Reusable Temperature Probe, Disposable Temperature Probe
K223807 · Shenzhen Sino-K Medical Technology Co.,Ltd · FLL · Aug 25, 2023 · General Hospital
Device Facts
Record ID
K223807
Device Name
Reusable Temperature Probe, Disposable Temperature Probe
Applicant
Shenzhen Sino-K Medical Technology Co.,Ltd
Product Code
FLL · General Hospital
Decision Date
Aug 25, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Reusable Temperature Probes: Reusable Temperature Probes are intended to be used for monitoring temperature for multi-patient use. The temperature probes are reusable and designed for use with monitor of GE-Marquette Model DASH3000. These devices are used by qualified medical professionals only. Disposable Temperature Probes: Disposable Temperature Probes are intended to be used for monitoring temperature for single patient use. The temperature probes are non-sterile and designed for use with monitors of GE- Marquette Model DASH3000. These devices are used by qualified medical professionals only.
Device Story
Device consists of temperature probes (reusable or disposable) with a connector for GE-Marquette DASH3000 monitors and a thermistor at the patient end. Principle of operation: resistance-based sensing using a Negative Temperature Coefficient (NTC) thermistor; resistance increases as temperature decreases, providing linear temperature measurement. Used in healthcare facilities by qualified medical professionals for patient temperature monitoring. Output is displayed on the compatible monitor to assist clinicians in patient assessment. Benefits include accurate, continuous temperature monitoring for clinical decision-making.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-5, ISO 10993-10), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and performance testing (ISO 80601-2-56).
Technological Characteristics
NTC thermistor-based sensing. Reusable models: PVC cable, steel/epoxy/PA66 probe end. Disposable models: PVC cable, epoxy/medical pressure-sensitive adhesive probe end. Measurement range 25-45°C; accuracy ±0.1°C. Compatible with GE-Marquette DASH3000. Non-sterile. Complies with IEC 60601-1, IEC 60601-1-2, ISO 80601-2-56, and ASTM D4169-22.
Indications for Use
Indicated for monitoring patient body temperature in clinical settings. Suitable for adult and pediatric populations. Reusable probes for multi-patient use; disposable probes for single-patient use. Designed for use with GE-Marquette Model DASH3000 monitors by qualified medical professionals.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Med-link Reusable Temperature Probes, Med-link Disposable Temperature Probes (K193338)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 25, 2023
Shenzhen SINO-K Medical Technology Co., Ltd. % Kevin Wang Consultant Shenzhen Chonconn Medical Consulting Co., Ltd. Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District Shenzhen, Guangdong 518067 China
Re: K223807
Trade/Device Name: Reusable Temperature Probe, Disposable Temperature Probe Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: July 25, 2023 Received: July 26, 2023
Dear Kevin Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Porsche Bennett
Porsche Bennett For David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223807
Device Name
Reusable Temperature Probe, Disposable Temperature Probe
Indications for Use (Describe)
Reusable Temperature Probes:
Reusable Temperature Probes are intended to be used for monitoring temperature for multi- patient use. The temperature probes are reusable and designed for use with monitor of GE-Marquette Model DASH3000. These devices are used by qualified medical professionals only.
Disposable Temperature Probes:
Disposable Temperature Probes are intended to be used for monitoring temperature for single patient use. The temperature probes are non-sterile and designed for use with monitors of GE- Marquette Model DASH3000. These devices are used by qualified medical professionals only.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary - K223807
### Prepared in accordance with the requirements of 21 CFR Part 807.92
## Prepared Date: August 25, 2023
### 1. Submission sponsor
Name: Shenzhen SINO-K Medical Technology Co., Ltd. Room401, Bldg2, Veteran Ind. City, Gongle Community, Xixiang Street, Baoan District, Shenzhen, Guangdong, China Contact person: Lao Chengxin Title: General Manager E-mail: boss(@sino-k.com Tel: +86 137 15333326
### Submission correspondent 2.
Name: Shenzhen Chonconn Medical Device Consulting Co., Ltd. Address: Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District, Shenzhen, Guangdong, P.R. China Contact person: Kevin Wang E-mail: kevin(@chonconn.com Tel: +86-755 33941160
### 3. Subject Device Information
| Trade/Device Name | Reusable Temperature Probe, Disposable Temperature Probe |
|---------------------------|----------------------------------------------------------|
| Regulatory Class | Class II |
| Regulation Number | 21CFR 880.2910 |
| Regulation/Common<br>Name | Clinical Electronic Thermometer |
| Product Code | FLL |
| Submission type | Traditional 510(K) |
### Predicate Device 4.
| Manufacturer Name | Shenzhen Med-link Electronics Tech Co., Ltd. |
|------------------------|------------------------------------------------------------------------------|
| 510(k) Number | K193338 |
| Trade/Device Name | Med-link Reusable Temperature Probes, Med-link Disposable Temperature Probes |
| Regulatory Class | Class II |
| Regulation Number | 21CFR 880.2910 |
| Regulation/Common Name | Clinical Electronic Thermometer |
| Product Code | FLL |
| Submission type | Traditional 510(K) |
{4}------------------------------------------------
### 5. Device Description
The proposed devices are used for patient temperature measurement. The probes are reusable or disposable depending on the model. These probes consist of a connector on the monitor end and a thermistor on the patient end. The working principle is resistance based on the metal conductor increasing with temperature decrease, and the linear changes to the characteristics of the temperature measurement. The proposed devices are designed to be used in a healthcare facility and compatible with a monitor, GE-Marquette Model DASH3000.
The six models have two types of structure designs corresponding to reusable and disposable use which consists of different materials. The Negative Temperature Coefficient (NTC) of the six models are identical. Reusable models ST1304, ST2304, ST3304, ST4304 have a similar structure design with two different sensor shapes for different measure sites and consist of the same materials. Disposable models ST5105A and ST6105A have a similar structure design with two different sensor shapes for different measure sites and consist of the same materials.
| No. | Model | Description |
|-----|---------|----------------------------------------------------------------|
| 1 | ST1304 | Skin contact Temperature Probe, adult, reusable |
| 2 | ST2304 | Body cavity Temperature Probe, Esophageal, adult, reusable |
| 3 | ST3304 | Skin contact Temperature Probe, pediatric, reusable |
| 4 | ST4304 | Body cavity Temperature Probe, Esophageal, pediatric, reusable |
| 5 | ST5105A | Skin contact Temperature Probe, adult, disposable |
| 6 | ST6105A | Body cavity Temperature Probe, Esophageal, adult, disposable |
## 6. Indication for use
Reusable Temperature Probes:
Reusable Temperature Probes are intended to be used for monitoring temperature for multipatient use. The temperature probes are reusable and designed for use with monitor of GE-Marquette Model DASH3000.
These devices are used by qualified medical professionals only.
Disposable Temperature Probes:
Disposable Temperature Probes are intended to be used for monitoring temperature for single patient use.
The temperature probes are non-sterile and designed for use with monitors of GE-Marquette Model DASH3000.
These devices are used by qualified medical professionals only.
{5}------------------------------------------------
| Item | Subject Device,<br>K223807 | Predicate Device,<br>K193338 | Comparison |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Trade name | Reusable Temperature<br>Probe, Disposable<br>Temperature Probe | Med-link Reusable<br>Temperature Probes,<br>Med-link Disposable<br>Temperature Probes | / |
| 510(k) submitter | Shenzhen SINO-K<br>Medical Technology<br>Co., Ltd. | Shenzhen Med-link<br>Electronics Tech Co.,<br>Ltd. | / |
| 510(k) Number | K223807 | K193338 | / |
| Models | Reusable<br>adult/pediatric<br>models: ST1304,<br>ST2304, ST3304,<br>ST4304<br>Disposable<br>adult/pediatric<br>models:<br>ST5105A, ST6105A | Reusable pediatric<br>models:<br>W0003D,W0001C,W000<br>1D,<br>W00008C,W00008D<br>Disposable<br>adult/pediatric models:<br>W0001E, W0001F,<br>W0099F, W0003E,<br>W0003F | SE<br>Note 1 |
| Type of use | Prescription/Rx | Prescription/Rx | Same |
| Indication for<br>Use | Reusable Temperature<br>Probes:<br>Reusable Temperature<br>Probes are intended to<br>be used for monitoring<br>temperature for multi-<br>patient use. The<br>temperature probes are<br>reusable and designed<br>for use with monitor of<br>GE-Marquette Model<br>DASH3000.<br>These devices are used<br>by qualified | Med-link Reusable<br>Temperature Probes:<br>Med-link Temperature<br>Probes are intended to be<br>used for monitoring<br>temperature for multi-<br>patient use. The<br>temperature probes are<br>reusable and designed for<br>use with monitors of<br>Philips Model<br>IntelliVueMP50,<br>Mindray Model PM-9000<br>and Drager Model | SE<br>Note 1 |
| Item | Subject Device,<br>K223807 | Predicate Device,<br>K193338 | Comparison |
| | professional only.<br>Disposable<br>Temperature Probes:<br>Disposable<br>Temperature Probes<br>are intended to be used<br>for monitoring<br>temperature for single<br>patient use. The<br>temperature probes are<br>non-sterile and<br>designed for use with<br>monitors of GE-<br>Marquette Model<br>DASH3000.<br>These devices are used<br>by qualified medical<br>professional only. | These devices are used<br>by qualified medical<br>professional only.<br>Med-link Disposable<br>Temperature Probes:<br>Med-link Disposable<br>Temperature Probes are<br>intended to be used for<br>monitoring temperature<br>for single patient use.<br>The temperature probes<br>are<br>non-sterile and designed<br>for use with monitors of<br>GE Model B20 and<br>Philips Model<br>IntelliVueMP50.<br>These devices are used<br>by qualified medical<br>professional only. | |
| Operating<br>Principle | Resistance of<br>thermistor based on<br>the metal conductor<br>increases with<br>temperature decrease,<br>and the linear changes<br>to the characteristics<br>of the temperature<br>measurement. | Resistance of thermistor<br>based on the metal<br>conductor increases with<br>temperature decrease,<br>and the linear changes to<br>the characteristics of the<br>temperature<br>measurement. | Same |
| Measure site | Skin, Esophagus | Skin, Esophageal and<br>Rectal | SE<br>Note 2 |
| Usage | Reusable, Disposable | Reusable, Disposable | Same |
| Measurement<br>Range | 25-45°C | 25-45°C | Same |
| Measurement<br>Time | 60s | 180s | SE Note 3 |
| Accuracy | ±0.1°C | ±0.1°C | Same |
| Components | plug, cable and<br>temperature sensing | plug, cable and<br>temperature sensing | Same |
| Item | Subject Device,<br>K223807 | Predicate Device,<br>K193338 | Comparison |
| | probe | probe | |
| Thermistor<br>resistance | 2.25KΩ@25°C | 2.25KΩ@25°C | Same |
| Device<br>Dimension | Reusable models:3m<br>Disposable models:<br>0.8m | Unknown | SE<br>Note 4 |
| Material | Materials of reusable<br>probe:<br>Cable: PVC<br>Probe end: Steel,<br>Epoxy, PA66<br>Materials of<br>disposable probe:<br>Cable: PVC<br>Probe end: Epoxy,<br>Medical pressure-<br>sensitive adhesive | Materials of reusable<br>probe:<br>Cable: TPU<br>Probe end: Epoxy,<br>S304 Stainless Steel;<br>Materials of disposable<br>probe:<br>Cable: PVC<br>Probe end: Epoxy, PVC | SE<br>Note 5 |
| Compatible<br>Monitors | GE-Marquette Model<br>DASH3000 | Reusable temperature<br>probes are compatible<br>with Philips Model<br>IntelliVue MP50,<br>Mindray Model PM-<br>9000, Drager Model<br>Infinity Gamma XL, and<br>disposable temperature<br>probes are compatible<br>with GE Model B20 and<br>Philips Model IntelliVue<br>MP50. | SE<br>Note 1 |
| Operating<br>Environment | Temperature: +5<br>to+40°C<br>Humidity: ≤80%<br>(non-condensing)<br>Atmospheric pressure<br>: 86kPa ~<br>106kPa | Temperature:<br>+5°C~+40°C;<br>Atmospheric Pressure: 86<br>kPa to 106 kPa<br>Relative humidity<br>range:0 % to 80 %,<br>non-condensing (% RH) | Same |
| Storage<br>Environment | Temperature: -10°C to<br>+40°C | Temperature: -10°C to<br>+40°C | Same |
| Electrical Safety | Comply with IEC 60601-1 | Comply with IEC 60601-1 | Same |
| Performance | ISO 80601-2-56 | ISO 80601-2-56 | Same |
| Biocompatibility | All the patient-contacting materials are evaluated by the biocompatibility standard ISO10993-1 | All the patient-contacting materials are evaluated by the biocompatibility standard ISO10993-5, ISO 10993-10. | Same |
| Sterilization | Non-sterile | Non-sterile | Same |
| Operational Type | Continual | Continual | Same |
| Shelf life | 2 years | Unknown | SE<br>Note 6 |
### 7. Comparison to the Predicate Device
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
Note 1
The subject device is compatible with a different monitor than the predicate device. In addition, the intended patient population of both the subject device and the predicate device include adult and pediatric patient populations with the only difference being that The core component of the temperature probes in the subject devices is Negative Temperature Coefficient (NTC) which is identical to the NTC used in the predicate devices. The NTC determines the accuracy and range of temperature measurement. The subject device performance was tested in accordance with ISO 80601-2-56 for all models, using the compatible monitor. All results met the requirements. Therefore, the differences do not raise new questions of safety and effectiveness.
## Note 2
The subject device measurement sites are a subset of the predic sites. The subject device performance and biocompatibility was tested in accordance to ISO 80601-2-56 and ISO10993-1, respectively. Therefore, the difference does not raise new questions of safety and effectiveness.
## Note 3
The meas ne of the subject device is 60 seconds; he time of the predicate device is 180 seconds. The subject device's measurement time is le than that of the predicate and the device performance of all models have been tested to ISO80601-2-56. Therefore, the differences do not raise new or different questions of safet and effectiveness.
## Note 4
ns of the predicate device are unknown. The ide flexibility for clinical use. The length e of the subject devices. The performance o
{9}------------------------------------------------
been verified with ISO 80601-2-56; the difference does not raise new questions of safety and effectiveness.
## Note 5
Although patient-contacting materials are different between the subject devices and predicate devices, all of the subject device models were tested in accordance to and complied with ISO 10993-5 and ISO 10993-10. Therefore, the differences do not raise new questions of safety and effectiveness.
## Note 6
The shelf life of the subject device is 2 years; however, the shelf life of the predicate device is unknown. The performance of the subject device has been verified for the claimed shelf life. Therefore, the difference does not raise new questions of safety and effectiveness.
### 8. Non- Clinical Testing
The following testing was conducted to demonstrate substantial equivalence.
# Biocompatibility testing
The biocompatibility evaluation for the subject devices were conducted in accordance with the FDA Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The biocompatibility testing included the following tests:
- Cytotoxicity
- o ISO10993-5 Third edition 2009-06-01 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- . Sensitization
- ISO10993-10 Third Edition 2010-08-01 Biological evaluation of medical 0 devices - Part 10: Tests for irritation and skin sensitization
- Irritation .
- ISO10993-10 Third Edition 2010-08-01 Biological evaluation of medical O devices - Part 10: Tests for irritation and skin sensitization
- Oral mucosa Irritation (Body cavity Temperature Probe only)
# Performance Testing
Non-clinical testing has been conducted to verify that the subject devices meet all design specifications which support the conclusion that it's Substantially Equivalent (SE) to the predicate devices. The testing results demonstrate that the targeted device complies with the following standards:
- IEC 60601-1-2005+CORR.1:2006+CORR.2:2007+A1:2012, Medical Electrical
{10}------------------------------------------------
Equipment- Part 1: General requirements for basic safety and essential performance
- . IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- ISO 80601-2-56:2017+A1:2018 Medical electrical equipment – Particular requirements for the basic safety and essential performance of clinical thermometers for body temperature measurement.
- ASTM D4169-22 Standard Practice for Performance Testing of Shipping Containers and Systems
### 9. Clinical Testing
No clinical testing was performed as this device does not require clinical studies to demonstrate substantial equivalence with the predicate device.
## 10. Conclusion
It can be concluded that the differences between the subject devices, Reusable Temperature Probe, Disposable Temperature Probe and predicate devices do not raise new or different questions of safety or effectiveness. The subject device is substantially equivalent to the predicate device. Med-link Reusable Temperature Probes, Med- link Disposable Temperature Probes, cleared under K193338.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.