Aimanfun Lumea Comfort, Model: A-2788, A-2789 and A-3588
K230107 · Kam Yuen Plastic Products , Ltd. · OHT · Apr 5, 2023 · General, Plastic Surgery
Device Facts
Record ID
K230107
Device Name
Aimanfun Lumea Comfort, Model: A-2788, A-2789 and A-3588
Applicant
Kam Yuen Plastic Products , Ltd.
Product Code
OHT · General, Plastic Surgery
Decision Date
Apr 5, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Aimanfun Lumea Comfort is an over-the-counter device intended for removal of unwanted body and/or facial hair.
Device Story
Hand-held IPL device for OTC hair removal; targets underarms, bikini line, arms, and legs. Uses Xenon lamp to emit pulsed light (475-1200nm) to heat hair roots; 5 energy levels. Includes skin proximity sensor to prevent pulse emission unless in full contact with skin. Operated by user at home. Output is light pulse; effect is hair growth reduction. Leverages existing predicate technology; no clinical data provided.
Clinical Evidence
No clinical data provided. Safety and efficacy established via bench testing, electrical safety/EMC compliance, and a usability study confirming lay user comprehension of labeling and device operation.
Technological Characteristics
IPL technology using Xenon Arc Flashlamp; 475-1200nm wavelength; 3.0 cm2 spot size; 4.5 J/cm2 max fluence; 3ms pulse duration. 5 energy levels (7-13.5 J). External power supply. Hand-held form factor. Complies with IEC 60601-1, IEC 60601-1-11, IEC 60601-2-57, IEC 60601-1-2, ISO 10993-5, and ISO 10993-10.
Indications for Use
Indicated for removal of unwanted body and/or facial hair in an over-the-counter setting.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
April 5, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Kam Yuen Plastic Products Ltd. % Iris Fung Regulation manager Guangzhou KEDA Biological Tech Co., Ltd. 6F, No.1 TianTai road, Science City, LuoGang District Guangzhou, Guangdong China
Re: K230107
Trade/Device Name: Aimanfun Lumea Comfort, Model: A-2788, A-2789 and A-3588 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: January 13, 2023 Received: January 13, 2023
Dear Iris Fung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Colin K. C
for
Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230107
Device Name
Aimanfun Lumea Comfort, Model: A-2788, A-2789 and A-3588
Indications for Use (Describe)
Aimanfun Lumea Comfort is an over-the-counter device intended for removal of unwanted body and/or facial hair.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
#### K230107
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
#### 1 Submitter Information
Company Name: Kam Yuen Plastic Products Ltd.
Address: No. 2, Heng Feng Two Road, Pu Jin Industrial, Konghou Town, ZhongShan,
Guangdong, China
Phone: 86- 400- 962- 1668
Fax: 86- 0760- 8841- 3080
Contact Person (including title): Anna Dan
E-mail: kamyuen@kyplastic.com
Application Correspondent: Jet Li Company: Guangzhou KEDA Biological Technology Co., Ltd Tel: +86-13512755282 Email: jianda-lee@foxmail.com Address: 6F, No.1 TianTai road, Science City, LuoGang District, GuangZhou City, China
### 2 Subject Device Information
Trade Name: Aimanfun Lumea Comfort Model: A-2788, A-2789 and A-3588 Common Name: Light Based Over-The-Counter Hair Removal Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Review Panel: General & Plastic Surgery Product Code: OHT Requlation Number: 878.4810
{4}------------------------------------------------
Requlation Class: 2
#### 3 Predicate device Information
#### Legally marketed device K190820
Trade Name: Aimanfun Lumea Comfort
Model: A-2788
Common Name: Light based hair removal devices
Classification Name: Powered Light Based Non-Laser Surgical Instrument With Thermal
Effect
Review Panel: General & Plastic Surgery
Product Code: ONF
Regulation Number: 878.4810
Regulation Class: 2
#### Leqally marketed device K212907
Trade Name: Aimanfun Lumea Comfort Model: A-2789, A-3588 Common Name: Light based hair removal devices Classification Name: Powered Light Based Non-Laser Surgical Instrument With Thermal Effect Review Panel: General & Plastic Surgery Product Code: ONF Requlation Number: 878.4810
Regulation Class: 2
#### Reference device K211185
Common Name: IPL Home Use Hair Removal Device
Model: D-1150, D-1171, D-1153, D-1155, D-1156, D-1126, D-1178, D-1187
Common Name: Light Based Over-The-Counter Hair Removal
{5}------------------------------------------------
Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Review Panel: General & Plastic Surgery Product Code: OHT Regulation Number: 878.4810 Regulation Class: 2
#### 4 Device Description
Aimanfun Lumea Comfort, Model: A-2788, A-2789 and A-3588 is an over-the-counter use device for the reduction of hair growth. Ideal body areas include the underarms, bikini line, arms and legs. The device used the Intense Pulsed Light (IPL) technology with lower energy level, including 5 Levels of output energy. Intense Pulsed light technology is able to achieve hair removal results at a fraction of the energy level used in other light-based hair removal equipment.
The hand-held device package includes main unit, adaptor and user manual and it uses a Xenon Lamp to emit specified wavelength pulsed light to heat the root where the hair grows, and a skin proximity sensor to detect appropriate skin contact. If the device is not properly applied to the treatment area (in full contact with the skin), the device cannot be triggered a pulse emitting.
#### 5 Intended Use
Aimanfun Lumea Comfort is an over-the-counter device intended for removal of unwanted body and/or facial hair.
#### 6 Complied Standards
Aimanfun Lumea Comfort, Model: A-2788, A-2789 and A-3588 complies with the following FDA recognized consensus standards:
- . IEC 60601-1: 2005+A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
{6}------------------------------------------------
- . IEC 60601-1-11:2015 Medical Electrical Equipment –Part 1: General Requirements for Basic Safety and Essential Performance -Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
- . IEC 60601-2-57:2011 Medical electrical equipment –Part 2-57: Particular requirements for the basic safety and essential performance of non-laser source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
- . IEC 60601-1-2: 2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests
- ISO 10993-5:2009/(R) 2014, Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity.
- ISO 10993-10:2010, Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization.
#### 7 Device modification description
The only modification of subject device is Intended Use. The device has changed to overthe-counter use. Therefore, the User Manual and Box labelling has been adding corresponding use instructions for lay person user.
#### 8 Performance Testing
As the modification of subject device as above, results in no technological characteristics changes, the tests and data utilized to demonstrate safety and efficacy of the predicate device (legally existing device) are suitable for use in the assessment of the subject devices except for usability study verification. In usability study, the testing result demonstrates that the intended users can understand the package labeling, correctly choose the device and use it for the indicated OTC use, based on reading the lableling materials.
As there have been no changes to the performance of the subject device from the predicate
{7}------------------------------------------------
device, this submission leverages performance and electrical testing provided in previous submission.
#### 9 Biocompatibility
All the modified device materials which come in direct contact with the patient skin are biocompatible and identical to the materials used in the predicate device manufacturing. No biocompatibility test report is provided in this submission.
#### 10 Clinical performance
Clinical performance is not deemed necessary.
{8}------------------------------------------------
#### 11 Comparison with predicate device
Compare with predicate device (Aimantun Lumea Comfort, K190820 and K212907), the subject device is same in design principle, functions, material and
the applicable standards. The differences between subject devices do not raise any new questions of safety or effectiveness.
| Elements of<br>Comparison | Subject Device | Predicate Device 1 | Predicate Device 2 |
|-----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Kam Yuen Plastic Products Ltd. | Kam Yuen Plastic Products Ltd. | Kam Yuen Plastic Products Ltd. |
| Product Name | Aimanfun Lumea Comfort | Aimanfun Lumea Comfort | Aimanfun Lumea Comfort |
| Model | A-2788, A-2789 and A-3588 | A-2789<br>A-3588 | A-2788 |
| 510(K) No. | K230107 | K212907 | K190820 |
| Indications for<br>Use | Aimanfun Lumea Comfort is an over-<br>the-counter device intended for<br>removal of unwanted body and/or<br>facial hair. | The Aimanfun Lumea Comfort (Model: A-<br>2789, A-3588) is indicated for patient<br>removal of unwanted hair by using a<br>selective photothermal treatment under the<br>direction of a physician, after training by a<br>healthcare professional. The Aimanfun<br>Lumea Comfort is also intended for<br>permanent reduction in unwanted hair. | The Aimanfun Lumea Comfort (Model: A-<br>2788) is indicated for patient removal of<br>unwanted hair by using a selective<br>photothermal treatment under the direction<br>of a physician, after training by a<br>healthcare professional. The Aimanfun<br>Lumea Comfort is also intended for<br>permanent reduction in unwanted hair. |
| Elements of<br>Comparison | Subject Device | Predicate Device 1 | Predicate Device 2 |
| | | Permanent hair reduction is defined as the<br>long-term stable reduction in the number of<br>hairs regrowing when measured at 6. 9.<br>and 12 months after the completion of a<br>treatment regimen. | Permanent hair reduction is defined as the<br>long-term stable reduction in the number of<br>hairs regrowing when measured at 6. 9.<br>and 12 months after the completion of a<br>treatment regimen. |
| Type of Use | Over-The-Counter Use | Prescription use | Prescription use |
| Classification | OHT | ONF | ONF |
| Product Code | | | |
| Technology | Intense Pulsed Light | Intense Pulsed Light | Intense Pulsed Light |
| Device Type | Hand-held | Hand-held | Hand-held |
| Status display | A-2788: LED indicators | | |
| | A-2789: LCD screen | LCD screen | LED indicators |
| | A-3588: LCD screen | | |
| Work mode | A-2788: Single flash mode | Single flash mode | Single flash mode |
| | A-2789: Single flash mode, Sliding | Sliding flash mode | |
| Elements of<br>Comparison | Subject Device | Predicate Device 1 | Predicate Device 2 |
| | flash mode<br>A-3588: Single flash mode, Sliding<br>flash mode | | |
| Ice-sensing<br>function | A-2788: No<br>A-2789: No<br>A-3588: Yes | No | No |
| Power source | External power supply | External power supply | External power supply |
| Light source | Xenon Arc Flashlamp | Xenon Arc Flashlamp | Xenon Arc Flashlamp |
| Wavelength | 475~1200nm | 475~1200nm | 475~1200nm |
| Spot Size | $3.0 cm^2$ | $3.0 cm^2$ | $3.0 cm^2$ |
| Max. Fluence | $4.5 J/cm^2$ | $4.5 J/cm^2$ | $4.5 J/cm^2$ |
| Pulse duration | 3 milliseconds | 3 milliseconds | 3 milliseconds |
| Output energy | 7-13.5 J | 7-13.5 J | 7-13.5 J |
| Pulsing Control | Finger switch | Finger switch | Finger switch |
| Delivery Device | Direct Illumination to Tissue | Direct Illumination to Tissue | Direct Illumination to Tissue |
| | | Yes | |
| Elements of<br>Comparison | Subject Device | Predicate Device 1 | Predicate Device 2 |
| Software Control | Yes | Yes | Yes |
| Electrical safety,<br>EMC, Biological<br>Evaluation | IEC 60601-1 | IEC 60601-1 | IEC 60601-1 |
| | IEC 60601-1-2 | IEC 60601-1-2 | IEC 60601-1-2 |
| | IEC 60601-1-11 | IEC 60601-1-11 | IEC 60601-1-11 |
| | IEC 60601-2-57 | IEC 60601-2-57 | IEC 60601-2-57 |
| | ISO 10993-5 | ISO 10993-5 | ISO 10993-5 |
| | ISO 10993-10 | ISO 10993-10 | ISO 10993-10 |
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
## 12 Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the comparison of intended use, design, materials and performance, the subject device Aimanfun Lumea Comfort is to be concluded substantial equivalent to its predicate device.
#### 13 Summary Prepared Date
Apr 04, 2023
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.