K223779 · Encore Medical, LLC · LPH · Feb 13, 2023 · Orthopedic
Device Facts
Record ID
K223779
Device Name
EMPOWR™ blade stem
Applicant
Encore Medical, LLC
Product Code
LPH · Orthopedic
Decision Date
Feb 13, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The hip joint metal uncemented prosthesis is intended to replace a hip joint. The device is intended for primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intraoperative instability and for whom all other options to constrained acetabular components have been considered. Joint replacement is indicated for patients suffering from disability due to: - noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis of the . natural femoral head; - . rheumatoid arthritis; - . correction of functional deformity; - femoral fracture; and . - . revision procedures where other treatments or devices have failed. This stem is to be press-fit. This stem is intended for cementless use.
Device Story
EMPOWR™ blade stem is a press-fit, porous-coated, wedge-style femoral stem for total hip arthroplasty. It serves as a prosthetic replacement for the femoral head. Designed for cementless implantation, it features enhancements over the predicate Linear® Hip Stem, including coxa vara offset, updated neck angle, and reduced distal angle. The device is implanted by surgeons in a clinical setting using a reproducible surgical technique compatible with various surgical approaches. It is compatible with existing Enovis™ femoral heads and acetabular systems. By replacing the diseased or damaged hip joint, the device aims to restore patient mobility and reduce pain. It is intended for permanent implantation.
Clinical Evidence
No clinical data submitted. Substantial equivalence is supported by bench testing, including CAD-based range of motion analysis (ISO 21535:2007), distal stem fatigue testing (ISO 7206-4:2010), and neck stem fatigue testing (ISO 7206-6:2013). Biocompatibility was evaluated per ISO 10993-1.
Technological Characteristics
Press-fit, wedge-style femoral stem; porous titanium coating; permanent implant; cementless use. Materials are qualified for permanent tissue/bone contact. Testing standards: ISO 7206-4 (distal fatigue), ISO 7206-6 (neck fatigue), ISO 21535 (ROM).
Indications for Use
Indicated for primary or revision total hip arthroplasty patients, including those with high risk of dislocation (prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, intraoperative instability), noninflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, or femoral fracture.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 13, 2023
Encore Medical LLC Trev Thorsen Regulatory Affairs 9800 Metric Blvd Ste 200 Austin, Texas 78758
Re: K223779
Trade/Device Name: EMPOWR™ blade stem Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: LPH, LZO, KWY, KWZ Dated: December 15, 2022 Received: December 16, 2022
Dear Trey Thorsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Christopher Ferreira -S
for
Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known) K223779
Device Name EMPOWR™ blade stem
#### Indications for Use (Describe)
The hip joint metal uncemented prosthesis is intended to replace a hip joint. The device is intended for primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intraoperative instability and for whom all other options to constrained acetabular components have been considered.
Joint replacement is indicated for patients suffering from disability due to:
- noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis of the . natural femoral head;
- . rheumatoid arthritis;
- . correction of functional deformity;
- femoral fracture; and .
- . revision procedures where other treatments or devices have failed.
This stem is to be press-fit. This stem is intended for cementless use.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><div><span style="font-size:20px">☒</span></div><div>Prescription Use (Part 21 CFR 801 Subpart D)</div></div> | <div style="display:flex; align-items:center;"><div><span style="font-size:20px">☐</span></div><div>Over-The-Counter Use (21 CFR 801 Subpart C)</div></div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Page 1 of 3
Image /page/3/Picture/1 description: The image shows the logo for "enovis". The word "enovis" is written in a stylized font, with the "e", "n", and the first part of the "o" in red. The second part of the "o" is a black circle with a gradient effect, and the "vis" is in black. There is a trademark symbol to the bottom right of the logo.
# 510(k) Summary
14DEC2022
#### I. SUBMITTER
Encore® Medical, L.P. (dba Enovis™) 9800 Metric Blvd. Austin, TX 78758
Contact Person: Trey Thorsen Email: trey.thorsen@enovis.com Phone: 850-450-3932
### II. DEVICE
Name of Device: EMPOWR™ blade stem Common or Usual Name: Total Hip Implant Classification Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis – 21 CFR 888.3358 Regulatory Class: II Primary Product Code: LPH
Secondary Product Code: LZO, KWY, KWY, KWZ
#### III. PREDICATE DEVICE
Primary Predicate: Enovis™ Linear® Hip Stem, K120241 Reference Predicate: Zimmer Biomet™ TaperLoc® Complete Stem, K120030
### IV. DEVICE DESCRIPTION
The EMPOWR™ blade stem represents the latest generation of a press-fit, porous coated, wedge style femoral stem. Building on the design principles of its predicate, Linear® Hip Stem, the EMPOWR™ blade stem system features enhancements such as addition of coxa vara offset, neck angle update, and distal angle reduction for wider application.
The EMPOWR™ blade stem is applicable for use in total hip arthroplasty and the prosthetic replacement of femoral neck fractures in all age groups. It offers a stem design with a reproducible surgical technique carefully developed to enable consistent and accurate implantation. The EMPOWR™ blade stem features streamlined and accurate instrumentation that is highly adaptable to any surgical approach.
### V. INTENDED USE/INDICATIONS FOR USE
Intended Use:
Enovis™ hip devices are intended for treatment of patients who are candidates for total hip arthroplasty per the indications for use. While hip replacements are not intended to withstand activity levels and loads of normal healthy bone, they are a means of restoring mobility and reducing pain for many patients.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for "enovis". The "eno" portion of the logo is red, while the "o" is stylized as a black and grey ring. The "vis" portion of the logo is black, and the trademark symbol is in the lower right corner.
| Page 2 of 3
Indications for Use:
The hip joint metal uncemented prosthesis is intended to replace a hip joint. The device is intended for primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intraoperative instability and for whom all other options to constrained acetabular components have been considered.
Joint replacement is indicated for patients suffering from disability due to:
- noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis of the ● natural femoral head;
- rheumatoid arthritis;
- correction of functional deformity;
- femoral fracture; and
- revision procedures where other treatments or devices have failed.
This stem is to be press-fit. This stem is intended for cementless use.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The EMPOWR™ blade stem and accessories are technologically the same when compared to the predicate Linear Hip Stem (cleared via K120241). Comparative testing demonstrates substantial equivalence between the subject and predicate device. The stem trunnions and the porous titanium coating are identical between the subject and predicate device.
There is a slight difference between the subject device ("EMPOWR™ blade stem") indications for use and the predicate device, Linear® stem ("Linear®") (K120241) indications for use statement. The added statements are for clarity and do not change the overall intent of the indications.
The subject EMPOWR™ blade stem has one size smaller and one size larger than the predicate Enovis™ Linear® stem. The subject EMPOWR™ blade stem has two fewer sizes than the predicate Biomet® TaperLoc® Complete stem. The smallest size is equivalent between the two systems. The largest subject stem is not larger than the largest Biomet® TaperLoc® Complete stem. The size offering for the subject device is within the size offering of the predicate Biomet® TaperLoc® Complete stem offering.
The component compatibility of the EMPOWR™ blade stem is identical to the Linear® stem (K120241) as femoral stems are compatible with all Enovis™ femoral heads and acetabular systems, including Encore® previously cleared FMP® and EMPOWR™ semi constrained acetabular shells, previously cleared femoral heads, previously cleared unipolar heads/sleeves, EMPOWR™ dual mobility and mobility system, bipolar systems, and FMP® constrained system.
#### Biocompatibility testing
The biocompatibility evaluation for the EMPOWR™ blade stem was conducted in accordance with the FDA guidance, Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing, May 1, 1995, and International Standard ISO 10993-1 Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process, as recognized by FDA.
{5}------------------------------------------------
The EMPOWR™ blade stem is intended for permanent implantation, contacting tissue/bone. Biocompatibility was ensured through the use of qualified materials and contact agents.
#### Performance Testing
Performance testing was completed on the subject EMPOWR™ blade stem and is detailed in the Performance Testing section. The following testing was performed to FDA recognized standards and internal protocols that is specific to the EMPOWR™ blade stem:
- CAD Based Range of Motion (ROM) ISO 21535:2007 ●
- Distal Stem Fatigue Testing – ISO 7206-4:2010
- Neck Stem Fatigue Testing ISO 7206-6:2013 ●
#### Animal Studies
No animal data submitted.
Clinical Studies No clinical data submitted.
### VIII. CONCLUSIONS
All testing and evaluations demonstrate that the subject device is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.