← Product Code [FBN](/productcode/FBN) · K223741

# Pusen Single Use Flexible Video Cystoscope/Choledochoscope System (Single Use Flexible Video Cystoscope/Choledochoscope:PC200-AS, PC200-AR, PC200-S and PC200-R;HD Medical Video Endoscope Image Processor: PV300) (K223741)

_Zhuhai Pusen Medical Technology Co., Ltd. · FBN · Jun 22, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K223741

## Device Facts

- **Applicant:** Zhuhai Pusen Medical Technology Co., Ltd.
- **Product Code:** [FBN](/productcode/FBN.md)
- **Decision Date:** Jun 22, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Pusen Single Use Flexible Video Cystoscope/Choledochoscope System is intended to be used for endoscopic access to and examination of the lower urinary tract. This system is also indicated for the examination and therapeutic applications during endoscopic procedures in bile ducts. The Cystoscope/Choledochoscope is intended to provide visualization via video processor and can be used with endoscopic accessories. This system is intended for use in a hospital environment or medical office environment.

## Device Story

System comprises single-use flexible video cystoscope/choledochoscope (models PC200-AS, PC200-AR, PC200-S, PC200-R) and reusable HD medical video endoscope image processor (PV300). Device captures visual data via distal CMOS sensor and LED illumination; processor transforms signals into real-time video displayed on LCD screen. Operated by clinicians in hospital or medical office environments. Provides visualization for diagnostic and therapeutic procedures in lower urinary tract and bile ducts; compatible with endoscopic accessories. Output enables visual examination of target sites, supporting clinical decision-making during endoscopic interventions. Benefits include single-use sterility, reducing cross-contamination risk.

## Clinical Evidence

No clinical or animal studies were performed. Substantial equivalence is supported by non-clinical performance data, including verification reports for direction of view and insertion portion width, and reliance on performance data from the reference device (K222602).

## Technological Characteristics

Flexible, single-use endoscope; CMOS digital video technology; LED illumination. Sterilized via Ethylene Oxide (SAL 10^-6). Powered by PV300 processor (line power/battery). Models feature varying deflection (standard/reverse) and function buttons. Working length 380mm; working channel 2.3mm; insertion width 5.6mm. Field of view 90°; depth of field 3-80mm.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Flexible Video-Choledo-Cysto-Ureteroscope System ([K211686](/device/K211686.md))

## Reference Devices

- Pusen Single Use Flexible Video Cystoscope System ([K222602](/device/K222602.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 22, 2023

Zhuhai Pusen Medical Technology Co., Ltd. Changshen Wang Regulatory Affairs Director 5/F, Building 1, No 33, Ke Ji San Road High-tech Zone, Tangjiawan Town Zhuhai, Guangdong 519085 China

Re: K223741

Trade/Device Name: Pusen Single Use Flexible Video Cystoscope/Choledochoscope System (Single Use Flexible Video Cystoscope/Choledochoscope:PC200-AS, PC200-AR, PC200-S and PC200-R:HD Medical Video Endoscope Image Processor: PV300 Regulation Number: 21 CFR§ 876.1500

Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FBN, FAJ Dated: December 12, 2022 Received: May 18, 2023

Dear Changshen Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Mark R. Kreitz -S

for Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K223741

#### Device Name

Pusen Single Use Flexible Video Cystoscope/Choledochoscope System (Single Use Flexible Video Cystoscope/Choledochoscope:PC200-AR. PC200-S and PC200-R:HD Medical Video Endoscope Image Processor: PV300

#### Indications for Use (Describe)

The Pusen Single Use Flexible Video Cystoscope/Choledochoscope System is intended to be used for endoscopic access to and examination of the lower urinary tract. This system is also indicated for the examination and therapeutic applications during endoscopic procedures in bile ducts. The Cystoscope/Choledochoscope is intended to provide visualization via video processor and can be used with endoscopic accessories.

This system is intended for use in a hospital environment or medical office environment.

| Type of Use (Select one or both, as applicable)                                                                              |                                                                                                                             |
|------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><span>☒</span><span>Prescription Use (Part 21 CFR 801 Subpart D)</span></div> | <div style="display:flex; align-items:center;"><span>☐</span><span>Over-The-Counter Use (21 CFR 801 Subpart C)</span></div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

# I. SUBMITTER

Submitter's name: Zhuhai Pusen Medical Technology Co, Ltd. Submitter's address: 5/F, Building 1, No 33, Keji San Road, High-tech Zone, Tangjiawan Town, 519085 Zhuhai, Guangdong, People's Republic of China.

Phone: +86 756 688 0865

Contact Person: Ellen Wang Date Prepared: May 16, 2023

## II. DEVICE

Name of Device: Pusen Single Use Flexible Video Cystoscope/Choledochoscope System (Single Use Flexible Video Cystoscope/Choledochoscope: PC200-AR, PC200-AR, PC200-S and PC200-R; HD Medical Video Endoscope Image Processor: PV300 Classification Name: Endoscope and Accessories (21 CFR 876.1500) 510(k) number: N/A Regulatory Class: Class II Product Code: FBN, FAJ

#### III. PREDICATE DEVICE

Predicate device Name of Device: Flexible Video-Choledo-Cysto-Ureteroscope System 510(k) number: K211686 Classification Name: Endoscope and Accessories Product code: FGB, FBN, FET, FAJ, FGA This predicate device has not been subject to a design-related recall.

#### Reference device

Name of Device: Pusen Single Use Flexible Video Cystoscope System 510(k) number: K222602 Classification regulation: Endoscope and Accessories Product code: FAJ This reference device has not been subject to a design-related recall.

#### IV. DEVICE DESCRIPTION

{4}------------------------------------------------

The Pusen Single Use Flexible Video Cystoscope/Choledochoscope System includes a Single Use Flexible Video Cystoscope/Choledochoscope and an HD Medical Video Endoscope Image Processor.

This system is intended to be used for endoscopic access to and examination of the lower urinary tract. This system is also indicated for the examination and therapeutic applications during endoscopic procedures in bile ducts. This Cystoscope/Choledochoscope is intended to provide visualization via video processor and can be used with endoscopic accessories. This system is intended for use in a hospital environment or medical office environment.

- The Pusen Single Use Flexible Video Cystoscope/Choledochoscope is provided sterile and has the following 4 models:

| Model    | Description of difference                                                 |
|----------|---------------------------------------------------------------------------|
| PC200-AS | 'A' means this model has function buttons; 'S' means standard deflection. |
| PC200-AR | 'A' means this model has function buttons; 'R' means reverse deflection.  |
| PC200-S  | 'S' means standard deflection.                                            |
| PC200-R  | 'R' means reverse deflection.                                             |

- The Pusen Single Use Flexible Video Cystoscope/Choledochoscope needs to be connected with the HD Medical Video Endoscope Image Processor PV300 as a system, and it is powered by the latter.
- The HD Medical Video Endoscope Image Processor PV300 is a reusable device, used to process the image signal from the Pusen Single Use Flexible Video Cystoscope/Choledochoscope and display the real-time video on its LCD screen, which enables visual examination of target site.

This change is to expand the indications for the currently cleared reference Pusen Single Use Flexible Video Cystoscope System (K222602) to allow it to also be marketed as a Choledochoscope used for examination and therapeutic applications during endoscopic procedures in bile ducts. No changes were made to the device design. Only the product name, indications for use and applicable labeling were updated.

### V. INDICATIONS FOR USE

The subject device: The Pusen Single Use Flexible Video Cystoscope/Choledochoscope System is intended to be used for endoscopic access to and examination of the lower urinary tract. This system is also indicated for the examination and therapeutic applications during endoscopic procedures in bile ducts. This Cystoscope/Choledochoscope is intended to provide visualization via video processor and can be used with endoscopic accessories. This system is intended for use in a hospital environment or medical office environment.

{5}------------------------------------------------

# Zhuhai Pusen Medical Technology Co., Ltd.

Traditional 510(k)

The predicate device: The Flexible Video-Choledo-Cysto-Ureteroscope System is indicated for endoscopic examination in the urinary tract and can be used percutaneously to examine the interior of the kidney, and using additional accessories, to perform various diagnostic and therapeutic procedures. The Flexible Video-Choledo-Cysto-Ureteroscope System is also indicated for the examination of bile ducts, and using additional accessories, to perform various diagnostic and therapeutic procedures during cholecystectomy.

The indications for use of the subject device fall within the intended use of the predicate device and, therefore, the two devices have the same intended use.

#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

| Item                                  | Subject device                                                             | Predicate device                                        | Reference device                                     |
|---------------------------------------|----------------------------------------------------------------------------|---------------------------------------------------------|------------------------------------------------------|
| Trade name                            | Pusen Single Use Flexible<br>Video<br>Cystoscope/Choledochoscope<br>System | Flexible Video-Choledo-<br>Cysto-Ureteroscope<br>System | Pusen Single Use Flexible<br>Video Cystoscope System |
| 510(K)<br>number                      | N/A                                                                        | K211686                                                 | K222602                                              |
| Scope type                            | Flexible                                                                   | Flexible                                                | Flexible                                             |
| Scope<br>reusability                  | Single-use                                                                 | Single-use                                              | Single-use                                           |
| Energy used                           | Powered by chargeable<br>battery or line power.                            | Powered by chargeable<br>battery or line power.         | Powered by chargeable<br>battery or line power.      |
| Digital video<br>technology           | CMOS                                                                       | CMOS                                                    | CMOS                                                 |
| Illumination<br>source                | LED                                                                        | LED                                                     | LED                                                  |
| Field of view                         | 90 °                                                                       | 110 °                                                   | 90 °                                                 |
| Direction<br>of<br>view               | 0°                                                                         | 0°                                                      | 30°                                                  |
| Depth of field                        | 3~80 mm                                                                    | 5~50mm                                                  | 3~80 mm                                              |
| Maximum<br>insertion<br>portion width | 5.6mm                                                                      | 5.3mm                                                   | 6.0mm                                                |
| Working<br>length                     | 380 mm                                                                     | 385 mm                                                  | 380 mm                                               |

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image contains the word "PUSEN" in a stylized, sans-serif font. The letters are a light blue color. Below the word "PUSEN" is the text "Zhuhai Pusen Medical Technology Co." in a smaller, sans-serif font. The text is a dark gray color.

| Zhuhai Pusen Medical Technology Co., Ltd. |                                  |                        | Traditional 510(k)               |
|-------------------------------------------|----------------------------------|------------------------|----------------------------------|
| Working<br>channel size                   | 2.3 mm                           | 2.6 mm                 | 2.3 mm                           |
| Up/down<br>deflection                     | Up: 210°<br>Down: 210°           | Up: 210°<br>Down: 130° | Up: 210°<br>Down: 210°           |
| Suction                                   | Provided                         | Not Provided           | Provided                         |
| Sterility                                 | Ethylene Oxide (EO)<br>SAL: 10-6 | EO<br>SAL: 10-6        | Ethylene Oxide (EO)<br>SAL: 10-6 |

The subject and the predicate device have different dimensions (e.g., working length, channel size) and optical specifications (e.g., depth of field, field of view). However, these differences do not raise different questions of safety and effectiveness.

#### VII. PERFORMANCE DATA

#### Non-Clinical Performance Data

The Pusen Single Use Flexible Video Cystoscope/Choledochoscope System and reference Pusen Single Use Flexible Video Cystoscope System (K222602) share the same fundamental technology and physical characteristics except for the Direction of view and the Maximum insertion portion width, verification reports on these two performances are provided. The other performance data described in the cleared Pusen Single Use Flexible Video Cystoscope System, K222602, is also applicable to the Pusen Single Use Flexible Video Cystoscope/Choledochoscope System. The labelling was updated to reflect the new indications and new specifications.

#### Animal Study and Clinical study

No animal study or clinical study is included in this submission.

#### VIII. CONCLUSIONS

The Pusen Single Use Flexible Video Cystoscope/Choledochoscope System is substantially equivalent to its predicate device. Performance testing and compliance with voluntary standards, demonstrate that the subject device is substantially equivalent to the relevant aspects of the predicate device in terms of design, components, materials, principals of operation, biocompatibility, performance characteristics, and intended use. Therefore, the subject devices are determined to be substantially equivalent to the referenced predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K223741](https://fda-staging.innolitics.com/device/K223741)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com