← Product Code [GCJ](/productcode/GCJ) · K223759

# Arthrex SynergyID Endoscopic Imaging System (K223759)

_Arthex, Inc. · GCJ · Feb 24, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K223759

## Device Facts

- **Applicant:** Arthex, Inc.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Feb 24, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Arthrex SynergyID Endoscopic Imaging System is intended to be used as an endoscopic video camera to provide visible light imaging in a variety of endoscopic diagnostic and surgical procedures, including but not limited to: orthopedic, spine, laparoscopic, urologic, sinuscopic, plastic surgical procedures within the thoracic cavity. The device is also intended to be used as an accessory for microscopic surgery. The Arthrex SynergyID Endoscopic Imaging System is indicated for use to provide real time endoscopic visible and near-infrared fluorescence imaging. Upon intravenous administration and use of ICG consistent with its approved label, the system enables surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visualization of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct or common hepatic duct), using near-infrared imaging. Fluorescence imaging of biliary ducts with the Arthrex SynergyID Endoscopic Imaging System is intended for use with standard of care white light, and when indicated, intraoperative cholangiography. The device is not intended for standalone use for billary duct visualization. Upon interstitial administration and use of ICG consistent with its approved label, the Arthrex SynergyID Endoscopic Imaging System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

## Device Story

Endoscopic video imaging system; includes camera control unit (CCU), camera heads, laser light source, and laparoscopes. Inputs: visible light and NIR fluorescence signals from surgical site following ICG administration. Operation: NIR laser excites ICG; fluorescence response captured by CMOS sensor; processed by CCU; displayed on monitor. Used in OR by surgeons for minimally invasive procedures. Output: real-time visible and NIR fluorescence video. Benefits: intraoperative visualization of perfusion, biliary anatomy, and lymphatic structures to guide surgical decision-making.

## Clinical Evidence

Bench testing only. Verification activities included engineering evaluation, functional analysis, and software validation/verification (moderate level of concern). Biocompatibility testing per ISO 10993-5 and 10993-10. Electrical safety and EMC testing per AAMI/ANSI ES60601-1 and IEC 60601-1-2.

## Technological Characteristics

System includes CCU, camera heads, laser light source, and laparoscopes. CMOS image sensor; 3840 x 2160 resolution; 60 FPS. NIR excitation at 785nm; detection bandwidth 810-940nm. Connectivity via integrated console with tablet interface. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, IEC 60825-1.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Arthrex Synergy® Endoscopic Imaging System ([K202582](/device/K202582.md))
- PINPOINT Endoscopic Fluorescence Imaging System ([K182606](/device/K182606.md))
- TIPCAM1 Rubina Video Endoscope System ([K201526](/device/K201526.md))
- Arthrex NanoScope System ([K201134](/device/K201134.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/10 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

February 24, 2023

Arthex Inc. Troy Brooks Regulatory Affairs Team Lead 1370 Creekside Boulevard Naples, Florida 34108-1945

Re: K223759

Trade/Device Name: Arthrex SynergyID Endoscopic Imaging System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ, IZI Dated: February 16, 2023 Received: February 17, 2023

Dear Troy Brooks:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Colin K. Chen -S

Colin Kejing Chen Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K223759

#### Device Name

Arthrex SynergyID Endoscopic Imaging System

#### Indications for Use (Describe)

The Arthrex SynergyID Endoscopic Imaging System is intended to be used as an endoscopic video camera to provide visible light imaging in a variety of endoscopic diagnostic and surgical procedures, including but not limited to: orthopedic, spine, laparoscopic, urologic, sinuscopic, plastic surgical procedures within the thoracic cavity. The device is also intended to be used as an accessory for microscopic surgery.

The Arthrex SynergyID Endoscopic Imaging System is indicated for use to provide real time endoscopic visible and nearinfrared fluorescence imaging. Upon intravenous administration and use of ICG consistent with its approved label, the system enables surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visualization of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct or common hepatic duct), using near-infrared imaging. Fluorescence imaging of biliary ducts with the Arthrex SynergyID Endoscopic Imaging System is intended for use with standard of care white light, and when indicated, intraoperative cholangiography. The device is not intended for standalone use for billary duct visualization.

Upon interstitial administration and use of ICG consistent with its approved label, the Arthrex SynergyID Endoscopic Imaging System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

Type of Use (Select one or both, as applicable)

| <span style="font-family: sans-serif;"> <svg height="15" width="15"> <rect fill="none" height="15" stroke="black" width="15" x="0" y="0"></rect> <line stroke="black" x1="2" x2="13" y1="2" y2="13"></line> <line stroke="black" x1="13" x2="2" y1="2" y2="13"></line> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;"> <svg height="15" width="15"> <rect fill="none" height="15" stroke="black" width="15" x="0" y="0"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C)                                                                                                                        |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| Date Prepared                   | January 27, 2023                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter                       | Arthrex Inc.<br>1370 Creekside Boulevard<br>Naples, FL 34108-1945                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Contact Person                  | Troy Brooks, RAC<br>Regulatory Affairs Team Lead<br>1-239-643-5553<br>Troy.Brooks@Arthrex.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Name of Device                  | Arthrex Synergy® Endoscopic Imaging System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Common Name                     | Endoscopic Video Imaging System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Product Code                    | GCJ, IZI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Classification Name             | 21 CFR 876.1500: Endoscope and accessories<br>21 CFR 892.1600: Angiographic x-ray systems                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Regulatory Class                | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Primary Predicate<br>Device     | K202582 - Arthrex Synergy® Endoscopic Imaging System, Arthrex Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Additional Predicate<br>Devices | K182606 - PINPOINT Endoscopic Fluorescence Imaging System, Novadaq Technologies<br>K201526 - TIPCAM1 Rubina Video Endoscope System, Karl Storz Endoscopy America<br>K201134 - Arthrex NanoScope System, Arthrex Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Purpose of Submission           | This Traditional 510(k) premarket notification is submitted to obtain clearance for<br>expanded indications for use for the existing Arthrex Synergy® Endoscopic Imaging<br>System cleared under K202582.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Device Description              | The Arthrex Synergy® Endoscopic Imaging System includes a non-sterile camera<br>control unit (CCU) console, camera heads, a laser light source, and laparoscopes.<br>The system integrates ultra-high-definition camera technology, LED lighting, and<br>an image management system into a single console with a tablet interface. The<br>system provides real-time visible and near-infrared light illumination and imaging.                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                 | The Arthrex Synergy® Endoscopic Imaging System CCU interacts with the laser light<br>source to be able to provide near-infrared (NIR) imaging to visualize the presence of<br>Indocyanine Green (ICG). The ICG fluoresces when illuminated through a laparoscope<br>with NIR excitation light from the laser light source and the fluorescence response is<br>then imaged with the camera, processed and displayed on a monitor.                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Indications for Use             | The Arthrex Synergy® Endoscopic Imaging System is intended to be used as an<br>endoscopic video camera to provide visible light imaging in a variety of endoscopic<br>diagnostic and surgical procedures, including but not limited to: orthopedic, spine,<br>laparoscopic, urologic, sinuscopic, plastic surgical procedures, and procedures within<br>the thoracic cavity. The device is also intended to be used as an accessory for<br>microscopic surgery.                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                 | The Arthrex Synergy® Endoscopic Imaging System is indicated for use to provide real<br>time endoscopic visible and near-infrared fluorescence imaging. Upon intravenous<br>administration and use of ICG consistent with its approved label, the system enables<br>surgeons to perform minimally invasive surgery using standard endoscope visible<br>light as well as visualization of vessels, blood flow and related tissue perfusion, and<br>at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct or<br>common hepatic duct), using near-infrared imaging. Fluorescence imaging of biliary<br>ducts with the Arthrex Synergy® Endoscopic Imaging System is intended for use with<br>standard of care white light, and when indicated, intraoperative cholangiography.<br>The device is not intended for standalone use for biliary duct visualization. |
|                                 | Upon interstitial administration and use of ICG consistent with its approved label,<br>the Arthrex Synergy® Endoscopic Imaging System is used to perform intraoperative<br>fluorescence imaging and visualization of the lymphatic system, including lymphatic<br>vessels and lymph nodes.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |

## 510(k) Summary

{4}------------------------------------------------

### Technological Comparison

|                                                 | Arthrex<br>SynergyID<br>Endoscopic<br>Imaging System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Arthrex<br>SynergyID<br>Endoscopic<br>Imaging System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Novadaq<br>PINPOINT<br>Endoscopic<br>Fluorescence<br>Imaging System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Karl Storz<br>TIPCAM1 Rubina<br>Video Endoscope<br>System                                                                                                                                                                                                                                                                                                                                                         | Arthrex<br>NanoScope<br>System                                                                                                                                                                                                                                                                                                                                                                                                       | Comparison                                                                                                                                                                                                                                                                                                                                                                                                     |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device                                          | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | K182606<br>Additional<br>Predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | K201526<br>Additional<br>Predicate                                                                                                                                                                                                                                                                                                                                                                                | K201134<br>Additional<br>Predicate                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                |
| Classification                                  | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Class II                                                                                                                                                                                                                                                                                                                                                                                                          | Class II                                                                                                                                                                                                                                                                                                                                                                                                                             | Equivalent                                                                                                                                                                                                                                                                                                                                                                                                     |
| Product Code                                    | GCJ, IZI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | GCJ, IZI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | GCJ, IZI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | GCJ, FGB, HET                                                                                                                                                                                                                                                                                                                                                                                                     | GCJ, HRX                                                                                                                                                                                                                                                                                                                                                                                                                             | Equivalent                                                                                                                                                                                                                                                                                                                                                                                                     |
| 21 CFR                                          | 876.1500<br>Endoscope and<br>Accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 876.1500<br>Endoscope and<br>Accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 876.1500<br>Endoscope and<br>Accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 876.1500<br>Endoscope and<br>Accessories                                                                                                                                                                                                                                                                                                                                                                          | 876.1500<br>Endoscope and<br>Accessories                                                                                                                                                                                                                                                                                                                                                                                             | Equivalent                                                                                                                                                                                                                                                                                                                                                                                                     |
| Combination<br>Product                          | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | No                                                                                                                                                                                                                                                                                                                                                                                                                | No                                                                                                                                                                                                                                                                                                                                                                                                                                   | Equivalent                                                                                                                                                                                                                                                                                                                                                                                                     |
| Indications<br>For Use                          | The Arthrex<br>SynergyID<br>Endoscopic<br>Imaging System<br>is intended to<br>be used as an<br>endoscopic video<br>camera to<br>provide visual<br>imaging in a<br>variety of<br>endoscopic<br>diagnostic and<br>surgical<br>procedures,<br>including but<br>not limited to:<br>orthopedic,<br>spine,<br>laparoscopic,<br>urologic,<br>sinuscopic,<br>plastic surgical<br>procedures,<br>and procedures<br>within the<br>thoracic cavity.<br>The device is also<br>intended to be<br>used as an<br>accessory for<br>microscopic<br>surgery.<br><br>The Arthrex<br>SynergyID<br>Endoscopic<br>Imaging System<br>is indicated for<br>use to provide<br>real time<br>endoscopic<br>visible and<br>near-infrared<br>fluorescence<br>imaging. Upon<br>intravenous<br>administration<br>and use of ICG<br>consistent with<br>its approved                                                                                                                                                                                                                        | The Arthrex<br>SynergyID<br>Endoscopic<br>Imaging System<br>is intended to<br>be used as an<br>endoscopic video<br>camera in a<br>variety of<br>endoscopic<br>surgical<br>procedures,<br>including but<br>not limited to:<br>orthopedic,<br>laparoscopic,<br>urologic,<br>sinuscopic, and<br>plastic surgical<br>procedures. The<br>device is also<br>intended to be<br>used as an<br>accessory for<br>microscopic<br>surgery.<br><br>The Arthrex<br>SynergyID<br>Endoscopic<br>Imaging System<br>is indicated for<br>use to provide<br>real time<br>endoscopic<br>visible and<br>near-infrared<br>fluorescence<br>imaging. Upon<br>intravenous<br>administration<br>and use of ICG<br>consistent with<br>its approved<br>label, the system<br>enables surgeons<br>to perform<br>minimally<br>invasive surgery<br>using standard<br>endoscope | Upon<br>intravenous<br>administration<br>of TRADENAME<br>(ICG drug<br>product), the<br>PINPOINT<br>Endoscopic<br>Fluorescence<br>Imaging System<br>is used with<br>TRADENAME to<br>perform<br>intraoperative<br>fluorescence<br>angiography,<br>and it is also<br>indicated for use<br>in fluorescence<br>imaging of biliary<br>ducts, and when<br>indicated, during<br>intraoperative<br>cholangiography.<br><br>The PINPOINT<br>Endoscopic<br>Fluorescence<br>Imaging System<br>is indicated for<br>use to provide<br>real time<br>endoscopic<br>visible and near-<br>infrared<br>fluorescence<br>imaging. The<br>PINPOINT<br>System enables<br>surgeons to<br>perform<br>minimally<br>invasive surgery<br>using standard<br>endoscope<br>visible light as<br>well as visual<br>assessment of<br>vessels, blood<br>flow and related<br>tissue perfusion, | The TIPCAM1<br>Rubina Video<br>Endoscope<br>System is<br>intended to be<br>used together<br>with the camera<br>control unit<br>during diagnostic<br>and/or surgical<br>procedures<br>when endoscopic<br>video assistance<br>is required. For<br>use in all<br>endoscopy and<br>endoscopic<br>surgery within<br>the peritoneal<br>and thoracic<br>cavity, including<br>gynecological<br>and urological<br>anatomy. | The Arthrex<br>NanoScope<br>System is<br>intended to be<br>used as an<br>endoscopic video<br>camera in a<br>variety of<br>endoscopic<br>diagnostic and<br>surgical<br>procedures,<br>including but<br>not limited to:<br>orthopedic,<br>spine,<br>laparoscopic,<br>urologic,<br>sinuscopic, and<br>plastic surgical<br>procedures. The<br>device is also<br>intended to be<br>used as an<br>accessory for<br>microscopic<br>surgery. | The subject<br>device includes<br>the following<br>additional<br>indications<br>for use:<br>- fluorescence<br>imaging and<br>visualization<br>of the<br>lymphatic<br>system,<br>including<br>lymphatic<br>vessels and<br>lymph nodes<br>- use as an<br>endoscopic<br>video camera<br>(i.e. visual<br>imaging) in<br>diagnostic<br>and surgical<br>procedures<br>for spine and<br>within the<br>thoracic cavity |
|                                                 | to perform<br>minimally<br>invasive surgery<br>using standard<br>endoscope<br>visible light as<br>well as<br>visualization of<br>vessels, blood<br>flow and related<br>tissue perfusion,<br>and at least one<br>of the major<br>extra-hepatic<br>bile ducts (cystic<br>duct, common<br>bile duct or<br>common hepatic<br>duct), using<br>near-infrared<br>imaging.<br>Fluorescence<br>imaging of biliary<br>ducts with the<br>Arthrex Synergy®<br>Endoscopic<br>Imaging System<br>is intended for<br>use with<br>standard of care<br>white light, and<br>when indicated,<br>intraoperative<br>cholangiography.<br>The device is not<br>intended for<br>standalone use<br>for biliary duct<br>visualization.<br>Upon interstitial<br>administration<br>and use of ICG<br>consistent with<br>its approved<br>label, the Arthrex<br>SynergyD<br>Endoscopic<br>Imaging System<br>is used to<br>perform<br>intraoperative<br>fluorescence<br>imaging and<br>visualization of<br>the lymphatic<br>system, including<br>lymphatic vessels<br>and lymph<br>nodes. | visualization of<br>vessels, blood<br>flow and related<br>tissue perfusion,<br>and at least one<br>of the major<br>extra-hepatic<br>bile ducts (cystic<br>duct, common<br>bile duct or<br>common hepatic<br>duct), using<br>near-infrared<br>imaging.<br>Fluorescence<br>imaging of biliary<br>ducts with the<br>Arthrex Synergy®<br>Endoscopic<br>Imaging System<br>is intended for<br>use with<br>standard of care<br>white light, and<br>when indicated,<br>intraoperative<br>cholangiography.<br>The device is not<br>intended for<br>standalone use<br>for biliary duct<br>visualization.                                                                                                                                                                                                                                                | extra-hepatic<br>bile ducts (cystic<br>duct, common<br>bile duct or<br>common hepatic<br>duct), using<br>near-infrared<br>imaging.<br>Fluorescence<br>imaging of biliary<br>ducts with the<br>PINPOINT<br>System is<br>intended for use<br>with standard of<br>care white light,<br>and when<br>indicated,<br>intraoperative<br>cholangiography.<br>The device is not<br>intended for<br>standalone use<br>for biliary duct<br>visualization.<br>Upon interstitial<br>administration<br>of TRADENAME<br>(ICG drug<br>product), the<br>PINPOINT<br>System is used<br>to perform<br>intraoperative<br>fluorescence<br>imaging and<br>visualization of<br>the lymphatic<br>system, including<br>lymphatic vessels<br>and lymph<br>nodes.                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                |
| System<br>Components                            | Camera Control<br>Unit<br>Light Source<br>Camera Heads<br>Scopes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Camera Control<br>Unit<br>Light Source<br>Camera Heads<br>Scopes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Video Processor<br>Illuminator (VPI)<br>Camera Head<br>Scopes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Camera Control<br>Unit<br>Camera/Scope<br>(integrated)                                                                                                                                                                                                                                                                                                                                                            | Console<br>Camera/Scope<br>(integrated)                                                                                                                                                                                                                                                                                                                                                                                              | Equivalent to<br>the primary<br>predicate<br>K202582.<br>Similar to<br>predicates<br>K182606<br>and K201526.                                                                                                                                                                                                                                                                                                   |
| Imaging<br>Modes                                | White Light<br>NIR Fluorescence                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | White Light<br>NIR Fluorescence                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | White Light<br>NIR Fluorescence                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | White Light                                                                                                                                                                                                                                                                                                                                                                                                       | White Light                                                                                                                                                                                                                                                                                                                                                                                                                          | Equivalent<br>per white light<br>indications and<br>NIR fluorescence<br>indications.                                                                                                                                                                                                                                                                                                                           |
| Safety<br>Standards                             | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-18<br>IEC 60825-1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-18<br>IEC 60825-1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-18<br>IEC 60825-1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-18                                                                                                                                                                                                                                                                                                                                                                    | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-18                                                                                                                                                                                                                                                                                                                                                                                       | Equivalent                                                                                                                                                                                                                                                                                                                                                                                                     |
| Light<br>Source                                 | Integrated<br>(Visible)<br>External (NIR)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Integrated<br>(Visible)<br>External (NIR)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Integrated                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Integrated                                                                                                                                                                                                                                                                                                                                                                                                        | Integrated                                                                                                                                                                                                                                                                                                                                                                                                                           | Equivalent<br>to primary<br>predicate<br>K202582.                                                                                                                                                                                                                                                                                                                                                              |
| Image Sensor                                    | CMOS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | CMOS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | CMOS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | CMOS                                                                                                                                                                                                                                                                                                                                                                                                              | CMOS                                                                                                                                                                                                                                                                                                                                                                                                                                 | Equivalent                                                                                                                                                                                                                                                                                                                                                                                                     |
| Image<br>Resolution                             | 3840 x 2160…

---

**Source:** [https://fda-staging.innolitics.com/device/K223759](https://fda-staging.innolitics.com/device/K223759)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
