# LaserME (K230077)

_Neauvia North America · GEX · Jul 17, 2023 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K230077

## Device Facts

- **Applicant:** Neauvia North America
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Jul 17, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The LaserMe is intended for use in dermatological procedures requiring coagulation of soft tissue and skin resurfacing procedures. The LaserMe is further indicated for treatment of benign pigmented lesions, such as, but not limited to lentigines (age spots), solar lentigos (sun spots), melasma, dyschromia, and for treatment of facial wrinkles and fine lines.

## Device Story

LaserME is a non-ablative diode laser surgical instrument for dermatological procedures. Device delivers 1470nm wavelength laser energy via micro-beams to epidermis; energy levels adjustable 5mJ-50mJ; spacing 1mm-4mm. Used in clinical settings by trained professionals for soft tissue coagulation, skin resurfacing, and treatment of pigmented lesions or wrinkles. Physician operates device to create controlled thermal micro-injuries; thermal effects stimulate tissue healing/remodeling. Benefits include improved skin texture and reduction of pigmented lesions/wrinkles.

## Clinical Evidence

Clinical study conducted on 18 healthy volunteers. Subjects received single laser treatment; histology analysis performed at 0, 3, and 10 days post-treatment to evaluate thermal effects and tissue healing. Results demonstrated device behaved as intended.

## Technological Characteristics

Non-ablative diode laser; 1470 +/- 20nm wavelength; max power 2W; 12x12mm applicator window. Energy 5-50mJ; pulse time 30ms; repetition frequency up to 18Hz. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-2-22 (laser safety), and ISO 10993-1 (biocompatibility).

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Emerge Fractional Laser ([K111840](/device/K111840.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 17, 2023

Neauvia North America Joy Willard Director of Quality, Regulatory, and Clinical Affairs 8480 Honevcutt Road Raleigh, North Carolina 27615

Re: K230077

Trade/Device Name: LaserME Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX. Dated: June 15, 2023 Received: June 16, 2023

Dear Joy Willard:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/5 description: The image shows the name "Jianting Wang -S" in a large, clear font. The text is horizontally oriented and appears to be the primary focus of the image. The background is plain, ensuring the text is easily readable.

Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K230077

Device Name LaserME

## Indications for Use (Describe)

The LaserMe is intended for use in dermatological procedures requiring coagulation of soft tissue and skin resurfacing procedures. The LaserMe is further indicated for treatment of benign pigmented lesions, such as, but not limited to lentigines (age spots), solar lentigos (sun spots), melasma, dyschromia, and for treatment of facial wrinkles and fine lines.

| Type of Use (Select one or both, as applicable)                                                          |                                                                                                         |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;">     Prescription Use (Part 21 CFR 801 Subpart D)  </span> | <span style="font-family: Arial, sans-serif;">     Over-The-Counter Use (21 CFR 801 Subpart C)  </span> |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

This 510(K) Summary of safety and effectiveness for the LaserME is submitted in accordance with the requirements of 21 CFR 807.92.

Applicant: Neauvia North America Address: Neauvia North America 8480 Honeycutt Road Raleigh, NC 27615 Contact Person: Joy Willard Contact Information: 984-777-5296 joy@neauvia-us.com Preparation Date: January 23, 2022 Device Trade Name: LaserME Powered laser surgical instrument Common Name: Regulation Name: 21 CFR 878.4810, Laser Surgical Instrument for use in general and plastic surgery and in dermatology Product Codes: GEX Emerge Fractional Laser K111840 Legally Marketed Predicate Device: Regulatory Class: Class II Prescription Use Description of the LaserME Laser LaserME is a non-ablative diode laser. The device is equipped with a diode laser with a wavelength of 1470 System: nm and a maximum power of 2W, intended for nonablative skin resurfacing. The device works by punctuating the skin (epidermis) with several micro laser beams. The points are evenly spaced with adjustable spacing from 1mm to 4mm. The energy delivered to a single point is regulated from 5mJ to 50mJ.

{4}------------------------------------------------

Intended use of LaserME

The LaserMe is intended for use in dermatological procedures requiring coagulation of soft tissue and skin resurfacing procedures. The LaserMe is further indicated for treatment of benign pigmented lesions, such as, but not limited to lentigines (age spots), solar lentigos (sun spots), melasma, dyschromia, and for treatment of facial wrinkles and fine lines.

| Specifications          | LaserME                                                                                                                                                                                                                                                                                                                                                                                                   | Emerge Fractional Laser K111840                                                                                                                                                                                                                                                                                                                                                                                                  |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Laser type              | Diode                                                                                                                                                                                                                                                                                                                                                                                                     | Diode                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Power Supply            | 112-240 VAC, 50/60Hz                                                                                                                                                                                                                                                                                                                                                                                      | 110-240 VAC, 50/60Hz                                                                                                                                                                                                                                                                                                                                                                                                             |
| Output wavelength       | 1470 +/- 20nm                                                                                                                                                                                                                                                                                                                                                                                             | 1410nm                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Pulse Time              | 30ms                                                                                                                                                                                                                                                                                                                                                                                                      | Max 20ms                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Applicator Window       | 12 x 12mm                                                                                                                                                                                                                                                                                                                                                                                                 | 12mm x 8mm                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Energy                  | 5mJ-50mJ                                                                                                                                                                                                                                                                                                                                                                                                  | 10mJ to 30mJ                                                                                                                                                                                                                                                                                                                                                                                                                     |
| No. of microbeams       | Up to 70                                                                                                                                                                                                                                                                                                                                                                                                  | Up to 70                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Repetition<br>Frequency | Up to 18Hz                                                                                                                                                                                                                                                                                                                                                                                                | Up to 30Hz                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Weight                  | 4 Kg                                                                                                                                                                                                                                                                                                                                                                                                      | Handpiece: 250 g<br>Base: 3 kg                                                                                                                                                                                                                                                                                                                                                                                                   |
| Dimensions              | 156 x 335 x 421 mm                                                                                                                                                                                                                                                                                                                                                                                        | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Indications for use     | The LaserMe is intended for use<br>in dermatological procedures<br>requiring coagulation of soft<br>tissue and skin resurfacing<br>procedures. The LaserMe is<br>further indicated for treatment<br>of benign pigmented lesions,<br>such as, but not limited to<br>lentigines (age spots), solar<br>lentigos (sun spots), melasma,<br>dyschromia, and for treatment of<br>facial wrinkles and fine lines. | The Emerge Fractional Laser is<br>intended for use in dermatological<br>procedures requiring coagulation<br>of soft tissue and skin resurfacing<br>procedures. The Emerge Fractional<br>Laser is further indicated for<br>treatment of pigmented lesions,<br>such as, but not limited to lentigos<br>(age spots), solar lentigos (sun<br>spots), melasma, dyschromia, and<br>for treatment of facial wrinkles and<br>fine lines. |
| Non-clinical Testing:   |                                                                                                                                                                                                                                                                                                                                                                                                           | The following performance data was provided in<br>support of the substantial equivalence determination:<br>IEC 60601-1 Test for Medical Electrical equipment<br>was performed for General Requirements for basic<br>safety and essential performance.                                                                                                                                                                            |

Comparison to the Predicate Device:

{5}------------------------------------------------

|                   | IEC 60601-1-2 Test for Medical Equipment for<br>General Requirements for basic safety and essential<br>performance: electromagnetic compatibility                                                                                                                                                                                                        |
|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                   | IEC 60601-2-22 Test for Particular requirements for<br>basic safety and essential performance of surgical,<br>cosmetic, therapeutic and diagnostic laser equipment                                                                                                                                                                                       |
|                   | ISO 10993-1 Biological Evaluation of Medical<br>Devices - Part 1: Evaluation and Testing Within a Risk<br>Management Process                                                                                                                                                                                                                             |
|                   | Histological assessment in ex vivo tissue to<br>demonstrate that the extent of the damage induced due<br>to the worst-case pulse energy of the subject device is<br>comparable to the damage induced due to the worst-<br>case pulse energy of the predicate device.                                                                                     |
| Clinical Testing: | The thermal effects of the LaserME and tissue healing<br>process were evaluated in 18 healthy volunteers. The<br>study was conducted by exposing human subjects to a<br>single laser treatment followed by histology analysis<br>performed immediately, 3 and 10 days post treatment.<br>The study results show that the LaserMe behaved as<br>intended. |

Conclusion: Based on the comparison and analysis above, the proposed subject device LaserME is determined to be Substantially Equivalent (SE) to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K230077](https://fda-staging.innolitics.com/device/K230077)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).