The F&P 820 System is intended to provide therapeutic levels of heat and humidity to a patient's inspired respiratory gases when using a continuous or intermittent ventilator system or a continuous gas flow. This system is intended for both noninvasive and invasive therapies. The addition of heat and humidity to the supply of cold and dry respiratory gases provided through noninvasive ventilation is beneficial to prevent drying of the patient's airways. The F&P 820 System is designed for adult and pediatric patients (excluding neonate), requiring a flow range ≥ 5 L/min. It is designed for use in hospitals, long term care facilities and homes, under the prescription of qualified medical professionals.
Device Story
F&P 820 Humidification System provides heated, humidified respiratory gases to patients via ventilator systems or continuous gas flow. Device consists of a humidifier base, F&P 820A10 breathing tubes, and MR325 humidification chamber. Operation involves passing gas through a heated chamber and heated breathing tube to add moisture and heat, preventing airway drying. Used in hospitals, long-term care facilities, and homes by clinicians or lay users under prescription. Output is conditioned gas delivered to patient interface (invasive or non-invasive). System includes auditory alarms for safety. Benefits include maintenance of airway hydration and patient comfort during respiratory therapy.
Clinical Evidence
Bench testing only. No clinical data. Testing included humidification output, thermal overshoot, surface temperature (ISO 80601-2-74), resistance to flow, compliance, and gas leak testing (ISO 5367). Biocompatibility evaluated per ISO 10993-1 and ISO 18562-1. Electrical safety and EMC testing performed per ANSI AAMI ES 60601-1 and IEC 60601-1-2. Human factors and usability engineering study confirmed safety and effectiveness for intended users.
Technological Characteristics
Respiratory gas humidifier; provides heat/humidity via heated chamber and heated breathing tube. Class II electrical device. Complies with ISO 80601-2-74, ISO 5367, IEC 60601-1, and IEC 60601-1-2. Features auditory alarms (>45 dbA). Operating temperature range 18–26°C. Service life 7 years. Software-controlled temperature regulation.
Indications for Use
Indicated for adult and pediatric patients (excluding neonates) requiring a flow range ≥ 5 L/min who need heated and humidified respiratory gases during continuous or intermittent ventilation or continuous gas flow therapy. Used in hospital, long-term care, and home settings under prescription.
Regulatory Classification
Identification
A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 5, 2023
Fisher & Paykel Healthcare Ltd Reena Daken Regulatory Affairs Manager - North America 15 Maurice Paykel Place, East Tamaki Auckland. 2013 New Zealand
Re: K223684
Trade/Device Name: F&P 820 Humidification System Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory Gas Humidifier Regulatory Class: Class II Product Code: BTT Dated: December 8, 2022 Received: December 8, 2022
Dear Reena Daken:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Ethan L. Nyberg -S
Ethan Nyberg, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K223684
Device Name F&P 820 Humidification System
### Indications for Use (Describe)
F&P 820 System Humidifier Base:
The F&P 820 System is intended to provide therapeutic levels of heat and humidity to a patient's inspired respiratory gases when using a continuous or intermittent ventilator system or a continuous gas flow.
This system is intended for both noninvasive and invasive therapies. The addition of heat and humidity to the supply of cold and dry respiratory gases provided through noninvasive ventilation is beneficial to prevent drying of the patient's airways.
The F&P 820 System is designed for adult and pediatric patients (excluding neonate), requiring a flow range ≥ 5 L/min. It is designed for use in hospitals, long term care facilities and homes, under the prescription of qualified medical professionals.
F&P 820A10 Breathing Tubes:
The F&P 820 series breathing tubes are an accessory to the F&P 820 System and are compatible with F&P 820 series humidifiers.
The F&P 820 System is intended to provide therapeutic levels of heat and humidity to a patient's inspiratory gases when using a continuous or intermittent ventilator system or a continuous gas flow.
This system is intended for both noninvasive and invasive therapies. The addition of heat and humidity to the supply of cold and dry respiratory gases provided through noninvasive ventilation is beneficial to prevent drying of the patient's airways.
This system is designed for adult and pediatric (excluding neonate) use in hospitals, long term care facilities and homes, under the prescription of qualified medical professionals.
F&P MR325 Humidification Chamber:
The MR325 humidification chamber is an accessory to the F&P 820 System and is compatible with F&P 820 series humidifiers.
The F&P 820 System is intended to provide therapeutic levels of heat and humidity to a patient's inspired respiratory gases when using a continuous or intermittent ventilator system or a continuous gas flow.
This system is intended for both noninvasive and invasive therapies. The addition of heat and humidity to the supply of cold and dry respiratory gases provided through noninvasive ventilation is beneficial to prevent drying of the patient's airways.
This system is designed for adult and pediatric (excluding neonate) use in hospitals, long term care facilities and homes, under the prescription of qualified medical professionals.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
As required by 21 CFR 807.92
# I. SUBMITTER
| Company Name and Address | Fisher & Paykel Healthcare Limited<br>15 Maurice Paykel Place<br>East Tamaki<br>Auckland 2013, New Zealand<br>Telephone: +64 9 574 0100 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Prepared and Submitted by | Arisha Samad<br>Regulatory Affairs Specialist |
| Contact Person | Reena Daken<br>Regulatory Affairs Manager<br>Telephone: +64 9 574 0100<br>Email: reena.daken@fphcare.co.nz |
| Date Prepared | 05 September 2023 |
# II. DEVICE
| Name of Device | F&P 820 Humidification System |
|------------------------|-------------------------------|
| Common/Usual Name | Respiratory Humidifier |
| Classification Name | Respiratory Gas Humidifier |
| Regulatory Class | Class II |
| Primary Classification | BTT (21 CFR §868.5450) |
| Product Code | |
{5}------------------------------------------------
#### PREDICATE DEVICE III.
- Primary Predicate device: .
| 510(k) Number | Device Name |
|---------------|----------------------------------|
| K143646 | F&P MR810 Respiratory Humidifier |
- Secondary Predicate devices: .
| 510(k)<br>Number | Device Name | Reason for secondary predicate |
|------------------|---------------------------------------|------------------------------------------------------------------------------|
| K983112 | Adult breathing circuit (RT102) | Used as a predicate for the F&P 820A10J<br>accessory breathing circuit kits. |
| K003973 | F&P humidification chamber<br>(HC325) | Used as a predicate for the F&P MR325<br>accessory humidification chamber. |
- . Reference device(s):
| 510(k)<br>Number | Device Name | Reason for reference |
|------------------|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K152029 | Hamilton-BC4022- Adult single<br>limb breathing set | Used to address the technological differences<br>between the subject and predicate device and<br>support claims of safety and effectiveness with<br>respect to breathing circuit use in patient<br>population |
| K143646 | F&P Adult breathing circuit<br>(900MR810) | Used to address the technological differences<br>between the subject and predicate device. |
#### DEVICE DESCRIPTION IV.
The Fisher & Paykel 820 Humidification (also referred to as F&P 820 System) is designed to provide respiratory humidification by providing heat, humidity, and delivering respiratory gases to patients.
The F&P 820 System is a revision of the predicate device platform, the Fisher & Paykel MR810 Respiratory Humidifier (K143646).
{6}------------------------------------------------
## V. INDICATIONS FOR USE STATEMENT
### F&P 820 System Humidifier Base:
The F&P 820 System is intended to provide therapeutic levels of heat and humidity to a patient's inspired respiratory gases when using a continuous or intermittent ventilator system or a continuous gas flow.
This system is intended for both noninvasive and invasive therapies. The addition of heat and humidity to the supply of cold and dry respiratory gases provided through noninvasive or invasive ventilation is beneficial to prevent drying of the patient's airways.
The F&P 820 System is designed for adult and pediatric (excluding neonate), requiring a flow range ≥ 5 L/min. It is designed for use in hospitals, long term care facilities and homes, under the prescription of qualified medical professionals.
### F&P 820A10 Breathing Tubes:
The F&P 820 series breathing tubes are an accessory to the F&P 820 System and are compatible with F&P 820 series humidifiers.
The F&P 820 System is intended to provide therapeutic levels of heat and humidity to a patient's inspired respiratory gases when using a continuous or intermittent ventilator system or a continuous gas flow.
This system is intended for both noninvasive and invasive therapies. The addition of heat and humidity to the supply of cold and dry respiratory gases provided through noninvasive or invasive ventilation is beneficial to prevent drying of the patient's airways.
This system is designed for adult and pediatric (excluding neonate) use in hospitals, long term care facilities and homes, under the prescription of qualified medical professionals.
### F&P MR325 Humidification Chamber:
The MR325 humidification chamber is an accessory to the F&P 820 System and is compatible with F&P 820 series humidifiers.
The F&P 820 System is intended to provide therapeutic levels of heat and humidity to a patient's inspired respiratory gases when using a continuous or intermittent ventilator system or a continuous gas flow.
This system is intended for both noninvasive and invasive therapies. The addition of heat and humidity to the supply of cold and dry respiratory gases provided through noninvasive or invasive ventilation is beneficial to prevent drying of the patient's airways.
This system is designed for adult and pediatric (excluding neonate) use in hospitals, long term care facilities and homes, under the prescription of qualified medical professionals.
{7}------------------------------------------------
# VI.COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| Design /<br>Technological<br>Characteristics | Subject Device<br>F&P 820 System - Humidifier Base | Predicate Device<br>F&P MR810 Respiratory Humidifier (K143646) | Comments |
|--------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use<br>statement | The Fisher & Paykel 820 System is intended to provide<br>therapeutic levels of heat and humidity to a patient's inspired<br>respiratory gases, when using a continuous or intermittent<br>ventilator system or a continuous gas flow.<br>The 820 System is intended for both non-invasive and<br>invasive therapies. Addition of heat and humidity to the<br>supply of cold and dry respiratory gases provided through<br>non-invasive or invasive ventilation is beneficial to prevent<br>drying of the patient airways.<br>The 820 System is for adult and pediatric (excluding<br>Neonatal) patients requiring a flow range $≥5$ L/min. For use in<br>hospitals, long term care facilities and homes under the<br>prescription of a qualified medical professional. | The Fisher & Paykel MR810 System is intended to provide<br>therapeutic levels of heat and humidity to a patient's inspired<br>respiratory gases, when using a continuous or intermittent<br>ventilator system or a continuous gas flow system.<br>The MR810 System is intended for both non-invasive and<br>invasive therapies. Addition of heat and humidity to the supply<br>of cold and dry respiratory gases provided through non-<br>invasive or invasive ventilation is beneficial to prevent drying of<br>the patient airways.<br>The MR810 System is for use for adult and pediatric patients<br>requiring a flow range $≥5$ L/min. The MR810 System is<br>designed for use in hospitals, long term care facilities and<br>homes under the prescription of a qualified medical<br>professional. | Identical |
| Application | Non-invasive<br>Invasive | Non-invasive<br>Invasive | Identical |
| Use Environment | Hospital<br>Home | Hospital<br>Home | Identical |
| Patient Population | Adult and Paediatric (Excluding Neonatal) | Adult and Paediatric (Excluding Neonatal) | Identical |
| Intended User Group | Medical or clinical professionals.<br>Technical healthcare professionals.<br>Professional Care givers.<br>Lay users. | Medical or clinical professionals.<br>Technical healthcare professionals.<br>Professional Care givers.<br>Lay users. | Identical |
| Operating Principle | Provides heat and humidity to the respiratory gases by<br>passing the gas through a humidification chamber and heated<br>breathing tube. | Provides heat and humidity to the respiratory gases by passing<br>the gas through a humidification chamber and heated<br>breathing tube. | Identical |
| Design /<br>Technological<br>Characteristics | Subject Device<br>F&P 820 System - Humidifier Base | Predicate Device<br>F&P MR810 Respiratory Humidifier (K143646) | Comments |
| Humidity<br>Performance | ≥ 12 mg/L in Settings 1, 2 or 3<br>For 5 - 70 L/min Flow Range<br>≥ 33 mg/L in Setting 4<br>For 5 - 40 L/min Flow Range | ≥ 10 mg/L in Setting Low, Med or High:<br>For 5 - 60 L/min Flow Range<br>≥ 33 mg/L in Setting High<br>For 5 - 30 L/min Flow Range | Equivalent<br>The 820 System meets the<br>applicable requirements of ISO<br>80601-2-74:2017 |
| Gas temperature at<br>patient end of<br>breathing tube for<br>flow range | Setting 1: 26-30 °C<br>Setting 2: 29-37 °C<br>Setting 3: 32-40 °C<br>For 5 - 70 L/min Flow Range<br>Setting 4: 35-43 °C<br>For 5 - 40 L/min Flow Range | Setting Low: 28 - 29 °C<br>Setting Medium: 28 - 31 °C<br>For 5 - 60 L/min Flow Range<br>Setting High: 33 - 36°C<br>For 5 - 30 L/min Flow Range | Equivalent<br>The 820 System meets the<br>applicable requirements of ISO<br>80601-2-74:2017 |
| Time to reach set<br>temperature | <60 mins | <60 mins | Identical |
| Ambient Temperature<br>Range | 18 – 26 °C | 18 - 26 °C | Identical |
| Alarms | Alarms exceed 45 dbA @1m<br>Auditory alarm pause: 120 seconds | No audio alarm functionality | Different<br>The subject device conforms to IEC<br>60601-1-8:2006 + A1:2012 |
| Transport Conditions | -25 to 70 Deg °C | -10 to +50 Dec °C | Equivalent<br>The subject device has broader<br>storage conditions than the primary<br>predicate device. |
| Service Life | 7 Years | 7 Years | Identical |
| IEC 60601-1<br>Classification | CLASS II | CLASS I | Different<br>The subject device conforms to<br>AAMI/IEC 60601-1:2005 +<br>AMD1:2012 |
| Characteristic for | Subject Device | Secondary Predicate Device | Comments |
| comparison | F&P 820A10J | F&P RT102 (K983112) | |
| Intended Use | Deliver heated and humidified respiratory gases. | Deliver heated and humidified respiratory gases. | Identical |
| Indications for Use | The breathing set is an accessory to the F&P 820 System. The | The heated-wire breathing circuits are intended as conduits of | Equivalent |
| | F&P 820 System is intended to provide therapeutic levels of | breathing gas for ventilation of patients, and to maintain the | Both the subject and secondary |
| | heat and humidity to a patient's inspired respiratory gases, | temperature of humidified inspired gas, to reduce | predicate device are indicated to |
| | when using a continuous or intermittent ventilator system or a | condensation. They are accessories for the Fisher & Paykel | provide heat and humidity to |
| | continuous gas flow. | MR850 Respiratory Gas Humidifier. | respiratory gases delivered to |
| | The 820 System is intended for both non-invasive and invasive | | patients. |
| | therapies. Addition of heat and humidity to the supply of cold | | |
| | and dry respiratory gases provided through non-invasive or | | |
| | invasive ventilation is beneficial to prevent drying of the patient | | |
| | airways. | | |
| | The 820 System is for adult and pediatric (excluding Neonatal) | | |
| | for use in hospitals, long term care facilities and homes under | | |
| | the prescription of a qualified medical professional. | | |
| Operating Principle | The heated breathing tube is kept warm by the heater wire | The heated breathing tube is kept warm by the heater wire | Identical |
| | (controlled by the Humidification Base) to minimize | (controlled by the Humidification Base) to minimize | |
| | condensation and maintain the desired temperature and level…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.