← Product Code [GAT](/productcode/GAT) · K223822

# ACTIVBRAID™ (K223822)

_Embody, Inc. · GAT · Jun 2, 2023 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K223822

## Device Facts

- **Applicant:** Embody, Inc.
- **Product Code:** [GAT](/productcode/GAT.md)
- **Decision Date:** Jun 2, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.5000
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

ACTIVBRAID is indicated for use in soft tissue approximation and/or ligation. These sutures may be incorporated, as components, into surgeries where constructs including those with allograft or autograft tissues are used for repair.

## Device Story

ACTIVBRAID is a partially absorbable surgical co-braid suture; constructed of nonabsorbable Ultra High Molecular Weight Polyethylene (UHMWPE) and absorbable bovine-derived type I collagen. Suture ends are stiffened with cyanoacrylate. Provided in pre-cut lengths with or without needles. Used by surgeons in clinical settings for soft tissue approximation and ligation; can be incorporated into surgical constructs involving allograft or autograft tissues. Operates via standard surgical technique. Benefits include long-term soft tissue approximation with a controlled resorption profile; collagen component absorbs between 8 and 16 weeks post-implantation while maintaining approximately 85% tensile strength at 16 weeks.

## Clinical Evidence

Bench testing only. Biocompatibility testing (cytotoxicity, sensitization, systemic toxicity, pyrogenicity, genotoxicity, implantation) per ISO 10993-1. Non-clinical performance testing per USP <861>, <871>, and <881> (diameter, needle attachment, tensile strength). Comparative testing against predicate included abraded tensile strength, knot profile, knot security, tissue abrasiveness, and physiological fluid absorption. Animal implantation studies characterized resorption profile, showing 85% strength retention at 16 weeks and collagen absorption between 8-16 weeks.

## Technological Characteristics

Multifilament braided suture composed of UHMWPE and bovine-derived type I collagen. Ends stiffened with cyanoacrylate. Available in various USP sizes and tape configurations. Sterilized via Ethylene Oxide (SAL 10^-6). Meets USP Monograph for nonabsorbable surgical sutures (except diameter).

## Regulatory Identification

Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

## Predicate Devices

- Arthrex Bio-Suture ([K112899](/device/K112899.md))

## Reference Devices

- Arthrex BioWire ([K091018](/device/K091018.md))
- ORTHOCORD® Braided Composite Suture ([K040004](/device/K040004.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized image of a human figure, and the text on the right is in blue. The text is arranged in three lines, with "FDA" on the first line, "U.S. FOOD & DRUG" on the second line, and "ADMINISTRATION" on the third line.

June 2, 2023

Embody, Inc. Brianna Schehr Vice President, Quality, Regulatory & Clinical Affairs 4211 Monarch Way, Suite 500 Norfolk, Virginia 23508

Re: K223822

Trade/Device Name: ActivbraidTM Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable polyethylene terephthalate surgical suture Regulatory Class: Class II Product Code: GAT Dated: December 21, 2022 Received: December 21, 2022

Dear Brianna Schehr:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Cynthia Chang -S

Cynthia J. Chang, Ph.D. Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K223822

Device Name ACTIVBRAID™

### Indications for Use (Describe)

ACTIVBRAID™ is indicated for use in soft tissue approximation and/or ligation. These sutures may be incorporated, as components, into surgeries where constructs including those with allograft tissues are used for repair.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for "emBODY". The word "em" and "DY" are in a dark blue font. The "BODY" portion of the logo has an orange square with diagonal lines inside of it. 

## 510(k) Summary - K223822

#### SUBMITTER l.

| Owner/Submitter:      | Embody, Inc.<br>4211 Monarch Way<br>Suite 500<br>Norfolk, VA 23508       |
|-----------------------|--------------------------------------------------------------------------|
| Phone:                | 236-994-7912                                                             |
| Contact Person:       | Brianna Schehr<br>Vice President, Quality, Regulatory & Clinical Affairs |
| Date Prepared:        | May 24, 2023                                                             |
| II. DEVICE            |                                                                          |
| Name of Device:       | ACTIVBRAID™                                                              |
| Common or Usual Name: | Suture                                                                   |
| Classification Name:  | Suture, Nonabsorbable, Synthetic, Polyethylene                           |
| Regulation Number:    | 21 CFR 878.5000                                                          |
| Regulatory Class:     | Class II                                                                 |
| Product Code:         | GAT                                                                      |

#### lll. PREDICATE DEVICE

Medical Specialty:

Arthrex Bio-Suture, K112899. This predicate has not been subject to a design-related recall.

General & Plastic Surgery

Reference Devices: Arthrex BioWire, K091018 ORTHOCORD® Braided Composite Suture, K040004

#### IV. DEVICE DESCRIPTION

ACTIVBRAID is a partially absorbable surgical co-braid suture constructed of nonabsorbable Ultra High Molecular Weight Polyethylene (UHMWPE) and absorbable bovine-derived type I collagen. ACTIVBRAID suture ends are stiffened with cyanoacrylate.

ACTIVBRAID is available in several sizes (sutures meet U.S. Pharmacopeia standards for nonabsorbable sutures, except diameter). Sutures are oversized in diameter. ACTIVBRAID is provided in pre-cut lengths with and without needles. Suture strands that are dyed (Chromium Cobalt-Aluminum Oxide or D&C Black No. 4) are made of UHMWPE.

#### V. INDICATION FOR USE

ACTIVBRAID is indicated for use in soft tissue approximation and/or ligation. These sutures may be incorporated, as components, into surgeries where constructs including those with allograft or autograft tissues are used for repair.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "embody" in a stylized font. The letters are a dark teal color, except for the "o", which is an orange square with diagonal lines inside. The word is written in all lowercase letters.

### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

ACTIVBRAID is substantially equivalent in terms of indication for use, technological characteristics, and performance characteristics to the predicate device. Both the subject and predicate devices meet the USP Monograph for nonabsorbable surgical sutures, except diameter.

The following technological differences exist between the subject and predicate devices:

- The subject device is composed of nonabsorbable UHMWPE fiber braided with . absorbable bovine type I collagen fiber, whereas the predicate device is composed of nonabsorbable UHMWPE fiber braided with polyester fibers coated with absorbable type I bovine collagen, and
- The subject device is available in sizes 2-0, 2, 1.5mm tape, and 2.5mm tape, whereas . the predicate device is available in sizes 2 and 2mm tape.

The differences between the subject and predicate device are considered minor and do not raise questions concerning safety and effectiveness. Biocompatibility and nonclinical performance testing data demonstrate that ACTIVBRAID is as safe, as effective, and performs as well as or better than the predicate device, K112899.

### VII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

### Biocompatibility Testing

The biocompatibility evaluation for the ACTIVBRAID device was conducted in accordance with the FDA guidance document "Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process" dated September 4, 2020, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests: cytotoxicity, sensitization, systemic toxicity, pyrogenicity, genotoxicity, and implantation.

### Non-Clinical Performance Testing

Nonclinical performance testing was performed to demonstrate that ACTIVBRAID meets the current USP Monograph for nonabsorbable surgical sutures, except diameter. Testing was performed in accordance with FDA's guidance document, "Surgical Sutures – Class II Special Controls Guidance Document for Industry and FDA Staff" and included the following tests:

- Nonabsorbable Suture Diameter USP <861> Sutures Diameter .
- Nonabsorbable Suture Needle Attachment USP <871> Sutures Needle Attachment .
- . Nonabsorbable Suture Tensile Strength USP <881> Sutures – Tensile Strength

ACTIVBRAID meets or exceeds USP performance standards for nonabsorbable surgical sutures, except for an oversize in diameter. Bacterial Endotoxin Test (BET) was performed to demonstrate that the subject device meets pyrogen limit specifications.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "emBODY" in a stylized font. The letters "em" and "DY" are in a dark blue color, while the "BOD" is represented by an orange square with diagonal lines inside. The logo appears to be for a company or product named "emBODY."

Additional performance testing was conducted in comparison to the predicate device and included the following tests:

- Abraded Tensile Strength .
- . Knot Profile
- Knot Security .
- Abrasiveness to Tissue .
- Physiological Fluid Absorption .

ACTIVBRAID performs as well as or better than the predicate device for all performance tests listed. Both the subject and predicate devices have suitable characteristics for their indication for use.

### Resorption Profile

ACTIVBRAID is partially absorbable; therefore, resorption profile testing was conducted in accordance with FDA's special controls guidance, "Surgical Sutures – Class II Special Controls Guidance Document for Industry and FDA Staff" to characterize the device's rate of absorption, time to complete absorption, and residual tensile strength over time. The results of implantation studies of ACTIVBRAID in animals indicate that approximately 85% of its original strength remains 16 weeks after implantation and absorption of the collagen component is complete between 8 and 16 weeks post-implantation. The ACTIVBRAID degradation profile combined with the tensile strength retention over time demonstrates the resorption profile is consistent with the intended use of the subject device, for the long-term approximation of soft tissue.

| Characteristic                | Subject Device                                                                                                                                                                                                                        | Predicate Device<br>Arthrex Bio-Suture<br>(K112899)                                                                                                                                                                                   | Comparison                                            |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|
| Product Code                  | GAT                                                                                                                                                                                                                                   | GAT                                                                                                                                                                                                                                   | Same                                                  |
| Regulation                    | 21 CFR 878.5000                                                                                                                                                                                                                       | 21 CFR 878.5000                                                                                                                                                                                                                       | Same                                                  |
| Classification<br>Name        | Suture, Nonabsorbable,<br>Synthetic, Polyethylene                                                                                                                                                                                     | Suture, Nonabsorbable,<br>Synthetic, Polyethylene                                                                                                                                                                                     | Same                                                  |
| Intended Use                  | Long-term approximation of soft<br>tissue                                                                                                                                                                                             | Long-term approximation of soft<br>tissue                                                                                                                                                                                             | Same                                                  |
| Indications for<br>Use        | For use in soft tissue<br>approximation and/or ligation.<br>These sutures may be<br>incorporated, as components,<br>into surgeries where constructs<br>including those with allograft or<br>autograft tissues are used for<br>repair. | For use in soft tissue<br>approximation and/or ligation.<br>These sutures may be<br>incorporated, as components,<br>into surgeries where constructs<br>including those with allograft or<br>autograft tissues are used for<br>repair. | Same                                                  |
| Prescription Use              | Yes                                                                                                                                                                                                                                   | Yes                                                                                                                                                                                                                                   | Same                                                  |
| Material(s) of<br>Composition | UHMWPE, bovine-derived type I<br>collagen                                                                                                                                                                                             | UHMWPE, polyester, bovine-<br>derived type I collagen                                                                                                                                                                                 | Similar, no<br>new issues of<br>safety or<br>efficacy |
| Resorbable                    | Partially absorbable (i.e.,<br>nonabsorbable UHMWPE fiber<br>and absorbable collagen fiber<br>co-braid)                                                                                                                               | Nonabsorbable (i.e.,<br>nonabsorbable braided<br>UHMWPE and polyester fibers<br>coated with absorbable type I<br>bovine collagen)                                                                                                     | Similar, no<br>new issues of<br>safety or<br>efficacy |

### Substantial Equivalence

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "EMBODY" in a stylized font. The letters are primarily dark blue, except for the "B", which is enclosed in an orange square with diagonal lines inside. The word appears to be a logo or brand name, with a modern and clean design.

| Characteristic                            | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Predicate Device<br>Arthrex Bio-Suture<br>(K112899)                                                                                                                                                               | Comparison                                            |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|
| Suture<br>Configuration(s)                | Configured as multifilament<br>braided strands of multiple sizes<br>(per USP size system)<br>ACTIVBRAID suture:<br>• Size 2-0 (blue) w/ tapered<br>needle<br>• Size 2 (blue) w/ tapered<br>needle<br>• Size 2 (white/black) w/ tapered<br>needle<br>ACTIVBRAID suture tape:<br>• Flat, tape-style suture, 1.5mm<br>(white/black) w/ tapered<br>needle<br>• Flat, tape-style suture, 1.5mm<br>(blue) w/ tapered needle<br>• Flat, tape-style suture, 2.5mm<br>(white/black)<br>• Flat, tape-style suture, 2.5mm<br>(blue) | Configured as multifilament<br>braided strands of multiple sizes<br>(per USP size system)<br>Bio-FiberWire:<br>• Size 2 (blue) w/ tapered<br>needle<br>Bio-FiberTape:<br>• Flat, tape-style suture, 2mm<br>(blue) | Similar, no<br>new issues of<br>safety or<br>efficacy |
| Residual Tensile<br>Strength Over<br>Time | 85% of original strength remains<br>16 weeks after implantation                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Not known to have significant<br>change in tensile strength in vivo                                                                                                                                               | Similar, no<br>new issues of<br>safety or<br>efficacy |
| Technological<br>Principle                | Long-term approximation of soft<br>tissue                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Long-term approximation of soft<br>tissue                                                                                                                                                                         | Same                                                  |
| Operating<br>Principles                   | Standard surgical technique                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Standard surgical technique                                                                                                                                                                                       | Same                                                  |
| How Supplied                              | Sterile, single use only, with or<br>without needles attached                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Sterile, single use only, with or<br>without needles attached                                                                                                                                                     | Same                                                  |
| Sterility                                 | Sterile, SAL 10-6                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Sterile, SAL 10-6                                                                                                                                                                                                 | Same                                                  |
| Sterilization<br>Method                   | Ethylene Oxide                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Ethylene Oxide                                                                                                                                                                                                    | Same                                                  |
| Single Use                                | Single Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Single Use                                                                                                                                                                                                        | Same                                                  |

# VIII. CONCLUSIONS

Results of the biocompatibility testing and nonclinical performance testing demonstrate that ACTIVBRAID is as safe, as effective, and performs as well as or better than the legally marketed predicate device, K112899.

---

**Source:** [https://fda-staging.innolitics.com/device/K223822](https://fda-staging.innolitics.com/device/K223822)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).