← Product Code [KGC](/productcode/KGC) · K223916

# PUMA-G System (K223916)

_Coaptech, Inc. · KGC · Mar 29, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K223916

## Device Facts

- **Applicant:** Coaptech, Inc.
- **Product Code:** [KGC](/productcode/KGC.md)
- **Decision Date:** Mar 29, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5980
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K223916 · Mar 29, 2023 | PUMA-G System | Coaptech, Inc. | Post-market analysis data | Post-market analysis was utilized as part of the performance data to support the safety and effectiveness profile of the device in comparison to the predicate. | Post-market analysis; Performance data |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Post-Market Analysis; Post-market surveillance/analysis | — | Not applicable for this study | Safety and effectiveness |

## Indications for Use

The CoapTech PUMA-G System (PUMA-G) is intended to affix the stomach to the anterior abdominal wall facilitating the initial percutaneous placement of a gastrostomy feeding tube in adults and adolescents of sufficient size ("Transitional Adolescent B" patients 18 to 21 years of age with no special considerations compared to adults).

## Device Story

PUMA-G System aids percutaneous gastrostomy tube placement; replaces traditional endoscopic or fluoroscopic visualization with ultrasound. System components: balloon catheter with internal magnet, guidewire, external magnet. Procedure: internal magnet placed orogastrically into stomach; external magnet pulls stomach to anterior abdominal wall; balloon inflated; ultrasound used to visualize tract and guide needle placement; guidewire inserted and snared by balloon; paired unit withdrawn through mouth; gastrostomy tube placed via Seldinger technique. Used in clinical settings by physicians. Benefits: ultrasound-guided placement; avoids radiation exposure associated with fluoroscopy.

## Clinical Evidence

Bench testing only. Testing included magnetic force characterization, coupling strength, and functional performance to verify the system's ability to affix the stomach to the abdominal wall and capture the guidewire. No clinical data presented.

## Technological Characteristics

Materials: ISO 10993 compliant. Energy: Magnetic attraction. Sterilization: Ethylene Oxide (ISO 11135). Components: Balloon catheter with internal magnet, external magnet, guidewire. Connectivity: None (standalone).

## Regulatory Identification

A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).

## Special Controls

*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

## Predicate Devices

- PUMA-G System ([K183057](/device/K183057.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.

March 29, 2023

CoapTech Inc. Jack Kent Chief Commercialization Officer 101 W. Dickman Street, Suite 700 Baltimore, MD 21230

> K223916 Trade/Device Name: PUMA-G System Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: Class II Product Code: KGC Dated: December 28, 2022 Received: December 29, 2022

Dear Jack Kent:

Re:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the 

{1}------------------------------------------------

Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medicaldevice-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@tda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Sivakami Venkatachalam -S

for

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K223916

Device Name PUMA-G System

Indications for Use (Describe)

The CoapTech PUMA-G System (PUMA-G) is intended to affix the stomach to the anterior abdominal wall facilitating the initial percutaneous placement of a gastrostomy feeding tube in adults and adolescents of sufficient size ("Transitional Adolescent B" patients 18 to 21 years of age with no special considerations compared to adults).

| Type of Use (Select one or both, as applicable)                                                                                                                                                                                                                             |                                                                                     |                                                                                    |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <table border="0"><tr><td><div style="text-align: left;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</div></td></tr><tr><td><div style="text-align: left;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</div></td></tr></table> </span> | <div style="text-align: left;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="text-align: left;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
| <div style="text-align: left;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</div>                                                                                                                                                                                         |                                                                                     |                                                                                    |
| <div style="text-align: left;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</div>                                                                                                                                                                                          |                                                                                     |                                                                                    |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a logo for a company called COAPTECH. The logo features the company name in a bold, sans-serif font, with the letters stacked vertically. Above the name is a stylized graphic that resembles two magnets facing each other. The magnets are blue with green ends.

# 510(k) Summary

(per 21 CFR 807.92)

# Submitter Information

| Name           | CoapTech Inc.                                           |
|----------------|---------------------------------------------------------|
| Address        | 101 W. Dickman Street, Suite 700<br>Baltimore, MD 21230 |
| Phone Number   | (443) 574-6981                                          |
| Contact Person | Jack Kent, MBA MPH<br>Chief Commercialization Officer   |
| Date Prepared  | March 18, 2023                                          |

# Device Information

| Trade Name     | PUMA-G System                                                               |
|----------------|-----------------------------------------------------------------------------|
| Common Name    | Gastrostomy Aid                                                             |
| Classification | Gastrointestinal tube and accessories<br>21 CFR 876.5980 (Product Code KGC) |

# Predicate Device Information

| Device Name | PUMA-G System |
|-------------|---------------|
| K Number    | K183057       |

## Device Description

The PUMA-G System is a medical device designed to aid in the initial placement of permanent feeding tubes, which are placed via percutaneous gastrostomy. Historically, gastrostomy tubes have been placed using either endoscopy or radiography (e.g., fluoroscopy) for visualization. The PUMA-G System provides visualization by ultrasound, allowing the user to assess the future gastrostomy tract prior to placement. The PUMA-G System contains three main custom components: a balloon catheter, a guidewire, and an external magnet. An internal magnet within the balloon catheter is placed orogastrically into the stomach, where it is pulled up to the anterior abdominal wall by the external magnet. After filling the balloon with fluid, the user's existing ultrasound is used for visualization of the future gastrostomy tract and for ultrasound-guided needle placement. The guidewire is then inserted and snared by the balloon, and the paired unit (balloon catheter and guidewire) is then pulled out the mouth. A gastrostomy tube is then placed using the guidewire and Seldinger technique.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for COAPTECH. The logo features a stylized, interconnected blue design resembling two magnets facing each other, with green squares at the ends. Below the design, the word "COAPTECH" is written in a bold, sans-serif font, also in blue.

### Intended Use

To Aid in Percutaneous Access to the Stomach During Gastrostomy Tube Placement

# Indications for Use

The CoapTech PUMA-G System (PUMA-G) is intended to affix the stomach to the anterior abdominal wall facilitating the initial percutaneous placement of a gastrostomy feeding tube in adults and adolescents of sufficient size ("Transitional Adolescent B" patients 18 to 21 years of age with no special considerations compared to adults).

### Comparison to Predicate

The PUMA-G System indications for use includes adolescents of sufficient size, meaning patients that are similarly sized to adults.

# Technological Characteristics

As shown in Table 1, many technological similarities exist between the subject and predicate devices. The magnetic force was characterized in performance testing to ensure it can adequately create temporary affixture of the tissue (effectiveness) without harming the tissue (safety). Other differences in technological characteristics are minor and/or common to other gastrostomy tube placement devices (e.g., endoscopy).

| Characteristic          | Subject Device<br>PUMA-G System                                         | Predicate Device (K183057)<br>PUMA-G System                                                               | Comparison                                                                                                            |
|-------------------------|-------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Mechanism (General)     | Mechanical (affix stomach to<br>anterior abdominal wall)                | Mechanical (affix stomach to<br>anterior abdominal wall)                                                  | Same                                                                                                                  |
| Mechanism (Detail)      | Magnetic Attraction                                                     | Magnetic Attraction                                                                                       | Same                                                                                                                  |
| Time                    | < 10 Minutes                                                            | < 10 Minutes                                                                                              | Same                                                                                                                  |
| Materials               | ISO 10993 Compatible                                                    | ISO 10993 Compatible                                                                                      | Same                                                                                                                  |
| Sterilization           | Ethylene Oxide<br>ISO 11135 Validated                                   | Ethylene Oxide<br>ISO 11135 Validated                                                                     | Same                                                                                                                  |
| Procedural Requirements | Orogastric<br>Over-the-Wire and Pull Tubes<br>Ultrasound Visualization  | Orogastric<br>Over-the-Wire Tubes<br>Ultrasound Visualization                                             | Minor Differences<br>No Impact on SE Decision                                                                         |
| Performance Testing     | Functional Testing<br>Magnetic Characterization<br>Post-Market Analysis | Biocompatibility<br>Sterilization<br>Functional Testing<br>Magnetic Characterization<br>GLP Animal Safety | Equivalent<br>Performance Data Results<br>Show Do Not Raise New or<br>Different Issues of Safety and<br>Effectiveness |

#### Table 1: Comparison of Technological Characteristics

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for COAPTECH. The logo features a stylized design resembling two magnets facing each other, with blue curved shapes forming the core and green rectangles at the ends representing the magnetic poles. Below the symbol, the company name "COAPTECH" is written in a blue, sans-serif font.

# Performance Data

Bench performance data were collected to support a substantial equivalence determination. This testing included magnetic force characterization and coupling strength information. Performance test results demonstrate reasonable assurance that the PUMA-G System can effectively bring together the stomach and anterior abdominal wall while maintaining healthy tissue. The PUMA-G System also reliably captures and retains the guidewire for eventual completion of the gastrostomy tube placement. These data suggest the PUMA-G System is as safe and as effective as the identified predicate device.

## Conclusions

The PUMA-G System and predicate device have the same intended use and similar indications, both serving as medical devices to aid in the placement of gastrostomy tubes. Based upon analysis and valid scientific evidence, reasonable assurance of safety and effectiveness is apparent, therefore concluding that the PUMA-G System is substantially equivalent to its predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K223916](https://fda-staging.innolitics.com/device/K223916)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).