Baxter SIGMA Spectrum Infusion Pump with Master Drug Library
K230022 · Baxter Healthcare Corporation · FRN · Mar 31, 2023 · General Hospital
Device Facts
Record ID
K230022
Device Name
Baxter SIGMA Spectrum Infusion Pump with Master Drug Library
Applicant
Baxter Healthcare Corporation
Product Code
FRN · General Hospital
Decision Date
Mar 31, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used for the controlled administration of fluids. These may include pharmaceutical drugs, blood, blood products and mixtures of required patient therapy. The intended routes of administration consist of the following clinically accepted routes; intravenous, arterial, subcutaneous, epidural or irrigation of fluid space. The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used in conjunction with legally marketed and compatible intravenous administration sets and medications provided by the user. The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is suitable for a variety of patient care environments such as, but not limited to hospitals and outpatient care areas. The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to reduce operator interaction through guided programming, thereby helping to reduce errors. The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used by trained healthcare professionals.
Device Story
Large volume infusion pump; delivers fluids/medications via linear peristaltic mechanism; uses inlet/outlet valves for flow control. Operated by trained healthcare professionals in hospitals/outpatient areas. Features Master Drug Library (MDL) software for preprogrammed delivery profiles, advisories, and limits to reduce medication errors. Pump accepts drug library via wireless network or IrDA. Primary and secondary processors ensure safe operation. Updates include modified upstream occlusion alarm algorithm and associated labeling. Device provides precision volumetric accuracy and smooth fluid dynamics. Benefits include reduced operator interaction and minimized risk of medication errors.
Clinical Evidence
Bench testing only. No clinical data presented. Performance testing was conducted to validate the updated upstream occlusion alarm algorithm.
Technological Characteristics
Electromechanical large volume infusion pump; linear peristaltic pumping mechanism; primary and secondary processor architecture. Software-controlled. Connectivity via wireless network or IrDA. Operates on Windows-based MDL Editor platform. No patient-contact materials in fluid path.
Indications for Use
Indicated for controlled administration of fluids (pharmaceuticals, blood, blood products) via intravenous, arterial, subcutaneous, epidural, or irrigation routes in hospital and outpatient settings. Intended for use by trained healthcare professionals.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (K133801)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.
March 31, 2023
Baxter Healthcare Corporation Meaghan Bonn Principal Specialist, Regulatory Affairs 25212 W. Illinois Route 120 Round Lake. Illinois 60073
## Re: K230022
Trade/Device Name: Baxter SIGMA Spectrum Infusion Pump with Master Drug Library Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: FRN Dated: December 21, 2022 Received: January 4, 2023
#### Dear Meaghan Bonn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
# Courtney H. Lias -S
Courtney H. Lias, Ph.D. Office Director OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
#### Device Name
Baxter SIGMA Spectrum Infusion Pump with Master Drug Library
#### Indications for Use (Describe)
The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used for the controlled administration of fluids. These may include pharmaceutical drugs, blood, blood products and mixtures of required patient therapy. The intended routes of administration consist of the following clinically accepted routes; intravenous, arterial, subcutaneous, epidural or irrigation of fluid space. The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used in conjunction with legally marketed and compatible intravenous administration sets and medications provided by the user.
The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is suitable for a variety of patient care environments such as, but not limited to hospitals and outpatient care areas.
The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to reduce operator interaction through guided programming, thereby helping to reduce errors. The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used by trained healthcare professionals.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size:10px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span style="font-size:10px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
|----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif and slightly italicized. The word is horizontally oriented and takes up most of the frame.
# Section 5. 510(k) Summary
January 4, 2023
### OWNER:
Baxter Healthcare Corporation One Baxter Parkway Deerfield, Illinois 60015
## CONTACT PERSON:
Meaghan Bonn Principal Specialist, Global Regulatory Affairs 32650 N Wilson Road Round Lake, IL 60073 Telephone: (224)270-6470 Fax: (224)270-4119
# IDENTIFICATION OF THE DEVICE:
Common Name: Infusion Pump Trade/Device Name: Baxter SIGMA Spectrum Infusion Pump with Master Drug Library Classification Panel: 80 General Hospital Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: FRN, PHC
# Table 1. Product Codes for Baxter SIGMA Spectrum Infusion Pump with Master Drug Library
| Code Number | Name |
|-------------|--------------------------------|
| 35700BAX2 | SIGMA Spectrum Infusion System |
| 35723v080 | Master Drug Library |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Baxter" in a bold, blue, sans-serif font. The letters are slightly slanted to the right, giving the word a sense of movement. The word is the logo for Baxter International, a global healthcare company.
# PREDICATE DEVICE:
The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is substantially equivalent to the following predicate device: See Table 2 for predicate device.
| Device | Company | Predicate 510(k) | Clearance Date |
|-----------------------------------------------------------------------|----------------------------------|------------------|----------------|
| Baxter SIGMA<br>Spectrum Infusion<br>Pump with Master Drug<br>Library | Baxter Healthcare<br>Corporation | K133801 | 04/07/2014 |
Table 2. Predicate Device(s)
# DESCRIPTION OF THE DEVICE:
The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is a large volume infusion pump system that provides for safe and effective delivery of fluids into a patient in a controlled manner, as identified in 21 CFR, 880.5725. The pump is a software controlled, electromechanical device used for the infusion of pharmaceutical drugs, blood, blood products and mixtures of required patient therapy through administration sets at user selectable rates and volumes. The feedback-controlled, motorized pumping mechanism is of linear peristaltic design and uses inlet and outlet valves for flow control. The pump utilizes a primary and secondary processor to assure safe operation while providing infusion pump capabilities for a wide range of applications.
The pump is specifically manufactured and calibrated for the application of a manufacturer's brand of standard gravity administration sets, as indicated in the device labeling. For use, the administration set is loaded into the infusion pump. After acceptance of program parameters, the pump is started and fluid is propelled by the peristaltic action of the pumping mechanism against the outside surface of the administration set tubing. The pump is controlled to create smooth fluid dynamics, precision volumetric accuracy and uniformity of flow rate. None of the pump materials contact the administration set's fluid path.
The infusion pump is small in comparison to the traditional Large Volume Parenteral (LVP) infusion pumps currently on the market. It is designed to be used in a variety of patient care environments such as, but not limited to hospitals and outpatient care areas using an IV pole mounted configuration.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be italicized, slanting slightly to the right. The background is plain white, which makes the blue text stand out.
The Master Drug Library (MDL) Editor is a software application that allows the generation, configuration and management of a downloadable drug library into a SIGMA Spectrum infusion pump.
The drug library can be loaded directly into the SIGMA Spectrum infusion pump through a wireless network host or through an Infrared Data Association (IrDA) device. The MDL Editor software operates on a Microsoft Windows® platform.
Using the MDL Editor software application, a facility can provide preprogrammed delivery profiles, advisories and limits for a corresponding drug that is intended for a specific use classification or clinical care area, thus reducing the risk of medication errors.
The MDL Editor software application allows the ability to generate both standard or customized drug and fluid reports by clinical care area. The MDL Editor software application also provides a feature to restrict/limit the access of data to only appropriate personnel, providing additional security and rights to specific users.
# INDICATIONS FOR USE:
The SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used for the controlled administration of fluids. These may include pharmaceutical drugs, blood, and blood products. The intended routes of administration consist of the following clinically accepted routes: intravenous, arterial, subcutaneous, epidural or irrigation of fluid space. The SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used in conjunction with legally marketed and compatible intravenous administration sets and medications provided by the user.
The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is suitable for a variety of patient care environments such as, but not limited to hospitals and outpatient care areas.
The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to reduce operator interaction through guided programming, thereby helping to reduce errors. The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used by trained healthcare professionals.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the word "Baxter" in a bold, blue font. The font appears to be sans-serif and slightly italicized. The word is the primary focus of the image and is positioned horizontally.
# TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE:
The intended use and function of the proposed device remains the same as the predicate device. The 510(k) premarket notification, K133801, of the predicate device was cleared on May 7, 2014.
# DISCUSSION OF NONCLINICAL TESTS:
The scope of this notification is to update the algorithm and labeling related to the upstream occlusion alarm of the pump. Performance testing for the software was completed. Biocompatibility is not affected by this notification. The intended use and function of the proposed device is the same as the predicate device.
# CONCLUSION:
The scope of this notification is to update the algorithm and labeling related to the upstream occlusion alarm of the pump per the performance testing. The intended use and function of the proposed device is equivalent to the predicate 510(k), currently commercialized device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.