K230082 · Ge Medical Systems, LLC · QKB · May 4, 2023 · Radiology
Device Facts
Record ID
K230082
Device Name
Auto Segmentation
Applicant
Ge Medical Systems, LLC
Product Code
QKB · Radiology
Decision Date
May 4, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K230082 · May 4, 2023
Auto Segmentation
Ge Medical Systems, LLC
Retrospective CT radiation therapy planning exams from multiple clinical sites
Retrospective clinical CT exams were used to perform bench testing and evaluate the device's segmentation performance (DICE similarity coefficient) against ground truth annotations, demonstrating substantial equivalence.
Retrospective clinical data; Performance evaluation; Bench testing; Deep learning validation
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Performance testing of Auto Segmentation; Retrospective performance evaluation
Adult patients (18-89 years old) undergoing radiotherapy planning; Sample Size: 302 unique patient exams (2552 contours); Number of Sites: Multiple clinical sites in North America, Asia, and Europe
Not applicable for this study
DICE similarity coefficient (DSC) compared to ground truth annotations
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Organ at risk segmentation
Deep learning algorithms
Dice Mean per organ (various thresholds)
Dice Mean per organ (various results)
911 CT exams from several clinical sites from multiple countries.
—
302 retrospective CT radiation therapy planning exams from multiple clinical sites in North America, Asia, and Europe.
3 (radiotherapy practitioners)
Indications for Use
Auto Segmentation is intended to be used as a workflow tool for initial anatomy segmentation of organs at risk on CT images as an aid in radiation therapy planning after user confirmation.
Device Story
Auto Segmentation is a server-based software tool for radiation therapy planning. It takes CT images (contrast or non-contrast) as input and uses deep learning algorithms to automatically segment 40 organ subregions across head, neck, thorax, abdomen, and pelvis. The device outputs a DICOM Radiotherapy Structure Set (RTSS) containing the generated contours. These contours are networked to a review workstation where a clinician (dosimetrist, medical physicist, or radiation oncologist) reviews, verifies, and edits the output as necessary. By providing initial contours, the device streamlines the delineation workflow, potentially reducing planning time and improving efficiency for the healthcare provider. It is intended for use in adult patients.
Clinical Evidence
Bench testing used 302 independent, retrospective CT radiation therapy planning exams (adults, global sources). Primary endpoint was DICE similarity coefficient (DSC) compared to ground truth established by three practitioners. Results showed high performance across 40 organs (e.g., Body 99.5%, Lung 97.7-97.9%, Liver 94.28%). A qualitative preference study with three blinded practitioners confirmed generated contours are adequate for clinical use. Subgroup analysis confirmed consistency across vendors, demographics, and imaging parameters.
Technological Characteristics
Server-based software application. Uses deep learning algorithms for automated image segmentation. Compatible with DICOM CT images (single energy). Output is RTSS DICOM. Developed under ISO 13485 and 21 CFR 820. Software level of concern: Major.
Indications for Use
Indicated for adult patients undergoing radiation therapy planning. Generates a Radiotherapy Structure Set (RTSS) DICOM with segmented organs at risk (OARs) from CT images to accelerate contouring workflow. Users (dosimetrists, medical physicists, radiation oncologists) must verify and correct output contours and labels.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
AdvantageSim MD With CT Atlas-Based Contouring and Re-Planning Options (K132045)
Submission Summary (Full Text)
{0}------------------------------------------------
May 4, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
GE Medical Systems, LLC % Niki Mavrodieva Regulatory Affairs Leader 3000 N. Grandview Blvd. WAUKESHA WI 53188
Re: K230082
Trade/Device Name: Auto Segmentation Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QKB Dated: April 7, 2023 Received: April 7, 2023
Dear Niki Mavrodieva:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Image /page/1/Picture/5 description: The image shows a digital signature. The signature is from Lora D. Weidner -S. The date of the signature is 2023.05.04, and the time is 13:08:39-04'00'.
Lora D. Weidner, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230082
Device Name
Auto Segmentation
#### Indications for Use (Describe)
Auto Segmentation generates a Radiotherapy Structure Set (RTSS) DICOM with segmented organs at risk which can be used by dosimetrists, medical physicists, and radiation oncologists as initial contours to accelerate workflow for radiation therapy planning. It is the responsibility of the user to verify the processed output contours and user-defined labels for each organ at risk and correct the contours/labels as needed. Auto Segmentation may be used with images acquired on CT scanners, in adult patients.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="white-space: nowrap;"> <span style="text-decoration: underline;"></span>Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="white-space: nowrap;"> <span style="text-decoration: underline;"></span>Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. There are decorative flourishes resembling water droplets or stylized leaves surrounding the circle, adding a touch of elegance to the design. The logo is simple, recognizable, and represents the multinational conglomerate.
#### K230082 510(k) SUMMARY
This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.92:
- Date: January 06, 2023 GE Medical Systems, LLC Submitter: 3000 North Grandview Blvd Waukesha, WI 53188 Primary Contact: Niki Mavrodieva
- Regulatory Affairs Leader GE Healthcare Phone: +1 (902) 302-3913 Email: nikolina.mavrodieva@ge.com
- Camille Vidal Secondary Contacts: Senior Director Regulatory Affairs GE Healthcare Phone: +1 (240) 280-5356 Email: camille.vidal@ge.com
| Subject Device Name: | Auto Segmentation |
|-----------------------|----------------------------------------------------------------|
| Device Classification | Class II |
| Regulation Number: | 21 CFR 892.2050 Medical image management and processing system |
| Product Code: | QKB |
#### Predicate Device Information
| Device Name: | AccuContour |
|--------------------|----------------------------------------------------------------|
| Manufacturer: | Xiamen Manteia Technology LTD. |
| 510(k) Number: | K191928 |
| Regulation Number: | 21 CFR 892.2050 Medical image management and processing system |
| Product Code: | QKB |
#### Reference Devices Information
| Device Name: | AdvantageSim MD With CT Atlas-Based Contouring and Re-Planning Options |
|--------------------|------------------------------------------------------------------------|
| Manufacturer: | GE Hungary KFT |
| 510(k) Number: | K132045 |
| Regulation Number: | 21 CFR 892.5840 Radiation therapy simulation system |
| Product Code: | KPQ |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. The circle is surrounded by several white, teardrop-shaped elements, giving the impression of motion or energy emanating from the center.
## Device Description
Auto Segmentation is a post-processing software designed to automatically generate contours of organ(s) at risk (OARs) from Computed Tomography (CT) images in the form of a DICOM Radiotherapy Structure Set (RTSS) series. The application is intended as a workflow tool for initial segmentation of OARs to streamline the process of organ at risk delineation. The Auto Segmentation is intended to be used by radiotherapy (RT) practitioners after review and editing, if necessary, and confirming the accuracy of the contours for use in radiation therapy planning.
Auto Segmentation uses deep learning algorithms to generate organ at risk contours for the head and neck, thorax, abdomen and pelvis regions from CT images across 40 organ subregion(s). The automatically generated organ at risk contours are networked to predefined DICOM destination(s), such as review workstations supporting RTSS format, for review and editing, as needed.
The organ at risk contours generated with the Auto Segmentation are designed to improve the contouring workflow by automatically creating contours for review by the intended users. The application is compatible with CT DICOM images with single energy acquisition modes and may be used with both GE and non-GE CT scanner acquired images (contrast), in adult patients.
#### Intended Use
Auto Segmentation is intended to be used as a workflow tool for initial anatomy segmentation of organs at risk on CT images as an aid in radiation therapy planning after user confirmation.
#### Indications for Use
Auto Segmentation generates a Radiotherapy Structure Set (RTSS) DICOM with segmented organs at risk which can be used by dosimetrists, medical physicists, and radiation oncologists as initial contours to accelerate workflow for radiation therapy planning. It is the responsibility of the user to verify the processed output contours and user-defined labels for each organ at risk and correct the contours/labels as needed. Auto Segmentation may be used with images acquired on CT scanners, in adult patients.
## Technology:
The proposed device, Auto Segmentation, employs similar fundamental scientific technology as its predicate device.
#### Comparisons
The Auto Segmentation software is substantially equivalent to the predicate device, AccuContour (K191928). The proposed device is based on the same fundamental technology as the predicate device using deep learning algorithms for organ at risk segmentation. The proposed device is intended for automatic segmentation only, while the predicate has manual contouring, registration, and other general capabilities. The table below summarizes the substantive feature/technological similarities and differences between the predicate device and the proposed device:
{5}------------------------------------------------
# GE Healthcare 510(k) Premarket Notification Submission – Auto Segmentation
Image /page/5/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized script, enclosed within a blue circle. The circle has decorative white swirls around the edges, giving it a classic and recognizable appearance.
| Specification | Predicate Device<br>AccuContour (K191928) | Proposed Device<br>Auto Segmentation |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | It is used by radiation oncology<br>department to register<br>multimodality images and segment<br>(non-contrast) CT images, to<br>generate needed information for<br>treatment planning, treatment<br>evaluation and treatment<br>adaptation. | Auto Segmentation generates a Radiotherapy<br>Structure Set (RTSS) DICOM with segmented<br>organs at risk which can be used by dosimetrists,<br>medical physicists, and radiation oncologists as<br>initial contours to accelerate workflow for<br>radiation therapy planning. It is the responsibility<br>of the user to verify the processed output contours<br>and user-defined labels for each organ at risk and<br>correct the contours/labels as needed. Auto<br>Segmentation may be used with images acquired<br>on CT scanners, in adult patients. |
| Contra-indications | None | Same |
| Patient Population | Adults only | Same |
| Algorithm | Deep Learning | Same |
| Compatible Modality | Non-contrast CT images | CT (contrast and non-contrast) images |
| OAR Segmentation<br>Anatomic Regions | Head & Neck<br>Thorax<br>Abdomen<br>Pelvis | Same |
| Workflow | Automated | Same |
| User Interface | Basic result preview of automatic<br>segmentation results. Manual<br>segmentation is possible.<br>Configuration menu. | Automated execution of the software with no user<br>interaction, other than configuration settings.<br>Generated contours are automatically transmitted<br>to review workstation(s) supporting RTSS objects<br>for review and editing, as needed. |
| Compatible Scanner<br>Models | No Limitation on scanner model,<br>DICOM 3.0 compliance required | Same |
| Deployment Platform | Cloud and server-based<br>deployment | Server-based deployment |
5-3
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. The circle is surrounded by decorative swirls, also in blue, giving the logo a classic and recognizable appearance.
#### Determination of Substantial Equivalence
#### Summary of Non-Clinical Testing
Auto Segmentation has successfully completed the design control testing per GE Healthcare's quality system. It was designed under the Quality System Regulations of 21CFR 820 and ISO 13485. No new questions of safety and effectiveness and no unexpected test results were observed.
The following quality assurance measures have been applied to the development of the system:
- Requirement Definition
- Risk Analysis and Control
- Technical Design Reviews
- Formal Design Reviews
- . Software Development Lifecycle
- Safety Testing (Verification)
- Performance Testing (Verification, Validation)
- Software Release
Auto Segmentation has been successfully verified. The testing and results did not raise any new issues of safety and effectiveness. Software documentation provided is for a "Major" Level of Concern.
The Auto Segmentation algorithms were developed and trained using a dataset of 911 different CT exams from several clinical sites from multiple countries. The original development and training data was used for radiotherapy planning, and so is representative of typical clinical practice for the subject device.
Performance testing to evaluate the device's performance in segmenting organs-at-risk was performed using a database of 302 retrospective CT radiation therapy planning exams, from multiple clinical sites in North America, Asia, and Europe, that is representative of the clinical scenarios where Auto Segmentation is intended to be used. This bench testing dataset was segregated, completely independent and not used in any stage of algorithm development, including training.
The data was acquired using a variety of CT scanners and scanner protocols from different manufacturers. The demographic distribution of the dataset consists of:
- -Gender: 87 Female, 160 Male, 55 Unknown
- -Age: Adult (18 - 89 years old)
- Ethnicity: Dataset was collected from 9 global sources, including USA, EU, and Asia. -
Note: due to anonymization, gender, age, and ethnicity information is not available for all exams.
Ground truth annotations were established following RTOG and DAHANCA clinical guidelines manually by three independent, qualified radiotherapy practitioners. The annotation process was designed to reflect segmentation practices following international clinical guidelines.
The evaluation used the DICE similarity coefficient (DSC) as a primary metric to compare 2552 Auto Segmentation generated contours to ground truth contours, and evaluate performance against predefined acceptance DICE values on a per organ basis. These 2552 contours were generated from the 302 unique patient exams in the bench testing dataset.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized script, enclosed within a blue circle. There are also two white swirls on either side of the circle.
The acceptance criteria were defined individually for each organ as the target quantitative performance requirement that the segmentation model must reach in order to establish the performance for that specific organ. Acceptance criteria were established from:
- -Published performance metrics from deep-learning based or atlas-based FDA cleared products;
- -Estimation of expected performance based on organ specific characteristics and clinical justification, where a DICE value for deep-learning or atlas-based FDA cleared devices was not accessible.
The Auto Segmentation device performance results are shown in Table 1. The following is a summary of the overall performance evaluation:
| OAR | Auto Segmentation<br>(subject device) | | Acceptance Criteria | |
|----------------------|---------------------------------------|------------|---------------------|-----------|
| | Dice Mean | Lower C195 | Type | Dice Mean |
| Adrenal Left | 78.68% | 76.63% | Estimated | 68.0% |
| Adrenal Right | 72.48% | 69.78% | Estimated | 68.0% |
| Bladder | 81.50% | 78.33% | Deep learning | 80.0% |
| Body | 99.50% | 99.38% | Atlas-based | 98.1% |
| Brainstem | 87.69% | 87.15% | Deep learning | 88.4% |
| Chiasma | 43.81% | 41.03% | Atlas-based | 11.7% |
| Esophagus | 81.69% | 80.38% | Atlas-based | 45.8% |
| Eye Left | 91.32% | 89.77% | Deep learning | 90.1% |
| Eye Right | 90.25% | 88.23% | Deep learning | 89.9% |
| Femur Left | 97.65% | 97.18% | Atlas-based | 71.6% |
| Femur Right | 97.92% | 97.78% | Atlas-based | 70.8% |
| Kidney Left | 92.53% | 90.30% | Deep learning | 86.8% |
| Kidney Right | 94.82% | 93.48% | Deep learning | 85.6% |
| Lacrimal Gland Left | 59.79% | 57.65% | Deep learning | 50.0% |
| Lacrimal Gland Right | 58.09% | 55.81% | Deep learning | 50.0% |
| Lens Left | 76.86% | 74.80% | Deep learning | 73.3% |
| Lens Right | 79.09% | 77.40% | Deep learning | 75.6% |
| Liver | 94.28% | 92.27% | Deep learning | 91.1% |
| Lung Left | 97.70% | 97.38% | Deep learning | 97.4% |
| Lung Right | 97.99% | 97.81% | Deep learning | 97.8% |
| Mandible | 92.70% | 92.36% | Deep learning | 94.0% |
#### Table 1: Summary of Auto Segmentation performance
{8}------------------------------------------------
# GE Healthcare 510(k) Premarket Notification Submission – Auto Segmentation
Image /page/8/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. The letters are also blue. There are three white swirls around the letters, evenly spaced around the circle.
| OAR | Auto Segmentation<br>(subject device) | Auto Segmentation<br>(subject device) | Acceptance Criteria | |
|-------------------------------------------------|---------------------------------------|---------------------------------------|---------------------|-----------|
| | Dice Mean | Lower CI95 | Type | Dice Mean |
| Optic Nerve Left | 79.22% | 77.99% | Deep learning | 71.1% |
| Optic Nerve Right | 80.20% | 78.94% | Deep learning | 71.2% |
| Oral Cavity | 87.43% | 86.20% | Deep learning | 91.0% |
| Pancreas | 80.34% | 78.50% | Estimated | 73.0% |
| Parotid Left | 84.35% | 83.27% | Deep learning | 65.0% |
| Parotid Right | 85.55% | 84.48% | Deep learning | 65.0% |
| Proximal Bronchial Tree<br>(PBtree) | 84.94% | 83.71% | Atlas-based | 54.8% |
| Inferior PCM (Pharyngeal<br>Constrictor Muscle) | 70.51% | 68.72% | Estimated | 68.0% |
| Middle PCM | 67.09% | 65.21% | Estimated | 68.0% |
| Superior PCM | 59.57% | 57.85% | Estimated | 50.0% |
| Pericardium | 93.58% | 92.00% | Atlas-based | 84.4% |
| Pituitary | 75.62% | 74.12% | Deep learning | 78.0% |
| Prostate | 79.67% | 77.60% | Atlas-based | 52.1% |
| Spinal Cord | 88.55% | 87.43% | Deep learning | 87.0% |
| Submandibular Left | 86.85% | 85.95% | Deep learning | 77.0% |
| Submandibular Right | 85.70% | 84.79% | Deep learning | 78.0% |
| Thyroid | 85.37% | 84.27% | Deep learning | 83.0% |
| Trachea | 91.02% | 90.47% | Atlas-based | 69.2% |
| Whole Brain | 98.53% | 98.46% | Estimated | 93.0% |
Note: in "Type" column above, Deep learning means similar device using deep learning technology, while Atlas-based means similar devices using Atlas-based technology. Estimated means acceptance criteria were estimated based on organ specific characteristics and clinical justification.
The subject device performance was superior for all organs with atlas-based predicates. The evaluation of the Dice mean for the Auto Segmentation algorithms demonstrates that the algorithm performance is in line with the performance of the predicate, as well as state of the art, recently cleared similar automated contouring devices.
A subgroup analysis found that the algorithms' performance is consistent across multiple CT system vendors, pixel spacing, slice distance, gender, and geographical subgroups. Known limitations are described in the user documentation.
The results of the algorithm testing demonstrate that Auto Segmentation performs as expected.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. The circle has decorative swirls or flourishes around the letters, giving it a classic and recognizable appearance.
## Summary of Clinical Testing
A qualitative preference study evaluation comparing to a Likert scale was conducted using a database of sample clinical CT images to demonstrate that the contours generated by the Auto Segmentation application are adequate for radiotherapy planning use. Each contour used in the evaluation was generated using the Auto Segmentation application, and reviewed by three qualified radiotherapy practitioners, who provided an assessment of the adequacy of the subject device generated contours. The evaluators completed their assessments independently and were blinded to the results of the other evaluators' assessments. The results of the algorithm clinical testing shows that the Auto Segmentation generated organ contours are adequate for use in radiotherapy planning.
## Substantial Equivalence Conclusion
Auto Segmentation and the predicate have substantially equivalent indications for use, and represent equivalent technological characteristics, including the use of deep learning algorithms.
Auto Segmentation was developed under GE Healthcare's quality system. Design verification and validation, along with bench testing and the clinical reader study provided in this submission demonstrate that the Auto Segmentation software is substantially equivalent and, hence, as safe and effective as the legally marketed predicate device. GE Healthcare's quality system design, verification, and risk management processes did not identify any unexpected results or new questions of safety and effectiveness.
GE Healthcare believes that Auto Segmentation is substantially equivalent to the predicate device and, hence, is safe and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.