Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH)
K223810 · Sb-Kawasumi Laboratories, Inc. · JKA · Feb 24, 2023 · Clinical Chemistry
Device Facts
Record ID
K223810
Device Name
Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH)
Applicant
Sb-Kawasumi Laboratories, Inc.
Product Code
JKA · Clinical Chemistry
Decision Date
Feb 24, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1675
Device Class
Class 2
Indications for Use
Female luer type · The Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH) is a sterile, non-invasive device used for connection with syringes when transferring blood specimen from a syringe to blood culture bottle. Male luer type · The Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH) is a sterile, non-invasive device used for connection with a female luer system and non-needle devices in order to collect blood specimen to blood collection tube.
Device Story
Sterile, single-use, non-invasive blood collection adapter; consists of plastic holder, stainless steel cannula, rubber sheath, and luer adapter. Two configurations: male luer (connects to vascular access devices) and female luer (connects to syringes). Used in hospitals/healthcare facilities by clinicians. For female luer type, clinician draws blood into syringe, connects device to syringe, and inserts blood collection tube/culture bottle; vacuum in tube/bottle draws blood through cannula. Facilitates safe blood transfer and collection; minimizes exposure risk.
Clinical Evidence
Bench testing only. Testing included luer connector evaluation (ISO 80369-7), simulation testing, physical testing (ISO 1135-3), particulate contamination (USP <788>), chemical testing (ISO 1135-3), usability testing (ISO 14971, IEC 62366-1), EtO residuals (ISO 10993-7), bacterial endotoxin (USP <85>), sterility (USP <71>), and transportation/packaging validation (ISO 11607, ASTM D4169). All tests passed.
Indicated for use in hospitals or healthcare facilities for blood specimen collection. Female luer type: used to transfer blood from a syringe to a blood collection tube or blood culture bottle. Male luer type: used to collect blood into a blood collection tube by connecting to female luer connectors of vascular access devices (e.g., peripheral/central catheters, Huber needles).
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
Predicate Devices
Kawasumi Multiple Sample Adapter with Pre-attached Holder (K190485)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 24, 2023
SB-Kawasumi Laboratories, Inc. % Valerie Followell Regulatory Consultant Regulatory Compliance Associates, Inc. (RCA) 10411 Corporate Drive, Suite 102 Pleasant Prairie, Wisconsin 53158
Re: K223810
Trade/Device Name: Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH) Regulation Number: 21 CFR 862.1675 Regulation Name: Blood Specimen Collection Device Regulatory Class: Class II Product Code: JKA Dated: January 27, 2022 Received: January 27, 2022
Dear Valerie Followell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Danil Walloschek
David Wolloscheck, Ph.D. Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
### 510(k) Number (if known) K223810
### Device Name
Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH)
## Indications for Use (Describe)
Female luer type
· The Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH) is a sterile, non-invasive device used for connection with syringes when transferring blood specimen from a syringe to blood culture bottle.
### Male luer type
· The Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH) is a sterile, non-invasive device used for connection with a female luer system and non-needle devices in order to collect blood specimen to blood collection tube.
Type of Use (Select one or both, as applicable)
| <span> <b> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</b> </span> |
|-------------------------------------------------------------------------------------------------------------|
| <span> <b> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</b> </span> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K223810 - 510(K) Summary
Manufacturer: SB-KAWASUMI LABORATORIES, INC.
## Address:
3-25-4, Tonomachi, Kawasaki-ku Kawasaki-shi. Kanagawa 210-8602, JAPAN
Corresponding Official/Contact: Valerie Followell Regulatory Consultant Telephone Number: 847-400-6187 Email: v.followell@rcainc.com
#### Summary Date: Februarv 24, 2023
Trade Name: Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH)
Common or Usual Name: Blood Collection Tubes, Vials, Systems, Serum Separators
Regulation Number: 21 CFR 862.1675
Regulation Name: Blood specimen collection device
Product Code: JKA
Class: Class 2
Panel: Clinical Chemistry
Predicate Device: Kawasumi Multiple Sample Adapter with Pre-attached Holder has only one configuration with Male Luer Adapter (K190485)
# Device Description:
The Kawasumi Multiple Sample Adapter with Pre-attached Holder (MBCH) is a sterile, single use device consisting of a plastic holder with a non-patient contacting stainless steel cannula covered with a rubber sheath and a luer adapter. There are two types of the device with difference in luer adapter type; male luer adapter type and female luer adapter type. The predicate device (K190485) is the male luer type and the proposed device is the female luer type, and the subject of this special 510(k) premarket notification submission.
- The male luer adapter type is used to collect blood specimen into blood collection tube by . connecting to female luer connectors of vascular access devices such as peripheral, central catheter, Huber needle, etc. (predicate [Item number MBCH-01])
- The female luer adapter type is used to transfer blood from a syringe into blood collection tube . or blood culture bottle. After drawing patient blood into a syringe, connect the female luer adapter of the device to the syringe. When inserting a tube, the blood is transferred into the tube using tube's vacuum. (subject [Item number MBCH-02])
The device is sterilized using Ethylene Oxide. The device is a prescription-use device intended to be used in hospitals or healthcare facilities.
{4}------------------------------------------------
# Indications for Use:
### Female luer type
· The Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH) is a sterile, non-invasive device used for connection with syringes when transferring blood specimen from a syringe to blood collection tube or blood culture bottle.
### Male luer type
· The Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH) is a sterile, non-invasive device used for connection with a female luer system and non-needle devices in order to collect blood specimen to blood collection tube.
| 510(k) Summary Device Comparison Table | | | |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature of the Device | Subject Device: | Predicate Device: | Discussion/Comment: |
| | SB-KAWASUMI LABORATORIES, INC.<br><br>Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH) | Kawasumi Laboratories, Inc.<br><br>Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH) | The company name has been changed. Statement included in submission. |
| Item number (Model number) | MBCH-01 and MBCH-02 | MBCH-01 | Different<br>The new item number MBCH-02 is added. |
| 510(k) Number | K223810 | K190485 | - |
| Product Code | JKA | JKA | Same |
| Classification | 21 CFR 862.1675 | 21 CFR 862.1675 | Same |
| Classification Name | Tubes, Vials, Systems, Serum Separators, Blood Collection | Tubes, Vials, Systems, Serum Separators, Blood Collection | Same |
| Common Name | Blood Specimen Collection Device | Blood Specimen Collection Device | Same |
| Rx or OTC | Rx only | Rx Only | Same |
| Number of Uses | Single-Use | Single-Use | Same |
| Use Environment | Hospital and/or Healthcare Facility | Hospital and/or Healthcare Facility | Same |
| Indications for Use | Female luer type The Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH) is a sterile, non-invasive device used for connection with syringes when transferring blood specimen from a syringe to blood collection tube or blood culture bottle. Male luer type The Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH) is a sterile, non-invasive device used for connection with a female luer system and non-needle devices in order to collect blood specimen to blood collection tube. | The Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH) is a sterile, non-invasive device used for connection with a female luer system and non-needle devices in order to collect blood specimen to blood collection tube. | Different<br>Although there is a difference in where to connect the device, both types are used for collecting blood specimen to blood collection tube or blood culture bottle. |
| Intended Use | Same as the Indications for Use | Same as the Indications for Use | Same |
| Design Feature | Gauge: 20G<br>Length (mm): $64.1 \pm 0.3$<br>Width (mm): $31 \pm 0.3 \times 23.5 \pm 0.2$ | Gauge: 20G<br>Length(mm): $64.1 \pm 0.3$<br>Width(mm): $31 \pm 0.3 \times 23.5 \pm 0.2$ | Same |
| Luer Adapter<br>Type | Female | Male | Different<br>Differences do not<br>change assertion of<br>equivalence of devices |
| Number of Uses | Single-Use Rx Only | Single-Use Rx Only | Same |
| Materials<br>Hub | Polycarbonate (PC) | Polycarbonate (PC) | Different<br>Material is identical with<br>the exception of its color.<br>Supportive information<br>included in submission. |
| Cannula | Stainless Steel | Stainless Steel | Same |
| Glue | Epoxy | Epoxy | Same |
| Silicone | Silicone Oil | Silicone Oil | Different<br>Supportive information<br>included in submission. |
| Sheath | Isoprene Rubber<br>Not made with Natural Rubber<br>Latex | Isoprene Rubber<br>Not made with Natural Rubber<br>Latex | Same |
| Holder | Polypropylene (PP) | Polypropylene (PP) | Same |
| Biocompatibility | Complies with ISO 10993 | Complies with ISO 10993 | Same |
| Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide | Same |
| Shelf Life | 3 years | 3 years | Same |
{5}------------------------------------------------
{6}------------------------------------------------
# Technological Characteristics:
The Subject Device and the Predicate Device have similar technological characteristics. Both devices are sterile, single-use, non-invasive devices with a Pre-Attached Holder used for connection with a female luer system and non-needle devices in order to collect blood specimen to blood collection tube.
The Kawasumi Multiple Sample Adaptor with Pre-Attached Holder (Subject Device [MBCH-02]) is substantially equivalent to the Predicate Device [MBCH-01] with regard to technologic characteristics, materials, performance and intended use.
# Non-Clinical Testing:
Non-Clinical Performance Testing was performed to ensure that the device meets design requirements and specifications and to confirm performance of the Kawasumi Multiple Sample Adaptor with Pre-Attached Holder [MBCH-02].
| Test Item | Referenced standard | Result |
|-------------------------------------------|-------------------------------------------------------|--------|
| Performance & Chemical Testing | | |
| Luer connector evaluation | ISO 80369-7:2021 | PASS |
| Simulation test | In-house | PASS |
| Physical test | ISO 1135-3:2016, in-house | PASS |
| Particulate contamination | USP<788> | PASS |
| Chemical test | ISO 1135-3:2016 | PASS |
| Usability test | ISO 14971:2019<br>IEC 62366-1:2015 + Amd.1:2020 | PASS |
| Sterilization, Transportation & Packaging | | |
| EtO residual | ISO 10993-7:2008/Amd.1:2019 | PASS |
| Bacterial endotoxin | USP <85> | PASS |
| Sterility test | USP <71> | PASS |
| Transportation & Package | ISO 11607-1:2019<br>ISO 11607-2:2019<br>ASTM D4169-16 | PASS |
### Conclusion:
Based on the Non-Clinical Performance Testing conducted on the subject device, intended use, and principles of operation it may be concluded that the Kawasumi Multiple Sample Adapter with Pre-Attached Holder (Subject Device [MBCH-02]) is substantially equivalent to the legally marketed Predicate Device [MBCH-01].
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.