← Product Code [LZA](/productcode/LZA) · K223752

# Black Non Sterile Powder Free Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs, with Chloroquine, Cyclosporin A, Retrovir, Gastric Acid and Fentanyl (K223752)

_Central Medicare Sdn Bhd · LZA · Jul 24, 2023 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K223752

## Device Facts

- **Applicant:** Central Medicare Sdn Bhd
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Jul 24, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

Black Non Sterile Powder Free Nitrile Examination Gloves, Tested for Use with Chloroquine, Cyclosporin A, Retrovir, Gastric Acid and Fentanyl is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs, with chloroquine, cyclosporin A, retrovir, gastric acid and fentanyl citrate in accordance with ASTM D6978-05(2019).

## Device Story

Disposable, non-sterile, powder-free nitrile examination gloves; worn by healthcare examiners to prevent cross-contamination between patient and examiner. Device provides a physical barrier; tested for permeation resistance against various chemotherapy drugs and specific non-chemotherapy agents (chloroquine, cyclosporin A, retrovir, gastric acid, fentanyl citrate) per ASTM D6978-05(2019). Used in clinical settings; provides protection during handling of hazardous substances. Healthcare providers use the gloves as personal protective equipment; breakthrough times inform safe handling durations for specific drugs. Warning: Do not use with Carmustine or Thiotepa due to short breakthrough times.

## Clinical Evidence

No clinical data. Bench testing only. Permeation testing performed per ASTM D6978-05(2019) to determine breakthrough detection times for 33 chemotherapy drugs and 5 non-chemotherapy agents. Results show >240 minutes for most agents, with specific exceptions for Carmustine (13.2 minutes) and Thiotepa (34.7 minutes).

## Technological Characteristics

Nitrile rubber material; powder-free; non-sterile; disposable. Barrier protection principle. Tested per ASTM D6978-05(2019) for chemical permeation resistance. No electronic components, software, or energy sources.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

July 24, 2023

Central Medicare Sdn. Bhd. Chua Kah Ying Product Executive PT 2609-2620, Batu 8, Jalan Changkat Jong Teluk Intan, Perak 36000 Malaysia

Re: K223752

Trade/Device Name: Black Non Sterile Powder Free Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs, with Chloroquine, Cyclosporin A, Retrovir, Gastric Acid and Fentanyl Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, ODO, OPJ Dated: June 22, 2023 Received: June 22, 2023

Dear Chua Kah Ying:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Bifeng Qian -S

Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K223752

#### Device Name

Black Non Sterile Powder Free Nitrile Examination Gloves, Tested for Use with Chloroquine, Cyclosporin A, Retrovir, Gastric Acid and Fentanyl

#### Indications for Use (Describe)

Black Non Sterile Powder Free Nitrile Examination Gloves, Tested for Use with Chloroquine, Cyclosporin A, Retrovir, Gastric Acid and Fentanyl is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs, with chloroquine, cyclosporin A, retrovir, gastric acid and fentanyl citrate in accordance with ASTM D6978-05(2019).

Tested chemotherapy drugs and average breakthrough detection time (minutes) are as follows:

| Chemotherapy Drug           | Concentration | Breakthrough Detection Time in Minutes |
|-----------------------------|---------------|----------------------------------------|
| Bleomycin Sulfate           | 15.0 mg/ml    | > 240 mins                             |
| Busulfan                    | 6.0 mg/ml     | > 240 mins                             |
| Carboplatin                 | 10.0 mg/ml    | > 240 mins                             |
| Carmustine (BCNU)           | 3.3 mg/ml     | 13.2 mins                              |
| Cisplatin                   | 1.0 mg/ml     | > 240 mins                             |
| Cyclophosphamide (Cytoxan)  | 20.0 mg/ml    | > 240 mins                             |
| Cytarabine (Cytosine)       | 100.0 mg/ml   | > 240 mins                             |
| Dacarbazine                 | 10.0 mg/ml    | > 240 mins                             |
| Daunorubicin HCl            | 5.0 mg/ml     | > 240 mins                             |
| Docetaxel                   | 10.0 mg/ml    | > 240 mins                             |
| Doxorubicin HCl             | 2.0 mg/ml     | > 240 mins                             |
| Epirubicin HCl (Ellence)    | 2.0 mg/ml     | > 240 mins                             |
| Etoposide (Toposar)         | 20.0 mg/ml    | > 240 mins                             |
| Fludarabine                 | 25.0 mg/ml    | > 240 mins                             |
| Fluorouracil                | 50.0 mg/ml    | > 240 mins                             |
| Gemcitabine (Gemzar)        | 38.0 mg/ml    | > 240 mins                             |
| Idarubicin HCl              | 1.0 mg/ml     | > 240 mins                             |
| Ifosfamide                  | 50.0 mg/ml    | > 240 mins                             |
| Irinotecan                  | 20.0 mg/ml    | > 240 mins                             |
| Mechlorethamine HCl         | 1.0 mg/ml     | > 240 mins                             |
| Melphalan                   | 5.0 mg/ml     | > 240 mins                             |
| Methotrexate                | 25.0 mg/ml    | > 240 mins                             |
| Mitomycin C                 | 0.5 mg/ml     | > 240 mins                             |
| Mitoxantrone HCl            | 2.0 mg/ml     | > 240 mins                             |
| Oxaliplatin                 | 5.0 mg/ml     | > 240 mins                             |
| Paclitaxel                  | 6.0 mg/ml     | > 240 mins                             |
| Paraplatin                  | 10.0 mg/ml    | > 240 mins                             |
| Rituximab                   | 10.0 mg/ml    | > 240 mins                             |
| Thiotepa                    | 10.0 mg/ml    | 34.7 mins                              |
| Topotecan HCl               | 1.0 mg/ml     | > 240 mins                             |
| Trisenox (Arsenic Trioxide) | 1.0 mg/ml     | > 240 mins                             |
| Velcade (Bortezomib)        | 1.0 mg/ml     | > 240 mins                             |
| Vincristine Sulfate         | 1.0 mg/ml     | > 240 mins                             |
|                             |               |                                        |

{3}------------------------------------------------

| Tested non-chemotherapy drugs and average breakthrough detection time (minutes) are as follows: |               |                                        |
|-------------------------------------------------------------------------------------------------|---------------|----------------------------------------|
| Non-Chemotherapy Drug                                                                           | Concentration | Breakthrough Detection Time in Minutes |
| Chloroquine                                                                                     | 50.0 mg/ml    | > 240 mins                             |
| Cyclosporin A                                                                                   | 100.0 mg/ml   | > 240 mins                             |
| Retrovir                                                                                        | 10.0 mg/ml    | > 240 mins                             |

Fentanyl Permeation - Under the testing conditions of ASTM D6978-05(2019), Fentanyl Citrate Injection (100mcg/2mL) was found to have no breakthrough detected up to 240 minutes.

Gastric Acid Permeation - Under the testing conditions of ASTM D6978-05(2019), was found to have no breakthrough detected up to 240 minutes.

CAUTION: Testing showed an average breakthrough time of 13.2 minutes with Carmustine and 34.7 minutes with Thiotepa.

WARNING: Do not use with Carmustine and Thiotepa.

Type of Use (Select one or both, as applicable)

__ Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda-staging.innolitics.com/device/K223752](https://fda-staging.innolitics.com/device/K223752)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com