K230043 · Anhui Hongyu Wuzhou Medical Manufacturer Co., Ltd. · FMI · Sep 20, 2023 · General Hospital
Device Facts
Record ID
K230043
Device Name
Insulin Pen Needle
Applicant
Anhui Hongyu Wuzhou Medical Manufacturer Co., Ltd.
Product Code
FMI · General Hospital
Decision Date
Sep 20, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Insulin Pen Needle is intended for use with pen injector devices for the subcutaneous injection of insulin. The intended user population of insulin pen needle is for "Adult Use Only" (ages 21yrs old and above).
Device Story
Insulin Pen Needle is a single-use, sterile, hypodermic needle designed for attachment to insulin pen injectors via a threaded connection. It consists of a needle tube (304 stainless steel), needle hub, needle container, and needle shield (all polypropylene). The device is used by adult patients to deliver insulin subcutaneously. It is available in various gauges (28G-34G) and lengths (4mm-12mm). The device is manually operated by the patient. It facilitates insulin delivery, which is essential for diabetes management. The device is EO sterilized.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via ISO 11608-2, ISO 9626, and ISO 7864. Biocompatibility testing (cytotoxicity, sensitization, irritation, systemic toxicity, hemolysis, pyrogen, subacute toxicity) performed per ISO 10993-1. Particulate matter testing per USP <788>. Packaging integrity verified via ASTM D3078-02, ASTM F1929-15, DIN 58953-6, and ASTM F88/F88M-21. Sterilization validated per ISO 11135 with EO residuals per ISO 10993-7. Shelf-life validated per ASTM F1980-16.
Technological Characteristics
Materials: 304 stainless steel needle tube, polypropylene hub/container/shield. Energy: Manual. Connectivity: None. Sterilization: EO (SAL 10^-6). Dimensions: 28G-34G; 4mm-12mm lengths. Standards: ISO 11608-2, ISO 7864, ISO 9626, ISO 10993, ISO 11135, ASTM F1980-16, USP <788>.
Indications for Use
Indicated for subcutaneous insulin injection using pen injector devices in adults (ages 21 and older).
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" and the words "U.S. Food & Drug Administration".
September 20, 2023
Anhui Hongyu Wuzhou Medical Manufacturer Co., Ltd. % Alice Huang RA Manager Shanghai Mind-link Business Consulting Co., Ltd. Room 8208, Second Floor, No 1399, Jiangyue Road, Minhang District Shanghai, 201114 China
Re: K230043
Trade/Device Name: Insulin Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: August 20, 2023 Received: August 23, 2023
Dear Alice Huang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/7 description: The image shows the text "Courtney Evans -S" in a large, sans-serif font. The text is arranged vertically, with "Courtney" on the top line, "Evans" on the second line, and "-S" on the third line. The background of the image is white, with a faint, light blue watermark pattern.
Digitally signed by Courtney Evans -S Date: 2023.09.20 11:28:53 -04'00'
For CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230043
Device Name Insulin Pen Needle
Indications for Use (Describe)
The Insulin Pen Needle is intended for use with pen injector devices for the subcutaneous injection of insulin. The intended user population of insulin pen needle is for "Adult Use Only" (ages 21yrs old and above).
Type of Use (*Select one or both, as applicable*)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K230043- 510(K) Summary
{4}------------------------------------------------
#### I. SUBMITTER:
Manufacture Name: Anhui Hongyu Wuzhou Medical Manufacturer Co., Ltd. Address : No.2 Guanyin Road, Taihu Economic Development Zone, Anqing City, Anhui Province, China. Tel: +86-556 5129666 Fax: +86-556 4249999 Contact Person: Bingyi Xiang Title:General Manager Phone: +86-556 5129666 Email:hwj1@hongyu-wuzhou.cn
Submission Correspondent: Alice Huang Email: alice.huang@mind-link.net Tel:+86 15618536177 Shanghai Mind-link Business Consulting Co., Ltd. Room A08, Floor 14th, No 699, Jiaozhou Road, Jingan District, Shanghai
Summary prepared: September 15, 2023
#### II. DEVICE
Name of Device:Insulin Pen Needle Regulation Number: 21 CFR 880.5570 Common Name: Insulin Pen Needle for single use Classification Panel: General Hospital Regulatory Class: II Product Code: FMI
#### III. PREDICATE DEVICE
Primary predicate device: Insulin Pen Needle (Ordinary Type), Insulin Pen Needle (Safety Type) (K202319)
{5}------------------------------------------------
#### DEVICE DESCRIPTION IV.
The proposed device, Insulin pen needle and insulin injection pen are use cooperatively to inject insulin into the human body. The Insulin Pen Needle is composed of a needle container, a needle shield, a needle hub, a needle tube.
The insulin pen needle for single use is offered in various gauge sized and length.
The proposed device is available in EO sterilized to achieve a Sterility Assurance Level (SAL) of 106.
| Designated<br>metric size<br>(mm) | Gauge | Bevel<br>angle | | Outer<br>diameter of<br>stylet (mm) | | Nominal length of needle tube<br>(mm) | | | | | | ID<br>(mm) | | OD<br>(mm) | | Color |
|-----------------------------------|-------|----------------|----|-------------------------------------|------|---------------------------------------|---|---|---|----|----|------------|-------|------------|-------|------------|
| | | LB | SB | RW | TW | 4 | 5 | 6 | 8 | 10 | 12 | RW | TW | MIN | MAX | |
| 0.18 | 34G | √ | √ | / | / | √ | √ | √ | – | – | – | 0.064 | 0.091 | 0.178 | 0.191 | Orange |
| 0.20 | 33G | √ | √ | / | / | √ | √ | √ | √ | – | – | 0.089 | 0.105 | 0.203 | 0.216 | Black |
| 0.23 | 32G | √ | √ | / | / | √ | √ | √ | √ | √ | – | 0.089 | 0.105 | 0.229 | 0.241 | Deep green |
| 0.25 | 31G | √ | √ | / | / | √ | √ | √ | √ | √ | – | 0.114 | 0.125 | 0.254 | 0.267 | White |
| 0.30 | 30G | √ | √ | 0.11 | 0.13 | √ | √ | √ | √ | √ | √ | 0.133 | 0.165 | 0.298 | 0.320 | Yellow |
| 0.33 | 29G | √ | √ | 0.11 | 0.15 | √ | √ | √ | √ | √ | √ | 0.133 | – | 0.324 | 0.351 | Red |
| 0.36 | 28G | √ | √ | 0.11 | 0.15 | √ | √ | √ | √ | √ | √ | 0.133 | – | 0.349 | 0.370 | Blue-green |
Table 1 Specification of Insulin Pen Needle
#### V. INDICATIONS FOR USE
The Insulin Pen Needle is intended for use with pen iniector devices for subcutaneous injection of insulin. The intended user population of insulin pen needle is for "Adult Use Only" (ages 21yrs old and above).
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE
Insulin pen needle and insulin injection pen are use cooperatively to inject insulin into the human body. Additionally, the Insulin Pen Needle is similar to the Insulin Pen Needle (Ordinary Type) (K202319) in regard to insertion, design, size ranges, and material.
Table 2 General Comparison for Insulin Pen Needle
{6}------------------------------------------------
| Item | Proposed Device<br>Insulin Pen Needle | Predicate Device<br>Insulin Pen Needle<br>(Ordinary Type) | Remark |
|-----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| K number | TBD | K202319 | |
| Classification | Class II | Class II | Same |
| Product Code | FMI | FMI | Same |
| Regulation No. | 880.5570 | 880.5570 | Same |
| Intended use | Insulin pen needle and insulin<br>injection pen are use<br>cooperatively to inject insulin<br>into the human body. | The Insulin Pen Needle<br>(Ordinary Type) is intended<br>for use with pen injector<br>devices for the<br>subcutaneous injection<br>of insulin. | Similar<br>Note 1 |
| Indications for use | The Insulin Pen Needle is<br>intended for use with pen<br>injector devices for<br>subcutaneous injection of<br>insulin. The intended user<br>population of insulin pen<br>needle is for “Adult Use Only”<br>(ages 21yrs old and above). | The Insulin Pen Needle<br>(Ordinary Type) is intended<br>for use with pen injector<br>devices for the<br>subcutaneous injection<br>of insulin. | Similar<br>Note 1 |
| User population | “Adult Use Only” (ages 21yrs<br>old and above) | Adult and Pediatric | Different<br>Note 2 |
| Type-of-use | Prescription only | Over the counter | Different<br>Note 3 |
| Configuration and<br>materials of<br>construction for all<br>components | Needle tube 304 stainless<br>steel | Needle tube 304 stainless<br>steel | |
| | Needle hub Polypropylene | Needle hub Polypropylene | |
| | Needle container Polypropylene | Outer Sheath Polypropylene | Same |
| | Needle shield Polypropylene | Inner Sheath Polypropylene | |
| | Sealed Paper Paper | Sealed Paper Paper | |
| Operation Mode | For Manual Use Only | For Manual Use Only | Same |
| Method of<br>attachment to pen<br>injector | Through threaded connection | Through threaded<br>connection | Same |
| Item | Proposed Device<br>Insulin Pen Needle | Predicate Device<br>Insulin Pen Needle (Ordinary<br>Type) | Remark |
| Needle Gauge | 28G, 29G, 30G, 31G, 32G, 33G,<br>34G | 29G, 30G, 31G, 32G, 33G | Different<br>Note 4 |
| Needle Length | 4mm, 5mm, 6mm, 8mm, 10mm,<br>12mm | 4mm, 5mm, 6mm, 8mm, 10mm,<br>12mm | Same |
| Patient-contact materials | | | |
| Needle Tube | 304 stainless steel | 304 stainless steel | Different<br>Note 5 |
| Needle Hub | PP | Polypropylene | Note 5 |
| Lubricant | Silicone oil | MDX4-4159 | |
| adhesive | Uv-curable adhesive | UV glue | |
| Design<br>specification of<br>the inner sheath<br>(dimensions,<br>color, materials<br>and strength) | 14.5*5.8*5.8mm;<br>Transparent color; Polypropylene;<br>Tensile stress at yield>24MPa;<br>Flexural Modulus:>1000MPa;<br>Charpy Notched Impact<br>Strength (23°C) >7kJ/m2 | 14.3*5.8*5.8mm;<br>Transparent color; Polypropylene;<br>Tensile stress at yield>20MPa;<br>Flexural Modulus:>800MPa;<br>Charpy Notched Impact<br>Strength (23°C) >2.8kJ/m2 | Similar |
| Design<br>specification of<br>the outer sheath<br>(dimensions,<br>color, materials<br>and strength) | 30*16*16mm;<br>Transparent color; Polypropylene;<br>Tensile stress at yield>24MPa;<br>Flexural Modulus:>1000MPa;<br>Charpy Notched Impact<br>Strength (23°C) >7kJ/m2 | 30.3*13.4*13.4mm;<br>Transparent color; Polypropylene;<br>Tensile stress at yield>20MPa;<br>Flexural Modulus:>800MPa;<br>Charpy Notched Impact<br>Strength (23°C) >2.8kJ/m2 | Similar |
| Bond between<br>hub and needle<br>tube | Clause 4.4 of ISO 11608-2:2012;<br>Clause 4.12 of ISO 7864:2016 | Clause 4.4 of ISO 11608-2:2012;<br>Clause 4.12 of ISO 7864:2016 | Same |
| Needle tip<br>configuration | Conform with ISO 7864 standards | Conform with ISO 7864 standards | Same |
| Cytotoxicity | No Cytotoxicity | No Cytotoxicity | Same |
| Intracutaneous<br>reactivity | No Irritation to Skin | No Irritation to Skin | |
| Skin<br>Sensitization | No Skin Sensitization | No Skin Sensitization | |
| Acute Systemic<br>Toxicity | No Systemic Toxicity | No Systemic Toxicity | |
| Pyrogen | No pyrogen | No pyrogen | |
| Hemolysis | No Hemolysis | No Hemolysis | |
| | | | |
| Subacute<br>Systemic Toxicity | No Subacute Systemic Toxicity | No Subacute Systemic Toxicity | |
| | | | |
| Method | EO Sterilized | Irradiation Sterilized | Different<br>Note 6 |
| SAL | 10-6 | 10-6 | Same |
| Endotoxin Limit | 20 EU per device | 20 EU per device | Same |
| Shelf life | 5 years | 5 years | Same |
| Label/ Labeling | Complied with 21 CFR part 801 | Complied with 21 CFR part 801 | Same |
{7}------------------------------------------------
# Table 3 Safety and Performance Comparison for Insulin Pen Needle
{8}------------------------------------------------
### Discussion in details:
### Note 1: Intended Use and Indications for use
Insulin pen needle and insulin pen needle (Ordinary Type) have similar intended use and indications for use, which are intended to be used with Insulin Pen for injection of insulin.
### Note 2: User population
The subject device' s user population is covered by the range of user population in the predicate devices. This different will do not raise different questions of safety and effectiveness than the predicate device because all necessary information will be labeled on the packaging and the Instruction for Use.
## Note 3: Type-of-use
The subject device' s type-of use is more rigorous than predicate devices. This different will do not raise different questions of safety and effectiveness than the predicate device because all necessary information will be labeled on the packaging and Instruction for Use.
## Note 4: Needle gauge
The needle gauge of proposed device is more than that of predicate devices. The difference in needle gauges is made of the same raw materials and the same manufacturing process. Different gauge of needle tube will be selected by users. This difference does not affect intended use and not affect substantially equivalence on safety and effectiveness. In addition, all the needle size of proposed device has been tested. The test results comply with ISO 7864 and ISO 9626 standards requirements. Therefore, this difference will not affect the Substantially Equivalency (SE) between the proposed and predicate device.
## Note 5: Patient-contact Material
Although the material of proposed and the predicate device is different, the MSDS of Silicone oil and UV-curable adhesive is provided, and the patient-contact material of the proposed device material conforms to the ISO 10993 series of standards. Therefore, this difference will not affect the Substantial Equivalence (SE) between the proposed and predicate device.
{9}------------------------------------------------
### Note 6: Sterilize Method
Although the Sterilize Method of proposed and the predicate device is different, the sterilize of the proposed device conforms to the ISO 11135:2014 of standards. And the EO residue result was comply with ISO 10993-7 standard requirement. Therefore, this difference will not affect the Substantial Equivalence (SE) between the proposed and predicate device.
#### VII. PERFORMANCE DATA
The non-clinical tests of the proposed device are tested in conformance with the following standards.
#### Bench Testing
The chemical performance and physical performance have been tested in compliance with the following standard:
- . ISO11608-2: 2012,
- ISO 9626: 2016 and ●
- ISO 7864: 2016. ●
The testing results are in conformance with the corresponding standard.
#### Sterility and Shelf-life
The sterilization process of the proposed device has been validated in compliance with ISO 11135. The EO residual doesn't exceed the limit according to ISO 10993-7. The shelf-life validation study was conducted under the accelerated ageing condition in compliance with ASTM F1980-16 to verify the claimed 5 years shelf-life.
#### Package Testing
The packaging testing has been performed in compliance with the following FDA recognized consensus standards.
- Vacuum Leak Test, ASTM D3078-02;
- Dye Penetration Test ASTM F1929-15.
- Microbial Barrier Properties Test DIN 58953-6: 2016;
- Seal Strength Test ASTM F88/F88M-21.
#### Biocompatibility Testing
The biocompatibility evaluation for the insulin pen needles were conducted in accordance with the guidance that is "Use of International Standard ISO 10993-1,"Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" June 16, 2016, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within Risk Management Process," as
{10}------------------------------------------------
recognized by FDA. The battery of testing included the following tests:
- Cytotoxicity
- · Sensitization
- · Irritation
- Systemic Toxicity
- Hemolysis
- Pyrogen
- Subacute Toxicity
- Particulate Matter
Particulate testing was conducted in accordance with USP <788>. The testing results and met the USP acceptance criteria.
#### Simulated Clinical Use Testing
Simulated clinical use testing of the proposed device has been conducted in compliance with the FDA Guidance for industry and FDA Stuff, Medical Devices with Sharps Injury Prevention Features, August 9, 2005.
### Clinical testing
Not Applicable.
#### VIII. CONCLUSION
The differences between the predicate and the proposed device do not raise any new or different questions of safety or effectiveness. The proposed device is substantially equivalent to the predicate device with respect to indications for use, technological characteristics, and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.