← Product Code [LZO](/productcode/LZO) · K223745

# Novae® Dual Mobility System, Hype® Hip System (K223745)

_Serf · LZO · Dec 22, 2023 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K223745

## Device Facts

- **Applicant:** Serf
- **Product Code:** [LZO](/productcode/LZO.md)
- **Decision Date:** Dec 22, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3353
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Novae® Dual Mobility and Hype® Hip Systems are indicated for total hip replacement, which includes: - · Osteoarthritis, - · Femoral neck fracture, - · Dislocation risk, - · Osteonecrosis of the femoral head, - · Revision procedures where other treatments or devices have failed and if bone reconstruction so permits. Sunfit TH, Novae® E TH, and Coptos TH acetabular cups are intended for press-fit use and Novae® Stick acetabular cup is indicated for cemented use. Hype® SCS, SCC, SCV, SCC mini, and SCLA mini hip stems are intended for press-fit use.

## Device Story

Novae® Dual Mobility and Hype® Hip Systems are hip replacement prostheses; components include acetabular cups, polyethylene liners, femoral heads, and hip stems. System uses dual mobility articulation; inner articular surface retains ceramic or metal femoral head; outer surface articulates within acetabular cup. Used in orthopedic surgery for total hip replacement; implanted by surgeons. Provides stable joint articulation; restores hip function; reduces dislocation risk. Line extension includes vitamin E stabilized UHMWPE liners, CoCrMo/ceramic femoral heads, and titanium alloy press-fit hip stems.

## Clinical Evidence

Bench testing only. Testing included range of motion (ISO 21535), lever-out (ASTM F1820-13), impingement (ASTM F2582), wear (ISO 14242-1), fatigue (ISO 7206-4, -6), fretting corrosion (ASTM F1875), torque/pull-off (ISO 7206-10, -13), burst strength (ISO 7206-10, ASTM F2345-21), coating characterization (ASTM F1854-15, F1160-14, F1044-05, F1147-05, F1978-18, F1185-03, F1926-14, F2024-10, F1609, ISO 13779-3, -6), MRI compatibility (ASTM F2503, F2052, F2213, F2182A, F2119), biocompatibility (ISO 10993-1), and endotoxin testing.

## Technological Characteristics

Acetabular cups: stainless steel, titanium/HA coating. Liners: vitamin E stabilized highly crosslinked UHMWPE. Femoral heads: Biolox® delta ceramic or CoCrMo alloy. Hip stems: titanium alloy with titanium/HA coating. Mechanical fixation: press-fit or cemented. Standards: ISO 21535, ISO 14242-1, ISO 7206, ASTM F1820-13, F2582, F1875, F2345-21, F1854-15, F1160-14, F1044-05, F1147-05, F1978-18, F1185-03, F1926-14, F2024-10, F1609, ISO 13779-3, -6, ASTM F2503, F2052, F2213, F2182A, F2119, ISO 10993-1.

## Regulatory Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

## Predicate Devices

- Serf BI-MENTUM™ dual mobility system ([K181744](/device/K181744.md))
- Serf NOVAE® dual mobility acetabular cup ([K111572](/device/K111572.md))
- DePuy Corail Hip Prosthesis ([K190344](/device/K190344.md), [K123991](/device/K123991.md), [K042992](/device/K042992.md))
- DePuy C-Stem AMT ([K082239](/device/K082239.md), [K042959](/device/K042959.md))
- DePuy Actis Hip Prosthesis ([K160907](/device/K160907.md), [K150862](/device/K150862.md))
- DePuy CoCrMo and Ceramic femoral heads ([K980513](/device/K980513.md), [K011533](/device/K011533.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

December 22, 2023

Serf % Robert Poggie President BioVera, Inc. 65 Promenade Saint Louis Notre Dame de Llle Perrot, Quebec J7W 3J6 Canada

Re: K223745

Trade/Device Name: Novae® Dual Mobility System, Hype® Hip System Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: Class II Product Code: LZO, MEH Dated: November 30, 2023 Received: November 30, 2023

Dear Robert Poggie:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE

by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

### Christopher Ferreira -S

for

Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known) K223745

Device Name Novae® Dual Mobility System Hype® Hip System

#### Indications for Use (Describe)

The Novae® Dual Mobility and Hype® Hip Systems are indicated for total hip replacement, which includes:

- · Osteoarthritis,
- · Femoral neck fracture,
- · Dislocation risk,
- · Osteonecrosis of the femoral head,
- · Revision procedures where other treatments or devices have failed and if bone reconstruction so permits.

Sunfit TH, Novae® E TH, and Coptos TH acetabular cups are intended for press-fit use and Novae® Stick acetabular cup is indicated for cemented use.

Hype® SCS, SCC, SCV, SCC mini, and SCLA mini hip stems are intended for press-fit use.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|                                                 |  |

|X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

### 510(k) SUMMARY for K223745

In accordance with 21 CFR 807.92, the following information is a 510(k) summary for the Novae® Dual Mobility System and Hype® Hip System.

### A. SUBMITTERS INFORMATION

| Submitter Name:     | BioVera, Inc.                                           |
|---------------------|---------------------------------------------------------|
| Submitter Address:  | 65 Promenade Saint-Louis, NDIP, Québec, J7W 3J6, CANADA |
| Contact Person:     | Robert A. Poggie, PhD                                   |
| Phone & Fax Number: | 514-901-0796                                            |
| Date of Submission: | December 22, 2023                                       |

### B. DEVICE IDENTIFICATION & MANUFACTURER

| Manufacturer Name:           | Serf                                                                                         |
|------------------------------|----------------------------------------------------------------------------------------------|
| Manufacturer Address:        | 85 avenue des Bruyères,                                                                      |
|                              | 69150 Décines-Charpieu – France                                                              |
| Registration Number:         | 3008668801                                                                                   |
| Contact Name:                | Jean-Charles MONCENIS                                                                        |
| Title:                       | Regulatory and Clinical Affairs Manager                                                      |
| Device Trade Names:          | Novae® Dual Mobility System                                                                  |
|                              | Hype® Hip System                                                                             |
| Device Common Name:          | Hip replacement prosthesis                                                                   |
| Classification Name:         | Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis |
| Classification Code:         | LZO, MEH                                                                                     |
| Classification Panel:        | Orthopedic                                                                                   |
| Regulation Number:           | Primary regulation: 21 C.F.R. § 888.3353                                                     |
| C1. PRIMARY PREDICATE DEVICE |                                                                                              |
| K181744                      | Serf BI-MENTUM™ dual mobility system                                                         |
| C2. PREDICATE DEVICES        |                                                                                              |
| K111572                      | Serf NOVAE® dual mobility acetabular cup                                                     |
| K190344, K123991, K042992    | DePuy Corail Hip Prosthesis                                                                  |
| K082239, K042959             | DePuy C-Stem AMT                                                                             |
| K160907, K150862             | DePuy Actis Hip Prosthesis                                                                   |
| K980513, K011533             | DePuy CoCrMo and Ceramic femoral heads                                                       |

{5}------------------------------------------------

# D. DEVICE DESCRIPTION

The Novae® Dual Mobility System consists of cemented acetabular shells and a polyethylene liner. Novae® XPEO-E liners are made from highly crosslinked vitamin E stabilized polyethylene and are spherical in geometry for dual mobility articulation. The inner articular surface mates and retains a Biolox® delta ceramic or CoCrMo metal femoral head. The outer articular surface articulates within the highly polished inner diameter of Novae® cementless or cemented cups. Novae® cups are manufactured from stainless steel, with pressfit options coated with titanium and hydroxyapatite (HA). Pegs and cortical bone screws for optional ancillary fixation are made from stainless steel.

Hype® press-fit hip stems are compatible with the Novae® Dual Mobility System and Serf CoCrMo and Biolox® delta femoral heads. Hype® press-fit hip stems are made from titanium alloy coated with titanium and HA.

The primary purpose of this 510(k) notification is a line extensions to the 510(k) cleared Novae® acetabular cups (K111572) that are compatible with: (1) liner option made from highly crosslinked and vitamin E stabilized polyethylene (XPEO-E), (2) femoral heads made from CoCrMo alloy and Biolox® delta ceramic, and (3) Hype® press-fit hip stems.

# E. INDICATIONS FOR USE

The Novae® Dual Mobility and Hype® Hip Systems are indicated for total hip replacement, which includes:

- · Osteoarthritis,
- Femoral neck fracture,
- Dislocation risk,
- · Osteonecrosis of the femoral head,
- Revision procedures where other treatments or devices have failed and if bone reconstruction so permits.

Sunfit TH, Novae® E TH, and Coptos TH acetabular cups are intended for press-fit use and Novae® Stick acetabular cup is indicated for cemented use.

Hype® SCS, SCC, SCC, SCC mini, and SCLA mini hip stems are intended for press-fit use.

# F. TECHNOLOGICAL CHARACTERISTICS AND COMPARSION

The Novae® Dual Mobility System has the same intended use and similar indications for use as the primary predicate device, the Bi-Mentum™ Dual Mobility Acetabular Cup (K181744). The design and materials for Novae® acetabular cups, pegs, and bone screws are the same as those for the primary predicate device. The design of the XPEO-E polyethylene liner is the same

{6}------------------------------------------------

as the primary predicate, with the material for the subject device being made from highly crosslinked and vitamin E stabilized UHMWPE that is more resistant to wear and oxidation than conventional polyethylene. The CoCrMo and ceramic femoral heads are the same material, size, and design as previous compatible heads with the Bi-Mentum™ Dual Mobility System. Hype® cementless hip stems are made from the same materials and similar sizing and design principles as the predicate hip stem devices.

## G. PERFORMANCE DATA AND SUBSTANTIAL EQUIVALENCE

Laboratory testing and engineering analyses were conducted to verify that the performance characteristics of the Novae® Dual Mobility and Hype® Hip System are sufficient for the intended use and to establish substantial equivalence. The following bench tests and analyses were performed and met predetermined acceptance criteria where specified:

- Range of motion analysis for Hype hip stems and Novae acetabular cups (ISO 21535)
- Head-liner assembly / disassembly tests
- Lever-out testing for Novae® acetabular cups and liners (ASTM F1820-13)
- Impingement analyses and testing for Hype® hip stems and Novae" acetabular cups (ASTM F2582)
- · Articular surface wear for pristine, outer bearing only, and abrasive conditions (ISO 14242-1)
- · Fatique strength testing of Hype hip stems (ISO 7206-4, -6)
- Fretting corrosion of Hype hip stem tapers and CoCrMo femoral heads (ASTM F1875)
- · Torque-off and pull-off testing of femoral heads and Hype® hip tapers (ISO 7206-10, -13)
- · Fatigue, post-fatigue burst strength of femoral head from stem (ISO 7206-10 and ASTM F2345-21)
- Physical, chemical, and mechanical tests and analyses of titanium and HA coatings for cementless Hype Hip Stems (ASTM F1854-15, F1160-14, F1044-05, F1147-05, F1978-18, F1185-03, F1926-14, F2024-10, F1609, ISO 13779-3, -6)
- · Assessment of MRI compatibility and determination of conditional labeling requirements per ASTM F2503 and performance testing per ASTM F2052, F2213, F2182A, and F2119
- · Assessment of biocompatibility of Novae® Dual Mobility and Hype® devices per ISO 10993-1
- Bacterial Endotoxin Testing (BET) / LAL chromogenic method demonstrating acceptable levels of endotoxins

The Novae® Dual Mobility and Hype® Hip Systems have the same intended use, similar indications for use, the same or similar materials and design features to the predicate devices. The results of performance testing and engineering analyses support substantial equivalence.

# H. CONCLUSIONS

Comparison of the Novae® Dual Mobility System and Hype® Hip System components with the predicate devices demonstrate substantial equivalence of intended and indicated use, technological characteristics, materials, and principles of operation. The performance data demonstrate safety and equivalence of the subject and predicate.

---

**Source:** [https://fda-staging.innolitics.com/device/K223745](https://fda-staging.innolitics.com/device/K223745)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com