Smart double sides wearable breast pump, Electric bilateral breast pump
Applicant
Huizhou Lvb Maternal and Infant Supplies Co., Ltd.
Product Code
HGX · Obstetrics/Gynecology
Decision Date
Sep 21, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5160
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Wearable Breast Pump is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The device is intended for a single user.
Device Story
Wearable Breast Pump (Model ABP-1508) is an electrically powered, single-user device for lactating women to express and collect breast milk. Device uses a flange to create a seal around the nipple; applies and releases vacuum pressure to the breast. Features stimulation and expression modes, each with nine adjustable suction levels. Vacuum pressure ranges: 80-200 mmHg (stimulation) and 90-240 mmHg (expression). Powered by rechargeable Li-Ion battery or power adapter. User interface includes LED display showing mode, suction level, timer, and battery status; soft-touch buttons control power, mode, and suction. Backflow protection membrane prevents milk from entering the vacuum system. Device is intended for home use by the patient. Output is collected milk in a container. Benefits include efficient milk expression and collection.
Clinical Evidence
Bench testing only. Testing included electrical safety (IEC 60601-1, IEC 60601-1-11), electromagnetic compatibility (IEC 60601-1-2), and biocompatibility (ISO 10993-5, ISO 10993-23, ISO 10993-10). Performance testing verified vacuum levels, backflow protection, use life, and battery functionality.
Technological Characteristics
Powered breast pump; materials include silicone (valve) and polypropylene (breast shield, bottle, suction cap). Microcontroller-based cycling control. Energy source: Li-Ion battery or power adapter. Connectivity: None. Sterilization: Not specified. Software: Microcontroller-based control of vacuum and cycle timing.
Indications for Use
Indicated for lactating women to express and collect milk from their breasts. Intended for single-user use.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" on the bottom.
September 21, 2023
Huizhou LVB Maternal And Infant Supplies Co.Ltd. Rain Yip Registration Engineer Block A 16 Huifengxi 3rd, Pingnan Industry Zone, Zhongkai H-i-tech District Huizhou, Guangdong 516000 China
Re: K230102
> Trade/Device Name: Wearable Breast Pump Regulation Number: 21 CFR§ 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: August 17, 2023 Received: August 21, 2023
Dear Rain Yip:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Monica D. Garcia -S
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230102
Device Name Wearable Breast Pump
Indications for Use (Describe)
Wearable Breast Pump is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The device is intended for a single user.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) summary - K230102
Date Prepared: September 20, 2023
#### l Submitter
Applicant: HuiZhou LVB Maternal and Infant Supplies Co., Ltd. Address: NO. 114,Xin Xing North Road, Sandong Town, Huicheng District, Huizhou City, Guangdong Province, P.R. China
Contact person: Dongan Zhang Position: Head of Firm Phone: +86-13075210606 Fax:+86-0752-5311318 E-mail: zhda7311@163.com
#### II Proposed Device
| Device Name: | Wearable Breast Pump |
|--------------------|-----------------------------|
| | (Model: ABP-1508) |
| Common Name: | Powered breast pump |
| Regulation Number: | 21 CFR 884.5160 |
| Regulation Name: | Powered Breast Pump |
| Regulatory Class: | Class II |
| Product code: | HGX (Pump, Breast, Powered) |
#### III Predicate Devices
| 510(k) Number: | K212180 |
|----------------|-----------------------------------|
| Trade name: | Wearable Breast Pump (Model S12) |
| Manufacturer : | Shenzhen TPH Technology Co., Ltd. |
The predicate device has not been subject to a design-related recall.
#### IV Device description
The Wearable Breast Pump is an electrically powered, software controlled, digital, single user, wearable breast pump for lactating women to express and collect milk from the breast. The subject device is capable of expression and stimulation modes, with nine associated suction levels for each. The vacuum pressure ranges from 80-200 mmHg for stimulation mode and 90-240 mmHg for expression mode. The device is powered by a rechargeable battery or a power adapter.
The device has an LED display, which shows the power button, pumping mode, suction level, timer, and battery charge level. Surrounding the front panel display are soft-touch
{4}------------------------------------------------
buttons allowing the user to power the device on/off, switch between stimulation and expression modes, and control the vacuum strength.
The breast pump expresses by creating a seal around the nipple using the flange and applying and releasing suction to the breast. The milk is collected in a milk collection container. To prevent milk from flowing into the vacuum system, a backflow protection membrane physically separates the milk-contacting pathway from the vacuum system.
The subject device components are made of the following materials: ·Valve: Silicone ·Breast shield, bottle, suction cap: Polypropylene
#### V Indication for use
Wearable Breast Pump is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The device is intended for a single user.
#### VI Comparison of technological characteristics with the predicate device
| Item | Subject Device | Predicate<br>Device<br>(K212180) | Comparison |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Product name | Wearable Breast<br>Pump (ABP-1508) | Wearable Breast<br>Pump (Model S12) | N/A |
| Product code | HGX | HGX | same |
| Regulation No. | 21 CFR 884. 5160 | 21 CFR 884. 5160 | same |
| Class | Class II | Class II | same |
| Indication for use | Wearable Breast<br>Pump is a powered<br>breast pump to be<br>used by lactating<br>women to express<br>and collect milk from<br>their breasts. The<br>device is intended<br>for a single user. | The powered<br>Electric Double<br>Breast Pumps are<br>intended to express<br>and collect milk from<br>the breast of a<br>lactating woman.<br>They are double<br>pumps with a single<br>pumping option and<br>intended for single<br>user. | same |
| Pumping options | Single | Single | same |
| Item | Subject Device | Predicate<br>Device<br>(K212180) | Comparison |
| Cycling control<br>Mechanism | Microcontroller | Microcontroller | same |
| Suction modes | Stimulation Mode and<br>Expression Mode | Stimulation Mode and<br>Expression Mode | same |
| Suction levels | 9 | 9 | same |
| Adjustable suction<br>levels | Yes | Yes | same |
| Flange size | 21, 24, mm and 27 mm | 24 mm and 27 mm | Different |
| Vacuum<br>range:<br>Stimulation | -80 to -200 (±5) mmHg | -40 to -105 (±5)<br>mmHg | Different: The<br>difference in<br>vacuum range<br>do not raise<br>different<br>questions of<br>safety and<br>effectiveness. |
| Vacuum<br>range:<br>Expression | -90 to -240 (±5) mmHg | -40 to -245 (±5)<br>mmHg | Different: The<br>difference in<br>vacuum range<br>do not raise<br>different<br>questions of<br>safety and<br>effectiveness. |
| Cycle<br>speed:<br>stimulation | 64 to 108 cycles/minute | 70 to 114<br>cycles/minute | Different: The<br>difference in<br>cycle speed<br>do not raise<br>different<br>questions of<br>safety and<br>effectiveness. |
| Cycle<br>speed:<br>Expression | 26 to 48 cycles/minute | 23 to 90 cycles/minute | Different: The<br>difference in<br>cycle speed<br>do not raise<br>different<br>questions of<br>safety and<br>effectiveness. |
| Item | Subject Device | Predicate<br>Device<br>(K212180) | Comparison |
| | | | safety and<br>effectiveness. |
| Power supply | Li-Ion Battery | Li-Ion Battery | same |
| Indicators | Yes, LED | Yes, LED | same |
| Back flow protection | Yes | Yes | same |
| Materials | Valve: Silicone<br>Breast shield, bottle,<br>suction cap:<br>Polypropylene | Milk Container:<br>Polypropylene<br>Flange: Silicone<br>Pump Outer Housing:<br>Acrylonitrile Butadiene<br>Styrene (ABS) plastic | Similar |
{5}------------------------------------------------
{6}------------------------------------------------
The indications for use of the subject and predicate devices are identical and they have the same intended use (i.e., the collection of breast milk from the breasts of lactating women).
The subject and predicate devices have different technological features, including different device design, user interface, vacuum pressure range, cycle speeds, materials, flange sizes, and power source. The different technological characteristics of the subject device do not raise different questions of safety and effectiveness.
### VII Non-Clinical Testing
Non-clinical tests were conducted to verify that the subject device met all design specifications to be considered substantially equivalent to the predicate device.
Electrical Safety
- IEC 60601-1:2005+AMD 1:2012+AMD2:2020, Medical electrical equipment Part 1: . General requirements for basic safety and essential performance
- . IEC 60601-1-11:2015+AMD1:2020, Medical Electrical Equipment - Part 1-11: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in The Home Healthcare Environment
#### Electromagnetic Compatibility
- IEC 60601-1-2:2014+AMD1:2020, Medical electrical equipment Part 1-2: General .
{7}------------------------------------------------
requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
#### Biocompatibility Evaluation
Per the 2022 FDA guidance document "Use of international Standard ISO 10993-1, Biological evaluation of medical - Part 1: Evaluation and testing within a risk management process," the following tests performed on the material which contacts with human for Biocompatibility:
- Cytotoxicity (ISO 10993-5:2009)
- . Skin irritation (ISO 10993-23:2021)
- . Skin Sensitization (ISO 10993-10:2021)
#### Software
Software was evaluated as recommended in the 2005 FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
#### Performance Testing
Other performance testing was conducted to show that the device meets its design requirements and performs as intended. The performance tests include:
- . Vacuum level verification testing at each mode/cycle demonstrated that the devices meet mode/cycle specifications.
- . Backflow protection testing was conducted to verify liquid does not backflow into the tubing.
- . Use life testing was conducted to demonstrate that the device maintains its specifications throughout its proposed use life.
- . Battery performance testing was conducted to demonstrate that the battery remains functional during its stated battery use-life.
- . Battery status indicator testing was conducted to demonstrate that the battery status indicator remains functional during its stated battery life.
#### IX Conclusion
The results of the performance testing described above demonstrate that the Wearable Breast Pump is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.