IvoBase CAD Bond, and Modelling Liquid)
{4}------------------------------------------------
#### IV. DEVICE DESCRIPTION
The Denture Base Resin mainly consist of denture base resin (powder), polymethyl methacrylate (PMMA), methyl methacrylate, ethylene glycol dimethacrylate, titanium dioxide, ferric oxide, iron oxide yellow and ferric tetroxide. The resin block is milled using the CAD/CAM technique to form the removable dentures. The product combines the prosthetic denture teeth with the denture base.
#### V. AVAIABLE MODEL
| Model | Shade | Specification |
|--------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FDB-LC | A1P1、A2P1、A3P1、A3.5P1、A4P1、<br>B1P1、B2P1、C1P1、C2P1、OM1P1、<br>BL3P1、A1P2、A2P2、A3P2、<br>A3.5P2、A4P2、B1P2、B2P2、C1P2、<br>C2P2、OM1P2、BL3P2、A1P3、<br>A2P3、A3P3、A3.5P3、A4P3、B1P3、<br>B2P3、C1P3、C2P3、OM1P3、<br>BL3P3、A1P4、A2P4、A3P4、<br>A3.5P4、A4P4、B1P4、B2P4、C1P4、<br>C2P4、OM1P4、BL3P4、A1P5、<br>A2P5、A3P5、A3.5P5、A4P5、B1P5、<br>B2P5、C1P5、C2P5、OM1P5、<br>BL3P5、A1P6、A2P6、A3P6、<br>A3.5P6、A4P6、B1P6、B2P6、C1P6、<br>C2P6、OM1P6、BL3P6 | Disc shape: 95×20、95×25、<br>95×30、95×35、95×38、95×40、<br>98×20、98×25、98×30、98×35、<br>98×38、98×40 (Unit: mm)<br>Horseshoe: 101×20、101×25、<br>101×30、101×35、101×38、<br>101×40 (Unit: mm) |
| | FDB-LM | B3P2、B3P4、B3P5、B4P2、B4P4、<br>B4P5、C3P2、C3P4、C3P5、C4P2、<br>C4P4、C4P5、D2P2、D2P4、D2P5、<br>D3P2、D3P4、D3P5、D4P2、D4P4、<br>D4P5、OM2P2、OM2P4、OM2P5、<br>OM3P2、OM3P4、OM3P5、BL1P2、<br>BL1P4、BL1P5、BL2P2、BL2P4、<br>BL2P5、BL4P2、BL4P4、BL4P5 |
| | | A1P1、A2P1、A3P1、A3.5P1、A4P1、<br>B1P1、B2P1、C1P1、C2P1、OM1P1、<br>BL3P1、A1P2、A2P2、A3P2、<br>A3.5P2、A4P2、B1P2、B2P2、C1P2、<br>C2P2、OM1P2、BL3P2、A1P3、<br>A2P3、A3P3、A3.5P3、A4P3、B1P3、 |
| | B2P3、C1P3、C2P3、OM1P3、<br>BL3P3、A1P4、A2P4、A3P4、<br>A3.5P4、A4P4、B1P4、B2P4、C1P4、<br>C2P4、OM1P4、BL3P4、A1P5、<br>A2P5、A3P5、A3.5P5、A4P5、B1P5、<br>B2P5、C1P5、C2P5、OM1P5、<br>BL3P5、A1P6、A2P6、A3P6、<br>A3.5P6、A4P6、B1P6、B2P6、C1P6、<br>C2P6、OM1P6、BL3P6 | 101×40 (Unit: mm) |
| | B3P2、B3P4、B3P5、B4P2、B4P4、<br>B4P5、C3P2、C3P4、C3P5、C4P2、<br>C4P4、C4P5、D2P2、D2P4、D2P5、<br>D3P2、D3P4、D3P5、D4P2、D4P4、<br>D4P5、OM2P2、OM2P4、OM2P5、<br>OM3P2、OM3P4、OM3P5、BL1P2、<br>BL1P4、BL1P5、BL2P2、BL2P4、<br>BL2P5、BL4P2、BL4P4、BL4P5 | Disc shape: 98×30、98×35、<br>98×38 (Unit: mm) |
| | A1P1、A2P1、A3P1、A3.5P1、A4P1、<br>B1P1、B2P1、C1P1、C2P1、OM1P1、<br>BL3P1、A1P2、A2P2、A3P2、<br>A3.5P2、A4P2、B1P2、B2P2、C1P2、<br>C2P2、OM1P2、BL3P2、A1P3、<br>A2P3、A3P3、A3.5P3、A4P3、B1P3、<br>B2P3、C1P3、C2P3、OM1P3、<br>BL3P3、A1P4、A2P4、A3P4、<br>A3.5P4、A4P4、B1P4、B2P4、C1P4、<br>C2P4、OM1P4、BL3P4、A1P5、<br>A2P5、A3P5、A3.5P5、A4P5、B1P5、<br>B2P5、C1P5、C2P5、OM1P5、<br>BL3P5、A1P6、A2P6、A3P6、<br>A3.5P6、A4P6、B1P6、B2P6、C1P6、<br>C2P6、OM1P6、BL3P6 | Notched Disc: 98×20、98×25、<br>98×30、98×35、98×38、98×40<br>(Unit: mm) |
| FDB-CC | B2P5、C1P5、C2P5、OM1P5、<br>BL3P5、A1P6、A2P6、A3P6、<br>A3.5P6、A4P6、B1P6、B2P6、C1P6、<br>C2P6、OM1P6、BL3P6 | |
| | B3P2、B3P4、B3P5、B4P2、B4P4、<br>B4P5、C3P2、C3P4、C3P5、C4P2、<br>C4P4、C4P5、D2P2、D2P4、D2P5、<br>D3P2、D3P4、D3P5、D4P2、D4P4、<br>D4P5、OM2P2、OM2P4、OM2P5、<br>OM3P2、OM3P4、OM3P5、BL1P2、<br>BL1P4、BL1P5、BL2P2、BL2P4、<br>BL2P5、BL4P2、BL4P4、BL4P5 | |
| | Notched Disc: 98×30、98×35、<br>98×38 (Unit: mm) | |
| FDB-CM | A1P1、A2P1、A3P1、A3.5P1、A4P1 | Notched Disc: 98×20、98×25、 |
| | | |
| | B1P1、B2P1、C1P1、C2P1、OM1P1、<br>98×30、98×35、98×38、98×40<br>BL3P1、 A1P2、 A2P2 、 A3P2 、<br>(Unit: mm) | |
| | A3.5P2、A4P2、B1P2、B2P2、C1P2、 | |
| | C2P2、OM1P2、BL3P2、A1P3、 | |
| | A2P3、A3P3、A3.5P3、A4P3、B1P3、 | |
| | B2P3 、 C1P3 、 C2P3 、 OM1P3 、 | |
| | BL3P3、 A1P4、 A2P4、 A3P4、 | |
| | A3.5P4、A4P4、B1P4、B2P4、C1P4、 | |
| | C2P4、OM1P4、BL3P4、A1P5、 | |
| | A2P5、A3P5、A3.5P5、A4P5、B1P5、 | |
| | B2P5、C1P5、C2P5、OM1P5、 | |
| | BL3P5、 A1P6、 A2P6、 A3P6、 | |
| | A3.5P6、A4P6、B1P6、B2P6、C1P6、 | |
| | C2P6、OM1P6、BL3P6 | |
| | B3P2、B3P4、B3P5、B4P2、B4P4、 | |
| | B4P5、C3P2、C3P4、C3P5、C4P2、 | |
| | C4P4、C4P5、D2P2、D2P4、D2P5、 | |
| | D3P2、D3P4、D3P5、D4P2、D4P4、Notched Disc: 98×30、98×35、 | |
| | D4P5、OM2P2、OM2P4、OM2P5、98×38 (Unit: mm) | |
| | OM3P2、OM3P4、OM3P5、BL1P2、 | |
| | BL1P4、BL1P5、BL2P2、BL2P4、 | |
| | BL2P5、BL4P2、BL4P4、BL4P5 | |
| | Disc shape: 98×20、98×22、 | |
| | 98×25、98×28、98×30、98×35、 | |
| | 98×38、98×40、95×20、95×22、 | |
| | 95×25、95×28、95×30、95×35、 | |
| FDB-C | P1、P2、P3、P4、P5、P6<br>95×38、95×40 (Unit: mm) | |
| | Horseshoe: 101×20、101×22、 | |
| | 101×25、101×28、101×30、 | |
| | 101×35 、 101×38 、 101×40<br>(Unit: mm) | |
{5}------------------------------------------------
{6}------------------------------------------------
#### VI. INTENED USE per 21CFR 807.92(A)(5)
Denture Base Resin is used for the fabrication of removable dentures.
{7}------------------------------------------------
#### INDICATIONS for USE per Form FDA 3881 VII.
Denture Base Resin is used for the fabrication of removable dentures.
# VIII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Denture Base Resin are compared with the predicate device, IvoBase CAD for Zenotec and IvoBase CAD Bond and Modelling Liquid (K151142). The results are shown below in the Technological Characteristics Comparison Table:
| Item | Subject Device | Predicate Device | Remark |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| | Denture Base<br>Resin | IvoBase CAD for<br>Zenotec and<br>IvoBase CAD Bond<br>and Modelling<br>Liquid | |
| 510(k) number | K230115 | K151142 | -- |
| Regulation Number | 21 CFR PART 872.3690 | 21 CFR PART 872.3690 | SE |
| Product Code | EBI | EBI | SE |
| Common name | Denture Relining, Repairing,<br>or Rebasing Resin | Denture Relining, Repairing, or<br>Rebasing Resin | SE |
| Classification | II | II | SE |
| Manufacturer | Aidite<br>(Qinhuangdao)<br>Technology Co., Ltd | Ivoclar Vivadent, Incorporated | -- |
| Intended Use | Denture Base Resin is used for<br>the fabrication of removable<br>dentures. | IvoBase CAD for Zenotec, IvoBase<br>CAD Bond and Modelling Liquid<br>is a system used:<br><br>For the fabrication of removable<br>dentures,e.g.:<br>partial and complete<br>denture prosthetics hybrid denture prosthetics combined denture<br>prosthetics mouthguards implant-supported denture<br>prosthetics | SE |
| Type Use | Prescription (Rx Only) | Prescription (Rx Only) | SE |
| CAD CAM Technology | Denture Base Resin is<br>intended for use in the<br>fabrication of denture<br>prosthetics using CAD CAM<br>Technology. | IvoBase CAD for Zenotec is a<br>PMMA disc intended for use in<br>the fabrication of denture<br>prosthetics using CAD CAM<br>Technology. | SE |
| Monomer/polymer form | The resin block is formed<br>using CAD CAM technology.<br>The product combines the<br>prosthetic denture teeth with<br>the denture base. | The denture base is formed using<br>CAD CAM technology. The<br>prosthetic denture teeth are then<br>bonded to the denture base using<br>IvoBase CAD Bond and IvoBase<br>CAD Modelling Liquid. IvoBase<br>CAD Bond consists of a powder<br>polymer and monomer liquid<br>similar to a traditional denture<br>base device. | Similar |
| Principle of operation | Step-by-step:<br>- Design and<br>processing of<br>CAD/CAM<br>removable dentures<br>- Separate<br>After CAD/CAM<br>cutting, the<br>restoration is<br>separated from the<br>resin block with a<br>tungsten carbide or<br>diamond bur.<br>- Polishing<br>- Cleaning and<br>disinfection<br>- Insertion | Step-by-step:<br>- Anatomical impression<br>and pre- bite registration<br>- Lab makes scan of<br>impression and mills<br>individual tray<br>- Functional impression<br>and bite registration<br>- Lab makes scan and<br>mills try-in body (Tray<br>Disc for Zenotec)<br>- Try-in and esthetic,<br>functional check<br>- Lab makes final<br>denture. IvoBase CAD<br>for Zenotec can be<br>processed using the<br>Zenotec select milling<br>machine. First the lingual<br>side is milled, then the<br>teeth are bonded in<br>position using ivoBase<br>CAD Bond and then the<br>basal side is milled.<br>- After milling, the<br>denture base is separated<br>from the disc, shape | SE |
| | | adjustments are made and<br>then it is polished.<br>• Insertion | |
| Composition | The Denture Base Resin mainly<br>consist of denture base resin<br>(powder), polymethyl<br>methacrylate (PMMA), methyl<br>methacrylate, ethylene glycol<br>dimethacrylate, titanium<br>dioxide, ferric oxide, iron oxide<br>yellow and ferric tetroxide. | The chemical composition of the<br>new product and predicate are the<br>same, except for small changes in<br>pigments and the fact that the disc<br>product is industrially<br>polymerized. Therefore certain<br>ingredients (the methyl<br>methacrylate and initiators) are<br>no longer present. The result of<br>the biocompatibility assessment<br>is that the product is equivalent to<br>the predicate. | Similar |
| Biocompatibility | Biocompatibility in accordance<br>to 10993-1(Surface device in<br>contact with mucosal membrane<br>with permanent contact (>30 d)) | Biocompatibility in accordance<br>to 10993-1(Surface device in contact<br>with mucosal membrane with<br>permanent contact (>30 d)) | Same |
| Performance | Ultimate flexural strength 87<br>Mpa;<br>Flexural modulus 2340Mpa;<br>Residual methyl methacrylate<br>monomer 1.29%<br>Sorption 23.64 µg/mm3<br>Solubility 0.25 µg/mm3<br>Maximum stress intensity<br>factor for materials with<br>improved impact resistance<br>2.89MPa m1/2<br>Total fracture work 1609 J/m2 | Ultimate flexural strength82 Mpa;<br>Flexural modulus 2032Mpa;<br>Residual methyl methacrylate<br>monomer1.34%<br>Sorption 19.48 µg/mm3<br>Solubility 0.65 µg/mm3<br>Maximum stress intensity factor<br>for materials with improved<br>impact resistance<br>2.35 MPa m1/2<br>Total fracture work 1411 J/m2 | Similar |
{8}------------------------------------------------
{9}------------------------------------------------
#### IX. SUBSTANTIAL EQUIVALENCE DISCUSSION per 21 CFR 807.92(b)
Denture Base Resin has been conducted biocompatibility studies in accordance with ISO 10993 to demonstrate the device is as safe as its predicate device. The performance bench testing was conducted to demonstrate that the subject device is as effective as its predicate device.
#### X. PERFORMANCE DATA
# Non-Clinical Performance Test Conclusion
# Biocompatibility
Based on Table A.1 of ISO 10993-1 and Table A.1 of FDA Guidance "Use of
{10}------------------------------------------------
International Standard ISO 10993-1, Biological evaluation of medical devices – Part 1 Evaluation and testing within a risk management process", the subject device is categorized as a surface device in contact with Surface device in contact with mucosal membrane with permanent contact (>30 d). The subject device was evaluated for: Cytotoxicity Sensitization
- Irritation Acute systemic toxicity Subchronic systemic toxicity Implantation Material-mediated pyrogenicity Genotoxicitv
# Performance Bench Testing
Performance testing were conducted to verify that the subject device met all design specifications was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the Dental Base Resin complies with the following standards:
- . ISO20795-1: 2013 Dentistry—Base polymers —Part 1:Denture base polymers
- ISO 4049: 2019 Dentistry Polymer-based restorative materials; ●
- ISO 7491: 2000 Dental materials-Determination of color stability. ●
The subject device and predicate device were evaluated for:
1)Surface characteristics:
2)Shape capability;
3)Colour;
4)Colour stability;
5)Translucency:
- 6)Freedom from porosity;
7)Ultimate flexural strength;
8)Flexural modulus;
9)Bonding to synthetic polymer teeth
10)Residual methyl methacrylate monomer
11)Sorption and solubility
12) Maximum stress intensity factor for materials with improved impact resistance and total fracture work
{11}------------------------------------------------
# Clinical Test Conclusion
No clinical study is included in this submission.
#### XI. CONCLUSION
The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device IvoBase CAD for Zenotec and IvoBase CAD Bond and Modelling Liquid (K151142).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.