hyperion X9 pro, NewTom GiANO HR, X-RADiUS TRiO PLUS
Applicant
Cefla S.C.
Product Code
OAS · Radiology
Decision Date
Jan 17, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The device is a digital panoramic, cephalometric and tomographic extra-oral X-ray system, intended to: (I) produce orthopanoramic images of the maxillofacial region and carry out diagnostic examination on teeth, dental arches and other structures in the oral cavity; (II) produce radiographs of jaws, parts of the skull and carpus for the purpose of cephalometric examination, if equipped with tele-radiographic arm (CEPH); (III) the production of tomographic images of the head, including the ear, nose and throat (ENT), of the dento-maxillo-facial complex, teeth, mandible and maxilla, temporomandibular-articular joint (TMJ), other areas of the human skull and neck with sections of the cervical spine for use in diagnostic support, if equipped with the CBCT option. The device is operated and used by physicians, dentists, x-ray technologists and other legally qualified professionals.
Device Story
Extra-oral X-ray system for dental/maxillofacial imaging; produces panoramic, cephalometric, and CBCT images. System uses X-ray tube generator and solid-state sensors (CMOS or amorphous silicon flat panel) to acquire digital image data. Device rotates around patient; collimates X-ray beam to region of interest to minimize dose. Patient sits or stands; stabilized by craniostat/cephalostat. Class 1 laser pointers define positioning. Operator uses control panel for exposure parameters and system status. Digital data processed into reconstructed images via NNT/iRYS software; transferred to computer for clinician review. Used in dental/clinical settings by physicians, dentists, or technologists. Supports diagnostic examination of teeth, jaws, TMJ, and cervical spine.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), radiation protection (IEC 60601-1-3), and image quality verification.
Technological Characteristics
Digital X-ray system; X-ray tube generator; CMOS (scintillator/direct conversion) and amorphous silicon flat panel sensors. Fan-beam (PAN/CEPH) and cone-beam (CBCT) geometry. Class 1 laser positioning. Connectivity via NNT/iRYS software. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-63, IEC 60825-1, IEC 62304.
Indications for Use
Indicated for adult and pediatric patients requiring diagnostic imaging of the maxillofacial region, teeth, dental arches, jaws, skull, carpus, ENT structures, and cervical spine via panoramic, cephalometric, or CBCT radiography.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
hyperion X9 pro, NewTom GIANO HR, X-RADIUS TRIO PLUS (K214084)
Submission Summary (Full Text)
{0}------------------------------------------------
January 17, 2023
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box. To the right of the box is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
CEFLA S.C. % Lorenzo Bortolotti Regulatory Affairs Via Selice Provinciale 23/A Imola. BO 40026 ITALY
Re: K223794
Trade/Device Name: hyperion X9 pro, NewTom GiANO HR, X-RADiUS TRiO PLUS Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: OAS Dated: December 19, 2022 Received: December 19, 2022
# Dear Lorenzo Bortolotti:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
2023.01.17
*Lu Jiang* 12:32:19 -05'00'
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K223794
#### Device Name
hyperion X9 pro, NewTom GiANO HR, X-RADiUS TRiO PLUS
Indications for Use (Describe)
hyperion X9 pro, NewTom GiANO HR, X-RADiUS TRiO PLUS is digital panoramic, cephalometric and tomographic extra-oral system, intended to:
(1) produce orthopanoramic images of the maxillofacial region and carry out diagnostic examination on teeth, dental arches and other structures in the oral cavity;
(II) produce radiographs of jaws, parts of the skull and carpus for the purpose of cephalometric examination, if equipped with tele-radiographic arm (CEPH);
(III) the production of tomographic images of the head, including the ear, nose and throat (ENT), of the dentomaxillofacial complex, teeth, mandible and maxilla, temporomandibular-articular joint (TMD), other areas of the human skull and neck with sections of the cervical spine for use in diagnostic support, if equipped with the CBCT option. The device is operated and used by physicians, dentists, x-ray technologists and other legally qualified professionals.
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| | 510(k) SUMMARY AS REQUIRED BY 21 CFR 807.92 | | | | | K223794 |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|------------|--------------|--|---------|
| Submitter's Name: | CEFLA S.C. | | | | | |
| Address: | Via Selice Provinciale 23/a | | | | | |
| | Imola, BO 40026 ITALY | | | | | |
| | Tel. +39 0542 653111 | | | | | |
| | Fax +39 0542 653444 | | | | | |
| Establishment<br>Registration Number: | 3006610845 | | | | | |
| Contact Person: | Lorenzo Bortolotti, Regulatory Affairs | | | | | |
| Telephone Number: | +39 0542 653441 | | | | | |
| Email Address: | regulatory@cefla.it | | | | | |
| Date prepared: | December 19th, 2022 | | | | | |
| Device name: | hyperion X9 pro, NewTom GIANO HR, X-RADIUS TRIO PLUS | | | | | |
| Common Name: | Extra oral source, Panoramic, Cephalometric, Computed tomography X-ray system | | | | | |
| Device Classification<br>Name: | X-Ray, Tomography, Computed, Dental | | | | | |
| Regulation Number: | 21 CFR §892.1750 | | | | | |
| Device Class: | Class II | | | | | |
| Classification<br>Product Code: | OAS | | | | | |
| Subsequent Product<br>Code: | MUH | | | | | |
| Device Description: | The Proposed device is a panoramic, cephalometric and tomographic radiological system developed and manufactured by CEFLA S.C. The proposed device<br>is a change of the predicate device: hyperion X9 pro, NewTom GIANO HR, X-RADIUS TRIO PLUS (K214084).<br>Like the predicate device the proposed device can be sold under three different proprietary product name and brands for commercial needs, without<br>changing any of the basic safety, essential performances and functional features: | | | | | |
| | # | Device Name | Brand | Manufacturer | | |
| | 1 | hyperion X9 pro | myray | CEFLA S.C. | | |
| | 2 | NewTom GIANO HR | NewTom | CEFLA S.C. | | |
| | 3 | X-RADIUS TRIO PLUS | Castellini | CEFLA S.C. | | |
Wherever the Proposed device is mentioned the device with its three different trade/ proprietary names: hyperion X9 pro, New Tom GiANO HR and X-RADiUS TRiO PLUS.
{4}------------------------------------------------
| | Like the predicate device hyperion X9 pro, NewTom GIANO HR, X-RADIUS TRIO PLUS (K214084) the proposed device is equipped with X-ray tube generator<br>and X-ray sensors (Solid State X-ray Imaging Detectors) for dental panoramic (PAN), cephalometric radiography (CEPH) and computed<br>tomography (CBCT). The proposed device permits to acquire radiological images (panoramic images, cephalometric images and 3D volumes) at varying<br>radiographic angles by rotating around the patient following different trajectories depending on the selected examination. The exposed area can be<br>adapted to a specific region of interest to keep the radiation dose as low as possible for the patient. This is achieved by collimating the x-ray beam and the<br>adjustment of starting and ending points of the x-ray source and sensor movement. Furthermore, the radiation dose can be adapted by various parameters<br>such as examination types and exposure technique factors. Class I lasers pointers are utilized to define reference lines for the patient position. The patient<br>stabilized through adjustable patient supports, can sit or stand. Control panel allows user actions as: height adjustment, selection of examination, and<br>exposure parameters and delivers information about the unit status. The obtained digital image data are processed to provide a reconstructed image. The<br>images are transferred to a computer, in real time or later depending on the needs and choice of the operator. The software used to manage the images,<br>essential for CBCT acquisitions, is NNT/iRYS, a radiological imaging software developed by CEFLA S.C. |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use: | The device is a digital panoramic, cephalometric and tomographic extra-oral X-ray system, intended to:<br>(1) produce orthopanoramic images of the maxillofacial region and carry out diagnostic examination on teeth, dental arches and other structures in the<br>oral cavity;<br>(II) produce radiographs of jaws, parts of the skull and carpus for the purpose of cephalometric examination, if equipped with tele-radiographic arm (CEPH);<br>(III) the production of tomographic images of the head, including the ear, nose and throat (ENT), of the dento-maxillo-facial complex, teeth, mandible and<br>maxilla, temporomandibular-articular joint (TMJ), other areas of the human skull and neck with sections of the cervical spine for use in diagnostic support,<br>if equipped with the CBCT option.<br>The device is operated and used by physicians, dentists, x-ray technologists and other legally qualified professionals. |
| Identification of<br>Predicate Device | The predicate device identified relating the substantial equivalence of the proposed device is:<br>Device Names: hyperion X9 Pro, NewTom GIANO HR, X-RADIUS TRIO PLUS<br>510(k) Number: K214084<br>Device Classification Name: X-Ray, Tomography, Computed, Dental<br>Applicant: CEFLA S.C.<br>Via Selice Provinciale 23/A<br>Imola, IT 40026<br>Regulation Number: 21 CFR §892.1750<br>Device Class: Class II<br>Classification Product Code: OAS<br>Subsequent Product Code: MUH |
{5}------------------------------------------------
Substantial Both Proposed device and Predicate device (K214084) permit the aquisition of panoramic images (CEPH) and cone beam Equivalence computed tomography images (CBCT).
The proposed device is a change of the Cefla's own legally marketed predicate device: hyperion X9 Pro, NewTom GiANO HR, X-RADUS TRO PLUS (K214084).
This device change makes available for the acquisition of 2D images (PAN, CEPH) alternative solid-state X-ray detectors already used in the predicate device hyperion X9 Pro, NewTom GiANO HR, X-RADiUS TRiO PLUS (K214084).
The following cross reference table shows similarity and diversity aspects between the proposed device and the Predicate device.
| | Proposed Device | Predicate Device | Justification for Differences |
|-------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | hyperion X9 pro,<br>NewTom GiANO HR,<br>X-RADIUS TRIO PLUS | hyperion X9 pro,<br>NewTom GiANO HR,<br>X-RADIUS TRIO PLUS | |
| Manufacturer | CEFLA S.C. | CEFLA S.C. | |
| 510(K) No. | - | K214084 | |
| Figure | Image: Proposed Device | Image: Predicate Device | The external appearances and materials<br>between Proposed device and Predicate<br>device are identical. |
| Classification and indication for use | | | |
| Regulation Number | 21 CFR 892.1750 | 21 CFR 892.1750 | No difference. |
| Regulatory Class | Class II | Class II | No difference. |
| Classification Product<br>Code | OAS (Classification Product code)<br>MUH (Subsequent Product code) | OAS (Classification Product code)<br>MUH (Subsequent Product code) | No difference. |
| Classification Name: | Computed Tomography X-ray System | Computed Tomography X-ray System | No difference. |
| | | | |
| Indication for use | The device is a digital panoramic, cephalometric and tomographic extra-oral X-ray system, intended to: (I) produce orthopanoramic images of the maxillofacial region and carry out diagnostic examination on teeth, dental dental arches and other structures in the oral cavity; (II) produce radiographs of jaws, parts of the skull and carpus for the purpose of cephalometric examination, if equipped with tele-radiographic arm (CEPH); (III) the production of tomographic images of the head, including the ear, nose and throat (ENT), of the dento-maxillo-facial complex, teeth, mandible and maxilla, temporomandibular-articular joint (TMJ), other areas of the human skull and neck with sections of the cervical spine for use in diagnostic support, if equipped with the CBCT option. The device is operated and used by physicians, dentists, x-ray technologists and other legally qualified professionals. | The device is a digital panoramic, cephalometric and tomographic extra-oral X-ray system, intended to: (I) produce orthopanoramic images of the maxillofacial region and carry out diagnostic examination on teeth, dental arches and other structures in the oral cavity; (II) produce radiographs of jaws, parts of the skull and carpus for the purpose of cephalometric examination, if equipped with tele-radiographic arm (CEPH); (III) the production of tomographic images of the head, including the ear, nose and throat (ENT), of the dento-maxillo-facial complex, teeth, mandible and maxilla, temporomandibular-articular joint (TMJ), other areas of the human skull and neck with sections of the cervical spine for use in diagnostic support, if equipped with the CBCT option. The device is operated and used by physicians, dentists, x-ray technologists and other legally qualified professionals. | No difference. |
| Performance features | | | |
| Performance<br>specification | Panoramic<br>Computed tomography<br>Cephalometric | Panoramic<br>Computed tomography<br>Cephalometric | No difference. |
| Patient population | Adult, Pediatric | Adult, Pediatric | No difference. |
| Exposition selectable | 2D: PAN, BTW (bitewing), DENT, SIN,<br>TMJ, CEPH | 2D: PAN, BTW (bitewing), DENT, SIN,<br>TMJ, CEPH | No difference. |
| | | | |
| Rated input | 20A @ 115V~ | 20A @ 115V~ | No difference. |
| | 12A @ 240V~ | 12A @ 240V~ |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.