← Product Code [GWQ](/productcode/GWQ) · K230073

# Okti (K230073)

_Compumedics Limited · GWQ · Feb 9, 2023 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K230073

## Device Facts

- **Applicant:** Compumedics Limited
- **Product Code:** [GWQ](/productcode/GWQ.md)
- **Decision Date:** Feb 9, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.1400
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The Okti System is intended for use in the recording, displaying, analysis, printing and storage of human biological parameters such as heart and muscle activity, eye movement, breathing and body movements to assist in the diagnosis of various neurological disorders. The Okti is designed for use in a hospital or other clinical environment. The Okti is only to be used under the direction of a physician.

## Device Story

Okti is a physiological data acquisition device for EEG studies; captures electrical signals via electrodes (32, 64, or 128 channels); inputs include EEG, ECG, EMG, and EOG. Device uses 24-bit ADCs and FPGA-based FIR filters to digitize and process signals; data transmitted via Ethernet, Wi-Fi, or stored on internal SD card. Operated by clinicians in hospital/clinical settings; data viewed/analyzed on PC running Compumedics Profusion EEG software. Output assists physicians in diagnosing neurological disorders by providing high-fidelity physiological recordings. Benefits include flexible channel configurations and wireless/networked data transmission for clinical workflow efficiency.

## Clinical Evidence

No clinical data provided. Substantial equivalence is supported by bench testing only, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and electroencephalograph performance testing (IEC 80601-2-26). Testing verified accuracy of signal reproduction, input dynamic range, input noise, frequency response, and common mode rejection ratio against design specifications.

## Technological Characteristics

Physiological data acquisition system; 32/64/128 referential/bipolar channels; 24-bit ADCs; FPGA-based FIR filtering; DC-coupled inputs; >100 MΩ input impedance; >100dB CMRR. Power: Li-ion battery or PoE (IEEE 802.3af). Connectivity: Ethernet, Wi-Fi, Bluetooth (oximeter). Class CF patient safety. Dimensions vary by module. Software: Profusion EEG for configuration/acquisition.

## Regulatory Identification

An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.

## Predicate Devices

- Grael System (Grael EEG) ([K093223](/device/K093223.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

February 9, 2023

Image /page/0/Picture/1 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white. To the right of the square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.

Compumedics Limited % Dave Yungvirt CEO Third Party Review Group, LLC 25 Independence Blvd Warren, New Jersey 07059

Re: K230073

Trade/Device Name: Okti Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: GWQ Dated: January 6, 2023 Received: January 10, 2023

Dear Dave Yungvirt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Patrick Antkowiak -S

for Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

Device Name Okti

Indications for Use (Describe)

The Okti System is intended for use in the recording, displaying, analysis, printing and storage of human biological parameters such as heart and muscle activity, eye movement, breathing and body movements to assist in the diagnosis of various neurological disorders. The Okti is designed for use in a hospital or other clinical environment. The Okti is only to be used under the direction of a physician.

| Type of Use (Select one or both, as applicable)                                                 |
|-------------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) |
| Over-The-Counter Use (21 CFR 801 Subpart C)                                                     |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a logo for Compumedics. The logo features a blue circle with a stylized "m" inside of it. To the right of the circle is the word "COMPUMEDICS" in blue, all caps. Below the company name is the tagline "Defining Life's Signals" in a smaller font.

# 510(k) Summary

#### Device Name: Okti

The following 510(k) summary is being submitted in accordance with 21 CFR 807.92.

#### Submission Details

Fax number:

#### Applicant Information:

| Submitter:      | Compumedics Limited                                          |
|-----------------|--------------------------------------------------------------|
| Address:        | 30-40 Flockhart Street, Abbotsford 3067, Victoria, Australia |
| Phone number:   | +61 (0) 3 8420 7300                                          |
| Fax number:     | +61 (0) 3 8420 7399                                          |
| Contact person: | Michael Frischman                                            |
| Address:        | 30-40 Flockhart Street, Abbotsford 3067, Victoria, Australia |
| Phone number:   | +61 (0) 3 8420 7362                                          |

Email: michael.frischman@compumedics.com.au Submission Prepared: 02/11/2022

## Subject Device Information:

| Trade name:           | Okti                                        |
|-----------------------|---------------------------------------------|
| Common name:          | Okti                                        |
| Primary Product code: | GWQ                                         |
| Classification Names: | Full-Montage Standard Electroencephalograph |
| Panel:                | Neurology                                   |
| Device class:         | II                                          |
| Regulation numbers:   | 21 CFR 882.1400                             |

+ 61 (0) 3 8420 7399

## Predicate Device Information

| Trade Name:           | Grael System                                                             |
|-----------------------|--------------------------------------------------------------------------|
| Model:                | Grael EEG                                                                |
| Manufacturer:         | Compumedics Limited                                                      |
| 510(k) number:        | K093223                                                                  |
| 510(k) Decision Date: | 15th January 2010                                                        |
| Classification:       | Full-Montage Standard Electroencephalograph, Ventilatory Effort Recorder |
| Product code:         | GWQ, MNR                                                                 |
| Device class:         | II                                                                       |
| Regulation numbers:   | 21 CFR 882.1400                                                          |

## Intended Use / Indications For Use

The Okti System is intended for use in the recording, displaying, analysis, printing and storage of human biological parameters such as heart and muscle activity, eye movement, breathing and body movements to assist in the diagnosis of various neurological disorders. The Okti is designed for use in a hospital or other clinical environment. The Okti is only to be used under the direction of a physician.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Compumedics. The logo features a blue circle with a stylized "m" inside, followed by the word "COMPUMEDICS" in blue, sans-serif font. Below the company name is the tagline "Defining Life's Signals".

# Device Description

The Okti is a physiological data acquisition device intended for conducting EEG studies. It is used in conjunction with a PC running Compumedics Profusion EEG software. Power is provided by either an internal Lithium-Ion battery or a POE 100BASE-TX network connection.

Physically, the Okti comprises a base unit housing the battery and digital electronic circuits. A changeable module connects to the base unit and houses the analogue circuity and electrode connectors providing options for 32, 64 or 128 referential, or single ended, channels primarily intended for EEG. Each module includes several differential inputs for electrode-based signals such as ECG, EMG or EOG. All these inputs are class CF for patient safety.

Patient physiological data is digitised using multiple 24-bit ADCs and filtered using fixed point FIR filters implemented internally to an FPGA located in the analogue module. The data is transferred to a generalpurpose processor in the base unit over a synchronous serial link. From there the patient data may be saved to a device internal SD card or sent to a remote PC over an Ethernet or Wi-Fi LAN connection. A Bluetooth interface is also included for connection to external devices such as an oximeter.

The Okti system is available in Okti32, Okti64, or Okti128. These differ only in channel number with below characteristics.

| Model   | Inputs                                                                                                               |
|---------|----------------------------------------------------------------------------------------------------------------------|
| Okti32  | 32, DC-coupled, fully isolated, referential (monopolar); 8 user-defined, DC-coupled, fully isolated, bipolar inputs  |
| Okti64  | 64, DC-coupled, fully isolated, referential (monopolar); 4 user-defined, DC-coupled, fully isolated, bipolar inputs  |
| Okti128 | 128, DC-coupled, fully isolated, referential (monopolar); 8 user-defined, DC-coupled, fully isolated, bipolar inputs |

## Comparison to Predicate Device

The Okti's predicate device is the Grael (K093223) comprises two primary models: the Grael, is a full channel device suitable for conducting either PSG or EEG studies; and the Grael EEG, a reduced channel device suitable for EEG studies only.

The Grael EEG is optimized for neurological studies, and does not contain inputs used for respiratory monitoring during PSG which are not necessary for neurological studies. The Grael EEG model is the predicate device and claims relating to PSG are not being made for the Okti.

| Characteristic                   | Okti                                                                                                                                                                                                                                                           | Grael EEG                                                                                                                                                                                                                                                     |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Usage                            |                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                               |
| 510(k) Number                    | -                                                                                                                                                                                                                                                              | K093223                                                                                                                                                                                                                                                       |
| Classification                   | 882.1400                                                                                                                                                                                                                                                       | 882.1400                                                                                                                                                                                                                                                      |
| Product Code                     | GWQ                                                                                                                                                                                                                                                            | GWQ                                                                                                                                                                                                                                                           |
| Class                            | II                                                                                                                                                                                                                                                             | II                                                                                                                                                                                                                                                            |
| Intended Use                     | EEG studies to assist in the<br>diagnosis of various<br>neurological disorders                                                                                                                                                                                 | EEG studies to assist in the<br>diagnosis of various<br>neurological disorders                                                                                                                                                                                |
| Type of Use                      | Prescription Only                                                                                                                                                                                                                                              | Prescription Only                                                                                                                                                                                                                                             |
| Use Environment                  | Hospital / Clinical use only                                                                                                                                                                                                                                   | Hospital / Clinical use only                                                                                                                                                                                                                                  |
| Temperature                      | -10°C to 50°C storage/non-<br>operating<br>0°C to 40°C operating                                                                                                                                                                                               | -10°C to 50°C storage/non-<br>operating<br>0°C to 40°C operating                                                                                                                                                                                              |
| Relative Humidity                | 20 to 90% relative humidity<br>non-condensing                                                                                                                                                                                                                  | 20 to 90% relative humidity<br>non-condensing                                                                                                                                                                                                                 |
| Altitude                         | < 3000m                                                                                                                                                                                                                                                        | < 3000m                                                                                                                                                                                                                                                       |
|                                  | Physical Specifications                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                               |
|                                  | Dimensions                                                                                                                                                                                                                                                     | Base unit: 84mm x 43mm x 184mm<br>Amplifier module: 32ch 84mm x 26mm x 150mm<br>64-ch: 84mm x 25mm x 150mm<br>128-ch: 84mm x 33mm x 150mm                                                                                                                     |
|                                  | Mass                                                                                                                                                                                                                                                           | Base unit: < 225g battery not fitted<br>< 555g battery fitted<br>Amplifier module weight:<br>32ch < 145g<br>64-ch < 170g<br>128ch < 155g                                                                                                                      |
|                                  | Electrical Specifications                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                               |
|                                  | Power Supply                                                                                                                                                                                                                                                   | Power is provided by either rechargeable Lithium-ion battery (6400 mAH, > 300 charge/discharge cycle at 23°C, 11.25 volts) or Power Over Ethernet (either mid-span injector or PoE switch, per IEEE 802.3af standard)                                         |
| Operating Voltage                | 48 Volts                                                                                                                                                                                                                                                       | 48 Volts                                                                                                                                                                                                                                                      |
| Power Consumption                | < 4 Watts                                                                                                                                                                                                                                                      | < 10 Watts                                                                                                                                                                                                                                                    |
|                                  | Channel Specifications                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                               |
| Input Impedance                  | > 100 MΩ all channels                                                                                                                                                                                                                                          | > 100 MΩ channels 1-32<br>> 20 MΩ channels 33-40                                                                                                                                                                                                              |
| Bias Current                     | Typically 1nA                                                                                                                                                                                                                                                  | Typically 1nA                                                                                                                                                                                                                                                 |
| Input Noise                      | < 2 µV peak-to-peak typical<br>(referential)                                                                                                                                                                                                                   | < 2 µV peak-to-peak typical<br>(referential)                                                                                                                                                                                                                  |
| Input Range                      | User Selectable:<br>300mV peak-to-peak 600mV peak-to-peak 1200mV peak-to-peak 3000mV peak-to-peak                                                                                                                                                              | User selectable:<br>300mV peak-to-peak<br>600mV peak-to-peak<br>1200mV peak-to-peak<br>3000mV peak-to-peak                                                                                                                                                    |
| CMRR                             | > 100dB                                                                                                                                                                                                                                                        | > 100dB                                                                                                                                                                                                                                                       |
| Crosstalk                        | < -60dB                                                                                                                                                                                                                                                        | < -60dB                                                                                                                                                                                                                                                       |
| High Pass Filter                 | DC Coupled all channels                                                                                                                                                                                                                                        | DC Coupled on all channels                                                                                                                                                                                                                                    |
| Low Pass Filter                  | 3dB cut-off frequency (fc) at sampling rate (Fs):<br>fc = 71 Hz, Fs = 256 Hz<br>fc = 143 Hz, Fs = 512 Hz<br>fc = 284 Hz, Fs = 1024 Hz<br>fc = 580 Hz, Fs = 2048 Hz<br>fc = 1150 Hz, Fs = 4096 Hz                                                               | 3dB cut-off frequency (fc) at sampling rate (Fs):<br>fc = 71 Hz, Fs = 256 Hz<br>fc = 143 Hz, Fs = 512 Hz<br>fc = 284 Hz, Fs = 1024 Hz<br>fc = 580 Hz, Fs = 2048 Hz<br>fc = 1150 Hz, Fs = 4096 Hz                                                              |
| Notch Filter                     | Software based display filtering of 50Hz, 60Hz or off                                                                                                                                                                                                          | Software based filtering of 50Hz, 60Hz or off                                                                                                                                                                                                                 |
| Isolation Specifications         | Complies with IEC 60601-1                                                                                                                                                                                                                                      | Complies with IEC 60601-1                                                                                                                                                                                                                                     |
| Analogue to Digital<br>Converter | 24 bit resolution                                                                                                                                                                                                                                              | 24 bit resolution                                                                                                                                                                                                                                             |
| Sample and Data rates            | Data sampled at 16384<br>samples per second and<br>decimated to data rates<br>shown below. Actual data<br>rate sent is determined by<br>controlling PC software<br>application.<br>Output data rates of 4096,<br>2048, 1025, 512 or 256<br>samples per second. | Data sampled at 16384<br>samples per second and<br>decimated to data rates<br>shown below. Actual data<br>rate sent is determined by<br>controlling PC software<br>application.<br>Output data rates: 4096,<br>2048, 1024, 512, or 256<br>samples per second. |
| Communication Interface          |                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                               |
| Software                         | Profusion EEG for amplifier<br>configuration, data<br>acquisition, and display                                                                                                                                                                                 | Profusion EEG for amplifier<br>configuration, data<br>acquisition, and display                                                                                                                                                                                |
| Network Types                    | IEEE 802.3/802.3u Ethernet,<br>auto MDIX Requires an IEEE<br>802.3af compliant switch or<br>mid-span power injector and<br>network isolator<br><br>Wi-Fi interface IEEE 802.11a,<br>802.11b, 802.11g and<br>802.11n compatible                                 | 802.3/802.3u twisted pair<br>Ethernet with auto MDIX;<br>RJ45 connector                                                                                                                                                                                       |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is a logo for Compumedics. The logo features a blue circle with a stylized "m" inside of it. To the right of the circle is the word "COMPUMEDICS" in blue, and below that is the tagline "Defining Life's Signals" also in blue.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image is a logo for Compumedics. The logo features a blue circle with a stylized "m" inside of it. To the right of the circle is the word "COMPUMEDICS" in blue, with the trademark symbol in the upper right corner. Below the word is the phrase "Defining Life's Signals" in a smaller font.

# Summary of Performance Data

An extensive collection of tests have been conducted and successfully completed including,

- . Electrical Safety Testing as per IEC 60601-1:2005. These ensure there is no potential for detrimental effects on patients, other persons, animals or the surroundings. They relate particularly to aspects concerning electrical safety such as means of operator protection, protection against electric shock and other hazards; as well as mechanical aspects such as construction and protection against spillage.
- . Electromagnetic Compatibility Testing as per IEC 60601-1-2, and modification by the listed particular and collateral standards. These EMC tests were designed to verify that device emissions are in accordance with allowable limits, as well as device immunity against electrical and magnetic phenomena which may be expected to encounter in line with its intended use.
- . Electroencephalograph safety and performance testing as per IEC 80601-2-26. Testing against the particular standard for electroencephalograph includes validation against multiple essential performance requirements such as accuracy of amplitude and rate of variation signal reproduction, input dynamic range and differential offset voltage, input noise, frequency response, and common mode rejection ratio.
- Bench testing against Okti functional requirements to ensure that performance meets hardware and software design specifications

Further details such as device specifications, test requirements, acceptance criteria, results, and evaluation are available throughout this 510(k) submission. All tests passed with results equivalent to the Grael, and did not raise additional concerns of safety and effectiveness.

## Conclusion

Compumedics Ltd. considers the Okti to be substantially equivalent to its predicate devices the Grael System, specifically the Grael EEG model. The indications for use, technological characteristics, and

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Compumedics. The logo features a blue circle with a stylized "m" inside, followed by the company name "COMPUMEDICS" in blue, uppercase letters. Below the company name is the tagline "Defining Life's Signals" in a smaller font.

underlying principles of operation are the same. There are no questions of safety and effectiveness, and substantial equivalence is supported by extensive performance testing data.

---

**Source:** [https://fda-staging.innolitics.com/device/K230073](https://fda-staging.innolitics.com/device/K230073)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
