POLARSHEATH™ Steerable Sheath 12F (UPN:M004CRBS3150), POLARMAP™ Circular Mapping Catheter (UPN:M004CRBS7210), POLARMAP™ EP Electrical Cable (UPN:M004CRBS62000), SMARTFREEZE™ Esophageal Temperature Sensor (ETS) Cable: -SMARTFREEZE™ Esophageal Temperature Sensor (ETS) Cable (UPN: M004CRBS6310), -SMARTFREEZE™ 16 PIN Esophageal Temperature Sensor (ETS) Cable (UPN: M004CRBS6320), -SMARTFREEZE™ CIRCA Esophageal Temperature Sensor (ETS) Cable (UPN: M004CRBS6340), SMARTFREEZE™ Pressure Sensor
K223824 · Boston Scientific Corporation · DRA · Mar 22, 2023 · Cardiovascular
Device Facts
Record ID
K223824
Device Name
POLARSHEATH™ Steerable Sheath 12F (UPN:M004CRBS3150), POLARMAP™ Circular Mapping Catheter (UPN:M004CRBS7210), POLARMAP™ EP Electrical Cable (UPN:M004CRBS62000), SMARTFREEZE™ Esophageal Temperature Sensor (ETS) Cable: -SMARTFREEZE™ Esophageal Temperature Sensor (ETS) Cable (UPN: M004CRBS6310), -SMARTFREEZE™ 16 PIN Esophageal Temperature Sensor (ETS) Cable (UPN: M004CRBS6320), -SMARTFREEZE™ CIRCA Esophageal Temperature Sensor (ETS) Cable (UPN: M004CRBS6340), SMARTFREEZE™ Pressure Sensor
Applicant
Boston Scientific Corporation
Product Code
DRA · Cardiovascular
Decision Date
Mar 22, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1280
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
POLARSHEATH Steerable Sheath 12F: The POLARSHEATH steerable sheath is indicated for percutaneous catheter introduction into the vasculature and into the chambers of the heart. POLARMAP Circular Mapping Catheter: The POLARMAP Catheter is indicated for electrophysiological mapping (recording or stimulating only) of the cardiac structures of the heart. POLARMAP EP Electrical Cable: The EP Electrical Cable is designed for use with the POLARMAP Mapping Catheter and the hospital EP recording system. The EP Electrical Cable connects the POLARMAP Mapping Catheter to the hospital EP recording system. Use of the EP Electrical Cable is optional. SMARTFREEZE ETS Cables: The Esophageal Temperature Sensor (ETS) Cable (model M004CRBS6320) is designed for use with the SMARTFREEZE Console and a general purpose series 400 temperature sensor. The ETS Cable is used to connect a general purpose series 400 temperature sensor to the ICB. Use of the ETS Cable is optional. The ETS Cable (CIRCA) (model M004CRBS6340) is designed for use with the SMARTFREEZE Console and the CIRCA S-CATH™ Esophageal Temperature Probe. The ETS Cable (CIRCA) is used to connect the CIRCA S-CATH™ Esophageal Temperature Probe to the ICB (Model M004CRBS4130 only). Use of the ETS Cable (CIRCA) is optional. NOTE: The role of esophageal temperature monitoring using the risk of cardiac cryoablationrelated esophageal injury has not been established. The performance of the SMARTFREEZE ETS Cables and compatible temperature probe in detecting esophageal temperature changes as a result of energy delivery during cardiac cryoablation procedures has not been evaluated. SMARTFREEZE Pressure Sensor Cable: The Pressure Sensor Cable is designed for use with the SMARTFREEZE Console and the cryoablation balloon catheter. The Pressure Sensor Cable is used to connect an intravascular pressure transducer to the ICB to measure the ventricular pressure during ablation procedures to aid in determining vein occlusion. Use of the Pressure Sensor Cable is optional.
Device Story
System includes steerable sheath, circular mapping catheter, electrical cables, and sensor cables for cryoablation procedures. Sheath facilitates percutaneous access to cardiac chambers; mapping catheter records intracardiac electrograms and provides pacing stimulation. Cables connect mapping catheter to hospital EP recording systems, or connect temperature/pressure sensors to SMARTFREEZE console/ICB. Used in clinical electrophysiology labs by physicians. Pressure sensor cable aids in determining vein occlusion by measuring ventricular pressure; temperature cables monitor esophageal temperature during ablation. Devices are single-use (except reusable cables) and provided sterile. System provides mechanical access and signal transmission to support cardiac mapping and ablation monitoring.
Clinical Evidence
Bench testing only. Evidence includes sterilization validation, EO residual assessment, biocompatibility testing, electrical safety (IEC 60601-1, IEC 60601-1-2), design verification (nominal and aged timepoints), and design validation confirming conformance to user needs.
Technological Characteristics
Materials: Pebax, stainless steel, PTFE, silicone rubber, polycarbonate, polyethylene, Nitinol, PET. Components: Steerable sheath (unidirectional deflection), circular mapping catheter (8 ring electrodes), electrical/sensor cables. Connectivity: Wired connection to EP recording systems/consoles. Sterilization: Ethylene Oxide. Single-use (sheath/catheter) and reusable (cables).
Indications for Use
Indicated for percutaneous catheter introduction into vasculature/cardiac chambers (sheath); electrophysiological mapping (recording/stimulating) of cardiac structures (catheter).
Regulatory Classification
Identification
A steerable catheter is a catheter used for diagnostic and monitoring purposes whose movements are directed by a steering control unit.
CIRCA S-CATH M Esophageal Temperature Probe and Temperature Monitoring System (K200943)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 22, 2023
Boston Scientific Corporation Derek Charchuk Senior Regulatory Affairs Specialist 4100 Hamline Avenue North St. Paul, Minnesota 55112-5798
### Re: K223824
Trade/Device Name: POLARSHEATH™ Steerable Sheath 12F, POLARMAP™ Circular Mapping Catheter, POLARMAPIM EP Electrical Cable, SMARTFREEZE™ Esophageal Temperature Sensor (ETS) Cables (SMARTFREEZETM ETS Cable. SMARTFREEZE™ 16 PIN ETS Cable, SMARTFREEZE™ CIRCA ETS Cable), SMARTFREEZE™ Pressure Sensor Cable Regulation Number: 21 CFR 870.1280 Regulation Name: Steerable Catheter Regulatory Class: Class II Product Code: DRA, DRF Dated: December 20, 2022 Received: December 21, 2022
Dear Derek Charchuk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Aneesh S. Deoras -S
Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology. Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
### 510(k) Number (if known) K223824
#### Device Name
POLARSHEATH™ Steerable Sheath 12F, POLARMAP™ Circular Mapping Catheter, POLARMAP™ EP Electrical Cable, SMARTFREEZE™ Esophageal Temperature Sensor (ETS) Cables (SMARTFREEZE™ ETS Cable, SMARTFREEZE™ 16 PIN ETS Cable, SMARTFREEZE™ CIRCA ETS Cable), SMARTFREEZE™ Pressure Sensor Cable
#### Indications for Use (Describe)
POLARSHEATH Steerable Sheath 12F:
The POLARSHEATH steerable sheath is indicated for percutaneous catheter introduction into the vasculature and into the chambers of the heart.
#### POLARMAP Circular Mapping Catheter:
The POLARMAP Catheter is indicated for electrophysiological mapping (recording or stimulating only) of the cardiac structures of the heart.
### POLARMAP EP Electrical Cable:
The EP Electrical Cable is designed for use with the POLARMAP Mapping Catheter and the hospital EP recording system. The EP Electrical Cable connects the POLARMAP Mapping Catheter to the hospital EP recording system. Use of the EP Electrical Cable is optional.
#### SMARTFREEZE ETS Cables:
The Esophageal Temperature Sensor (ETS) Cable (model M004CRBS6320) is designed for use with the SMARTFREEZE Console and a general purpose series 400 temperature sensor. The ETS Cable is used to connect a general purpose series 400 temperature sensor to the ICB. Use of the ETS Cable is optional.
The ETS Cable (CIRCA) (model M004CRBS6340) is designed for use with the SMARTFREEZE Console and the CIRCA S-CATH™ Esophageal Temperature Probe. The ETS Cable (CIRCA) is used to connect the CIRCA S-CATH™ Esophageal Temperature Probe to the ICB (Model M004CRBS4130 only). Use of the ETS Cable (CIRCA) is optional.
NOTE: The role of esophageal temperature monitoring using the risk of cardiac cryoablationrelated esophageal injury has not been established. The performance of the SMARTFREEZE ETS Cables and compatible temperature probe in detecting esophageal temperature changes as a result of energy delivery during cardiac cryoablation procedures has not been evaluated.
### SMARTFREEZE Pressure Sensor Cable:
The Pressure Sensor Cable is designed for use with the SMARTFREEZE Console and the cryoablation balloon catheter. The Pressure Sensor Cable is used to connect an intravascular pressure transducer to the ICB to measure the ventricular pressure during ablation procedures to aid in determining vein occlusion. Use of the Pressure Sensor Cable is optional.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) SUMMARY
# 510(k) Summary in Compliance with 21 CFR 807.92
# I. SUBMITTER INFORMATION
Submitter Name: Boston Scientific Corporation
Submitter Address: 4100 Hamline Avenue North St. Paul, Minnesota USA 55112-5798
- Contact #1: Derek Charchuk Senior Regulatory Affairs Specialist Phone: (905) 696-1913 Email: derek.charchuk@bsci.com
- Contact #2: Jim Johnson Senior Regulatory Affairs Specialist Phone: (651) 582-5561 Email: jim.johnson(@bsci.com
Date Prepared: 19-December-2022
## II. DEVICE INFORMATION
Generic Name: Sheath, Mapping Catheter, EP Electrical Cable, Temperature Sensor Cable, Pressure Sensor Cable
Trade Name: POLARSHEATH™ Steerable Sheath 12F
POLARMAP™ Circular Mapping Catheter
POLARMAP™ EP Electrical Cable
SMARTFREEZE™ Esophageal Temperature Sensor (ETS) Cables:
- SMARTFREEZE™ ETS Cable .
- SMARTFREEZE™ 16 PIN ETS Cable ●
- . SMARTFREEZE™ CIRCA ETS Cable
SMARTFREEZE™ Pressure Sensor Cable
Product Codes:
DRA: POLARSHEATH™ Steerable Sheath 12F
DRF: POLARMAP™ Circular Mapping Catheter
POLARMAP™ EP Electrical Cable SMARTFREEZE™ Esophageal ETS Cables
{5}------------------------------------------------
Traditional 510(k) submission POLARSHEATH Steerable Sheath 12F, POLARMAP™ Circular Mapping Catheter, POLARMAP™ EP Electrical Cable, SMARTFREEZE™ Esophageal Temperature Sensor (ETS) Cables, and SMARTFREEZE™ Pressure Sensor Cable
SMARTFREEZE™ Pressure Sensor Cable
Device Class: Class II
Review Panel: Cardiovascular (DRA & DRF)
### Classification Regulations:
(21 CFR 870.1280): POLARSHEATH™ Steerable Sheath 12F
(21 CFR 870.1220): POLARMAPTM Circular Mapping Catheter POLARMAP™ EP Electrical Cable SMARTFREEZE™ ETS Cables SMARTFREEZE™ Pressure Sensor Cable
### III. PREDICATE DEVICE INFORMATION
### Predicate Device #1: Flexcath Advance Steerable Sheath (12 French)
Manufacturer: Medtronic Cryocath LP 510(k) #: K123591 FDA Clearance Date: 27 December 2012 Product Code: DRA Predicate Device for: POLARSHEATH™ Steerable Sheath 12F
### Predicate Device #2: Achieve Advance Mapping Catheter
Manufacturer: Medtronic Inc. 510(k) #: K162892 FDA Clearance Date: 15 November 2016 Product Code: DRF Predicate Device for: POLARMAP™ Circular Mapping Catheter, POLARMAP™ EP Electrical Cable, SMARTFREEZE™ Esophageal ETS Cables, and SMARTFREEZE™ Pressure Sensor Cable
### Reference Device: CIRCA S-CATH M Esophageal Temperature Probe and Temperature Monitoring System
Manufacturer: CIRCA Scientific, LLC 510(k) #: K200943 FDA Clearance Date: 17 December 2021 Product Code: FLL, IKD Reference Device for: SMARTFREEZE™ Esophageal ETS Cables
### IV. INDICATIONS FOR USE
POLARSHEATH™ Steerable Sheath 12F: The POLARSHEATH steerable sheath is indicated for percutaneous catheter introduction into the vasculature and into the chambers of the heart.
POLARMAP™ Circular Mapping Catheter: The POLARMAP Catheter is indicated for electrophysiological mapping (recording or stimulating only) of the cardiac structures of the heart.
{6}------------------------------------------------
Traditional 510(k) submission POLARSHEATH™ Steerable Sheath 12F, POLARMAP™ Circular Mapping Catheter, POLARMAP™ EP Electrical Cable, SMARTFREEZE™ Esophageal Temperature Sensor (ETS) Cables, and SMARTFREEZE™ Pressure Sensor Cable
POLARMAP™ EP Electrical Cable: The EP Electrical Cable is designed for use with the POLARMAP Mapping Catheter and the hospital EP recording system. The EP Electrical Cable connects the POLARMAP Mapping Catheter to the hospital EP recording system. Use of the EP Electrical Cable is optional.
SMARTFREEZE™ ETS Cables: The Esophageal Temperature Sensor (ETS) Cable (model M004CRBS6310 or M004CRBS6320) is designed for use with the SMARTFREEZE Console and a general purpose series 400 temperature sensor. The ETS Cable is used to connect a general purpose series 400 temperature sensor to the ICB. Use of the ETS Cable is optional.
The ETS Cable (CIRCA) (model M004CRBS6340) is designed for use with the SMARTFREEZE Console and the CIRCA S-CATH™ Esophageal Temperature Probe. The ETS Cable (CIRCA) is used to connect the CIRCA S-CATH™ Esophageal Temperature Probe to the ICB (Model M004CRBS4130 only). Use of the ETS Cable (CIRCA) is optional.
NOTE: The role of esophageal temperature monitoring using this device in reducing the risk of cardiac cryoablation-related esophageal injury has not been established. The performance of the SMARTFREEZE ETS Cables and compatible temperature probe in detecting esophageal temperature changes as a result of energy delivery during cardiac cryoablation procedures has not been evaluated.
SMARTFREEZE™ Pressure Sensor Cable: The Pressure Sensor Cable is designed for use with the SMARTFREEZE Console and the cryoablation balloon catheter. The Pressure Sensor Cable is used to connect an intravascular pressure transducer to the ICB to measure the ventricular pressure during ablation procedures to aid in determining vein occlusion. Use of the Pressure Sensor Cable is optional.
# V. DEVICE DESCRIPTION
The subject devices are described below.
| Device | UPN/Model | Description |
|---------------------------------------------------------------------|---------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| POLARSHEATH™<br>Steerable Sheath 12F | M004CRBS3150 | Single-use, steerable percutaneous introducer sheath designed<br>for additional maneuverability of diagnostic and therapeutic<br>catheters that are advanced through the POLARSHEATH™<br>Sheath and into cardiac chambers. |
| POLARMAP™ Circular<br>Mapping Catheter | M004CRBS7210 | Single-use, multi-electrode catheter designed to record<br>intracardiac electrograms and provide pacing stimulation during<br>electrophysiology procedures. |
| POLARMAP™ EP<br>Electrical Cable | M004CRBS62000 | Single-use electrical cable that connects the POLARMAP™<br>mapping catheter to the hospital EP recording system. |
| SMARTFREEZE™<br>Esophageal Temperature<br>Sensor (ETS) Cable | M004CRBS6310 | Re-usable cables used to connect a temperature probe to the<br>Inter Connection Box (ICB); used for continuous temperature<br>monitoring during ablation procedures (optional). |
| SMARTFREEZE™ 16 PIN<br>Esophageal Temperature<br>Sensor (ETS) Cable | M004CRBS6320 | Re-usable cables used to connect a temperature probe to the<br>Inter Connection Box (ICB); used for continuous temperature<br>monitoring during ablation procedures (optional). |
{7}------------------------------------------------
Traditional 510(k) submission POLARSHEATH™ Steerable Sheath 12F, POLARMAP™ Circular Mapping Catheter, POLARMAP™ EP Electrical Cable, SMARTFREEZE™ Esophageal Temperature Sensor (ETS) Cables, and SMARTFREEZE™ Pressure Sensor Cable
| SMARTFREEZE TM CIRCA<br>Esophageal Temperature<br>Sensor (ETS) Cable | M004CRBS6340 | |
|----------------------------------------------------------------------|--------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SMARTFREEZE TM Pressure<br>Sensor Cable | M004CRBS6600 | Re-usable cable that connects an intravascular pressure<br>transducer to the ICB. The pressure sensor is used to measure<br>the ventricular pressure during ablation procedures to aid in<br>determining vein occlusion. |
# VI. SUBSTANTIAL EQUIVALENCE & COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
Comparison of Subject POLARSHEATH™ Steerable Sheath 12F versus Predicate Flexcath Advance Steerable Sheath 12F (Medtronic)
| Characteristic | Predicate Device (K123591)<br>Flexcath Advance Steerable Sheath 12F<br>(Medtronic Cryocath LP) | Subject Device<br>POLARSHEATHTM Steerable Sheath 12F<br>(Boston Scientific Corporation) | |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|
| Indications for<br>Use | The FlexCath Advance Steerable<br>Sheath is intended for percutaneous<br>catheter introduction into the<br>vasculature and into the chambers of<br>the heart. | The POLARSHEATHTM steerable sheath<br>is indicated for percutaneous catheter<br>introduction into the vasculature and into<br>the chambers of the heart. | |
| Sheath Usable<br>Length | 65cm | 68cm | |
| Dilator Overall<br>Length | 87cm | 87cm | |
| Guidewire<br>Compatibility | 0.81mm and 0.89mm | 0.81mm and 0.89mm | |
| Radiopaque<br>Markers | Yes | Yes | |
| Distal End | Unidirectional deflection curve | Unidirectional deflection curve | |
| Shaft Design | Deflectable shaft with single lumen | Deflectable shaft with single lumen | |
| Hemostasis<br>Valve | Yes | Yes | |
| Compatible<br>Catheter Size | Facilitates use with catheters up to<br>10.5F in diameter | Facilitates use with catheters up to 11.8F<br>in diameter | |
| Materials | Pebax with stainless steel braid and<br>PTFE liner<br>Silicone Rubber<br>Polycarbonate<br>Polyethylene | Pebax with stainless steel braid and PTFE<br>liner<br>Silicone Rubber<br>Polycarbonate<br>Polyethylene | |
| | Packaging | Tyvek/nylon polyethylene pouch | Tyvek/nylon polyethylene pouch |
{8}------------------------------------------------
### Traditional 510(k) submission
POLARSHEATH™ Steerable Sheath 12F, POLARMAP™ Circular Mapping Catheter, POLARMAP™ EP Electrical Cable, SMARTFREEZE™ Esophageal Temperature Sensor (ETS) Cables, and SMARTFREEZE™ Pressure Sensor Cable
| Characteristic | Predicate Device (K123591)<br>Flexcath Advance Steerable Sheath 12F<br>(Medtronic Cryocath LP) | Subject Device<br>POLARSHEATHTM Steerable Sheath 12F<br>(Boston Scientific Corporation) |
|---------------------------|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
| Sterilization | Provided sterile (Ethylene Oxide) | Provided sterile (Ethylene Oxide) |
| Single Use or<br>Reusable | Single use | Single use |
### Comparison of Subject POLARMAP™ Circular Mapping Catheter versus Predicate Achieve Advance Mapping Catheter (Medtronic)
| Characteristic | Predicate Device (K162892)<br>Achieve Advance Mapping Catheter<br>(Medtronic, Inc.) | Subject Device<br>POLARMAP™ Circular Mapping Catheter<br>(Boston Scientific Corporation) |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The Achieve Advance mapping<br>catheter is indicated for multiple<br>electrode electrophysiological<br>mapping of the cardiac structures<br>of the heart, i.e. recording or<br>stimulation only. The Achieve<br>Advance mapping catheter is<br>designed to obtain electrograms<br>in the atrial regions of the heart. | The POLARMAP Catheter is indicated<br>for electrophysiological mapping<br>(recording or stimulating only) of the<br>cardiac structures of the heart. |
| Shaft Size | 3.3F (1.10mm) | 3.3F (1.10mm) |
| Effective Length | 146 cm | 149 cm |
| Ring Electrode<br>Width | 1 mm | 1 mm |
| Loop Diameter | 20 mm<br>(15mm and 25 mm also) | 20 mm |
| Number of Ring<br>Electrodes | 8 (20 mm loop diameter<br>configuration) | 8 |
| Distal End Shape | Circular Loop | Circular Loop |
| Steering | Non-steerable | Non-steerable |
| Materials | Pebax (distal body)<br>Polyimide (proximal)<br>Nitinol insulated with PET | Pebax (distal shaft)<br>Stainless steel (hypotube)<br>Nitinol insulated with PET |
| Packaging | Individually packaged(box) | Individually packaged (carton) |
| Sterilization | Provided sterile (Ethylene Oxide) | Provided sterile (Ethylene Oxide) |
| Single Use or<br>Reusable | Single use | Single use |
{9}------------------------------------------------
## VII. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING
The non-clinical performance testing included in this 510(k) application provides evidence that the subject devices conform to user needs and intended use and are thus substantially equivalent to the predicates previously cleared under K123591 and K162892. No new safety or effectiveness issues were raised during device performance testing. The following test data is provided in this 510(k) application:
- Sterilization validation and EO residual assessment for the sterile POLARSHEATH™ Steerable . Sheath 12F, POLARMAP™ Circular Mapping Catheter, and POLARMAP™ EP Electrical Cable
- . Biocompatibility assessment and testing for the patient-contacting components of POLARSHEATH™ Steerable Sheath 12F and POLARMAP™ Circular Mapping Catheter devices
- . Performance Testing per IEC 60601-1: Medical electrical equipment - Part 1 and IEC 60601-1-2: Medical electrical equipment Part 1-2
- . Design verification testing of the devices and device packaging, demonstrating that the devices (and device packaging, if required/applicable) meet their design input requirements as per the product and/or packaging specifications Design verification testing at nominal (T=0) timepoint
- . Design verification testing at accelerated and/or real-time aging timepoints for devices with a labeled shelf life:
- T=12 months for POLARSHEATH™ Steerable Sheath 12F I
- 트 T=24 months for POLARMAP™ Circular Mapping Catheter and POLARMAP™ EP Electrical Cable
- . Design validation testing summary demonstrating that the POLARSHEATH™ Steerable Sheath 12F and POLARMAP™ Circular Mapping Catheter (including the POLARMAP™ EP Electrical Cable) conform to user needs and intended uses
# VIII. CONCLUSION
Based on the intended use, fundamental technological characteristics, and performance testing, the proposed POLARSHEATH™ Steerable Sheath 12F, POLARMAP™ Circular Mapping Catheter, POLARMAP™ EP Electrical Cable, SMARTFREEZE™ Esophageal Temperature Sensor (ETS) Cables, and SMARTFREEZE™ Pressure Sensor Cable have been shown to be substantially equivalent to their respective predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.