K230091 · Purgo Biologics, Inc. · NPL · Oct 6, 2023 · Dental
Device Facts
Record ID
K230091
Device Name
THE Cover
Applicant
Purgo Biologics, Inc.
Product Code
NPL · Dental
Decision Date
Oct 6, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3930
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
THE Cover is recommended for: - Covering of immediate or delayed extraction socket to enhance preservation of the alveolar ridge.
Device Story
THE Cover is a resorbable, white, pliable collagen membrane derived from porcine tendon; functions as a physical barrier between gingival soft tissue and alveolar bone; prevents soft tissue ingrowth into bone defects; facilitates bone regeneration; used in dental surgery; applied by clinicians; no distinction between face and back; fully resorbable, eliminating need for removal; provides patient benefit by preserving alveolar ridge structure.
Clinical Evidence
No clinical data included. Evidence consists of bench testing (physicochemical properties, mechanical strength, degradation), biocompatibility testing (ISO 10993 series), sterilization validation (ISO 11137), and an in vivo beagle mandibular intraoral study comparing the subject device to the predicate over 24 weeks, showing equivalent resorption and barrier function.
Technological Characteristics
Material: Porcine-derived fibrous collagen matrix (tendon). Form: Pliable membrane, various sizes (15x20mm to 30x40mm), 0.3-0.5mm thickness. Principle: Cell-occlusive, bioresorbable scaffold. Energy: None. Sterilization: Gamma irradiation (SAL 10^-6). Standards: ISO 10993 (biocompatibility), ISO 11137 (sterilization), ISO 11607 (packaging), ISO 22442 (animal origin materials).
Indications for Use
Indicated for covering immediate or delayed extraction sockets to enhance alveolar ridge preservation in patients requiring dental bone grafting procedures.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
October 6, 2023
Purgo Biologics Inc. Byungsun Kim RA team manager #812, 27 Dunchon-daero 457beon-gil, Jungwon-gu Seongnam-si, Gyeonggi-do 13219 KOREA, SOUTH
Re: K230091
Trade/Device Name: THE Cover Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: Class II Product Code: NPL Dated: September 6, 2023 Received: September 7, 2023
Dear Byungsun Kim:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sherrill Lathrop Blitzer
for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230091
Device Name THE Cover
Indications for Use (Describe)
THE Cover is recommended for:
- Covering of immediate or delayed extraction socket to enhance preservation of the alveolar ridge.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Purgo Dental Biologics Solution. The logo features a gray atom-like symbol on the left, followed by the word "Purgo" in a bold, purple font. To the right of "Purgo" are the words "Dental Biologics Solution" stacked on top of each other, also in purple.
# 510(k) Summary
## 10/05/2023
## 1. Company
| | Submitter |
|----------------|------------------------------------------------------------------------------------------|
| Name | Purgo Biologics Inc. |
| Address | #812, 27 Dunchon-daero 457beon-gil, Jungwon-gu, Seongnam-si,<br>Gyeonggi-do, Korea 13219 |
| Phone/Fax | Tel. +82-70-4827-0451, Fax. +82-70-8673-0660 |
| Contact person | Byungsun Kim / RA<br>kimbs@purgobio.com |
| Summary Date | 10/05/2023 |
# 2. Device Name
| Trade name | : | THE Cover |
|------------------------|---|-----------------------------------|
| Regulation number | : | 21 CFR 872.3930 |
| Regulation Description | : | Bone Grafting Material |
| Product code | : | NPL |
| Classification name | : | Barrier, Animal Source, Intraoral |
| Device class | : | Class II |
| Classification Panel | : | Dental |
# 3. Predicate Device
| Primary predicate device | |
|--------------------------|-----------|
| K192042 | Bio-Gide® |
{5}------------------------------------------------
## 4. Description
THE Cover Resorbable Collagen Membrane is a white, pliable membrane consisting of fibrous collagen matrix purified from porcine tendon. THE Cover act as a barrier between the gingival and the new bone that can facilitate proper bone regeneration. As soft tissues tend to grow faster than the bone can regenerate, the membrane can help protect the bone graft particles from this faster growing connective tissue. As collagen membrane will be dissolved, no additional operation is necessary. THE Cover does not have a distinction between face and back.
THE Cover is available in various sizes.
| Size | Thickness |
|------------|----------------|
| 15 x 20 mm | |
| 15 x 30 mm | 0.3 mm, 0.5 mm |
| 25 x 30 mm | |
| 30 x 40 mm | |
## 5. Indication for use
THE Cover is recommended for:
- Covering of immediate or delayed extraction socket to enhance preservation of the alveolar ridge.
## 6. Summary of Data to Support Substantial Equivalence
The determination of substantial equivalence was based on an assessment of non-clinical performance data obtained from in vitro characterization studies, an in vivo animal study, and biocompatibility testing of the subject device.
· Performance testing was performed to evaluated the physicochemical properties of the device as followings:
- Appearance and dimension (Size, Thickness)
- Composition and purity (Amino acid analysis, Heavy metal, Fat content, Loss on drying, Collagen content)
- Enzymatic (Collagenase degradation, Trypsin resistance)
- Structural (Denaturing temperature)
#### 510(k) Summary
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Purgo Dental Biologics Solution. The logo features a gray atom-like symbol on the left, followed by the word "Purgo" in a bold, purple font. To the right of "Purgo" are the words "Dental Biologics Solution" stacked vertically, also in purple but in a smaller font size.
- Mechanical properties (Tensile strength, Tear strength)
- · Biocompatibility was evaluated in accordance with ISO 10993-1 as followings:
- Cytotoxicity per ISO 10993-5
- Irritation per ISO 10093-10
- Sensitization per ISO 10993-10
- Genotoxicity per ISO 10993-3
- Acute toxicity per ISO 10993-11
- Subchronic toxicity per ISO 10993-11
- Implantation per ISO 10993-6
- Material mediated pyrogenicity per ISO 10993-11
· Sterilization process validation was performed in accordance with ISO 11137-1 and ISO 11137-2 demonstrating a sterility assurance level (SAL) of 10-6.
• Real time aging shelf-life study was performed in accordance with ISO 11607-1 and ISO 11607-2 demonstrating product stability and packaging integrity.
- · The control of animal origin materials was performed as followings:
- Controls on sourcing, collection, and handling per ISO 22442-2
- Viral Inactivation per ISO 22442-3
An in vivo animal study conducted in a beagle mandibular intraoral model comparing THE Cover and the predicate device at 4, 8, 12, 16 and 24 weeks demonstrated that both devices were fully resorbable and acted as a barrier preventing ingrowth of soft tissue into the bone defect.
Membrane degradation rate and performance of THE Cover were no different to the predicate when evaluated by histological and histomorphometric analysis, providing in vivo evidence for substantial equivalence of the two devices.
No clinical data were included in this submission.
{7}------------------------------------------------
## 7. Technological Characteristics and Substantial Equivalence
The following comparison table of the technological characteristics of the subject device and the predicate device outlines and provides the substantial equivalency of the subject device and the predicate.
| | Subject device | Primary predicate device | Discussion |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| Device name | THE Cover | Bio-Gide® | - |
| Manufacturer | Purgo Biologics Inc. | Geistlich Pharma Ag | - |
| 510(k) Number | K230091 | K192042 | - |
| Product Code | NPL | NPL | Same |
| Indication for use | THE Cover is recommended for:<br>- Covering of immediate or delayed<br>extraction socket to enhance<br>preservation of the alveolar ridge. | Bio-Gide® is recommended for:<br>1) Augmentation around implants<br>placed in immediate extraction<br>sockets<br>2) Augmentation around implants<br>placed in delayed extraction sockets<br>3) Localized ridge augmentation for<br>later Implantation<br>4) Alveolar ridge reconstruction for<br>prosthetic treatment<br>5) Filling of bone defects after root<br>resection, cystectomy, removal of<br>retained teeth<br>6) Guided bone regeneration in<br>dehiscence Defects<br>7) Guided tissue regeneration<br>procedures in Periodontal defects | Equivalent<br>The indication for the<br>subject device is a<br>subset of that for the<br>predicate device. |
| Function | Functions as a barrier when applied<br>between Alveolar bone and soft<br>tissue. The membrane serves as a<br>bioresorbable scaffold | Functions as a barrier when applied<br>between Alveolar bone and soft<br>tissue. The membrane serves as a<br>bioresorbable scaffold | Same |
## Comparison of Characteristics
#### 510(k) Summary
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for Purgo Biologics Solution. The logo features a gray icon resembling a microscope or scientific instrument on the left. To the right of the icon, the word "Purgo" is written in a bold, purple font. Further to the right, the words "Dental Biologics Solution" are stacked vertically, also in purple, but in a smaller font size.
| | • Cell-occlusive | • Cell-occlusive | |
|-------------------------|---------------------------------------|---------------------------------------|------------------------------------------------------------------------------------------------------|
| Operating<br>Principles | • Implantable | • Implantable | Same |
| | • Resorbable | • Resorbable | |
| | • Biocompatible | • Biocompatible | |
| Resorption<br>time | Substantially resorbed<br>by 24 weeks | Substantially resorbed<br>by 24 weeks | Equivalent |
| Material | Porcine Derived Collagen | Porcine Derived Collagen | Same |
| | | | Equivalent |
| Collagen<br>source | Porcine tendon | Porcine | Using the collagen |
| | | | derived from porcine |
| | | | is the same, but the |
| | | | parts of tissue used |
| | | | are different. |
| Form | Membrane | Membrane | Same |
| Color | White to off-white | White to off-white | Same |
| Crosslink | Non-crosslinked | Non-crosslinked | Same |
| Size | 15 x 20, 15 x 30,<br>20 x 30, 30 x 40 | 13 x 25, 25 x 25,<br>30 x 40, 40 x 50 | Equivalent<br>Both devices are<br>provided in various<br>sizes for intra-oral<br>surgical procedures |
| Sterilization | Gamma Irradiation | Gamma Irradiation | Same |
| Sterility | Sterile, SAL 10-6 | Sterile, SAL 10-6 | Same |
| Shelf-life | 3 years | 3 years | Same |
| Use | Prescription | Prescription | Same |
| Single Use<br>Only | Yes | Yes | Same |
The subject device is substantially equivalent to the primary predicate device K192042 in indication for use, function, material, design and sterilization. The collagen composed both devices are derived from porcine, but the parts of animal tissue used are different. The safety of material and device were verified through biocompatibility tests and virus inactivation validation.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the logo for Purgo Dental Biologics Solution. The logo features a gray atom-like symbol on the left, followed by the word "Purgo" in a bold, purple font. To the right of "Purgo" are the words "Dental Biologics Solution" stacked vertically, also in purple, but in a smaller, less bold font.
# 8. Conclusion
Based on the information provided, the subject device is substantially equivalent to the primary predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.