Kyphoplasty Balloon Catheter is intended to be used for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine during balloon kyphoplasty (for use with cleared spinal polymethymethacrylate (PMMA) bone cements).
Device Story
Kyphoplasty Balloon Catheter is an inflatable bone tamp used in percutaneous kyphoplasty. Device consists of an outer tube, core tube/rod, and distal inflatable balloon with radiopaque markers for positioning. Operated by physicians in clinical settings. Device is inserted into the spine; balloon is inflated with contrast medium to create a void or reduce fractures in cancellous bone. Following inflation, the balloon is deflated and removed, allowing for subsequent injection of PMMA bone cement. Radiopaque markers assist the physician in visualizing balloon placement under fluoroscopy. Benefits include fracture reduction and stabilization of the vertebral body.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Materials conform to ISO 10993. Inflatable balloon catheter; contrast medium inflation. Dimensions: 10.0mm, 15.0mm, 20.0mm sizes. Burst pressure 400 psi. Sterilization: Ethylene oxide (ISO 11135). Packaging validated per ASTM F1886, F88, F1929, and ISTA-2A. Shelf life 3 years (ASTM F1980).
Indications for Use
Indicated for patients requiring reduction and fixation of spinal fractures or creation of a void in cancellous bone during balloon kyphoplasty procedures using cleared PMMA bone cements.
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
{0}------------------------------------------------
August 16, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Jiangsu Changmei Medtech Co., Ltd. % Amber Pang Landlink Healthcare Technology (Shanghai) Co., Ltd. Room 1308, Baohua International Plaza 555 West Guangzhong Road Shanghai. 200072 China
Re: K223709
Trade/Device Name: Kyphoplasty Balloon Catheter Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX, HXG Dated: July 31, 2023 Received: July 31, 2023
Dear Ms. Pang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jesse Muir -S
Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K223709
Device Name Kyphoplasty Balloon Catheter
Indications for Use (Describe)
Kyphoplasty Balloon Catheter is intended to be used for the reduction of fixation of fractures and/or creation of a void in cancellous bone in the spine during balloon kyphoplasty (for use with cleared spinal polymethymethacrylate (PMMA) bone cements).
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) summary
K223709
## l Submitter
Jiangsu Changmei Medtech Co., Ltd. No.27, Xinke West Road, Luoyang Town, Wujin District, Changzhou, Jiangsu, CN 213104.
Establishment Registration Number: 3016789487
Contact person: Yang Lifan Position: Regulation Manager Tel.: +86-18651969542 E-mail: ylf@czmed.com
Preparation date: Aug 8, 2023
#### II Proposed Device
| Trade Name of Device: | Kyphoplasty Balloon Catheter |
|-----------------------|------------------------------|
| Common name: | Inflatable Bone Tamp |
| Regulation Number: | 21 CFR 888.1100 |
| Regulatory Class: | Class II |
| Product code: | HRX, HXG |
| Review Panel: | Orthopedic |
#### III Predicate Devices
| 510(k) Number: | K192449 |
|-----------------|---------------------------------------------|
| Trade name: | Joline® Kyphoplasty System Allevo |
| Common name: | Vertebroplasty System; Inflatable Bone Tamp |
| Classification: | Class II |
| Product Code: | HRX, NDN |
| Manufacturer | Joline GmbH & Co. KG |
#### IV Device description
Kyphoplasty Balloon Catheter is an inflatable balloon catheter used in percutaneous kyphoplasty (PKP). The Balloon Catheter consists of an outer tube and core tube/rod, inflatable balloon located at the distal tip. The radiopaque markers located at the balloon tip to reflect the balloon position during positioning. The balloon catheter is supplied sterilized, single-use.
{4}------------------------------------------------
## V Indication for use
Kyphoplasty Balloon Catheter is intended to be used for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine during balloon kyphoplasty (for use with cleared spinal polymethymethacrylate (PMMA) bone cements).
## VI Comparison of technological characteristics with the predicate devices
The comparison and discussion between the subject device and the predicate device is listed in Table 1 below:
| Item | Proposed device | Predicate device<br>(K192449) | Discussion |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Product name | Kyphoplasty Balloon<br>Catheter | Joline® Kyphoplasty<br>System Allevo | same |
| Product Code | HRX, HXG | HRX, NDN | same |
| Regulation No. | 21 CFR 888.1100 | 21 CFR 888.1100 | same |
| Class | Class II | Class II | same |
| Indication for<br>use | Kyphoplasty Balloon<br>Catheter is intended to<br>be used for the<br>reduction and fixation of<br>fractures and/or creation<br>of a void in cancellous<br>bone in the spine during<br>balloon kyphoplasty (for<br>use with cleared spinal<br>polymethymethacrylate<br>(PMMA) bone cements). | The Joline® Kyphoplasty<br>System Allevo is intended<br>to be used for the reduction<br>and fixation of fractures<br>and/or creation of a void in<br>cancellous bone in the<br>spine during balloon<br>kyphoplasty (for use with<br>cleared spinal<br>polymethylmethacrylate<br>(PMMA) bone cements). | same |
| Balloon Size | 10.0mm, 15.0mm,<br>20.0mm | 10.0mm, 16.0mm, 22.0mm | Different¹ |
| Balloon Brust<br>Pressure | 400 psi | 400 psi | same |
| Balloon<br>Burst<br>Volume | ≥ 6 ml | ≥ 6 ml | same |
| Balloon<br>Inflation<br>Behavior | Balloon Diameter:<br>$18.3\pm3.0$ mm.<br>Balloon working length:<br>$28.0\pm3.0$ | Balloon Diameter $\leq$ 19 mm<br>Balloon working length $\leq$ 24mm | Different¹ |
| Tensile<br>Force<br>Balloon | $\geq$ 15 N | $\geq$ 15 N | same |
| Balloon | The balloon must not | The balloon must not burst | Different² |
| Repeated<br>Inflation | rupture or leak within 20<br>inflation/deflation cycles | within 3 inflation/deflation<br>cycles | |
| Balloon<br>Deflation Time | $\leq$ 3 s | $\leq$ 3 s | same |
| Shapes | Peanut, cylindrical | Cylindrical | Different3 |
| Inflation<br>medium | Contrast medium | Contrast medium | Same |
| Single/double<br>use of catheter | Single | Single and double | Different4 |
| Sterility | Yes | Yes | same |
| Biocompatibility | Confirm to the<br>requirements of ISO<br>10993 series standards | Confirm to the<br>requirements of ISO 10993<br>series standards | Same |
Table 1 General Comparison of Kyphoplasty Balloon Catheter
{5}------------------------------------------------
1 The difference in the size does not raise additional questions for safety and effectiveness of the device. The performance test of the subject devices has been performed on the final finished device. The test results show pass the requirements.
2 The Balloon repeated inflation in the subject device might be different from the predicate devices. The requirement of balloon with subject device must not rupture or leak within 20 inflation/deflation cycles, and predicate device is 3 cycles. Therefore, the differences on balloon repeated inflation do not raise new questions about safety and effectiveness.
3 The shape of proposed device has peanut type, and predicate device haven't. The peanut shape Balloon Catheter have no different in shape with the cylindrical type under rated filling volume state. Therefore, the differences on balloon shape do not raise new questions about safety and effectiveness.
4 The predicate device has single use and double use of catheter, and proposed device have single use of catheter only. The predicated device can cover proposed device. Therefore, the differences on use of catheter do not raise new questions about safety and effectiveness.
## VII Non-Clinical Testing
#### Bench test: A
The following tests were performed in support of the substantial equivalence determination.
| Test Performed | Acceptance Criteria |
|------------------------|----------------------------------------|
| Balloon Burst Pressure | Balloon Burst Pressure (Constrained) ≥ |
{6}------------------------------------------------
| | 400psi |
|---------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Balloon Burst Volume | Balloon Burst Volume (Constrained) ≥ 6ml |
| Balloon<br>Inflation<br>Behavior<br>(Unconstrained Balloon<br>Compliance) | Inflated with 6 ml, the balloon diameter<br>should be 18.3±3.0 mm, and the balloon<br>working length should be 28.0±3.0 mm. |
| Tensile Force | Tensile force ≥ 15N. |
| Balloon Repeated<br>Inflation | The balloon should withstand 20<br>inflation/deflation cycles at rated filling<br>volume without rupture or leak. |
| Balloon Deflation Time | Deflation time ≤ 3s. |
#### > Biocompatibility Testing:
In accordance with ISO 10993-1, the Kyphoplasty Balloon Catheter is classified as: Externally Communicating Device, Intact bone/tissue, Limited Contact (≤ 24hours). The following biocompatibility endpoints were addressed:
- Cytotoxicity ●
- . Sensitization
- Intracutaneous Reactivity .
- Acute Systemic Toxicity .
- Material-Mediated Pyrogenicity .
#### A Sterility, Shipping and Shelf Life:
The sterilization method has been validated per ISO 11135: 2014 Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices. The shelf life of the Kyphoplasty Balloon Catheter is 3 years, determined based on stability studies which includes accelerated aging and simulated shipping.
Sterilization shelf-life studies were conducted in compliance with the following standards:
| Item | Standard |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| EO and ECH residuals | ISO 10993-7:2008 Biological evaluation of medical devices – Part 7: Ethylene oxide sterilization residuals |
| Bacteria Endotoxin Limit | USP <85> Bacterial Endotoxins |
| Shelf-Life Evaluation | Shelf life of 3 years is validated using the FDA recognized standard ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices |
{7}------------------------------------------------
- Package integrity testing, after accelerated aging, environmental conditioning and simulated transportation in accordance with ISTA-2A:2011 68kg or less partial simulation performance test procedure for individual packaged-products, was conducted on the final, packaged, and sterile devices. All packaging deemed acceptable for protection of product and sterility maintenance.
- Sterile Barrier Packaging Testing performed on the proposed device:
- Visual Inspection in accordance with ASTM F1886 / F1886M-16
- Seal Strength in accordance with ASTM F88/F88M-15
- Dye Penetration in accordance with ASTM F1929-15
#### VIII Clinical Testing
No clinical study is included in this submission.
#### IX Conclusion
The proposed device has the same indication for use and has similar design features and technological characteristic as the predicate device. Performance testing data demonstrates that the proposed device is safety and effectiveness as the predicated device. Accordingly, the proposed device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.