K223775 · South Dakota Partners · GXY · Jan 3, 2023 · Neurology
Device Facts
Record ID
K223775
Device Name
EZ-STIK Electrodes
Applicant
South Dakota Partners
Product Code
GXY · Neurology
Decision Date
Jan 3, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1320
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The EZ-STIK Electrodes are intended for use as a disposable, conductive adhesive interface between the patient's skin and the electrical stimulator. The EZ-STIK Electrodes are designed and intended to be used with marketed electrical stimulators, i.e. TENS (Transcutaneous Electrical Nerve Stimulation), EMS (Electrical Muscle Stimulation), IF (Inferential), or PGF (Pulsed Galvanic Stimulation). The electrode is for OTC (Over-the-Counter) or Prescription Use.
Device Story
EZ-STIK Electrodes function as a conductive interface between patient skin and electrical stimulators (TENS, EMS, IF, PGF). Device consists of a 3-layer flexible structure: insulating backing (non-woven polyester or PVC foam), conductive carbon/silver film, and biocompatible self-adhesive hydrogel. Electrodes connect to stimulators via lead wire with standard .080" female socket. Used in clinical or home settings by patients or clinicians. Device delivers electrical current from stimulator to skin; adhesive hydrogel secures electrode to patient. Benefits include non-invasive delivery of electrical stimulation therapy for muscle or nerve stimulation.
Clinical Evidence
No clinical data were necessary for this submission. Bench testing included biocompatibility (cytotoxicity, sensitization, irritation per ISO 10993-1), electrical performance, pull/peel strength, and shelf-life testing. Subject device performance was compared directly to predicate K050469, with results confirming equivalence.
Technological Characteristics
3-layer construction: non-woven polyester or soft PVC foam backing, carbon/silver conductive film, and biocompatible adhesive hydrogel. Connection via .080" female socket lead wire. Non-sterile. Complies with 21 CFR Part 898 for lead wire safety. Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for patients requiring TENS, EMS, IF, or PGF electrical stimulation therapy. Intended for single-patient, multiple-application use by healthcare professionals or lay persons in OTC or prescription settings.
Regulatory Classification
Identification
A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in a blue box. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 3, 2023
South Dakota Partners % Dave Yungvirt Official Correspondent Third Party Review Group, LLC 25 Independence Blvd Warren, New Jersey 07059
Re: K223775
Trade/Device Name: EZ-STIK Electrodes Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous electrode Regulatory Class: Class II Product Code: GXY Dated: December 14, 2022 Received: December 16, 2022
Dear Dave Yungvirt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Heather L. Dean -S
Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K223775
Device Name EZ-STIK Electrodes
## Indications for Use (Describe)
The EZ-STIK Electrodes are intended for use as a disposable, conductive adhesive interface between the patient's skin and the electrical stimulator. The EZ-STIK Electrodes are designed and intended to be used with marketed electrical stimulators, i.e. TENS (Transcutaneous Electrical Nerve Stimulation), EMS (Electrical Muscle Stimulation), IF (Inferential), or PGF (Pulsed Galvanic Stimulation). The electrode is for OTC (Over-the-Counter) or Prescription Use.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------|-----------------------------------------------|
| \[X] Prescription Use (Part 21 CFR 801 Subpart D) | |
| | 区 Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for South Dakota Partners, a Salona Global Medical Device Company. The words "South Dakota" and "Partners" are in a bold, blue font. Below the words is the phrase "A Salona Global Medical Device Company" in a smaller, green font. To the right of the text is a graphic of a sun with green leaves.
205 Hwy 22 E Clear Lake, SD 57226 www.sdpartnersinc.com T: (800) 626-3098
## 510(k) Summary
## 21 CFR 807.92(a)(1)
| Applicant Name | South Dakota Partners | |
|---------------------------------|---------------------------------|--|
| Applicant Address | 205 Hwy 22 E | |
| | Clear Lake, SD 57226 | |
| | United States | |
| Applicant Contact | Mike Plunkett | |
| Applicant Contact Email | Mike.plunkett@sdpartnersinc.com | |
| Correspondent Name | Simbex | |
| Correspondent Address | 10 Water St Suite 410 | |
| | Lebanon, NH 03766 | |
| | United States | |
| Correspondent Contact Telephone | 504-432-8171 | |
| Correspondent Contact | Amaris Ajamil, PhD, RAC | |
| Correspondent Contact Email | aajamil@simbex.com | |
## Device Name
Contact Details
Device Trade Name Common Name Classification Name Regulation Number Product Code
EZ-STIK Electrodes EZ-STIK Electrodes Electrode, Cutaneous 21 CR 882.1320 GXY
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
|------------------------------------|----------------------|---------------------|
| Predicate # | Predicate Trade Name | Product Code |
| K050469 | EZ-STIK Electrodes | GXY |
| K160138 | Adhesive Electrodes | GXY |
## Device Description Summary
21 CFR 807.92(a)(4)
EZ-STIK Electrodes have a basic 3-layer construction of non-woven polyester material or soft PVC foam, carbon film coated with silver and conductive hydrogel. EZ-STIK Electrodes are multi-layer reusable, flexible structures composed of laminated materials commonly used in this application.
First layer: Insulating backing material- non-woven polyester material or soft PVC foam coated with biocompatible medical adhesive tape
Second layer: Conductive carbon-carbon film or Carbon film/Carbon film coated with silver film
Third layer: Biocompatible self-adhesive conductive hydrogel
The electrodes are designed for single-patient/multiple application use. Because of the biocompatible conductive hydrogel's adhesive nature, no securing materials are required to secure the device to the
## 21 CFR 807.92(a)(2)
{4}------------------------------------------------
## 205 Hwv 22 E Clear Lake, SD 57226 www.sdpartnersinc.com T: (800) 626-3098
patient's skin. The electrode is connected to the electrical stimulator by lead wire, with a standard .080" female socket connector. By design, the insulated outer jacket prevents the conductive connection to earth or hazardous voltages. Wire assembly complies with FDA performance standard 21 CFR Part 898.
## Intended Use/Indications For Use
The EZ-STIK Electrodes are intended for use as a disposable, conductive adhesive interface between the patient's skin and the electrical stimulator. The EZ-STIK Electrodes are designed and intended to be used with marketed electrical stimulators, i.e. TENS (Transcutaneous Electrical Nerve Stimulation), EMS (Electrical Muscle Stimulation), IF (Inferential), or PGF (Pulsed Galvanic Stimulation). The electrode is for OTC (Over-the-Counter) or Prescription Use.
## Indications For Use Comparison
The subject and the predicate devices (K050469 and K160138) have the same intended use. The subject and predicate devices are cutaneous electrodes intended for use as a disposable, reusable, conductive adhesive interface between the patient's skin and the electrical stimulator. The subject and predicate devices are intended to be single patient reusable. The subject EZ-STIK electrode and predicate Adhesive Electrode (K160138) are indicated as OTC and prescription devices intended for use by healthcare professionals and lay persons, whereas the predicate EZ-STIK Electrode is indicated for prescription use only. As such, the subject EZ-STIK Electrode is substantially equivalent to the predicate device Adhesive Electrode with respect to intended use and indications for use.
## Technological Comparison
At a high level, the subject and predicate devices (K050469 and K160138) are based on the same technological elements:
- Use of three layers (insulating material, conductive carbon film with or without silver layer, biocompatible adhesive hydrogel)
- Use of Leadwire for electrical connections
- Provided non-sterile
- Greater than 6lbf pull strength
The following technological differences exist between the subject and predicate devices:
- Use of different hydrogel material
## Non-Clinical and/or Clinical Tests Summary & Conclusions
The following performance data were provided in support of the substantial equivalence determination.
The biocompatibility evaluation for the EZ-STIK Electrodes was conducted in accordance with International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:
Cytotoxicity, Sensitization, and Irritation. The EZ-STIK Electrode is considered direct tissue contacting for a duration of less than 24 hours.
21 CFR 807.92(a)(5)
21 CFR 807.92(a)(5)
## 21 CFR 807.92(a)(6)
# Use
tended for use as a disposable, reu
ne electrical stimulator. The EZ-S
Clear La
<www.sdparks.org>
T: (8
21 CFR 807.92(b)
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for South Dakota Partners, a Salona Global Medical Device Company. The words "South Dakota" and "Partners" are in a bold, blue font. Below the words is the phrase "A Salona Global Medical Device Company" in a smaller font. To the right of the words is a graphic of a sun with green leaves.
To verify the safety and performance of the EZ-STIK Electrode, the subject device was tested in the same method as the previously cleared EZ-STIK Electrode (K050469), including electrical performance test, pull and peel test, and shelf-life test.
No clinical data were necessary for this submission.
For all electrical and physical performance tests conducted, the subject device results were shown to be within the acceptance range generated from direct testing of predicate device. The test results and information contained in this submission demonstrate that any differences in technological characteristics do not raise any new questions of safety or effectiveness. Thus, the subject EZ-STIK Electrodes are substantially equivalent to the predicate EZ-STIK Electrodes.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.