A-Model: The F&P Solo mask is intended to be used by adults weighing ≥66lbs (30 kgs) who have been prescribed non-invasive positive airway pressure therapy such as CPAP or Bi-Level by a physician. The F&P Solo mask is intended for single-patient use in the home. SL A-Model: The F&P Solo mask is intended to be used by adults weighing ≥66lbs (30 kgs) who have been prescribed non-invasive positive airway pressure therapy such as CPAP or Bi-Level by a physician. The F&P Solo mask is intended for single-patient use in the home and for multiple patient use in the hospital or other clinical setting where proper disinfection of the device can occur between patient uses.
Device Story
Non-invasive patient interface; seals against nose via nasal or pillows cushions; held by headgear; connects to CPAP/Bi-Level device via 22mm male swivel adaptor. Delivers pressurized air to airway; exhaust bias holes on frame flush exhaled air. Used in home (A-Model) or home/hospital (SL A-Model) by patients; prescription only. Healthcare providers use device to deliver prescribed pressure therapy; patient benefits from improved compliance/comfort during sleep therapy. No electronic processing or software.
Clinical Evidence
Bench testing only. No clinical data presented. Testing included ISO 17510:2015 (CO2 rebreathing, pressure-flow, exhaust flow, resistance to flow, vibration/noise), ISO 5356-1:2015 (connectors), and extensive biocompatibility testing (ISO 10993 series and ISO 18562 series). Human factors/usability engineering and mechanical integrity testing were also performed.
Technological Characteristics
Non-invasive nasal mask; includes nasal and pillows cushion options. Materials evaluated per ISO 10993 and ISO 18562. Features 22mm ISO taper connection. Designed for 4-20 cm H2O pressure range. Reusable; supports thermal disinfection (75°C/30min or 80°C/10min). Conforms to ISO 17510:2015 and ISO 5356-1:2015. No software or electronic components.
Indications for Use
Indicated for adults ≥30kg prescribed non-invasive positive airway pressure therapy (CPAP or Bi-Level). Contraindicated for patients who are not responsive or unable to remove the mask.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
May 17, 2023
Fisher & Paykel Healthcare Ltd Reena Daken Regulatory Affairs Manager 15 Maurice Paykel Place, East Tamaki Auckland, 2013 New Zealand
Re: K223696
Trade/Device Name: F&P Solo Mask Range Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous ventilator (IPPB) Regulatory Class: Class II Product Code: BZD Dated: May 2, 2023 Received: May 2, 2023
Dear Reena Daken:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Rachana Visaria -S
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223696
Device Name F&P Solo Mask Range
### Indications for Use (Describe)
### A-Model
The F&P Solo mask is intended to be used by adults weighing ≥661bs (30 kgs) who have been prescribed non-invasive positive airway pressure therapy such as CPAP or Bi-Level by a physician. The F&P Solo mask is intended for singlepatient use in the home.
### SL A-Model
The F&P Solo mask is intended to be used by adults weighing ≥661bs (30 kgs) who have been prescribed non-invasive positive airway pressure therapy such as CPAP or Bi-Level by a physician. The F&P Solo mask is intended for singlepatient use in the home and for multiple patient use in the hospital or other proper disinfection of the device can occur between patient uses.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# As Required by 21 CFR 807.92
# I. SUBMITTER
| Company Name and Address | Fisher & Paykel Healthcare Limited<br>15 Maurice Paykel Place<br>East Tamaki<br>Auckland 2013, New Zealand<br>Telephone: +64 9 574 0100 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Prepared and Submitted by | Alec O'Connell-Biddlecombe<br>Regulatory Affairs Associate |
| Contact Person | Reena Daken<br>Regulatory Affairs Manager<br>Telephone: +64 9 574 0100<br>Email: reena.daken@fphcare.co.nz |
| Date Prepared | 16 May 2023 |
#### == DEVICE
| Name of Device | F&P Solo Mask Range |
|---------------------|----------------------------------|
| Common/Usual Name | Nasal Mask |
| Classification Name | Non Continuous Ventilator (IPPB) |
| Regulatory Class | Class II (21 CFR §868.5905) |
| Product Code | BZD |
#### lll. PREDICATE DEVICE
- . Predicate device:
- o F&P Evora™ Nasal Mask, K200089
- . Reference devices:
- o F&P Brevida™ Nasal Mask, K161412
- Used to support claims of substantial equivalence with respect to design and performance
{4}------------------------------------------------
#### IV. DEVICE DESCRIPTION
The Solo Mask, is a non-invasive patient interface with cushions that seal against the airway entrance of the nose. The mask is held on the face with a headqear and connects to a single breathing tube by a 22mm male swivel adaptor to receive pressurized air from a continuous Positive Airway Pressure device (CPAP or Bi-level). The exhaust bias holes on the frame of the mask allow exhaled air to be flushed out while the system is in operation. The F&P Solo Mask is a prescription only device, provided in a non-sterile state.
| Model | Product Code | Product Description |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|
| A-Model | SLN1SA | Solo Nasal Mask Small A Model |
| The A Model is intended for single- patient use in a<br>home environment only. The mask is available in<br>seven different configurations which differ only in<br>the size and type of the cushions made available in<br>the packaging (Nasal: S, M, L and W; Pillows: S,<br>M and L). | SLN1MA | Solo Nasal Mask Medium A Model |
| | SLN1LA | Solo Nasal Mask Large A Model |
| | SLN1WA | Solo Nasal Mask Wide A Model |
| | SLP1SA | Solo Pillows Mask Small A Model |
| | SLP1MA | Solo Pillows Mask Medium A Model |
| | SLP1LA | Solo Pillows Mask Large A Model |
| | SLNSMLA | Solo Nasal Mask Fit Pack/SML A Model |
| | SLPSMLA | Solo Pillows Mask Fit Pack/SML A Model |
| | SL A-Model<br>The Sleep Lab A Model is intended for single<br>patient use in a home environment as well as for<br>multi-patient use in the hospital or other clinical<br>setting. The mask is available in seven different<br>configurations which differ only in the size and type<br>of cushions made available in the packaging<br>(Nasal: S, M, L and Wide; Pillows: S, M and L). | SLN1SSLA |
| SLN1MSLA | | Solo Nasal Mask Medium Sleep Lab A |
| SLN1LSLA | | Solo Nasal Mask Large Sleep Lab A |
| SLN1WSLA | | Solo Nasal Mask Wide Sleep Lab A |
| SLP1SSLA | | Solo Pillows Mask Small Sleep Lab A |
| SLP1MSLA | | Solo Pillows Mask Medium Sleep Lab A |
| SLP1LSLA | | Solo Pillows Mask Large Sleep Lab A |
| SLN1SMLSLA | | Solo Nasal Mask Fit Pack Sleep Lab A |
| SLP1SMLSLA | Solo Pillows Mask Fit Pack Sleep Lab A | |
A list of the subject device product codes can be found below:
# IV. INDICATIONS FOR USE
## A-Model
The F&P Solo mask is intended to be used by adults weighing ≥66lbs (30 kgs) who have been prescribed non-invasive positive airway pressure therapy such as CPAP or Bi-Level by a physician. The F&P Solo mask is intended for singlepatient use in the home.
## SL A-Model
The F&P Solo mask is intended to be used by adults weighing ≥66lbs (30 kgs) who have been prescribed non-invasive positive airway pressure therapy such as CPAP or Bi-Level by a physician. The F&P Solo mask is intended for singlepatient use in the home and for multiple patient use in the hospital or other clinical setting where proper disinfection of the device can occur between patient uses.
{5}------------------------------------------------
#### V. COMPARISON OF TECHNOLOGICAL CHARATCERISTICS WITH THE PREDICATE DEVICE
The comparison of features, performance data and intended use demonstrate that the F&P Solo Mask Range is substantially equivalent to the predicate device, F&P Evora™ Nasal Mask (K200089). Please see the table below.
| Design/<br>Technological<br>Characteristics | Subject Device F&P<br>Solo Mask | Predicate Device<br>F&P Evora Nasal Mask (K200089) | Comments |
|--------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | | | |
| Legal Manufacturer | Fisher & Paykel Healthcare Ltd. | Fisher & Paykel Healthcare Ltd. | Identical |
| Product Code | BZD | BZD | |
| Device classification | 21 CFR 868.5905 | 21 CFR 868.5905 | |
| Classification panel | Anaesthesiology | Anaesthesiology | |
| Indications for Use | | | |
| | A-Model: | A-Model: | |
| | The F&P Solo mask is intended to be used by<br>adults weighing ≥ 66 lbs (30 kgs) who have been<br>prescribed non-invasive positive airway pressure<br>therapy such as CPAP or Bi-Level by a<br>physician. The F&P Solo mask is intended for<br>single- patient use in the home.<br>SL A-Model: | The F&P Evora™ Nasal Mask is intended to be<br>used by adults weighing<br>≥66lbs (30kgs) who have been prescribed non-<br>invasive positive airway pressure therapy such<br>as CPAP or bilevel by a physician. The F&P<br>Evora Nasal Mask is intended for single patient<br>use in the home | |
| Indications for Use | The F&P Solo mask is intended to be<br>used by adults weighing ≥ 66lbs (30 kgs) who<br>have been prescribed non-invasive positive<br>airway pressure therapy such as CPAP or Bi-<br>Level by a physician. The F&P Solo mask is<br>intended for single-patient use in the home and for<br>multiple patient use in the hospital or other clinical<br>setting where proper disinfection of the device can<br>occur between patient uses. | SL Model:<br>The F&P Evora Nasal Mask is intended to be<br>used by adults weighing ≥66lbs (30kgs) who<br>have been prescribed non-invasive positive<br>airway pressure therapy such as CPAP or<br>bilevel by a physician. The F&P Evora Nasal<br>Mask is intended for single patient use in the<br>home and for multiple patient use in the hospital<br>or other clinical setting where proper<br>disinfection of the device can occur between<br>uses. | Identical |
| Design/<br>Technological<br>Characteristics | Subject Device F&P<br>Solo Mask | Predicate Device<br>F&P Evora Nasal Mask<br>(K200089) | Comments |
| Availability | Prescription use only | Prescription use only | Identical |
| Patient Population | PAP Therapy | PAP Therapy | Identical |
| Patient Population | Adult (>30kg) | Adult (>30kg) | Identical |
| Patient Consciousness | Responsive and able to remove mask | Responsive and able to remove mask | Identical |
| Operating<br>Environment | Home, hospital or other clinical setting | Home, hospital or other clinical setting | Identical |
| Technical Specifications and Features | | | |
| Breathing Circuit | Single Inspiratory Tube | Single Inspiratory Tube | Identical |
| Breathing Tube<br>connection to mask | 22mm ISO Taper | 22mm ISO Taper | Identical |
| Seal sizes | Nasal Cushion: Available in four sizes:<br>Small, Medium, Large, Wide<br>Pillows Cushion:<br>Available in three sizes: Small, Medium, Large | Predicate:<br>Available in four sizes: Small, Medium, Large,<br>Wide | While the subject device's Nasal<br>cushions are identical, the Pillows<br>cushion offers three additional<br>sizes.<br>The subject device is designed to<br>be in conformance to ISO<br>17510:2015. |
| Headgear Sizes | One size | One size | Identical |
| Exhalation Vent | Numerous vent holes in Mask Frame | Numerous radial vent holes in Mask Frame | Identical.<br>The subject device vent holes are<br>located on the frame, whilst the<br>predicate has vent holes organized<br>in a radial pattern, also located on<br>the frame.<br>The subject device is designed to<br>be in conformance to ISO<br>17510:2015. |
| Design/<br>Technological<br>Characteristics | Subject Device F&P<br>Solo Mask | Predicate Device<br>F&P Evora Nasal Mask<br>(K200089) | Comments |
| Technical Specifications | | | |
| Pressure Range | 4 to 20 cm H2O | 4 to 25 cm H2O | The subject device is designed to be<br>in conformance with ISO 17510:2015 |
| Resistance to Flow | Pressure drop through Nasal Cushion: | Nasal (F&P Evora Nasal, predicate device):<br>Pressure drop through small 50 L/min: 1.0 ±<br>0.1cmH2O | The pressure drop for the subject<br>device is disclosed in labelling in |
| | Small 50 L/min: 0.84 cmH2O<br>Medium 50 L/min: 0.59 cmH2O | Pressure drop through medium 50 L/min: 1.0<br>± 0.1 cmH2O | accordance with ISO 17510:2015. |
| | Large 50 L/min: 0.54 cmH2O | Pressure drop through large 50 L/min: 1.0 ±<br>0.1 cmH2O | The reference device (Brevida,<br>K161412) has been used to support<br>claims of substantial equivalence for |
| | Wide 50 L/min: 0.58 cmH2O | Pressure drop through wide 50 L/min: 1.0 ±<br>0.1 cmH2O | pressure drop values with respect to<br>the 'Pillows Cushion'. |
| | Small 100 L/min: 3.63 cmH2O<br>Medium 100 L/min: 2.73 cmH2O | Pressure drop through small 100 L/min: 1.4 ±<br>0.25 cmH2O<br>Pressure drop through medium 100 L/min: 1.2<br>± 0.25 cmH2O<br>Pressure drop through large 100 L/min: 1.2 ±<br>0.25 cmH2O<br>Pressure drop through wide 100 L/min: 1.3 ±<br>0.25 cmH2O | |
| Large 100 L/min: 2.50 cmH2O<br>Wide 100 L/min: 2.65 cmH2O | | | |
| Pressure drop through Pillows Cushion:<br>Small 50 L/min: 2.17 cmH2O | | | |
| Medium 50 L/min: 1.05 cmH2O<br>Large 50 L/min: 0.89 cmH2O | | | |
| Small 100 L/min: 8.56 cmH2O<br>Medium 100 L/min: 4.42 cmH2O<br>Large 100 L/min: 3.83 cmH2O | | | |
| Dead Space | Nasal:<br>Small: 29.6 cc<br>Medium: 26.8 cc | Small: 28 cc<br>Medium: 26 cc<br>Large: 28 cc<br>Wide: 34 cc | Measured dead space is disclosed in<br>labelling in accordance with ISO<br>17510:2015. |
| | Large: 31.1 cc<br>Wide: 36.9 cc | | |
| | Pillows:<br>Small: 25.8 cc<br>Medium: 27.7 cc<br>Large: 30.5 cc | | |
| Sound | Sound Power Level of the Mask is 31.5 dBA, with<br>uncertainty 2.5 dBA<br>Sound Pressure Level of the Mask 23.6 dBA, with<br>uncertainty 2.5 dBA | Sound Power Level of the Mask is 26.8 dBA,<br>with uncertainty 2.5 dBA<br>Sound Pressure Level of the Mask 18.8 dBA,<br>with uncertainty 2.5 dBA | Measured sound power level is<br>disclosed in labelling in accordance<br>with ISO 17510:2015. |
| Shelf-Life | 2 year | 1 year | The subject device claims a 2-year<br>shelf life. |
| Design/<br>Technological<br>Characteristics | Subject Device F&P<br>Solo Mask | Predicate Device<br>F&P Evora Nasal Mask<br>(K200089) | Comments |
| Operating and Storage<br>Conditions | Operating Temperature: 5 to 40 °C<br>Storage Temperature: -20 to 50°C | Operating Temperature: 5 to 40 °C<br>Storage Temperature: -20 to 50°C | Identical |
| Cleaning and High-Level Disinfection | | | |
| Sterility | Device not provided sterile | Device not provided sterile | Identical |
| Reusability | Reusable - Multi Patient Use | Reusable - Multi Patient Use | Identical |
| High Level Disinfection<br>Methods | Thermal Disinfection: 75°C (167°F) for 30 mins<br>80°C (176°F) for 10 mins | Thermal Disinfection: 75°C (167°F) for 30<br>mins 80°C (176°F) for 10 mins 90°C (194°F)<br>for 1 min…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.