CLEAR+BRILLIANT TOUCH® Laser System

K223647 · Solta Medical, Inc. · GEX · Dec 29, 2022 · General, Plastic Surgery

Device Facts

Record IDK223647
Device NameCLEAR+BRILLIANT TOUCH® Laser System
ApplicantSolta Medical, Inc.
Product CodeGEX · General, Plastic Surgery
Decision DateDec 29, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic, 3rd-Party Reviewed

Indications for Use

The CLEAR + BRILLIANT TOUCH® System is indicated for use in dermatological procedures requiring the coagulation of soft tissue and general skin resurfacing procedures.

Device Story

Non-ablative fractional laser system for dermatological skin resurfacing; utilizes 1440nm and 1927nm diode lasers. System comprises console, handpieces, treatment tips, and credit key. Operator (physician/clinician) uses touchscreen interface to control treatment parameters; system delivers fractional photothermolysis. Embedded processor monitors system status and ensures correct operation via self-diagnostic algorithms; laser energy output inhibited unless treatment tip contacts patient skin. Device used in clinical settings to improve skin appearance; benefits include controlled tissue coagulation and resurfacing with minimal maintenance due to solid-state design.

Clinical Evidence

Bench testing only. No clinical data presented. Device demonstrated compliance with ANSI/AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1. Biocompatibility testing performed per ISO 10993-1, -5, and -10. Usability engineering conducted per IEC 62366-1 and IEC 60601-1-6.

Technological Characteristics

Solid-state diode laser; 1440nm and 1927nm wavelengths. Passive conductance skin cooling. Console-based system with touchscreen, mechanical handpiece latches, and credit key authentication. Materials include polycarbonate (Americhem FRPC300 L32721). Software-controlled fractional output. Complies with IEC 60601-1, 60601-1-2, 60601-2-22, 60825-1, and ISO 10993 standards.

Indications for Use

Indicated for dermatological procedures requiring soft tissue coagulation and general skin resurfacing. Patient population not explicitly defined by age or gender.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 29, 2022 Solta Medical, Inc. % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114 Re: K223647 Trade/Device Name: CLEAR+BRILLIANT TOUCH® Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX, ONG Dated: December 5, 2022 Received: December 6, 2022 Dear Prithul Bom: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Jessica Carr -S for Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K223647 Device Name CLEAR+BRILLIANT TOUCH® Laser System Indications for Use (Describe) The CLEAR + BRILLIANT TOUCH® System is indicated for use in dermatological procedures requiring the coagulation of soft tissue and general skin resurfacing procedures. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over The Counter Use (21 CFR 80 | |------------------------------------------------------------|-----------------------------------------------| |------------------------------------------------------------|-----------------------------------------------| | X | Prescription Use (Part 21 CFR 801 Subpart D) __ Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## Summary Preparation Date: December 27, 2022 #### 1. 510(k) Owner Name and Address Solta Medical, Inc. 11720 North Creek Parkway N, Suite 100 Bothell, WA 98011 USA > Submitter & Primary Contact: Aditi Chaubal (primary contact) Phone: 425-420-2350 Fax: 1-425-420-2300 aditi.chaubal@solta.com ### 2. I dentification of Subject Device | Trade Name: | CLEAR+BRILLIANT TOUCH® Laser System | |------------------------|--------------------------------------------------------------------| | Common Name: | Powered Laser Surgical Instrument with Microbeam\Fractional Output | | Regulation Number: | 21 CFR 878.4810 | | Product code: | GEX, ONG | | Device Panel: | General & Plastic Surgery | | Device Classification: | Class II | #### 3. Predicate Device | Trade Name: | CLEAR+BRILLIANT® Laser System | |------------------------|--------------------------------------------------------------------| | Common Name: | Powered Laser Surgical Instrument with Microbeam\Fractional Output | | Regulation Number: | 21 CFR 878.4810 | | Product code: | GEX, ONG | | Device Panel: | General & Plastic Surgery | | Device Classification: | Class II | | 510(k) Number: | K120433, K110349 | #### 4. Device Description The CLEAR + BRILLIANT TOUCH® Laser System is the second generation of the Clear+Brilliant line of products. Clear+Brilliant products use non-ablative laser designed for use in dermatological procedures intended to improve the appearance of a patient's skin. The system incorporates self-diagnostic algorithms to ensure its correct operation. The CLEAR + {4}------------------------------------------------ BRILLIANT TOUCH® Laser System is part of the family of Solta Medical products using the fractional photothermolysis principle to resurface the skin. CLEAR + BRILLIANT TOUCH® Laser System is comprised of four sub-systems: - 1. Console - 2. Handpieces: 1440nm and 1927nm (Perméa) handpieces - 3. Treatment Tip - 4. Credit Key The console and up to two handpieces are connected mechanically and electrically by an umbilical cable. The treatment tip is mechanically and magnetically secured to the handpiece. All components must be properly connected and fully functional for the system to be operable. The system will not permit laser energy output until the treatment tip is in contact with the patient's skin. The embedded processor and software control the precise delivery of the fractional laser treatment by directing the electrical and mechanical components that determine the position, spacing, and focus of the optical output. The processor and software also monitor the status of a treatment and the system and present this information to the operator on the touchscreen. The Credit Key is the replaceable accessory allowing access to treatment credits and is attachable to the console. It is removable and is disposed of after the credit balance reaches zero. The solid-state design of CLEAR + BRILLIANT TOUCH® Laser System limits maintenance and minimizes utility requirements. ## 5. Indications for Use "The CLEAR + BRILLIANT TOUCH® Laser System is indicated for use in dermatological procedures requiring the coagulation of soft tissue and general skin resurfacing procedures." ## 6. Comparison of Technological Characteristics The CLEAR+BRILLIANT TOUCH® Laser System and accessories share similar device operation, overall technical and functional capabilities. The design, function and treatment delivery are equivalent to the predicate device. | | CLEAR+BRILLIANT<br>(Predicate – K120433) | CLEAR+BRILLIANT Touch®<br>(Subject Device) | | |-----------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------| | Feature | | | Remark | | Product Code &<br>Reg. Classification | GEX/ONG, Class II | GEX/ONG, Class II | Equivalent | | Laser Classification | Class III-R | Class III-R | Equivalent | | Indications for Use | The Clear + Brilliant Laser System<br>is intended for dermatological<br>procedures requiring the<br>coagulation of soft tissue and | The CLEAR + BRILLIANT TOUCH®<br>System is indicated for use in<br>dermatological procedures requiring | Equivalent | | | general skin resurfacing | the coagulation of soft tissue and<br>general skin resurfacing procedures. | | | HandPiece | | | | | Laser Type | Diode Laser | Diode Laser | Equivalent | | | 1440 | 1440 | | | Wavelength, nm | 1927 | 1927 | Equivalent | | Peak Laser Power,<br>W | 1440nm: 2.5 Watts<br>1927nm: 0.9 Watts | 1440nm: 2.5 Watts<br>1927nm: 0.9 Watts | Equivalent | | Electrical | 110-240V- AC | 110-240V- AC | Equivalent | | Skin Cooling<br>Method | Passive Conductance | Passive Conductance | Equivalent | | Main Components | Console, hand piece(s), Treatment<br>Tips | Console, hand piece(s), Vented<br>Treatment Tips | Equivalent | | Laser Activation in<br>Hand or Foot<br>Activated | Hand switch activated | Hand switch activated | Equivalent | | Treatment Tip | | | | | Tip Design | No vents | Vented | Difference | | | | | Difference, but<br>no new<br>biocompatibility<br>safety concerns. | | Material | Polycarbonate GE HPH4404,<br>translucent grey (Polycarbonate) | Americhem FRPC300 L32721, grey<br>tinted (Polycarbonate) | | | Console | | | | | User Interface | LCD | Touchscreen Technology for user<br>interaction | Difference | | Number of<br>handpiece cradles | 1 | 2 | Difference | | Handpiece to<br>console connection | Manually fastened screws | Mechanical Latch | Difference | | Number of<br>Handpieces<br>connected during<br>operation | 1 | 2 | Difference | | Treatment<br>Authentication | | | | | Authenticator | SmartCard | Credit Key | Difference | | Authenticator Use | Different SmartCard for different<br>Handpieces | Same credit key for both Handpieces | Difference | | Dimensions | | | | | Console<br>Dimensions<br>(Heights, Width,<br>Depth) | 17.62"x12.47"x 8.03" | 12.5" W x 11.25" H x 11" D<br>(dimensions of the system with two<br>stowed handpieces) | Difference | | Hand Piece<br>Dimensions<br>(Height, Width,<br>Depth) | 8.3"x3.1"x2.0" | 8.3"x3.1"x2.0" | Equivalent | | Weight | Approximately 15 lbs. | Approximately 15.6 lbs. (weight of<br>system with two stowed handpieces<br>and credit key installed on console) | Equivalent | | Hand Piece Weight | 330 grams or approximately 0.7 lbs. | 330 grams or approximately 0.7 lbs. | Equivalent | | Treatment | | | | | Density | 1927nm: 20-40 microthermal zone punctures<br>1440nm: 40-60 micro thermal zone punctures per cm2 per pass | 1927nm: 20-40 microthermal zone punctures<br>1440nm: 40-60 micro thermal zone punctures per cm2 per pass | Equivalent | | Treatment coverage based on Lesion<br>Cross Section and Density | Approximately 4% to 12% | Approximately 4% to 12% | Equivalent | | Treatment width | 10 mm | 10 mm | Equivalent | | Operating<br>Conditions | | | | | Operating<br>Conditions:<br>Temperature | 59 to 86-degree F (15 to 30 degree C) | 59 to 86-degree F (15 to 30 degree C) | Equivalent | | Operating<br>Conditions:<br>Relative Humidity | 30-75%, Non-condensing | 30-75%, Non-condensing | Equivalent | | Operating<br>Conditions:<br>Altitude | <10, 000 Feet (<3,075 Meters) | <10, 000 Feet (<3,075 Meters) | Equivalent | ## Summary of the technological characteristics | 510(k) Summary PREMARKET NOTIFICATION TRADITIONAL 510(K): CLEAR+BRILLIANT TOUCH® LASER SYSTEM {5}------------------------------------------------ 3 | *510(k) Summary* 3 | 510(k) Summary PREMARKET NOTIFICATION TRADITIONAL 510(K): CLEAR+BRILLIANT TOUCH® LASER SYSTEM {6}------------------------------------------------ The CLEAR+BRILLIANT TOUCH® Laser System is developed on the Solta System Controller (SC) platform. This platform comprises hardware, firmware, and software (SW) which are common across different Solta products. This architecture is consistent with the Solta New Product architecture. The software portion of this architecture is split into two portions. 1. System controller software (SC) - 2. CLEAR+BRILLIANT TOUCH® Laser System software The System Controller software is responsible for Operating System (OS) functions and common product functions (across products such as infrastructure). This is a software component that implements common functions across products implemented at Solta Medical. This is shared as source code between the various products and will be included in the CLEAR+BRILLIANT TOUCH® Laser System software build. ## 7. Non-Clinical Performance Testing As with the predicate device, safety tests of the CLEAR+BRILLIANT TOUCH® Laser System has also demonstrated its compliance with applicable requirements of the following standards: | Document No. | Document Title | |------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------| | ANSI/AAMI ES 60601-<br>1(2005) + AMD (2012) and<br>AAMI STD ES60601-1. | Medical electrical equipment – Part 1: General requirements for<br>basic safety and essential performance | {7}------------------------------------------------ | Document No. | Document Title | |-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | IEC/EN 60601-1-2 Ed.4.0<br>(2014) | Medical electrical equipment - Part 1-2: General requirements for<br>basic safety and essential performance - Collateral Standard:<br>Electromagnetic disturbances - Requirements and tests | | IEC 60601-2-22:2012,<br>Edition 3.1 | Medical Electrical Equipment—Part 2-2: Particular requirements<br>for basic safety and essential performance of surgical, cosmetic,<br>therapeutic and diagnostic laser equipment. | | IEC 60825-1:2014 Edition<br>3.0 | Safety of laser products - Part 1: Equipment classification and<br>requirements. | # Biocompatibility | Document No. | Document Title | |----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | ISO 10993-1:2018 | Biological evaluation of medical devices -- Part 1: Evaluation and<br>testing within a risk management process | | EN/ISO-10993-5: 2009 | International Organization for Standardization (ISO) 10993-5,<br>Biological evaluation of medical devices - Part 5: Tests for in vitro<br>cytotoxicity. | | ISO 10993-10:2013 | International Organization for Standardization (ISO) 10993-10,<br>Biological evaluation of medical devices - Part 10: Tests for irritation<br>and skin sensitization (2010). | ## Management | Document No. | Document Title | |-------------------|------------------------------------------------------------------------------------| | 21 CFR Part 820 | FDA Quality Systems Regulation | | ISO 13485 | Medical Devices – Quality Management Systems | | EN ISO 13485:2016 | Medical devices—Quality management systems—Requirements for<br>regulatory purposes | | EN ISO 14971:2019 | Medical Devices—Application of risk management to medical<br>devices | | EN 62304:2015 | Medical device software—Software life cycle processes | {8}------------------------------------------------ | Document No. | Document Title | |------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------| | EN/IEC 62366-1 b: 2015 Edition<br>1.0 b cor.1: 2016 | Medical devices - Application of usability engineering to<br>medical devices | | IEC 60601-1-6:2013, Edition 3.1 | Medical electrical equipment -- Part 1-6: General requirements<br>for basic safety and essential performance - Collateral standard:<br>Usability. | | ANSI/AAMI HE75: 2009<br>(R)2018 (Reference Standard) | Human factors engineering - Design of medical devices | ## Human Factors and Usability Engineering ### Safe Transit | Document No. | Document Title | |------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------| | ISTA testing standards | International Safe Transit Association; Packaging integrity performance<br>tests; General test standards are ISTA 1A, 2A, and 3A per transport<br>environment. | The CLEAR+BRILLIANT TOUCH® Laser System passed all the above applicable standards testing. The product performance testing demonstrates that the functional requirements have been met and that CLEAR+BRILLIANT TOUCH® Laser System is equivalent to the predicate device. ### 8. Conclusion The CLEAR+BRILLIANT TOUCH® Laser System shares the same design and functional features as the predicate device, including the laser type, wavelength and device design. Therefore, the CLEAR+BRILLIANT TOUCH® Laser System is substantially equivalent to the previously cleared version of the CLEAR+BRILLIANT Laser System.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...