← Product Code [LZA](/productcode/LZA) · K223895

# Disposable Medical Nitrile Examination Gloves (K223895)

_Hebei Titans Hongsen Medical Technology Co., Ltd. · LZA · Mar 23, 2023 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K223895

## Device Facts

- **Applicant:** Hebei Titans Hongsen Medical Technology Co., Ltd.
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Mar 23, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

Disposable Medical Nitrile Examination Gloves are disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

## Device Story

Disposable medical nitrile examination gloves; single-use; non-sterile; ambidextrous; orange color. Worn by healthcare examiners to provide barrier protection against contamination between patient and examiner. Manufactured from nitrile latex. Device meets ASTM D6319-19 specifications. Used in clinical settings; no specialized operator required. Provides physical barrier; prevents cross-contamination. Benefits patient and clinician by reducing exposure to contaminants during medical examinations.

## Clinical Evidence

No clinical data. Bench testing only. Testing included physical dimensions, physical properties (tensile strength, ultimate elongation), freedom from holes (ASTM D5151), residual powder (ASTM D6124), and biocompatibility (ISO 10993-10, 10993-11, 10993-23). All tests met acceptance criteria.

## Technological Characteristics

Material: Nitrile latex. Form factor: Ambidextrous, powder-free, orange color. Dimensions: Length 230mm; Width 70-120mm; Thickness 0.07-0.1mm. Standards: ASTM D6319-19 (specifications), ASTM D6124-06 (residual powder), ASTM D5151-06 (holes). Biocompatibility: ISO 10993-10, 10993-11, 10993-23. Sterilization: Non-sterile.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Medical Examination Gloves (Nitrile) (XS, S, M, L, XL) ([K222498](/device/K222498.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 23, 2023

Hebei Titans Hongsen Medical Technology Co., Ltd. % Ray Wang General Manager Beijing Believe Technology Service Co., Ltd. Rm. 912 Building #15. XiYueHui. No.5. YiHe North Rd., FangShan District Beijing, Beijing 102401 China

Re: K223895

Trade/Device Name: Disposable Medical Nitrile Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: December 28, 2022 Received: December 28, 2022

Dear Ray Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Bifeng Qian -S

Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K223895

Device Name

Disposable Medical Nitrile Examination Gloves

Indications for Use (Describe)

Disposable Medical Nitrile Examination Gloves are disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## K223895 510(k) Summary

- 1. Date of Preparation: 2023/03/22
- 2. Submitter

#### Hebei Titans Hongsen Medical Technology Co., Ltd.

Eastern Industrial Zone, Nangong City, Xingtai City, 051800, Hebei, China Contact Person: Xiaojing Zhao Position: Office clerk Tel: +86-13383292209 Email: sara@titans-cn.com

- 3. Submission Correspondent
#### Beijing Believe-Med Technology Service Co., Ltd.

Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing, China,102401 Contact Person: Ray Wang Position: General Manager Tel: +86-18910677558 Fax: +86-10-56335780 Email: information@believe-med.com

- 4. Subject Device Identification
510(k) Number: K223895

Trade Name: Disposable Medical Nitrile Examination Gloves

Common Name: Polymer Patient Examination Glove

Regulatory Information: Classification: I Product Code: LZA Regulation Number: 21 CFR 880.6250 Review Panel: General Hospital

{4}------------------------------------------------

- న్. Predicate Device Identification
510(k) Number: K222498

Product Name: Medical Examination Gloves (Nitrile) (XS, S, M, L, XL)

Manufacturer: Humanwell Healthcare Group Medical Supplies Co., Ltd

Regulatory Information:

Classification: I

Product Code: LZA

Regulation Number: 21 CFR 880.6250

Review Panel: General Hospital

Common Name: Polymer Patient Examination Glove

- 6. Device Description
The Disposable Medical Nitrile Examination Gloves are manufactured to meet the all current specifications listed under the ASTM Specification D6319 - 19, Standard Specification for Nitrile Examination Gloves for Medical Application. Disposable Medical Nitrile Examination Gloves are mainly composed of nitrile latex. It is mainly suitable for hand protection during medical examination. These gloves are powder free.

The proposed device(s) are orange color and sold non-sterile and are intended to be single-use, disposable devices.

- Indication For Use Statement 7.
Disposable Medical Nitrile Examination Gloves are disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

- Technological Characteristic Comparison Summary 8.

| ITEM           | Proposed Device (K223895)<br>Disposable Medical Nitrile Examination<br>Gloves                                                                                                                           | Predicate Device (K222498)<br>Medical Examination Gloves (Nitrile) (XS,<br>S, M, L, XL)                                                                                                             | Remark |
|----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Product Code   | LZA                                                                                                                                                                                                     | LZA                                                                                                                                                                                                 | SAME   |
| Regulation No. | 21 CFR 880.6250                                                                                                                                                                                         | 21 CFR 880.6250                                                                                                                                                                                     | SAME   |
| Class          | I                                                                                                                                                                                                       | I                                                                                                                                                                                                   | SAME   |
| Intended Use   | Disposable Medical Nitrile Examination<br>Gloves are disposable device intended for<br>medical purpose that is worn on the<br>examiner's hand to prevent contamination<br>between patient and examiner. | The Medical Examination Glove (Nitrile) is<br>a disposable device intended for medical<br>purposes that is worn on the examiner's<br>hand to prevent contamination between<br>patient and examiner. | SAME   |
| Powder free    | Yes                                                                                                                                                                                                     | Yes                                                                                                                                                                                                 | SAME   |

#### Table 1 General Comparison

{5}------------------------------------------------

| Design feature | Ambidextrous                                                                                                                           | Ambidextrous                                                                                                                         | SAME |
|----------------|----------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|------|
| Material       | Nitrile                                                                                                                                | Nitrile                                                                                                                              | SAME |
| OTC use        | Yes                                                                                                                                    | Yes                                                                                                                                  | SAME |
| Sterility      | Non-sterile                                                                                                                            | Non-sterile                                                                                                                          | SAME |
| Use            | Single use                                                                                                                             | Single use                                                                                                                           | SAME |
| Label          | Single-use, indication, powder free, device color, device name, glove size and quantity, Nitrile Glove Powder Free orange, Non-Sterile | Single-use, indication, powder free, device color, device name, glove size and quantity, Nitrile Glove Powder Free Blue, Non-Sterile | SAME |

Table 2 Device Dimensions Comparison

| Proposed Device (K223895)<br>Disposable Medical Nitrile<br>Examination Gloves           | Designation    | Size        |          |          |          |           | Tolerance  |
|-----------------------------------------------------------------------------------------|----------------|-------------|----------|----------|----------|-----------|------------|
|                                                                                         | Length, mm     | XS<br>230   | S<br>230 | M<br>230 | L<br>230 | XL<br>230 | min        |
|                                                                                         | Width, mm      | 70          | 80       | 95       | 110      | 120       | $\pm$ 10   |
|                                                                                         | Thickness, mm: |             |          |          |          |           |            |
|                                                                                         | Finger         | 0.1         |          |          |          |           | $\pm$ 0.03 |
|                                                                                         | Palm           | 0.08        |          |          |          |           | $\pm$ 0.03 |
|                                                                                         | Cuff           | 0.07        |          |          |          |           | $\pm$ 0.03 |
| Predicate Device (K222498)<br>Medical Examination Gloves<br>(Nitrile) (XS, S, M, L, XL) | Designation    | Size        |          |          |          |           | Tolerance  |
|                                                                                         | Length, mm     | XS<br>220   | S<br>220 | M<br>230 | L<br>230 | XL<br>230 | min        |
|                                                                                         | Width, mm      | 70          | 80       | 95       | 110      | 120       | $\pm$ 10   |
|                                                                                         | Thickness, mm: |             |          |          |          |           |            |
|                                                                                         | Finger         | 0.05        |          |          |          |           | min        |
|                                                                                         | Palm           | 0.05        |          |          |          |           | min        |
|                                                                                         | Cuff           | unavailable |          |          |          |           | min        |
| Remark                                                                                  | Similar        |             |          |          |          |           |            |

#### Analysis:

The proposed device has different size specification to the predicate device, but all proposed devices are meet the specifications of ASTM D 6319. So the difference does not affect the substantial equivalence in effectiveness and safety.

| ITEM                   | Proposed Device (K223895)<br>Disposable Medical Nitrile Examination Gloves | Predicate Device (K222498)<br>Medical Examination Gloves (Nitrile) (XS, S, M, L, XL)                         | Remark                                                                                                       |      |
|------------------------|----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|------|
| Colorant               | Orange                                                                     | Blue                                                                                                         | Different                                                                                                    |      |
| Physical<br>Properties | Before Aging<br>Tensile Strength                                           | 14 MPa, min                                                                                                  | 14 MPa, min                                                                                                  | SAME |
|                        | Ultimate Elongation                                                        | 500 % min                                                                                                    | 500 % min                                                                                                    | SAME |
|                        | After Aging<br>Tensile Strength                                            | 14 MPa, min                                                                                                  | 14 MPa, min                                                                                                  | SAME |
|                        | Ultimate Elongation                                                        | 400 % min                                                                                                    | 400 % min                                                                                                    | SAME |
| Freedom from Holes     |                                                                            | Comply with ASTM D6319<br>Be free from holes when tested<br>in accordance with ASTM<br>D5151<br>G-1, AQL 1.5 | Comply with ASTM D6319<br>Be free from holes when tested<br>in accordance with ASTM<br>D5151<br>G-1, AQL 1.5 | SAME |
| Residual Powder        |                                                                            | Meet the<br>requirements of<br>ASTM D6124                                                                    | Meet the<br>requirements of<br>ASTM D6124                                                                    | SAME |

Table 3 Performance Comparison

#### Analysis:

The subject device has different colour to the predicate device. The colour of the proposed device is orange and predicate device is Blue. But the Biocompatibility testing was successfully completed for the subject device, demonstrating that the difference does not affect the substantial equivalence in effectiveness and safety.

{6}------------------------------------------------

| ITEM               | Proposed Device (K223895)<br>Disposable Medical Nitrile<br>Examination Gloves                                                             | Predicate Device (K222498)<br>Medical Examination Gloves<br>(Nitrile) (XS, S, M, L, XL)                  | Remark |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|--------|
| Material           | Nitrile                                                                                                                                   | Nitrile                                                                                                  | SAME   |
| Biocompatibility   | Irritation<br>Under the conditions of the<br>study, not an irritant                                                                       | Under the conditions of the<br>study, not an irritant                                                    | SAME   |
|                    | Sensitization<br>Under conditions of the<br>study, not a sensitizer.                                                                      | Under conditions of the<br>study, not a sensitizer.                                                      | SAME   |
|                    | acute<br>systemic<br>toxicity<br>Under the conditions of the<br>study, there was no evidence<br>of systemic toxicity from the<br>extract. | Under the conditions of the<br>study, there was no evidence<br>of systemic toxicity from the<br>extract. | SAME   |
| Label and Labeling | Meet FDA's Requirements                                                                                                                   | Meet FDA's Requirements                                                                                  | SAME   |

#### Analysis:

We have conducted the testing according to the ISO10993-1, Acute Systemic Toxicity, irritation and sensitization. The test results shown that no biocompatibility risk would be raised. So, we consider that the proposed device has same biocompatibility performance with the predicate device.

#### Summary of Non-Clinical Testing 9.

Bench tests were conducted to demonstrate that the proposed device complies with the following standards:

ISO 10993-10: 2021 Biological Evaluation of Medical Devices - Part 10: Tests For Skin Sensitization.

ISO 10993-23: 2021 Biological evaluation of medical devices - Part 23: Tests for irritation ISO 10993-11 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application ASTM D6124-06 (R2017) Standard Test Method for Residual Powder on Medical Gloves ASTM D5151-06 Standard Test Method for Detection of Holes in Medical Gloves

| Title of the test                       | Purpose of the test                                                                                                                                           | Acceptance criteria and<br>the source of references                                                 | Results                                                                                                                                              |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Physical<br>Dimensions                  | Detection of the conformity<br>that whether the gloves<br>whether the dimension<br>requirements prescribed in<br>ASTM 6319-19 Table 2.                        | ASTM D6319-19<br>Standard Specification for<br>Nitrile Examination<br>Gloves.                       | The subject device complies<br>with the dimension<br>requirements prescribed in<br>ASTM D6319-19.                                                    |
| Physical<br>Property<br>Characteristics | Detection of the conformity that<br>whether the gloves meet the<br>requirements specified in ASTM<br>D6319-19 Table 3, before and<br>after accelerated aging. | ASTM D412 Standard Test<br>Methods for Vulcanized<br>Rubber and Thermoplastic<br>Elastomers-Tension | Before and after<br>accelerated aging, the<br>subject device complies<br>with the physical<br>requirements specified in<br>ASTM D6319-19 Table<br>3. |
| Freedom From<br>Holes                   | Detection of holes that allow<br>water leakage under the<br>conditions of the test.                                                                           | ASTM D5151-06,<br>Standard Test Method for<br>Detection of Holes in<br>Medical Gloves.              | No water leakage is<br>inspected.<br>The subject device<br>complies with the<br>requirements specified in<br>ASTM D5151.                             |
| Residual                                | Determine the amount of                                                                                                                                       | Residual Powder less                                                                                | Residual Powder less than                                                                                                                            |

#### Table 5 Bench Tests Summary

{7}------------------------------------------------

| Powder                        | residual powder found on a<br>sample of medical gloves.                                                                                                                                                                                                                                                         | than 2 mg per glove as<br>required in ASTM D<br>6124-06, R2017,<br>Standard Test Method for<br>Residual Powder on<br>Medical Gloves. | 2 mg per glove.<br>The subject device<br>complies with the<br>requirements specified in<br>ASTM D6124.             |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| Acute<br>Systemic<br>Toxicity | The test was designed to<br>evaluate the potential acute<br>system toxicity caused by test<br>article contact with the ICR<br>mice and extrapolating the<br>results to humans.                                                                                                                                  | ISO 10993-11 Biological<br>evaluation of medical<br>devices - Part 11: Tests<br>for systemic toxicity.                               | The test article has no<br>potential acute system<br>toxicity on ICR mice in<br>the extraction method.             |
| Skin Irritation               | To evaluate the potential skin<br>irritation caused by test article<br>contact with the skin surface<br>of rabbits and extrapolating<br>the results to humans, but it<br>does not establish the actual<br>risk of irritation.                                                                                   | ISO 10993-23: 2021<br>Biological evaluation of<br>medical devices - Part 23:<br>Tests for irritation.                                | Under the experimental<br>conditions, the test article<br>has no skin irritation on<br>rabbits.                    |
| Skin<br>Sensitization         | The test was designed to<br>evaluate the potential of a test<br>article to cause skin<br>sensitization. The test is used<br>as a procedure for screening<br>of contact allergens in guinea<br>pigs and extrapolating the<br>results to humans, but it does<br>not establish the actual risk of<br>sensitization | ISO 10993-10: 2021<br>Biological Evaluation Of<br>Medical Devices - Part<br>10: Tests For Skin<br>Sensitization.                     | The test article showed<br>no evidence of causing<br>delayed dermal contact<br>sensitization in the<br>guinea pig. |

## 10. Summary of Clinical Testing

Not applicable

### 11. Conclusions

 Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device, Medical Examination Gloves (Nitrile) (XS, S, M, L, XL), cleared under K222498.

---

**Source:** [https://fda-staging.innolitics.com/device/K223895](https://fda-staging.innolitics.com/device/K223895)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).