Peer-reviewed clinical literature; Retrospective clinical cohorts from medical records/institutional databases
Systematic literature review of 11 primary publications reporting clinical outcomes for transvesical prostatectomy, compared against 10 publications for extraperitoneal and 7 for transabdominal prostatectomy, to support substantial equivalence.
Prostatectomy; Literature Review; Retrospective Clinical Data; Substantial Equivalence
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Systematic Literature Review of Transvesical Prostatectomy; Systematic literature review of retrospective clinical cohorts; Study Period: January 1, 2018 – January 7, 2023
Patients undergoing robotic-assisted prostatectomy (TVSP, TVRP, EPRP, TARP); Sample Size: ~2,336 total patients across all cohorts; Number of Sites: Multiple (Multi-institutional and single-center studies)
Extraperitoneal Radical Prostatectomy (EPRP) and Transabdominal Radical Prostatectomy (TARP)
Operative time, EBL, conversion rate, complication rates, length of stay, readmission rate, continence rate, positive surgical margins
Indications for Use
The Intuitive Surgical Endoscopic Instrument Control System (da Vinci SP Surgical System, Model SP1098) is intended to assist in the accurate control of Intuitive Surgical EndoWrist SP Instruments during urologic surgical procedures that are appropriate for a single port approach and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use. Intuitive Surgical EndoWrist SP Instruments are controlled by the da Vinci SP Surgical System, Model SP1098, and include flexible endoscopes, blunt and sharp endoscopic dissectors, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, and suturing through a single port. The system is indicated for urologic surgical procedures that are appropriate for a single port approach and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
Device Story
Computer-controlled surgical system enabling minimally invasive surgery via single port. Components: Surgeon Console, Vision Cart, Patient Cart. Surgeon controls instruments/camera via hand controls/foot pedals while viewing 3D imagery. Vision Cart provides video/image processing and networking. Patient Cart holds instrument arm with four drives for instruments and camera. Used in operating rooms by trained physicians. System facilitates tissue manipulation (grasping, cutting, dissection, suturing, electrocautery). Labeling modification adds simple prostatectomy as a representative procedure under radical prostatectomy umbrella. Benefits include robotic-assisted access to deep pelvis and precise instrument control.
Clinical Evidence
No clinical trials conducted for this labeling change. Evidence consists of design validation in male cadaver (transvesical prostatectomy access/motion) and systematic literature review (Real-World Evidence) of 11 primary publications (approx. 428 patients) comparing transvesical prostatectomy outcomes to extraperitoneal and transabdominal approaches. Metrics included operative time, estimated blood loss, complication rates, and length of stay, demonstrating similar clinical outcomes across approaches.
Technological Characteristics
Computer-controlled surgical system; 3D vision system; instrument arm with four drives; flexible endoscopes; EndoWrist SP instruments. No technological changes from predicate. Connectivity via Vision Cart networking hardware. Sterilization and biocompatibility specifications remain unchanged from predicate.
Indications for Use
Indicated for adult patients undergoing urologic surgical procedures appropriate for single-port approach and transoral otolaryngology procedures in the oropharynx for benign tumors and malignant tumors (T1/T2).
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
Predicate Devices
da Vinci SP Surgical System, Model SP1098, EndoWrist SP Instruments, and Accessories (K202571)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 5, 2023
Intuitive Surgical Inc. Mike Yramategui Fellow Regulatory Engineer 1020 Kifer Rd. Sunnyvale, California 94086
Re: K230033
Trade/Device Name: Da Vinci SP Surgical System (SP1098) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: NA Y Dated: January 3, 2023 Received: January 5, 2023
Dear Mike Yramategui:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at
https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
The OHT4: Office of Surgical and Infection Control Devices has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the device's labeling:
{1}------------------------------------------------
The safety and effectiveness of this device for use in the performance of general laparoscopic surgerv procedures have not been established. This device is only intended to be used for single port urological procedures and for transoral otolaryngology surgical procedures in the oropharynx for benign tumors and malignant tumors classified as T1 and T2 with the da Vinci EndoWrist SP Instruments and the da Vinci SP Surgical System (SP1098).
Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.
The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation -emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# David Krause -S
for Binita Ashar, M.D., M.B.A., F.A.C.S. Director OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K230033
#### Device Name
da Vinci SP Surgical System, Model SP1098
#### Indications for Use (Describe)
### da Vinci SP Surgical System, Model SP1098:
The Intuitive Surgical Endoscopic Instrument Control System (da Vinci SP Surgical System, Model SP1098) is intended to assist in the accurate control of Intuitive Surgical EndoWrist SP Instruments during urologic surgical procedures that are appropriate for a single port approach and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant turnors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
### Endo Wrist SP Instruments:
Intuitive Surgical Endo Wrist SP Instruments are controlled by the da Vinci SP Surgical System, Model SP1098, and include flexible endoscopes, blunt and sharp endoscopic dissectors, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, and suturing through a single port. The system is indicated for urologic surgical procedures that are appropriate for a single port approach and transoral otolaryngology surgical procedures in the oropharynx restricted to benign turnors and malignant tumors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
| Type of Use (Select one or both, as applicable) | <table style="border:none;"><tr><td><span style="font-size: 12px;"> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span></td><td><span style="font-size: 12px;"> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span></td></tr></table> | <span style="font-size: 12px;"> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-size: 12px;"> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 12px;"> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-size: 12px;"> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> | | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
60
{4}------------------------------------------------
# 510(k) Summary [21 CFR § 807.92(c)]
#### SUBMITTER INFORMATION I.
| Submitter: | Intuitive Surgical, Inc.<br>1266 Kifer Road<br>Sunnyvale, CA 94086 |
|------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Mike Yramategui<br>Fellow Regulatory Engineer<br>Phone Number: 408-523-2145<br>Fax Number: 408-523-8907<br>Email: Mike.Yramategui@intusurg.com |
Date Summary Prepared: April 4, 2023
#### SUBMITTER INFORMATION II.
| Trade Name: | da Vinci SP® Surgical System, Model SP1098, EndoWrist SP® Instruments, and Accessories |
|----------------------|----------------------------------------------------------------------------------------|
| Common Name: | System, Surgical, Computer Controlled Instrument |
| Classification Name: | Endoscope and Accessories (21 CFR §876.1500) |
| Regulatory Class: | Class II |
| Product Code: | NAY (System, Surgical, Computer Controlled Instrument |
| Submission Type: | Traditional 510(k) |
## III. PREDICATE DEVICE INFORMATION
| Predicate Device: | da Vinci SP Surgical System, Model SP1098,<br>EndoWrist SP Instruments, and Accessories (K202571) |
|-------------------|---------------------------------------------------------------------------------------------------|
|-------------------|---------------------------------------------------------------------------------------------------|
{5}------------------------------------------------
### IV. DEVICE DESCRIPTION
The da Vinci SP Surgical System is designed to enable complex surgery using a minimally invasive approach. The system consists of a Surgeon Console, a Vision Cart, and a Patient Cart and is used with a camera, instruments, and accessories.
The surgeon seated at the Surgeon Console controls all movement of the instruments and camera by using two hand controls and a set of foot pedals. The surgeon views the camera image on a three-dimensional (3D) viewer, which provides a view of patient anatomy and instrumentation, along with icons and other user interface features.
The Vision Cart includes supporting electronic equipment, such as the camera light source, video and image processing, and the networking hardware. The Vision Cart also has a touchscreen to view the camera image and adjust system settings.
The Patient Cart is the operative component of the da Vinci SP Surgical System. Its primary function is to support the positioning of the surgical port and to manipulate the surgical instruments and camera. The Patient Cart is positioned at the operating room and contains an instrument arm that is positioned with respect to the target patient anatomy. The instrument arm contains four instrument drives that hold up to three surgical instruments and the camera. The patient-side assistant installs and removes the camera and instruments intra-operatively.
This 510(k) is for a labeling modification only, to add "simple prostatectomy" as a new representative, specific procedure in the Professional Instructions for Use. Simple prostatectomy is a "covered" procedure under radical prostatectomy as an "umbrella" procedure. This 510(k) also includes the addition of performance data (cadaver) for "transvesical prostatectomy" as validated approach for prostatectomy in the Performance Data section of the User Manual.
{6}------------------------------------------------
#### V. INDICATIONS FOR USE
### da Vinci SP Surgical System, Model SP1098:
The Intuitive Surgical Endoscopic Instrument Control System (da Vinci SP Surgical System, Model SP1098) is intended to assist in the accurate control of Intuitive Surgical EndoWrist SP Instruments during urologic surgical procedures that are appropriate for a single port approach and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
### EndoWrist SP Instruments:
Intuitive Surgical EndoWrist SP Instruments are controlled by the da Vinci SP Surgical System, Model SP1098, and include flexible endoscopes, blunt and sharp endoscopic dissectors, scissors, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, and suturing through a single port. The system is indicated for urologic surgical procedures that are appropriate for a single port approach and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional
#### COMPARISON OF INTENDED USE, INDICATIONS FOR USE AND VI. TECHNOLOGAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The da Vinci SP Surgical System, Model SP1098 and EndoWrist SP Instruments and Accessories are unchanged from the predicate device in terms of intended use, design, performance, and technological characteristics. The labeling has been changed to add "simple prostatectomy" as a new representative, specific procedure in the Professional Instructions for Use as a "covered" procedure under the existing clearance for radical prostatectomy as an "umbrella" procedure, and benign and radical transvesical prostatectomy as umbrella procedures.
{7}------------------------------------------------
### VII. PERFORMANCE DATA
This 510(k) is for a labeling modification only and there are no technological changes made to the subject device. Thus, previous bench testing related to electromagnetic compatibility, electrical safety, software, biocompatibility, sterilization, and shelf life were not impacted by this change.
Simple prostatectomy was added as a new representative, specific "covered" procedure in the Professional Instructions for Use under the cleared "umbrella" radical prostatectomy procedure. No performance testing was necessary to add simple prostatectomy as a representative, specific procedure.
Design validation testing was performed in a male cadaver to confirm the ability of the SP1098 for performing a transvesical prostatectomy. Evaluation of the ability to provide robotic and laparoscopic instrument access to the deep pelvis via an incision in the urinary bladder was confirmed and an assessment of instrument motion throughout the prostatectomy procedure was performed. Evaluation of SP Access Port Kit robustness as well as the ability of the SP Access Port Kit to provide adequate retraction for SP instrument access while remaining securely installed during use was also evaluated.
Real-World Evidence (RWE) supporting the substantial equivalence of transvesical prostatectomy to previously cleared extraperitoneal prostatectomy (EPRP) and transabdominal prostatectomy (TARP) was also provided. Peer-reviewed literature published between Jan 01, 2018 and January 7, 2023 on robotic-assisted surgery using the SP1098 were identified by systematically searching the Scopus. Embase, and PubMed databases. After systematic literature screening, 11 primary publications reporting relevant clinical outcomes for transvesical simple/radical prostatectomy performed using the da Vinci SP Surgical System were included for comparison. Publications for extraperitoneal and transperitoneal approaches to prostatectomy using the SP1098 were similarly screened, and all primary publications reporting relevant clinical outcomes with 50 or more patients in at least one study arm were included, resulting in 10 publications reporting EPRP data and 7 publications reporting TARP data. Table 1 includes the ranges of raw values extracted for each clinical outcome for comparison between approaches. The data in Table 1 demonstrate similar ranges of outcomes between the surgical approaches listed. Table 2 provides a listing of the publications.
{8}------------------------------------------------
### VII. CONCLUSION
The da Vinci SP Surgical System is unmodified from the predicate device, and there are no changes to the intended use or indications for use for these labeling changes. The addition of simple prostatectomy as a representative procedure does not represent a change or modification that requires clinical data to evaluate safety or effectiveness of the device. It does not introduce any different issues of safety or effectiveness as compared to the urological procedures previously cleared for the SP1098 system.
Design validation testing provided demonstrates the ability of the da Vinci SP Surgical System to perform transvesical prostatectomy. A review of the published literature (RWE) provides additional confirmation that a transvesical approach (TVSP / TVRP) is substantially equivalent to the cleared transabdominal approach (TARP) and extraperitoneal (EPRP) approaches for performing radical prostatectomy, and do not raise different questions of safety or effectiveness.
Thus, these labeling changes to the da Vinci SP Surgical System are substantially equivalent to the cleared predicate device.
{9}------------------------------------------------
| Outcome | Transvesical Prostatectomy (TVP)a | | Extraperitoneal Radical<br>Prostatectomy (EPRP)b<br>(K202571) | Transabdominal Radical<br>Prostatectomy (TARP)b<br>(K173906) |
|--------------------------------------------------|-----------------------------------------|---------------------------------------|---------------------------------------------------------------|--------------------------------------------------------------|
| | Simple (TVSP)<br>for benign indications | Radical (TVRP)<br>for prostate cancer | | |
| # Included Publications<br>(Total Patient Count) | 8 publications1-8,<br>~247 patients | 3 publications9-11,<br>~181 patients | 10 publications10,12-20,<br>~1,259 patients | 7 publications16,20-25,<br>~649 patients |
| Operative Time (Avg) | $153^4$ - 232.43 min | 1999 - 21211 min | 14715 - 203.213 min | 11425,24 - 248.216 min |
| EBL (Avg) | 1001,2,4,8 - 227.13 mL | 10010 - 1359 mL | 5015 - 197.213 mL | 5023-25 - 20021 mL |
| Conversion Rate | 0%4,7 | 0%9 | 0%12,13,16,17,19 | 0%16,21 |
| Intraoperative Complication Rate | 0%2,4,5,7 - 2.4%6 | 0%9,11 | 0%12,13,17,19 - 2.0%15 | 0%23-25 - 2.7%20 |
| 30-Day Postoperative Complication Rate | 0%5 - 12.5%4,c | 5.0%9,d - 12.8%10 | 3.8%17 - 18.3%19 | 0%23,24 - 15.2%16 |
| Major (C-D grade ≥ III) Complication Rate | 0%4,6,8 - 7.14%3,e | 0%9 - 2.6%10 | 0%15 - 11.7%19 | 0%23,24 - 6.5%16 |
| Length of Stay | <24 hrs2,4,6-8 - 2.5 days3 | <24 hrs9-11 | <24 hrs10,12-14,16,19,20 - 2 days17 | <24 hrs20,22,23 - 7 days21 |
| 30-Day Readmission Rate | 1.0%8 | 5.0%9,d - 5.1%10 | 2.0%15 - 8.6%14 | 0%23-25 - 6.4%22 |
| Urinary Retention Rate | 0%2,4 - 4.4%8 | 5.1%10 | 0.6%17 - 2.6%10 | 1.3%20 |
| 90-Day/3-Month Continence Rate | 90%5,f - 100%2 | 96.6%10 - 97%11 | 38.4%15 - 85%13 | 62.5%16 - 78%23 |
| Urinary Catheterization Time (Avg) | 1.95 - 10.53 days | 310 - 49,11 days | 7 days 10,14,15,20 | 5 days20,23,24,25 |
| Positive Surgical Margins | NA | 14.5%11 - 15.4%10 | 17.3%15 – 29%17 | 14%23 - 41.3%16,g |
| Overall Mortality | 0 (0%) at 30 days3 | NR | NR | NR |
### Table 1: Literature Summary Table: da Vinci SP Transvesical vs. Extraperitoneal and Transabdominal Prostatectomy*
«All primary publications from 01 Jan 2018 – 07 Jan 2023 included
6Primary publications from 01 Jan 2018 – 07 Jan 2023 with n ≥ 50 patients in at least one study arm included
Clow sample size (1/8 patients). Next highest postoperative complication rate reported for TVSP: 11% (10/91 patients)
4Low sample size (1/20 patients) representing a single minor (C-D grade I) complication
eNo other TVSP publication reported major (C-D grade ≥ III) complications
/Low sample size (9/10 patients). Next lowest continence rate reported for TVSP: 97% (41/42 patients)
816/19 had high risk features: n=15 patients with T3abN0-1, n=1 patient with T2N0 with GrGp4-5
Complication / Adverse Event, C-D : Estinated Blood Loss; EPRP : Extrapertoned Radical Prostectorny : 1QR: Interquarile Range; NA: Not Applicable;
NR: Not Reported; TARP: Transbominal Radical Prostatetomy; TVRP: Transvescal Radical Prostatectomy; TVSP: Transvescal Simple Prostatetony
* Literature summary table provides data as extracted from publications and were not statistically analyzed
{10}------------------------------------------------
| Author(s) | Title |
|-----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. Ganesan V, Steinberg RL,<br>Garbens A, et al. | Single-port robotic-assisted simple prostatectomy is associated with<br>decreased post-operative narcotic use in a propensity score matched<br>analysis. Journal of Robotic Surgery. 2021. |
| 2. Kaouk J, Sawczyn G,<br>Wilson C, et al. | Single-Port Percutaneous Transvesical Simple Prostatectomy Using the<br>SP® Robotic System: Initial Clinical Experience. Urology. 2020. |
| 3. Khalil MI, Chase A,<br>Joseph JV, Ghazi A. | Standard Multi-Port versus Single-Port Robot-Assisted Simple<br>Prostatectomy: A Single Center Initial Experience. Journal of<br>endourology. 2022. |
| 4. Sawczyn G, Aminsharifi<br>A, Garisto J, Valero R,<br>Kaouk J. | Single-port transvesical robotic simple prostatectomy using the novel SP<br>surgical system: Technical aspects. Urology Video Journal. 2020;5. |
| 5. Steinberg RL, Passoni N,<br>Garbens A, Johnson BA,<br>Gahan JC. | Initial experience with extraperitoneal robotic-assisted simple<br>prostatectomy using the da Vinci SP surgical system. Journal of robotic<br>surgery. 2019. |
| 6. Zeinab AM, Kaviani A,<br>Ferguson E, et al. | Single-port transvesical versus open simple prostatectomy: a perioperative<br>comparative study. Prostate Cancer Prostatic Dis. 2022. |
| 7. Zeinab AM, Kaviani A,<br>Ferguson E, Beksac T,<br>Eltemamy M, Kaouk J. | A Transition Towards A Faster Recovery in Single-Port Transvesical<br>Simple Prostatectomy. J Endourol. 2022. |
| 8. Zeinab MA, Beksac AT,<br>Corse T, et al. | The Multi-Institutional Experience in Single-Port Robotic Transvesical<br>Simple Prostatectomy for BPH Management. J Urol.<br>2022:101097ju0000000000002692. |
| 9. Kaouk J, Beksac AT,<br>Zeinab MA, Duncan A,<br>Schwen ZR, Eltemamy<br>M. | Single Port Transvesical Robotic Radical Prostatectomy: Initial Clinical<br>Experience and Description of Technique. Urology. 2021. |
| 10. Zeinab AM, Beksac AT,<br>Ferguson E, Kaviani A,<br>Kaouk J. | Transvesical versus extraperitoneal single-port robotic radical<br>prostatectomy: a matched-pair analysis. World J Urol. 2022;40(8):2001-<br>2008. |
| 11. Zeinab AM, Kaviani A,<br>Ferguson E, Beksac AT,<br>Kaouk J. | Single-port transvesical robotic radical prostatectomy: Description of<br>technique. Urology Video Journal. 2022;15. |
| 12. Aminsharifi A, Hemal S,<br>Aram P, Abou Zeinab M,<br>Beksac T, Kaouk J. | The performance and optimum cutoff value for pelvic cavity index as a<br>predictor of early continence after extraperitoneal single-port robotic<br>radical prostatectomy: Role of pelvic anatomical characteristics. J<br>Endourol. 2022. |
| 13. Aminsharifi A, Wilson<br>CA, Sawczyn G, Kim S,<br>Lenfant L, Kaouk J. | Predictors associated with a prolonged hospital stay after single-port<br>extraperitoneal robotic radical prostatectomy: A comparative analysis of<br>outpatient versus inpatient care. J Endourol. 2020. |
| 14. Beksac AT, Zeinab MA,<br>Ferguson E, Kaviani A,<br>Kaouk J. | Single-Port Extraperitoneal Robot Assisted Radical Prostatectomy -<br>Description of Technique. Urology Video Journal. 2022;15. |
| 15. Harrison R, Stifelman<br>M, Billah M, et al. | Propensity-Score Matched Analysis Between Extraperitoneal Single Port<br>and Intraperitoneal Multiport Radical Prostatectomy: A Single-<br>Institutional Experience. Urology. 2022. |
| 16. Kaouk J, Aminsharifi A,<br>Wilson CA, et al. | Extraperitoneal versus Transabdominal Single-Port Robotic Radical<br>Prostatectomy: A Comparative Analysis of Perioperative Outcomes. The<br>Journal of urology. 2019:101097JU0000000000000700. |
| 17. Kim J, Kaldany A,<br>Lichtbroun B, et al. | Single Port Robotic Radical Prostatectomy: Short-term Outcomes and<br>Learning Curve. J Endourol. 2022. |
| Author(s) | Title |
| 18. Lenfant L, Corrigan D,<br>Beksac AT, Schwen Z,<br>Kaouk J. | Learning curve analysis of single-port robot-assisted extraperitoneal<br>prostatectomy using the cumulative sum (CUSUM) method. BJU Int.<br>2021. |
| 19. Wilson CA, Aminsharifi<br>A, Sawczyn G, et al. | Outpatient Extraperitoneal Single-port Robotic Radical Prostatectomy.<br>Urology. 2020. |
| 20. Zeinab AM, Beksac AT,<br>Ferguson E, et al. | Single-port Extraperitoneal and Transabdominal Radical Prostatectomy: A<br>Multi-Institutional Propensity-Score Matched Study. Urology. 2022. |
| 21. Kim KH, Ahn HK, Kim<br>M, Yoon H. | Technique and perioperative outcomes of single-port robotic surgery using<br>the da Vinci SP platform in urology. Asian J Surg. 2022. |
| 22. Lenfant L, Sawczyn G,<br>Kim S, Aminsharifi A,<br>Kaouk J. | Single-institution Cost Comparison: Single-port Versus Multiport Robotic<br>Prostatectomy. European Urology Focus. 2020. |
| 23. Moschovas CM, Bhat<br>KRS, Onol F, Rogers T,<br>Patel V. | Early outcomes of Single Port Robotic Radical Prostatectomy.Lessons<br>Learned from The learning curve experience. BJU Int. 2020. |
| 24. Moschovas CM, Bhat<br>KRS, Sandri M, et al. | Comparing the Approach to Radical Prostatectomy Using the Multiport da<br>Vinci Xi and da Vinci SP Robots: A Propensity Score Analysis of<br>Perioperative Outcomes. Eur Urol. 2020. |
| 25. Moschovas CM, Kind S,<br>Bhat S, et al. | Implementing the da Vinci SP Without Increasing Positive Surgical<br>Margins: Experience and Pathologic Outcomes of a Prostate Cancer<br>Referral Center. Journal of Endourology. 2022;36(4):493-498 |
### Table 2: List of Publications Reviewed
{11}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.