K223608 · Instrumentation Laboratory Company · CHL · Aug 10, 2023 · Clinical Chemistry
Device Facts
Record ID
K223608
Device Name
GEM Premier 7000 with IQM3
Applicant
Instrumentation Laboratory Company
Product Code
CHL · Clinical Chemistry
Decision Date
Aug 10, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1120
Device Class
Class 2
Indications for Use
The GEM Premier 7000 with iQM₃ is a portable critical care system for use by health care professionals to rapidly analyze lithium heparinized whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory. The instrument provides quantitative measurements of pH, pCO2, pO2, sodium, potassium, chloride, ionized calcium, glucose, lactate, hematocrit, total bilirubin, and CO-Oximetry (tHb, O2Hb, COHb, MetHb, HHb, sO2*) parameters from arterial, venous, or capillary lithium heparinized whole blood. These parameters, along with derived parameters, aid in the diagnosis of a patient's acid/base status, electrolyte and metabolite balance and oxygen delivery capacity.
Device Story
Portable critical care system; analyzes lithium heparinized whole blood samples (arterial, venous, capillary) at point-of-care or central lab. Uses potentiometry, amperometry, electrical conductivity, and spectrophotometry to measure pH, blood gases, electrolytes, metabolites, hematocrit, bilirubin, and CO-Oximetry parameters. Includes iQM₃ (Intelligent Quality Management) with new photometric hemolysis detection feature. Device flags samples with hemolysis ≥116 mg/dL plasma-free hemoglobin to prevent reporting of inaccurate potassium results. Healthcare providers use output to diagnose acid-base disturbances, electrolyte imbalances, and oxygen delivery capacity. Benefits include rapid, accurate bedside testing and automated detection of pre-analytical sample interference.
Clinical Evidence
Bench testing only. Verification studies included internal method comparison, internal whole blood precision, hemolysis interference on potassium, hemolysis verification, shelf-life, and use-life studies. Testing followed CLSI guidelines (EP05-A3, EP07, EP09c, EP25-A, EP37). Results demonstrated that the addition of the hemolysis detection module does not impact performance and that the device is equivalent to the predicate.
Technological Characteristics
Portable critical care analyzer. Measurement principles: potentiometry (pH, pCO2, Na+, K+, Cl-, Ca++), amperometry (pO2, glucose, lactate), electrical conductivity (hematocrit), and spectrophotometry (bilirubin, CO-Oximetry). Features iQM₃ for automated quality management and photometric hemolysis detection. Operates on lithium heparinized whole blood. Connectivity and software architecture are consistent with the predicate GEM Premier 5000.
Indications for Use
Indicated for health care professionals to analyze lithium heparinized whole blood (arterial, venous, capillary) for pH, blood gases, electrolytes, metabolites, hematocrit, total bilirubin, and CO-Oximetry parameters to aid in diagnosis/treatment of acid-base, electrolyte, metabolite, and oxygenation status in clinical/laboratory settings.
Regulatory Classification
Identification
A blood gases (PCO2 , PO2 ) and blood pH test system is a device intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. Measurements of blood gases (PCO2 , PO2 ) and blood pH are used in the diagnosis and treatment of life-threatening acid-base disturbances.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K223608
B Applicant
Instrumentation Laboratory Company
C Proprietary and Established Names
GEM Premier 7000 with iQM₃
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| CHL | Class II | 21 CFR 862.1120 - Blood Gases (PCO2, PO2) And Blood Ph Test System | CH - Clinical Chemistry |
| JGS | Class II | 21 CFR 862.1665 – Sodium test system | CH - Clinical Chemistry |
| CEM | Class II | 21 CFR 862.1600 - Potassium test system | CH - Clinical Chemistry |
| CGZ | Class II | 21 CFR 862.1170 - Chloride test system | CH - Clinical Chemistry |
| JFP | Class II | 21 CFR 862.1145 - Calcium test system | CH - Clinical Chemistry |
| CGA | Class II | 21 CFR 862.1345 - Glucose test system | CH - Clinical Chemistry |
| KHP | Class I | 21 CFR 862.1450 - Lactic acid test system | CH - Clinical Chemistry |
| MQM | Class I, reserved | 21 CFR 862.1113 – Bilirubin (total and unbound) in the neonate test system | CH - Clinical Chemistry |
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| GKF | Class II | 21 CFR 864.5600 – Automated hematocrit instrument | HE – Hematology |
| GHS | Class II | 21 CFR 864.7425 - Carboxyhemoglobin assay | HE - Hematology |
| GKR | Class II | 21 CFR 864.5620 – Automated hemoglobin system | HE - Hematology |
| GLY | Class II | 21 CFR 864.7500 - Whole blood hemoglobin assays | HE - Hematology |
## II Submission/Device Overview:
### A Purpose for Submission:
Modification of an existing device
### B Measurand:
pH, pCO2, pO2, sodium, potassium, chloride, ionized calcium, glucose, lactate, hematocrit, total bilirubin, and CO-Oximetry (tHb, O2Hb, COHb, MetHb, HHb, sO2*)
### C Type of Test:
Quantitative, potentiometric for pH, pCO2, Na+, K+, Cl-, Ca++
Quantitative, amperometry for pO2, glucose and lactate
Quantitative, electrical conductivity for hematocrit
Quantitative, spectrophotometric method for total bilirubin, tHb, O2Hb, COHb, MetHb, HHb, sO2
## III Intended Use/Indications for Use:
### A Intended Use(s):
See Indications for Use below.
### B Indication(s) for Use:
The GEM Premier 7000 with iQM₃ is a portable critical care system for use by health care professionals to rapidly analyze lithium heparinized whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory. The instrument provides quantitative measurements of pH, pCO2, pO2, sodium, potassium, chloride, ionized calcium, glucose, lactate, hematocrit, total bilirubin, and CO-Oximetry (tHb, O2Hb, COHb, MetHb, HHb, sO2*) parameters from arterial, venous, or capillary lithium heparinized whole blood. These parameters, along with derived parameters, aid in the diagnosis of a patient's acid/base status, electrolyte and metabolite balance and oxygen delivery capacity.
K223608 - Page 2 of 10
{2}
*sO2 = ratio between the concentration of oxyhemoglobin and oxyhemoglobin plus deoxyhemoglobin.
pH, pCO2, and pO2 measurements in whole blood are used in the diagnosis and treatment of life-threatening acid-base disturbances.
Electrolytes in the human body have multiple roles. Nearly all metabolic processes depend on or vary with electrolytes:
- Sodium (Na+) measurements are used in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension, Addison’s disease, dehydration, inappropriate antidiuretic secretion, or other diseases involving electrolyte imbalance.
- Potassium (K+) measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels.
- Ionized calcium (Ca++) measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease, and tetany.
- Chloride (Cl-) measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders, such as cystic fibrosis and diabetic acidosis.
Hematocrit (Hct) measurements in whole blood of the packed red cell volume of a blood sample are used to distinguish normal from abnormal states, such as anemia and erythrocytosis (an increase in the number of red cells).
Glucose (Glu) measurement is used in the diagnosis, monitoring and treatment of carbohydrate metabolism disturbances including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and pancreatic islet cell carcinoma.
Lactate (Lac) measurement is used:
- to evaluate the acid-base status of patients suspected of having lactic acidosis;
- to monitor tissue hypoxia and strenuous physical exertion;
- in the diagnosis of hyperlactatemia.
Total Bilirubin (tBili) measurement is used to aid in assessing the risk of kernicterus and hyperbilirubinemia in neonates.
CO-Oximetry (tHb, COHb, MetHb, O2Hb, HHb, and sO2) evaluates the ability of the blood to carry oxygen by measuring total hemoglobin and determining the percentage of functional and dysfunctional hemoglobin species.
- Total Hemoglobin (tHb): Total hemoglobin measurements are used to measure the hemoglobin content of whole blood for the detection of anemia.
- COHb: Carboxyhemoglobin measurements are used to determine the carboxyhemoglobin content of human blood as an aid in the diagnosis of carbon monoxide poisoning.
K223608 - Page 3 of 10
{3}
- MetHb: Methemoglobin measurements are used to determine different conditions of methemoglobinemia.
- HHb: Deoxyhemoglobin, as a fraction of total hemoglobin, is used in combination with oxyhemoglobin to measure oxygen status.
- O2Hb: Oxyhemoglobin, as a fraction of total hemoglobin, is used in combination with deoxyhemoglobin to measure oxygen status.
- sO2: Oxygen saturation, more specifically the ratio between the concentration of oxyhemoglobin and oxyhemoglobin plus deoxyhemoglobin, is used to measure oxygen status.
## C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
## D Special Instrument Requirements:
GEM Premier 7000 with iQM₃
## IV Device/System Characteristics:
### A Device Description:
The GEM Premier 7000 system with iQM₃ is a modified GEM Premier 5000 (predicate cleared under K203079) that includes a hemolysis detection feature within the Intelligent Quality Management system (iQM₃).
### B Principle of Operation:
Same as the predicate cleared under K203079 except for the addition of the photometric hemolysis detection feature.
### C Instrument Description Information:
1. Instrument Name: GEM Premier 7000 with iQM₃
2. Specimen Identification: Same as the predicate cleared under K203790
3. Specimen Sampling and Handling: Same as the predicate cleared under K203790
4. Calibration: Same as the predicate cleared under K203790
5. Quality Control: iQM₃ has the same quality control functions as iQM2 in the predicate except for the addition of the hemolysis detection feature.
K223608 - Page 4 of 10
{4}
V Substantial Equivalence Information:
A Predicate Device Name(s): GEM Premier 5000
B Predicate 510(k) Number(s): K203790
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K223608 | K203790 |
| --- | --- | --- |
| Device Trade Name | GEM Premier 7000 with iQM3 | GEM Premier 5000 |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The instrument provides quantitative measurements of pH, pCO2, pO2, sodium, potassium, chloride, ionized calcium, glucose, lactate, hematocrit, total bilirubin, and CO-Oximetry (tHb, O2Hb, COHb, MetHb, HHb, sO2*). These parameters, along with derived parameters, aid in the diagnosis of a patient's acid/base status, electrolyte and metabolite balance and oxygen delivery capacity. | Same |
| Intended User | Central Laboratory and Point-of-Care | Same |
| Measurement Principle | Potentiometry: pH, pCO2, Na+, K+, Cl-, Ca++
Amperometry: pO2, Glucose, Lactate
Spectrophotometry: tHb, O2Hb, COHb, MetHb, HHb, sO2, bilirubin
Conductivity: hematocrit | Same |
| Sample Type | Venous, arterial and capillary lithium heparinized whole blood | Same |
| General Device Characteristic Differences | | |
| Hemolysis detection | iQM3 - hemolysis detection in whole blood samples | iQM2 without hemolysis detection feature |
K223608 - Page 5 of 10
{5}
VI Standards/Guidance Documents Referenced:
CLSI EP05-A3: Evaluation of Precision of Quantitative Measurement Procedures, Third Edition;
CLSI EP09c: Measurement Procedure Comparison and Bias Estimation Using Patient Samples, Third Edition;
CLSI EP07: Interference Testing in Clinical Chemistry, Third Edition;
CLSI EP37: Supplemental Tables for Interference Testing in Clinical Chemistry, First Edition;
Content of Premarket Submissions for Management of Cybersecurity in Medical Devices
Guidance for Industry and Food and Drug Administration Staff, October 2, 2014.
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
An internal precision study was performed in normal mode (syringe) and in micro mode (capillary) using lithium heparin venous whole blood samples at 5 different levels for each analytes and the results support that the precision is not changed by adding the hemolysis detection feature to the device.
2. Linearity:
Previously established in K160225, K160412, and K160415.
3. Analytical Specificity/Interference:
In addition to the studies described in K160225, K160412, and K160415, a new interference study was performed to evaluate the potential impact of hemolysis interference on potassium results. Venous whole blood was ultrasonically-lysed to generate a high hemolysis pool (100% = 1000 mg/dL plasma-free hemoglobin). The high hemolysis pool was mixed with native venous whole blood without hemolysis (0%) to generate 3 additional hemolysis levels (25%, 50%, 75%). Testing was performed using two concentrations of potassium: 3.8 mmol/L and 5.3 mmol/L. Each combination of potassium level and hemolysis level was tested in triplicate on 3 GEM Premier 7000 with iQM₃, using 3 different cartridge lots, for a total of 9 replicates of each hemolysis level/potassium level. Both the normal mode and the micro mode were evaluated. Interference was calculated as the bias between the average result with the interfering substance and the average control measurement. The sponsor considered no significant interference to be ≤7% bias for this study. The highest level of hemolysis with no significant interference was 115 mg/dL plasma-free hemoglobin. Hemolysis of ≥116 mg/dL was found to have significant interferences the with potassium result.
The labeling contains the following statement:
Do not use hemolyzed samples, as hemolyzed samples may cause falsely elevated potassium results. A GEM Premier 7000 with iQM₃ hemolysis flag (estimated hemolysis = 116 mg/dL) indicates hemolysis interference has been detected and the potassium result should not be
K223608 - Page 6 of 10
{6}
reported out of the laboratory. It is strongly recommended that hemolysis detection remain enabled in order to detect hemolysis in samples.
# 4. Assay Reportable Range:
Previously established in K160225, K160412, and K160415.
# 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
As described in K160225, K160412, and K160415.
# 6. Detection Limit:
Previously established in K160225, K160412, and K160415.
# 7. Assay Cut-Off:
Not applicable.
# 8. Accuracy (Instrument):
See method comparison study.
# 9. Carry-Over:
Not applicable
# B Comparison Studies:
# 1. Method Comparison with Predicate Device:
Lithium heparin venous and arterial whole blood and fingerstick capillary samples were assayed by 3 point-of-care (POC) operators at 1 simulated clinical site using 2 GEM Premier 7000 with iQM₃ and 2 GEM Premier 5000. Venous and arterial samples were assayed in normal mode and fingerstick capillary samples were assayed in normal mode and micro mode. Each sample was run in singlicate on 1 GEM Premier 7000 with iQM₃ and 1 GEM Premier 5000. Deming Regression was used for analytes with constant SD and Passing-Bablok was used for analytes with mixed variability. The results support that the addition of the hemolysis detection feature did not impact the accuracy of device results in normal mode or in micro mode.
The representative study results for venous and arterial samples from POC in normal mode are shown below:
| Analyte | N | Sample Range | Analysis | Slope | Intercept | R |
| --- | --- | --- | --- | --- | --- | --- |
| pH | 156 | 7.03-7.75 | Deming | 0.981 | 0.135 | 0.996 |
| pCO2 (mmHg) | 159 | 10-116 | Passing-Bablok | 1.000 | 0.000 | 0.997 |
| pO2 (mmHg) | 160 | 30-536 | Passing-Bablok | 1.000 | 1.000 | 0.999 |
| Sodium (mmol/L) | 159 | 105-169 | Deming | 0.964 | 5.256 | 0.994 |
K223608 - Page 7 of 10
{7}
| Analyte | N | Sample Range | Analysis | Slope | Intercept | R |
| --- | --- | --- | --- | --- | --- | --- |
| Potassium(mmol/L) | 152 | 2.2-18.6 | Passing-Bablok | 1.000 | 0.000 | 0.999 |
| Chloride (mmol/L) | 158 | 48-156 | Passing-Bablok | 1.000 | 0.000 | 0.997 |
| Ionized Calcium (mmol/L) | 159 | 0.36-3.70 | Passing-Bablok | 0.982 | 0.003 | 0.999 |
| Glucose (mg/dL) | 155 | 12-637 | Passing-Bablok | 1.000 | 0.000 | 0.999 |
| Lactate (mmol/L) | 156 | 0.7-16.6 | Passing-Bablok | 0.970 | -0.027 | 0.997 |
| Hematocrit (%) | 157 | 19-68 | Deming | 0.996 | -0.370 | 0.995 |
| Total hemoglobin (g/dL) | 157 | 7.0-22.2 | Deming | 0.994 | 0.055 | 0.998 |
| Oxyhemoglobin (%) | 157 | 6.0-98.4 | Deming | 0.995 | 0.377 | 1.000 |
| Carboxyhemoglobin (%) | 157 | 1.2-65.6 | Deming | 0.998 | 0.020 | 0.999 |
| Methemoglobin (%) | 128 | 0.7-26.6 | Passing-Bablok | 0.994 | 0.045 | 0.999 |
| Deoxyhemoglobin (%) | 124 | 1.1-92.8 | Deming | 0.999 | -0.082 | 1.000 |
| sO2 (%) | 157 | 6.1-100.0 | Deming | 0.997 | 0.294 | 1.000 |
| Total bilirubin (mg/dL) | 106 | 2.0-35.8 | Passing-Bablok | 0.985 | -0.003 | 0.999 |
The study results for capillary fingerstick samples from POC in normal mode are shown below:
| Analyte | N | Sample Range | Analysis | Slope | Intercept | R |
| --- | --- | --- | --- | --- | --- | --- |
| pH (pH units) | 140 | 7.07-7.71 | Deming | 0.987 | 0.089 | 0.982 |
| pCO2 (mmHg) | 140 | 10-113 | Passing-Bablok | 1.000 | 0.000 | 0.991 |
| pO2 (mmHg) | 140 | 32-554 | Passing-Bablok | 1.000 | -3.000 | 0.997 |
| Sodium (mmol/L) | 137 | 106-169 | Deming | 0.990 | 1.685 | 0.987 |
| Potassium (mmol/L) | 138 | 2.2-18.0 | Passing-Bablok | 1.000 | 0.100 | 0.995 |
| Chloride (mmol/L) | 137 | 48-148 | Passing-Bablok | 1.000 | 0.000 | 0.993 |
| Ionized Calcium (mmol/L) | 138 | 0.36-3.60 | Passing-Bablok | 1.000 | 0.000 | 0.997 |
| Glucose (mg/dL) | 136 | 35-626 | Passing-Bablok | 1.000 | 0.000 | 0.998 |
| Lactate (mmol/L) | 135 | 0.5-12.5 | Passing-Bablok | 0.974 | -0.100 | 0.993 |
| Hematocrit (%) | 138 | 17-67 | Deming | 0.977 | -1.135 | 0.979 |
| Total hemoglobin (g/dL) | 133 | 7.1-22.0 | Deming | 0.981 | -0.005 | 0.992 |
| Oxyhemoglobin (%) | 130 | 6.1-97.5 | Deming | 0.979 | 0.834 | 0.997 |
| Carboxyhemoglobin (%) | 131 | 0.7-65.2 | Deming | 1.003 | 0.040 | 0.999 |
| Methemoglobin (%) | 107 | 0.7-26.7 | Passing-Bablok | 1.000 | 0.000 | 1.000 |
| Deoxyhemoglobin (%) | 128 | 1.1-92.6 | Deming | 0.975 | 0.349 | 0.997 |
| sO2 (%) | 130 | 6.2-99.2 | Deming | 0.984 | 0.845 | 0.996 |
| Total bilirubin (mg/dL) | 109 | 2.2-35.9 | Passing-Bablok | 0.986 | -0.200 | 0.999 |
K223608 - Page 8 of 10
{8}
The study results for capillary fingerstick samples in micro mode are shown below:
| Analyte | N | Sample Range | Analysis | Slope | Intercept | R |
| --- | --- | --- | --- | --- | --- | --- |
| pH (pH units) | 116 | 7.08-7.71 | Deming | 0.985 | 0.096 | 0.980 |
| pCO2 (mmHg) | 116 | 11-99 | Passing-Bablok | 1.000 | 0.000 | 0.989 |
| pO2 (mmHg) | 116 | 33-567 | Passing-Bablok | 0.978 | -0.648 | 0.996 |
| Sodium (mmol/L) | 114 | 106-170 | Deming | 1.007 | -0.445 | 0.990 |
| Potassium (mmol/L) | 114 | 2.2-18.6 | Passing-Bablok | 0.995 | 0.000 | 0.995 |
| Chloride (mmol/L) | 112 | 48-157 | Passing-Bablok | 0.996 | -1.000 | 0.996 |
| Ionized Calcium (mmol/L) | 114 | 0.35-3.55 | Passing-Bablok | 0.987 | 0.006 | 0.997 |
| Glucose (mg/dL) | 113 | 36-662 | Passing-Bablok | 1.000 | -2.000 | 0.999 |
| Lactate (mmol/L) | 115 | 0.5-13.4 | Passing-Bablok | 1.000 | 0.000 | 0.997 |
| Hematocrit (%) | 114 | 19-72 | Deming | 0.963 | 1.838 | 0.978 |
2. Matrix Comparison:
Not applicable.
C Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable.
D Clinical Cut-Off:
Not applicable.
E Expected Values/Reference Range:
Previously established in K160225, K160412, and K160415.
F Other Supportive Instrument Performance Characteristics Data:
The GEM Premier 7000 with iQM₃ does not report a specific plasma free hemoglobin result to the user, but displays flags indicating different degrees of hemolysis (mild, moderate, gross) to represent the estimated hemolysis (plasma free hemoglobin).
K223608 - Page 9 of 10
{9}
To verify the hemolysis flag performance of the GEM Premier 7000 with iQM₃ in whole blood samples, a hemolysis verification study was performed using a volumetric spiking procedure by adjusting native (venous and arterial) whole blood samples with ultrasonically lysed whole blood to span 6 hemolysis index categories to verify the hemolysis flag performance in venous and arterial samples in normal mode. Hemolysate-contrived venous samples transferred to a capillary device were tested in micro mode to verify the hemolysis flag performance in capillary samples. All results were analyzed using a volumetric target value and overall agreement analysis was performed for the pooled dataset (venous, arterial, and transferred capillary samples) as well as for venous, arterial, and transferred capillary samples separately. All samples were correctly flagged, and 99.2% of venous, 97.6% of arterial, and 99.2% of transferred capillary samples were correctly flagged based on the target value of the spiked hemoglobin.
## VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
## IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K223608 - Page 10 of 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.