K223832 · Ge Medical System Ultrasound and Primary Care Diagnostics · IYN · Feb 16, 2023 · Radiology
Device Facts
Record ID
K223832
Device Name
Vivid S60N, Vivid S70N
Applicant
Ge Medical System Ultrasound and Primary Care Diagnostics
Product Code
IYN · Radiology
Decision Date
Feb 16, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Vivid S60N/Vivid S70N is a general-purpose ultrasound system, specialized for use in cardiac imaging. It is intended for use by, or under the direction of a qualified and trained physician for ultrasound imaging, measurement, display and analysis of the human body and fluid. The device is intended for use in a hospital environment including echo lab, other hospital settings, operating room, Cath lab and EP lab or in private medical offices. The systems support the following clinical applications: Fetal/Obsterics, Abdominal (including renal, GYN), Pediatric, Small Organ (breast, testes, thyroid), Neonatal Cephalic, Adult Cephalic, Cardiac (adult and pediatric), Peripheral Vascular, Musculo-skeletal Conventional, Musculo-skeletal Superficial, Urology (including prostate), Transvaginal, Transvaginal, Transrectal, Intra-cardiac and Intra-luminal, Interventional Guidance (including Biopsy, Vascular Access), Thoracic/Pleural, and Intraoperative (vascular). Modes of operation include 3D/4D Imaging mode, B, M, PW Doppler, Color Doppler, Color M Dopler, Power Doppler, Harmonic Imaging, Coded Pulse and Combined modes: B/M, B/Color M, B/PWD or CWD, B/ Color/PWD or CWD, B/Power/PWD.
Device Story
Vivid S60N/S70N is a mobile diagnostic ultrasound system for cardiac, vascular, and general radiology imaging. It acquires ultrasound data via electronic array transducers (linear, curved, sector/phased, matrix, dual, CW, and 3D/4D) and ICE catheters. The system processes signals, performs software-based analysis, and displays images on an integrated LCD monitor. Operated by trained healthcare professionals in clinical settings (OR, Cath lab, EP lab, echo lab, private office), the device supports digital acquisition, storage, and network connectivity (LAN/wireless). It now integrates with the Carto 3 EP Navigation System for digital video streaming of 2D/4D ICE data. Output assists clinicians in diagnostic assessment and interventional guidance, potentially improving patient management through real-time visualization and measurement.
Clinical Evidence
No clinical data required; substantial equivalence supported by bench testing, including acoustic output, biocompatibility, cleaning/disinfection effectiveness, and thermal/electrical/electromagnetic/mechanical safety testing.
Technological Characteristics
Mobile console with LCD touch panel; supports linear, curved, phased, matrix, and dual array transducers. Connectivity via LAN/wireless (USB adapter). Complies with AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1, IEC 62359, ISO 14971, and NEMA PS 3.1-3.20 (DICOM).
Indications for Use
Indicated for ultrasound imaging, measurement, and analysis of human body and fluid in patients requiring cardiac, fetal/obstetric, abdominal, pediatric, small organ, neonatal/adult cephalic, vascular, musculoskeletal, urological, transvaginal/transrectal, intra-cardiac/intra-luminal, and interventional procedures. Used by or under direction of qualified physicians in hospitals or private offices.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
CARTO 3 EP Navigation System with CARTOSOUND 4D Version V7.4 (K223733)
Submission Summary (Full Text)
{0}------------------------------------------------
February 16, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
GE Medical System Ultrasound and Primary Care Diagnostics % Lee Bush Regulatory Affairs Director 9900 W. Innovation Drive WAUWATOSA WI 53226
Re: K223832
Trade/Device Name: Vivid S60N / Vivid S70N Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: December 21, 2022 Received: December 22, 2022
Dear Lee Bush:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Yanna S. Kang -S
Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K223832
Device Name Vivid S60N / Vivid S70N
#### Indications for Use (Describe)
Vivid S60N/Vivid S70N is a general-purpose ultrasound system, specialized for use in cardiac imaging. It is intended for use by, or under the direction of a qualified and trained physician for ultrasound imaging, measurement, display and analysis of the human body and fluid. The device is intended for use in a hospital environment including echo lab, other hospital settings, operating room, Cath lab and EP lab or in private medical offices. The systems support the following clinical applications: Fetal/Obsterics, Abdominal (including renal, GYN), Pediatric, Small Organ (breast, testes, thyroid), Neonatal Cephalic, Adult Cephalic, Cardiac (adult and pediatric), Peripheral Vascular, Musculo-skeletal Conventional, Musculo-skeletal Superficial, Urology (including prostate), Transvaginal, Transvaginal, Transrectal, Intra-cardiac and Intra-luminal, Interventional Guidance (including Biopsy, Vascular Access), Thoracic/Pleural, and Intraoperative (vascular). Modes of operation include 3D/4D Imaging mode, B, M, PW Doppler, Color Doppler, Color M Dopler, Power Doppler, Harmonic Imaging, Coded Pulse and Combined modes: B/M, B/Color M, B/PWD or CWD, B/ Color/PWD or CWD, B/Power/PWD.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size:12pt;"> <span style="font-family:Arial;"> </span> </span> </div> <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:12pt;"> <span style="font-family:Arial;"> </span> </span> </div> <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a blue circle. Three droplet-like shapes surround the circle, giving the impression of motion or energy. The logo is simple, recognizable, and associated with a well-known multinational corporation.
# 510(k) Summary
# K223832
| Date:<br>Submitter: | In accordance with 21 CFR 807.92 the following summary of information is provided:<br>December 21, 2022<br>GE Medical Systems Ultrasound and Primary Care Diagnostics<br>9900 Innovation Drive<br>Wauwatosa, WI 53226 |
|----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary Contact Person: | Lee Bush<br>Regulatory Affairs Director<br>GE Healthcare<br>T:(262)309-9429 |
| Secondary Contact Person: | Nadine Gebhard- Höflinger<br>Regulatory Affairs Leader<br>GE Healthcare |
| Device<br>Trade Name:<br>Common/Usual Name:<br>Classification Names: | Vivid S60N / Vivid S70N<br>Diagnostic Ultrasound System<br>Class II |
| Product Code(s): | Ultrasonic Pulsed Doppler Imaging System, 21CFR 892.1550, 90-IYN;<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO;<br>Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX |
| Primary Predicate Device: | Vivid S60N, Vivid S70N (K220619), Diagnostic Ultrasound System |
| Classification Names: | Class II |
| Product Code(s): | Ultrasonic Pulsed Doppler Imaging System, 21CFR 892.1550, 90-IYN;<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO;<br>Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX |
| Reference Device: | NUVISION TM NAV (K223766), Diagnostic Intravascular Catheter,<br>Ultrasound |
| Classification Names: | Class II |
| Product Code(s): | OBJ |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized script in the center. The circle is surrounded by a white, swirling design that resembles water or wind. The logo is simple and recognizable, and it is often used to represent the company's history and innovation.
Reference device: CARTO 3 EP Navigation System with CARTOSOUND 4D Version V7.4 (K223733), Programable Diagnostic Computer
Classification Names: Class II
Product Code(s): DQK
## Device Description:
Vivid S60N / Vivid S70N is a Track 3, diagnostic ultrasound system for use by qualified and trained healthcare professionals, which is primarily intended for cardiac imaging and analysis but also includes vascular and general radiology applications. It is a full featured diagnostic ultrasound system that provides digital acquisition, processing, analysis and display capability.
The Vivid S60N / Vivid S70N consists of a mobile console with a height-adjustable control panel, color LCD touch panel. LCD display monitor and optional image storage and printing devices. It includes a variety of electronic array transducers operating in linear, curved, sector/phased array, matrix array or dual array format, including dedicated CW transducers and real time 3D transducer. System can also be used with compatible ICE (Intra Cardiac Echocardiography) transducers.
The system includes electronics for transmit and receive of ultrasound data, ultrasound signal processing, software computing, hardware for Image storage, hard copy printing, and network access to the facility through both LAN and wireless (supported by use of a wireless LAN USB-adapter) connection.
## Intended Use/Indication for Use:
Vivid S60N/Vivid S70N is a general-purpose ultrasound system, specialized for use in cardiac imaging. It is intended for use by, or under the direction of a qualified and trained physician for ultrasound imaging, measurement, display and analysis of the human body and fluid. The device is intended for use in a hospital environment including echo lab, other hospital settings, operating room, Cath lab and EP lab or in private medical offices. The systems support the following clinical applications: Fetal/Obstetrics, Abdominal (including renal, GYN), Pediatric, Small Organ (breast, testes, thyroid), Neonatal Cephalic, Adult Cephalic, Cardiac (adult and pediatric), Peripheral Vascular, Musculoskeletal Conventional, Musculo-skeletal Superficial, Urology (including prostate), Transesophageal, Transvaginal, Transrectal, Intra-cardiac and Intra-luminal, Interventional Guidance (including Biopsy, Vascular Access), Thoracic/Pleural, and Intraoperative (vascular). Modes of operation include: 3D/4D Imaging mode, B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse and Combined modes: B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" written in a stylized script in the center. The circle is surrounded by a series of small, teardrop-shaped elements, also in blue. The logo is simple and recognizable, and it is often used to represent the company's brand.
#### Technology:
The Vivid S60N/Vivid S70N employs the same fundamental scientific technology as its predicate device.
Determination of Substantial Equivalence:
The proposed Vivid S60N/Vivid S70N systems are substantially equivalent to the predicate Vivid S60N/ S70N and reference devices with regards to intended use, indications for use, imaging capabilities, technological characteristics, imaging modes, hardware, and safety and effectiveness.
The following is an overview of the differences between the proposed Vivid S60N / Vivid S70N and its predicate and reference devices.
Features/Functionality:
- Addition of Carto 3 Communication Options (K223733) ● - Digital Video Streaming for 2D, CF, MP, 4D modes (both 2D ICE and 4D ICE) with OEM (Original Equipment Manufacturer) Carto 3 EP Navigation System with CARTOSOUND 4D Version V7.4 in addition to previously available analog/VGA output.
Transducers and Modes:
- Addition of OEM ICE catheter (NUVISION™ NAV from Biosense Webster, Inc., . K223766) to the compatible accessories list. The clinical indications and modes are the same on the proposed Vivid S60N/S70N device as they are on the predicate Vivid S60N/S70N device which already included 4D ICE catheter support.
Instruction for Use:
- . Addition of addendum for added OEM ICE catheter
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circle. The circle is surrounded by three curved lines that resemble water droplets or stylized flames. The logo is presented in a light blue color against a white background.
## Summary of Non-Clinical Tests:
Vivid S60N / Vivid S70N were evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic, and mechanical safety, and have been found to comply with applicable medical device safety standards. The Vivid S60N / Vivid S70N complies with voluntary standards:
- AAMI/ANSI ES60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety and Essential Performance, 2005/ A2:2012
- IEC 60601-1-2, Medical Electrical Equipment Part 1-2: General Requirements ● for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbance - Requirements and Tests, 2014
- IEC 60601-2-37, Medical Electrical Equipment - Part 2-37: Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment, Ed. 2.1, 2015
- ISO 10993-1, Biological Evaluation of Medical Devices- Part 1: Evaluation and Testing Within A Risk Management Process, 2009
- IEC 62359, Ultrasonics Field characterization Test methods for the ● determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields, 2017
- ISO 14971, Application of risk management to medical devices, 2019
- . NEMA PS 3.1 - 3.20, Digital Imaging and Communications in Medicine (DICOM) Set. (Radiology), 2016
The following quality assurance measures are applied to the development of the system:
- Risk Analvsis
- Requirements Reviews
- Design Reviews
- Testing on unit level (Module verification)
- Integration testing (System verification)
- Performance testing (Verification) ●
- . Safety testing (Verification)
Transducer material and other patient contact materials are biocompatible.
#### Summary of Clinical Tests:
The subject of this premarket submission, Vivid S60N and Vivid S70N, did not require clinical studies to support substantial equivalence.
## Conclusion:
GE Healthcare considers the Vivid S60N / Vivid S70N to be as safe, as effective, and performance is substantially equivalent to the primary predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.