← Product Code [OHT](/productcode/OHT) · K230097

# IPL Hair Removal Device, Model(s): BSXT101, BSXT102, BSXT103, BSXT105, BSXT106, BSXT108 (K230097)

_Shenzhen Bsx Technology Electronics Co., Ltd. · OHT · Apr 6, 2023 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K230097

## Device Facts

- **Applicant:** Shenzhen Bsx Technology Electronics Co., Ltd.
- **Product Code:** [OHT](/productcode/OHT.md)
- **Decision Date:** Apr 6, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body and or facial hair.

## Device Story

IPL Hair Removal Device (models BSXT101-108) is an over-the-counter, home-use device for hair reduction. It utilizes Intense Pulsed Light (IPL) generated by a Xenon arc lamp to target hair follicles below the skin surface. The device is powered by an external AC adapter (100-240V). Operation involves manual activation via a finger switch; an integrated skin sensor ensures the light outlet is in full contact with the skin before pulse emission. The device features five intensity levels. Users apply the device directly to the skin to reduce hair growth with minimal pain. It is intended for personal use by consumers to achieve hair reduction, potentially benefiting patients by providing a convenient, non-invasive alternative to professional hair removal services.

## Clinical Evidence

Bench testing only. No clinical data provided. Compliance demonstrated via biocompatibility (ISO 10993-5, -10), electrical safety/EMC (ANSI/AAMI ES60601-1, IEC 60601-1-2, -1-11, -2-83), photobiological safety (IEC 62471), software verification/validation, and usability engineering.

## Technological Characteristics

Materials: ABS plastic (surface-contacting, intact skin). Energy: Xenon arc lamp (470-1200nm). Output: Max 5.0 J/cm², 3.0cm² spot size, 4-10ms pulse duration. Power: External AC adapter (100-240V, 50/60Hz). Connectivity: None. Sterilization: Not required. Software: Microprocessor-controlled with skin contact sensor and finger switch activation.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- IPL Salon Hair Reduction System (Model: F60001) ([K181568](/device/K181568.md))

## Reference Devices

- IPL HAIR REMOVAL HANDSET, Model: IPL-666 ([K220669](/device/K220669.md))
- Leaf Smooth ([K212697](/device/K212697.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

April 6, 2023

Shenzhen BSX Technology Electronics Co., Ltd. % Riley Chen Registration engineer Feiying Drug & Medical Consulting Technical Service Group Rm 2401 Zhenye International Business Center, No. 3101-90. Oianhai Road Shenzhen, Guangdong 518052 China

Re: K230097

Trade/Device Name: IPL Hair Removal Device, Model(s): BSXT101, BSXT102, BSXT103, BSXT105, BSXT106, BSXT108 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: OHT Dated: December 28, 2022 Received: January 13, 2023

Dear Riley Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Colin K. Chen -S

for

Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

Device Name IPL Hair Removal Device Model(s): BSXT101, BSXT102, BSXT103, BSXT105, BSXT106, BSXT108

#### Indications for Use (Describe)

The IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body and or facial hair.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary

#### "510(k) Summary" as required by 21 CFR Part 807.92.

#### I. Submitter

Shenzhen BSX Technology Electronics Co., Ltd.

Rm101&2/F~4/F, Building No.13, Ailian Industrial Park, Wulian Community, LongGang District, ShenZhen, 518116, GuangDong, P.R. China. Tel .: +86-0755-28719103

Hong Li Regulation Assistant Tel: +86 18133439986 Email: 1013540630@qq.com

#### II. Device

Name of Device: IPL Hair Removal Device Model(s): BSXT101, BSXT102, BSXT103, BSXT105, BSXT106, BSXT108 Common or Usual Name: Light Based Over-The-Counter Hair Removal Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: OHT Regulation Number: 21 CFR 878.4810

#### III. Predicate Device

Primary predicate device:

| Manufacturer                                                         | Predicate Device                                   | 510(k) Number | Approval Date |
|----------------------------------------------------------------------|----------------------------------------------------|---------------|---------------|
| Medical Device Branch of<br>Zhangzhou Easepal<br>Industrial Co.,Ltd. | IPL Salon Hair Reduction<br>System (Model: F60001) | K181568       | Sep. 11, 2018 |

Reference devices:

| Manufacturer                                 | Reference Device                            | 510(k) Number | Approval Date |
|----------------------------------------------|---------------------------------------------|---------------|---------------|
| Shenzhen Junbobeauty<br>Technology Co., Ltd. | IPL HAIR REMOVAL<br>HANDSET, Model: IPL-666 | K220669       | May 16, 2022  |
| Shenzhen Leaflife<br>Technology Co., Ltd     | Leaf Smooth                                 | K212697       | Nov. 19, 2021 |

### IV. Device Description

IPL Hair Removal Device (Model: BSXT101, BSXT102, BSXT103, BSXT105, BSXT106, BSXT108), is an over-the-counter, home-use and personal device for hair reduction by using 

{4}------------------------------------------------

Intense Pulsed Light (IPL). It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain.

The device is only powered by the external power adapter (a.c.100~240V 50/60Hz) and its IPL emission activation is by finger switch. It contains a skin sensor to detect appropriate skin contact, if the Light Outlet of the device is not in full contact with the skin, the device cannot trigger a light pulse. The device have six models and these models share the same performance, structure and operation, the only difference is their enclosure color.

#### V. Indications for Use

The IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body and/ or facial hair.

#### VI. Materials

| Component name          | Material of Component | Body Contact Category                  | Contact Duration   |
|-------------------------|-----------------------|----------------------------------------|--------------------|
| IPL Hair Removal Device | ABS                   | Surface-contacting device: Intact skin | Less than 24 hours |

We have tested the device for biocompatibility by a reliable third-party lab. For details, please refer to Section 16 "Biocompatibility Discussion".

#### VII. Comparison of Technological Characteristics With the Predicate Device

The IPL Hair Removal Device has the same intended use as the predicate and reference devices. The technological characteristics such as wavelength, energy density, spot size and pulse duration, are similar to the predicate device and reference devices. Any minor differences between the subject device and the listed predicate device and reference devices do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate device and reference devices for its intended use. Therefore, the IPL Hair Removal Device may be found substantially equivalent to its predicate device and reference devices.

| Comparison<br>Elements                                | Subject Device                                                                                                                         | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Reference Device 1                                                                                                                    | Reference Device 2                                                                                                     | Remark    |
|-------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|-----------|
| 510(k)<br>Number                                      | Pending                                                                                                                                | K181568                                                                                                                                                                                                                                                                                                                                                                                                                                                             | K220669                                                                                                                               | K212697                                                                                                                | /         |
| Trade name                                            | IPL Hair Removal<br>Device                                                                                                             | IPL Salon Hair<br>Reduction System,<br>Model: F60001                                                                                                                                                                                                                                                                                                                                                                                                                | IPL<br>HAIR<br>REMOVAL<br>HANDSET<br>Model: IPL-666                                                                                   | Leaf Smooth                                                                                                            | /         |
| Manufacturer                                          | Shenzhen<br>Technology<br>Electronics Co., Ltd.                                                                                        | BSX Medical Device Branch<br>of Zhangzhou Easepal<br>Industrial Co.,Ltd.                                                                                                                                                                                                                                                                                                                                                                                            | Shenzhen<br>Junbobeauty<br>Technology Co., Ltd.                                                                                       | Shenzhen Leaflife<br>Technology Co., Ltd                                                                               | /         |
| Regulation                                            | 21 CFR 878.4810                                                                                                                        | 21 CFR 878.4810                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 21 CFR 878.4810                                                                                                                       | 21 CFR 878.4810                                                                                                        | Same      |
| Comparison<br>Elements                                | Subject Device                                                                                                                         | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Reference Device 1                                                                                                                    | Reference Device 2                                                                                                     | Remark    |
| number                                                |                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                       |                                                                                                                        |           |
| Product code                                          | OHT                                                                                                                                    | OHT                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | OHT                                                                                                                                   | OHT                                                                                                                    | Same      |
| Device<br>classification                              | Class II                                                                                                                               | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Class II                                                                                                                              | Class II                                                                                                               | Same      |
| Indication for<br>use/ Intended<br>use                | The IPL Hair<br>Removal Device is<br>an over-the-counter<br>device intended for<br>removal of<br>unwanted body and/<br>or facial hair. | The IPL Salon Hair<br>Reduction System<br>(Model: F60001) is an<br>over the Counter device<br>intended for the removal<br>of unwanted body<br>and/or facial hair in<br>adults. It is also intended<br>for permanent reduction<br>in unwanted hair.<br>Permanent hair<br>reduction is defined as<br>the long-term stable<br>reduction in the numberof hairs re-growing<br>when measured at 6. 9.<br>and 12 months after the<br>completion of a<br>treatment regimen. | IPL HAIR<br>REMOVAL<br>HANDSET is an<br>over-the-counter<br>device intended for<br>removal of unwanted<br>body and/or facial<br>hair. | The Leaf Smooth is<br>an over-the-counter<br>device intended for<br>removal of<br>unwanted body<br>and/or facial hair. | Same      |
| Prescription<br>or OTC                                | OTC                                                                                                                                    | OTC                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | OTC                                                                                                                                   | OTC                                                                                                                    | Same      |
| Applicable<br>skin                                    | Fitzpatrick Skin<br>Types I-V                                                                                                          | /                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Fitzpatrick Skin<br>Types I-V                                                                                                         | /                                                                                                                      | Same      |
| Device design                                         |                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                       |                                                                                                                        |           |
| Source<br>energy                                      | Supplied by external<br>adapter                                                                                                        | Supplied by external<br>adapter                                                                                                                                                                                                                                                                                                                                                                                                                                     | Powered by external<br>power adapter                                                                                                  | Supplied by<br>external adapter                                                                                        | Same      |
| Power supply                                          | 100-240V, 50/60Hz                                                                                                                      | 100-240V AC; 50/60 Hz                                                                                                                                                                                                                                                                                                                                                                                                                                               | 100~240V AC Input<br>12V3A DC Output                                                                                                  | /                                                                                                                      | Same      |
| Dimension                                             | 190 x 70 x 45 mm                                                                                                                       | 143*69.5*43mm<br>(H*W*D)                                                                                                                                                                                                                                                                                                                                                                                                                                            | 124*83*48.5mm                                                                                                                         | /                                                                                                                      | Different |
| Weight                                                | Approx. 225 g                                                                                                                          | 650g                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 186g                                                                                                                                  | /                                                                                                                      | Different |
| Sterilization                                         | Not required                                                                                                                           | Not required                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Not required                                                                                                                          | Not required                                                                                                           | Same      |
| Output specification                                  |                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                       |                                                                                                                        |           |
| Light source                                          | Intense Pulsed Light                                                                                                                   | Intense Pulsed Light                                                                                                                                                                                                                                                                                                                                                                                                                                                | Intense Pulsed Light                                                                                                                  | Intense Pulsed Light                                                                                                   | Same      |
| Energy<br>medium                                      | Xenon Arc lamp                                                                                                                         | Xenon Arc Flashlamp                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Xenon Quartz Tube                                                                                                                     | Xenon Arc<br>Flashlamp                                                                                                 | Same      |
| Wavelength<br>range                                   | 470nm-1200nm                                                                                                                           | 475nm-1200nm                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 470nm~1100nm                                                                                                                          | 475-1100nm                                                                                                             | Similar   |
| Comparison<br>Elements                                | Subject Device                                                                                                                         | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Reference Device 1                                                                                                                    | Reference Device 2                                                                                                     | Remark    |
| Energy<br>density                                     | Max 5.0 J/cm²                                                                                                                          | Max 5.0 [J/cm²]                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 1.3~2.49J/cm²                                                                                                                         | 4-6J/cm²                                                                                                               | Similar   |
| Spot size                                             | 3.0cm² ± 0.5cm²                                                                                                                        | 1.72 cm² or 3.02 cm²                                                                                                                                                                                                                                                                                                                                                                                                                                                | 3cm²                                                                                                                                  | 3.8cm²                                                                                                                 | Same      |
| Pulse<br>duration                                     | 4-10ms                                                                                                                                 | 11-12 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 11.5-15ms                                                                                                                             | 2-10ms                                                                                                                 | Similar   |
| Pulsing<br>control                                    | Finger switch                                                                                                                          | Finger switch                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Finger switch                                                                                                                         | Finger switch                                                                                                          | Same      |
| Delivery<br>device                                    | Direct illumination<br>to tissue                                                                                                       | Direct illumination<br>to tissue                                                                                                                                                                                                                                                                                                                                                                                                                                    | Direct illumination to<br>tissue                                                                                                      | Direct illumination<br>to tissue                                                                                       | Same      |
| Output<br>intensity level                             | 5 Levels                                                                                                                               | 5 Levels                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 5 Levels                                                                                                                              | 6 Levels                                                                                                               | Same      |
| Software/<br>Firmware/<br>Microprocess<br>or Control? | Yes                                                                                                                                    | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                   | Yes                                                                                                                    | Same      |
| Additional features                                   |                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                       |                                                                                                                        |           |
| Electrical<br>safety                                  | ANSI<br>AAMI<br>ES60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-2-83                                                         | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-57                                                                                                                                                                                                                                                                                                                                                                                                                      | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-2-83                                                                      | ANSI/AAMI<br>ES60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-2-57                                            | Same      |
| Eye safety                                            | IEC 62471                                                                                                                              | /                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | IEC 62471                                                                                                                             | IEC 62471                                                                                                              | Same      |
| Biocompatibi<br>lity                                  | ISO 10993-5<br>ISO 10993-10                                                                                                            | ISO 10993-5<br>ISO 10993-10                                                                                                                                                                                                                                                                                                                                                                                                                                         | ISO 10993-5<br>ISO 10993-10                                                                                                           | ISO 10993-5<br>ISO 10993-10                                                                                            | Same      |

{5}------------------------------------------------

{6}------------------------------------------------

#### Non-clinical studies and performance data VIII.

Non-clinical tests were conducted to verify that the IPL Hair Removal Device meets all design specifications which supports the conclusion that it's Substantially Equivalent (SE) to the predicate device and reference devices. The test results demonstrated that the subject device complies with the following standards:

#### 1) Biocompatibility Testing

The device has been tested for biocompatibility, it complies with the following standards:

- > ISO 10993-5:2009, Biological Evaluation of Medical Devices -Par t 5: Tests for In Vitro Cytotoxicity
- > ISO 10993-10:2010, Biological Evaluation of Medical Devices -Par t 10: Tests for Irritation and Skin Sensitization

#### 2) Electrical Safety and EMC

{7}------------------------------------------------

Electrical safety and EMC testing was performed to, and passed, as per the following standards:

- > IEC 60601-1-2 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance -Collateral standard: electromagnetic compatibility
- > ANSI AAMI ES60601-1 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- > IEC 60601-1-11 Medical Electrical Equipment-Part 1: General Requirements for Basic Safety and Essential Performance-Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
- > IEC 60601-2-83 Medical electrical equipment Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment

#### 3) Eye Safety

- > IEC 62471 Photobiological safety of lamps and lamp systems
#### 4) Software Verification and Validation

Software verification and validation test according to the requirements of the FDA "Guidance for Pre Market Submissions and for Software Contained in Medical Devices"

#### 5) Usability

The product usability has been evaluated and verified according to the FDA guidance "Applying Human Factors and Usability Engineering to Medical Devices", issued on FEBRUARY 2016.

#### IX. Conclusions

Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device IPL Hair Removal Device is as safe, as effective, and performs as well as the legally marketed predicate device and reference devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K230097](https://fda-staging.innolitics.com/device/K230097)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).