← Product Code [GEI](/productcode/GEI) · K223805

# Agnes (K223805)

_Agnes Medical Co., Ltd. · GEI · May 4, 2023 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K223805

## Device Facts

- **Applicant:** Agnes Medical Co., Ltd.
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** May 4, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

AGNES is indicated for use in dermatological and general surgical procedures for electrocoagulation and hemostasis.

## Device Story

RF electrosurgical device; consists of RF generator, LCD screen, SMPS, footswitch, handpiece, and single-use RF electrodes (needle or rounded type). Device delivers 1MHz oscillating rectangular wave RF energy to target tissue via electrodes to induce coagulation/hemostasis. Used in clinical settings by healthcare professionals. Operator controls output power (2-46W) and duration (50-2000ms). System includes neutral electrode pad connection monitoring and camera-based electrode inspection. Output heat effect facilitates tissue coagulation; benefits include precise thermal control for dermatological/surgical applications.

## Clinical Evidence

No human clinical data. Bench testing performed per IEC 60601-2-2:2017. Animal study conducted on mini pig tissues (skin, liver, kidney, femoral muscle) to assess thermal tissue damage and spread via thermal imaging and histopathological analysis.

## Technological Characteristics

RF electrosurgical unit; 1MHz output frequency; oscillating rectangular waveform; 2-46W power range. Materials: SUS 304 (electrodes), Polycarbonate (cover tips). Connectivity: Standalone. Sterilization: Ethylene Oxide (EO) gas per ISO 11135:2014. Safety: Neutral electrode continuity monitor, electrode inspection camera. Standards: IEC 60601-2-2:2017, ISO 10993-1/5/10/7.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Agnes ([K203013](/device/K203013.md))

## Reference Devices

- Needle type RF electrode ([K171707](/device/K171707.md))
- Disposable neutral electrode pad ([K102372](/device/K102372.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 4, 2023

Agnes Medical Co., Ltd % Sanghwa Myung Regulatory Affair Consultant E&m D-1474, 230, Simin-Daero, Dongan-gu Gyeonggi-do. Gyeonggi-do 14067 Korea, South

Re: K223805

Trade/Device Name: Agnes Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI, KCW Dated: April 3, 2023 Received: April 3, 2023

Dear Sanghwa Myung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2023.05.04 13:42:12 -04'00'

Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K223805

Device Name AGNES

Indications for Use (Describe)

AGNES is indicated for use in dermatological and general surgical procedures for electrocoagulation and hemostasis.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary

This summary of 510(K) – safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

AGNES MEDICAL CO., LTD. Submitter: (Seohyeon-Donq, 5f Cocoplaza), 20, Seohyeon-Ro 210beon-Gil, Bundang-Gu, Seongnam-Si, Gyeonqgido, Korea (ZIP code: 13591) Telephone: +82-31-8020-9702 Fax: +82-31-701-5162 E-mail: jks@aqnesmedical.com

E&M Contact: Requlatory Affair/Terri Myung Telephone: +82-70-7807-0550 FAX: +82-31-388-9263 Cellphone: +82-10-4952-6638 E-mail: mshenmc@gmail.com Primary Contact: Terri Myung

Date 510(k) summary prepared: May 4, 2023

### 1. Device Information

- Trade Name: AGNES
- Common Name: RF Electrosurgical Device
- Classification Name: Electrosurgical cutting and coagulation device and accessories
- Classification Product Code and Regulation: GEI, 21CFR 878.4400
- Subsequent Product Code and Regulation: KCW, 21CFR 878.5350
- Device Class: II

### 2. Predicate Device

- Agnes, K203013
- 3. Description

AGNES is a RF electrosurgical device. It consists of LCD screen, radiofrequency generator and SMPS. The accessories are a footswitch, a hand-piece, single use RF electrodes and FDA cleared Disposable neutral electrode pad (K102372).

### 4. Indication for use

{4}------------------------------------------------

AGNES is indicated for use in dermatological and general surgical procedures for electrocoagulation and hemostasis.

### 5. Comparison of the modified device to the cleared device

The following comparison table is presented to demonstrate substantial equivalence.

| Descriptive Information                            |                                                                      | Predicated Device                                                                                                                                       | Subject Device                                                                                                                                       |
|----------------------------------------------------|----------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                                       |                                                                      | AGNES MEDICAL CO., LTD                                                                                                                                  | AGNES MEDICAL CO., LTD                                                                                                                               |
| Device Name                                        |                                                                      | AGNES                                                                                                                                                   | AGNES                                                                                                                                                |
| 510(k) number                                      |                                                                      | K203013                                                                                                                                                 | K223805                                                                                                                                              |
| Classification Product Code /<br>Regulatory Number |                                                                      | GEI / 878.4400                                                                                                                                          | GEI / 878.4400                                                                                                                                       |
| Subsequent Product Code                            |                                                                      | KCW / 878.5350                                                                                                                                          | KCW / 878.5350                                                                                                                                       |
| Regulatory Class                                   |                                                                      | II                                                                                                                                                      | II                                                                                                                                                   |
| Indications for Use                                |                                                                      | AGNES is indicated for use<br>in dermatological and<br>general surgical procedures<br>for electrocoagulation and<br>hemostasis.                         | AGNES is indicated for use in<br>dermatological and general<br>surgical procedures for<br>electrocoagulation and<br>hemostasis.                      |
| Prescription or OTC                                |                                                                      | Prescription                                                                                                                                            | Prescription                                                                                                                                         |
| Operation                                          |                                                                      | The device uses RF energy<br>delivered through micro<br>needle electrode to apply<br>heat to target tissue for<br>coagulating.                          | The device uses RF energy<br>delivered through micro needle<br>electrode to apply heat to target<br>tissue for coagulating.                          |
|                                                    | Monopolar or<br>Bipolar                                              | Monopolar                                                                                                                                               | Monopolar                                                                                                                                            |
|                                                    | Temperature<br>sensors                                               | None                                                                                                                                                    | None                                                                                                                                                 |
|                                                    | Impedance<br>monitor                                                 | None                                                                                                                                                    | None                                                                                                                                                 |
| Electrosurgical Unit<br>(ESU)                      | Continuity<br>monitor                                                | Checking the connection<br>between the neutral<br>electrode and the<br>electrosurgical unit.                                                            | Checking the connection<br>between the neutral electrode<br>and the electrosurgical unit.                                                            |
|                                                    | Electrode<br>monitor                                                 | Provide a camera to monitor<br>the electrode coating<br>condition and shape before<br>using by the user.                                                | Provide a camera to monitor the<br>electrode coating condition and<br>shape before using by the user.                                                |
|                                                    | Waveform                                                             | Oscillating rectangular wave                                                                                                                            | Oscillating rectangular wave                                                                                                                         |
|                                                    | Output<br>frequency                                                  | 1MHz                                                                                                                                                    | 1MHz                                                                                                                                                 |
|                                                    | Output<br>operating time                                             | Min 50ms / Max 2,000ms                                                                                                                                  | Min 50ms / Max 2,000ms                                                                                                                               |
|                                                    | Output power<br>levels                                               | 25 levels<br>(2 to 46 W)                                                                                                                                | 25 levels<br>(2 to 46 W)                                                                                                                             |
|                                                    | Max. output                                                          | 46 W at 200 ohm                                                                                                                                         | 46 W at 200 ohm                                                                                                                                      |
|                                                    | power                                                                |                                                                                                                                                         |                                                                                                                                                      |
|                                                    | Max. output<br>voltage                                               | 104V                                                                                                                                                    | 104V                                                                                                                                                 |
|                                                    | Dimensions                                                           | 290mm(W)x455mm(L)x271.7mm(H)                                                                                                                            | 290mm(W)x455mm(L)x271.7m(H)                                                                                                                          |
|                                                    | Weight                                                               | 5.8Kg                                                                                                                                                   | 5.8Kg                                                                                                                                                |
|                                                    | Power Input                                                          | 100-240VAC, 50-60Hz, 420VA                                                                                                                              | 100-240VAC, 50-60Hz, 420VA                                                                                                                           |
|                                                    | RF Handpiece<br>for the needle<br>type RF<br>electrode               | It is connected with the<br>electrosurgical unit.<br>It transfers the<br>radiofrequency energy<br>through the Single use RF<br>electrode (Needle type)  | It is connected with the<br>electrosurgical unit.<br>It transfers the radiofrequency<br>energy through the Single use<br>RF electrode (Needle type)  |
|                                                    | RF Handpiece<br>for the<br>rounded<br>electrode type<br>RF electrode | It is connected with the<br>electrosurgical unit.<br>It transfers the<br>radiofrequency energy<br>through the Single use RF<br>electrode (Rounded type) | It is connected with the<br>electrosurgical unit.<br>It transfers the radiofrequency<br>energy through the Single use<br>RF electrode (Rounded type) |
| Active accessory<br>(RF Electrode)                 | Needle type                                                          | 510(k) cleared by FDA<br>(K171707)                                                                                                                      | 510(k) cleared by FDA<br>(K171707)                                                                                                                   |
|                                                    | Rounded<br>electrode type                                            | Models: AG-CN-23G                                                                                                                                       | Add models (AG-CN21G,<br>AG-CN20G, AG-CN19G,<br>AG-CN18G)                                                                                            |
|                                                    | Neutral electrode pad                                                | 510(k) cleared by FDA<br>(K102372)                                                                                                                      | 510(k) cleared by FDA<br>(K102372)                                                                                                                   |
|                                                    | FDA Approval                                                         |                                                                                                                                                         |                                                                                                                                                      |
|                                                    | Functions                                                            | For emitting RF energy into<br>electrode.                                                                                                               | For emitting RF energy into<br>electrode.                                                                                                            |
| Miscellaneous accessory<br>(Foot switch)           | Performance<br>Specifications                                        | Single pole, single throw                                                                                                                               | Single pole, single throw                                                                                                                            |
|                                                    | Physical<br>Specification                                            | Single pedal, IPX8                                                                                                                                      | Single pedal, IPX8                                                                                                                                   |

{5}------------------------------------------------

| Rounded<br>electrode type | Model                         | AG-CN-23G     | AG-<br>CN21G | AG-<br>CN20G  | AG-<br>CN19G | AG-<br>CN18G |
|---------------------------|-------------------------------|---------------|--------------|---------------|--------------|--------------|
|                           | Diameter (ø)                  | 0.64          | 0.84         | 0.90          | 1.09         | 1.28         |
|                           | Rounded Electrode<br>Length   | 40mm ± 5%     |              | 40mm ± 5%     |              |              |
|                           | Rounded Electrode<br>material | SUS 304       |              | SUS 304       |              |              |
|                           | Cover tip length              | 38mm ± 5%     |              | 38mm ± 5%     |              |              |
|                           | Cover tip<br>thickness        | 9.14mm ± 5%   |              | 9.14mm ± 5%   |              |              |
|                           | Cover tip material            | Polycarbonate |              | Polycarbonate |              |              |
|                           | Total length                  | 77.5mm ± 5%   |              | 77.5mm ± 5%   |              |              |

{6}------------------------------------------------

| Sterility  | E.O gas<br>Sterilization | E.O gas Sterilization |
|------------|--------------------------|-----------------------|
| Single Use | Single Use               | Single Use            |

### - Differences between Subject and Predicates Device

Predicate device's electrode is AG-CN23G. Modified to add electrode AG-CN21G, AG-CN20G, AG-CN19G, AG-CN18G. The each of model's difference is diameter of electrode. Among the same products, the product with the smallest diameter was added to the previous submission, and the remaining electrodes are added this time.

- Discussion

The modifications expressed in this 510(k) Premarket Notification do not change the intended use, nor alter the fundamental scientific technology of the device. The modified devices are as safe and as effective as the predicate device.

### 6. Biocompatibility

Biocompatibility tests were conducted to ensure that no risks arise from biological hazards associated with materials of manufacture and the final device.

- 1) ISO 10993-1: 2009, Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process
- 2) ISO 10993-5:2009, Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity
- 3) ISO 10993-10:2010, Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization.

### 7. Sterilization

The EO gas sterilization method of the additional models RF electrode was validated according to the FDA recognized standards. In the validation, SAL (10th) and EO gas residuals also was verified.

- 1) ISO 11135:2014 + A1:2018, Sterilization of health-care products Ethylene oxide Requirements for the development, validation and routine control of a sterilization process for medical devices
- 2) ISO 10993-7: 2008+A1:2019, Biological evaluation of medical devices Part 7: Ethylene oxide sterilization residuals

### Performance Testing - Bench Testing 8.

The RF output power testing was performed in accordance with the FDA recognized standard, IEC 60601- 2-2:2017, Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories.

#### 9. Performance Testing - Animal Testing

{7}------------------------------------------------

Animal study was performed to assess thermal tissue damage/spread using mini pig tissues. The skin, liver, kidney, and femoral muscle dissected from 3 mini pigs were used for thermal tissue spread experiment. The skin from 1 mini pig was used in the histopathological analysis for thermal tissue damage experiment. Thermal imaging area analysis, the maximum temperature, the time to reach basal temperature, and histopathological analysis were evaluated.

### Conclusion

The major consideration such as intended use and principle of operation is not changed. Also, it is only adding more models of rounded electrode. The changes (modifications) of the device were verified and validated with biocompatibility of material. We conclude that the modified device is substantially equivalent to the legally marketed (existed) device.

---

**Source:** [https://fda-staging.innolitics.com/device/K223805](https://fda-staging.innolitics.com/device/K223805)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).