← Product Code [HIH](/productcode/HIH) · K223885

# HOPKINS Telescopes (K223885)

_Karl Storz Endoscopy America, Inc. · HIH · Sep 14, 2023 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K223885

## Device Facts

- **Applicant:** Karl Storz Endoscopy America, Inc.
- **Product Code:** [HIH](/productcode/HIH.md)
- **Decision Date:** Sep 14, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.1690
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Therapeutic

## Indications for Use

The HOPKINS Telescopes when used with sheaths are intended to be used to permit viewing of the cervical and uterine cavity for the purpose of performing diagnostic and surgical procedures.

## Device Story

HOPKINS Telescopes are rigid, rod-lens endoscopes used in hospital settings by physicians to visualize the cervical canal and uterine cavity. The device consists of a distal lens, a shaft containing optical glass rods for image transmission and magnification, and a proximal eyepiece. It is used in conjunction with sheaths to facilitate diagnostic and surgical procedures. The telescope provides optical images to the clinician, enabling direct visualization of the anatomy to guide clinical decision-making and surgical interventions. The device is available in various diameters (2mm, 2.9mm, 4mm), working lengths (26cm, 30cm, 36cm), and directions of view (0°, 12°, 30°).

## Clinical Evidence

No clinical data. Substantial equivalence was demonstrated through comparative bench testing against the predicate device, confirming adherence to design specifications and performance standards.

## Technological Characteristics

Rigid rod-lens endoscope; materials include surgical stainless steel. Available in 2mm, 2.9mm, 4mm diameters and 26cm-36cm lengths. Direction of view: 0°, 12°, 30°. External light source. Sterilization via steam (pre-vacuum), STERRAD (100NX, NX, 100S), or STERIS V-PRO. Complies with ISO 8600-1, 8600-3, 8600-5, 8600-6, and IEC 60601-2-18.

## Regulatory Identification

A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.

## Predicate Devices

- KARL STORZ Magnifying Hysteroscopes ([K935716](/device/K935716.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 14, 2023

Karl Storz Endoscopy America, Inc. Jordan Lydia Verla Senior Regulatory Affairs Specialist 2151 E. Grand Avenue El Segundo, CA 90245

Re: K223885

> Trade/Device Name: HOPKINS Telescopes Regulation Number: 21 CFR§ 884.1690 Regulation Name: Hysteroscope and Accessories Regulatory Class: II Product Code: HIH Dated: August 28, 2023 Received: August 29, 2023

Dear Jordan Lydia Verla:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Jason Roberts -S

Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K223885

Device Name HOPKINS Telescopes

Indications for Use (Describe)

The HOPKINS Telescopes when used with sheaths are intended to be used to permit viewing of the cervical and uterine cavity for the purpose of performing diagnostic and surgical procedures.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The logo is in blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the text "KARL STORZ - ENDOSKOPE" in a smaller font. 

## 7. 510(k) Summary

This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.

| Submitter:                                  | KARL STORZ Endoscopy-America, Inc.<br>2151 E. Grand Avenue<br>El Segundo, CA 90245                                                                                                                                                                                                                                                                                                                                                                                                                               |
|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:                                    | Jordan Lydia Verla<br>Senior Regulatory Affairs Specialist<br>Tel: (424) 218-8100 ext. 8382<br>Email: Jordan.Verla@karlstorz.com                                                                                                                                                                                                                                                                                                                                                                                 |
| Date of<br>Preparation:                     | September 11, 2023                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Type of 510(k)<br>Submission:               | Traditional                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Device<br>Identification:                   | Trade Name: HOPKINS Telescopes<br>Classification Name: Hysteroscope and accessories (21 CFR 884.1690)                                                                                                                                                                                                                                                                                                                                                                                                            |
| Regulatory Class:                           | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Product Code:                               | HIH                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Classification<br>Panel:                    | Obstetrics/Gynecology                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Predicate<br>Device(s):                     | KARL STORZ Magnifying Hysteroscopes (K935716)<br>The predicate device has not been subject to a design-related recall.                                                                                                                                                                                                                                                                                                                                                                                           |
| Device<br>Description:                      | The HOPKINS Telescopes are rigid telescopes that utilize the rod lens<br>technology. At the distal end of the telescope's shaft is the lens and the<br>other end of the shaft is attached to the eyepiece. Throughout the central<br>lumen of the HOPKINS Telescopes, optical glass rods are used to transmit<br>and magnify the image received from the lens. The HOPKINS Telescopes are<br>available with 0°, 12° and 30° direction of view, 2mm, 2.9mm, 4mm<br>diameter and 26cm, 30cm, 36cm working lengths. |
| Intended Use and<br>Indications for<br>Use: | Intended Use:<br>The HOPKINS Telescopes in conjunction with the sheath are intended to be<br>used to permit viewing of the cervical canal and uterine cavity for the<br>purpose of performing diagnostic and surgical procedures.                                                                                                                                                                                                                                                                                |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The logo is in blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the text "KARL STORZ - ENDOSKOPE" in a smaller font. 

| Technological<br>Characteristics: | Indications for Use:<br>The HOPKINS Telescopes when used with sheaths are intended to be used<br>to permit viewing of the cervical and uterine cavity for the purpose of<br>performing diagnostic and surgical procedures.<br>The subject and predicate devices have similar indications for use, operating<br>principle, and similar technological characteristics. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                   | HOPKINS Telescopes<br>Subject Device                                                                                                                                                                                                                                                                                                                                 | Magnifying Hysteroscopes<br>Predicate Device K935716                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   | Indications<br>for Use                                                                                                                                                                                                                                                                                                                                               | The HOPKINS Telescopes when<br>used with sheaths are intended to<br>be used to permit viewing of the<br>cervical and uterine cavity for the<br>purpose of performing diagnostic<br>and surgical procedures.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | I. Diagnostic Hysteroscopy<br>A. Indications for Use  Abnormal uterine<br>bleeding Infertility and pregnancy<br>wastage Evaluation of abnormal<br>hysterosalpingogram Intrauterine foreign body Pelvic pain              B. Absolute Contraindications<br>for Use  Acute Pelvic Inflammatory<br>Disease (PID)              C. Relative Contraindications<br>for Use  Inability to distend uterus Cervical/vaginal infection Uterine bleeding or<br>menses Known pregnancy Invasive carcinoma of the<br>cervix Recent uterine<br>perforation Medical contraindication Intolerance to anesthesia              D. Warnings  Suspicion of pregnancy<br>should suggest a |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | performance of diagnostic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | hysteroscopy.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   | E.                                                                                                                                                                                                                                                                                                                                                                   | Precautions                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | • Vaginal ultrasonography                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | prior to hysteroscopy may                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | identify clinical conditions                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | that will alter patient                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | management.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   | F.                                                                                                                                                                                                                                                                                                                                                                   | Operative Hysteroscopy                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | A.Indications for Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | • Directed biopsy                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | • Removal of submucuous                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | fibroids and large polyps                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | • Submucosous                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | myomectomy                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | • Transection of                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | intrauterine adhesions                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | • Transection of                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | intrauterine septa                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | • Endometrial ablation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | B. Absolute                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | Contraindications for Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | • Acute PID                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | C. Relative Contraindications                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | for Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | • Inability to distend uterus                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | • Cervical/vaginal infection                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | • Uterine bleeding or                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | menses                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                   |                                                                                                                                                                                                                                                                                                                                                                      | • Known pregnancy                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |…

---

**Source:** [https://fda-staging.innolitics.com/device/K223885](https://fda-staging.innolitics.com/device/K223885)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).