K223560 · Scientia Vascular, Inc. · QJP · Aug 21, 2023 · Cardiovascular
Device Facts
Record ID
K223560
Device Name
Plato 17 Microcatheter
Applicant
Scientia Vascular, Inc.
Product Code
QJP · Cardiovascular
Decision Date
Aug 21, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Plato 17 Microcatheter is intended for the introduction of therapeutic devices, infusion of diagnostic agents, such as contrast media, and delivery of embolization materials to the peripheral and neuro vasculature systems. The Plato 17 Microcatheter is not intended for use in the coronary vasculature.
Device Story
Plato 17 Microcatheter is a single-lumen, open-ended, flexible, axially stable catheter for accessing distal neuro and peripheral vasculature. Device features include nitinol/polymer shaft, hydrophilic coating on distal exterior, two radiopaque platinum-iridium tip markers, and clear hub with luer lock. Supplied sterile for single use; includes steam-shaping mandrel, peel-away introducer, and ruler. Operated by physicians in interventional/diagnostic procedures. Input: therapeutic devices, diagnostic agents, or embolization materials. Output: delivery of these materials to target vasculature. Physician visualizes device via fluoroscopy using radiopaque markers. Facilitates minimally invasive access to complex vascular anatomy; enables targeted therapy/embolization.
Clinical Evidence
Bench testing only. Biocompatibility testing per ISO 10993-1 (cytotoxicity, hemolysis, PTT, platelet/leukocyte counts). Functional testing per ISO 80369-7, ISO 80369-20, and ISO 10555-1. Simulated use testing in tortuous neurovascular models confirmed successful delivery of therapeutic devices and embolization materials. All tests passed.
Indicated for patients requiring introduction of therapeutic devices, infusion of diagnostic agents (e.g., contrast media), or delivery of embolization materials into peripheral and neuro vasculature. Not indicated for coronary vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
August 21, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters FDA in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Scientia Vascular, Inc. Max Alfonso Regulatory Affairs Specialist 3487 West 2100 South Suite 100 West Valley City, Utah 84119
Re: K223560
Trade/Device Name: Plato 17 Microcatheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: QJP, DQY Dated: July 20, 2023 Received: July 21, 2023
Dear Max Alfonso:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Naira Muradyan -S
Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K223560
Device Name Plato® 17 Microcatheter
#### Indications for Use (Describe)
The Plato 17 Microcatheter is intended for the introduction of therapeutic devices, infusion of diagnostic agents, such as contrast media, and delivery of embolization materials to the peripheral and neuro vasculature systems. The Plato 17 Microcatheter is not intended for use in the coronary vasculature.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="text-decoration: underline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Scientia Vascular, Inc. Traditional 510(k) Plato® 17 Microcatheter
Image /page/3/Picture/1 description: The image shows the logo for Scientia. The logo consists of a stylized, teal-colored symbol resembling an abstract letter 'S' or a curved, interconnected chain. Below the symbol, the word "SCIENTIA" is written in a modern, sans-serif font with a gray color scheme. The font appears to be slightly stylized, with rounded edges and a futuristic aesthetic.
# 510(K) SUMMARY K223560
## (Per 21 CFR 807.92)
## SCIENTIA VASCULAR, INC. Plato® 17 Microcatheter
Submitter Name and Address: Scientia Vascular, Inc. 3487 West 2100 South Suite 100 West Valley City, UT 84119
| Contact Person: | Max Alfonso |
|-----------------|----------------------------------------|
| | Regulatory Affairs Specialist |
| | Phone: 1 (888) 385-9016 |
| | Email: regulatory@scientiavascular.com |
Date Prepared: 18 August 2023
| Trade Name: | Plato® 17 Microcatheter |
|-------------------------|-------------------------------------------------------|
| Common Name: | Microcatheter |
| Classification Name: | Catheter, Percutaneous |
| Primary Product Code: | QJP |
| Secondary Product Code: | DQY |
| Review Panel: | Neurology |
| Device Class: | Class II device per 21 CFR 870.1250 |
| Predicate Device: | Plato® 17 Microcatheter (K210601, cleared 04/28/2021) |
{4}------------------------------------------------
Scientia Vascular, Inc. Traditional 510(k) Plato® 17 Microcatheter
### DEVICE DESCRIPTION
Scientia Vascular's Plato® 17 Microcatheter is a single lumen, open-ended catheter designed to be flexible and axially stable, to aid the physician in the accessing of the distal neuro and peripheral vasculature. The Plato® 17 Microcatheter is supplied sterile and is for single use only. The Plato® 17 Microcatheter is supplied in various tip configurations, including straight and pre-shaped (45° and 90°). The microcatheter shaft design includes nitinol, polymers of varying durometer, and an internal liner providing lubricity for therapeutic device, diagnostic agent, and embolization material delivery. The distal exterior section of the catheter shaft is hydrophilic coated to reduce friction during manipulation in vessels. The microcatheter includes two radiopaque tip markers to facilitate fluoroscopic visualization and a clear hub with luer lock and a strain relief.
The microcatheter is provided with a steam shaping mandrel, peel-away introducer, and a ruler. The shaping mandrel, peel-away introducer, and ruler are accessories included to facilitate use of the microcatheter and are not intended to contact the patient's body.
The Plato® 17 Microcatheter is substantially equivalent with respect to intended use, technological characteristics, design, and materials to Scientia Vascular's currently marketed Plato® 17 Microcatheter cleared under K210601.
#### INTENDED USE
To assist in delivery of therapeutic devices and embolization materials and infusion of diagnostic agents to the neuro and peripheral vasculature during interventional or diagnostic procedures.
#### INDICATIONS FOR USE
The Plato 17 Microcatheter is intended for the introduction of therapeutic devices, infusion of diagnostic agents, such as contrast media, and delivery of embolization materials to the peripheral and neuro vasculature systems. The Plato 17 Microcatheter is not intended for use in the coronary vasculature.
#### TECHNOLOGICAL CHARACTERISTICS
The Plato® 17 Microcatheter is equivalent or similar to the predicate device in the following ways:
- Functionality ●
- Intended use ●
- Materials ●
- Design ●
- . Biological safety
Shown in the table below is the comparison of technological characteristics for the Plato® 17 Microcatheter to those of the predicate device, the Plato® 17 Microcatheter (K210601).
{5}------------------------------------------------
| Table 1: Comparison between Subject & Predicate Device Technological Characteristics: | |
|---------------------------------------------------------------------------------------|--|
|---------------------------------------------------------------------------------------|--|
| Note: Differences are <b>bolded</b> . | | | |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Characteristic | Subject Device<br>Plato® 17 Microcatheter (K223560) | Predicate Device<br>Plato® 17 Microcatheter (K210601) | Comparison<br>Analysis |
| Anatomical<br>Location | Neuro and peripheral vasculature | Neuro and peripheral vasculature | Same |
| Indications for Use | The Plato 17 Microcatheter is<br>intended for the introduction of<br><b>therapeutic devices</b> , infusion of<br>diagnostic agents, such as contrast<br>media, and <b>delivery of<br/>embolization materials</b> to the<br>peripheral and neuro vasculature<br>systems. The Plato 17 Microcatheter<br>is not intended for use in the<br>coronary vasculature. | The Plato 17 Microcatheter is<br>intended for the introduction of<br>embolic coils and infusion of<br>diagnostic agents, such as contrast<br>media, to the peripheral and neuro<br>vasculature systems. The Plato 17<br>Microcatheter is not intended for<br>use in the coronary vasculature. | Similar |
| Materials | Hub:<br>Clear polymer | Hub:<br>Clear polymer | Same |
| | Strain Relief:<br>Thermoplastic Elastomer | Strain Relief:<br>Thermoplastic Elastomer | Same |
| | Catheter Shaft:<br>Polyamide, Nitinol, Stainless Steel,<br>and PTFE | Catheter Shaft:<br>Polyamide, Nitinol, Stainless Steel,<br>and PTFE | Same |
| | Radiopaque Marker Bands:<br>Platinum Iridium | Radiopaque Marker Bands:<br>Platinum Iridium | Same |
| Proximal ID | 0.017" (0.43mm) | 0.017" (0.43mm) | Same |
| Proximal OD | 2.1F (0.70mm) | 2.1F (0.70mm) | Same |
| Distal ID | 0.017" (0.43mm) | 0.017" (0.43mm) | Same |
| Distal OD | 1.7F (0.60mm) | 1.7F (0.60mm) | Same |
| Effective Length | 160cm | 160cm | Same |
| Tip Design | Straight and pre-shaped tips | Straight and pre-shaped tips | Same |
| Steam-Shapeable<br>tip | Yes | Yes | Same |
{6}------------------------------------------------
| Table 1: Comparison between Subject & Predicate Device Technological Characteristics: | | | | | |
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------|--------------------------------------------------------------------|------------------------|--|--|
| Note: Differences are bolded. | | | | | |
| Characteristic | Subject Device<br>Plato® 17 Microcatheter (K223560) | Predicate Device<br>Plato® 17 Microcatheter (K210601) | Comparison<br>Analysis | | |
| Distal Coating | Hydrophilic coating | Hydrophilic coating | Same | | |
| Coated Length | 90cm | 90cm | Same | | |
| Radiopaque<br>Markers | 2 radiopaque markers located at<br>distal tip of the microcatheter | 2 radiopaque markers located at<br>distal tip of the microcatheter | Same | | |
| Sterilization<br>Method | 100% Ethylene Oxide (EO) | 100% Ethylene Oxide (EO) | Same | | |
| Guidewire<br>Compatibility | 0.014" | 0.014" | Same | | |
| Accessories:<br>Introducer | Included (ID: 0.060") | Included (ID: 0.045") | Different | | |
| Accessories:<br>Tip Shaping<br>Mandrel | 0.015" diameter stainless steel tip<br>shaping mandrel included | 0.015" diameter stainless steel tip<br>shaping mandrel included | Same | | |
| Accessories:<br>Ruler | Included | Included | Same | | |
|Table 1: Comparison between Subject & Predicate Device Technological Characteristics:
The subject device, Plato® 17 Microcatheter, has similar intended use and the same fundamental technological characteristics (design, materials, and device features) when compared to the predicate device. The subject device completed and passed required nonclinical testing to support the expanded indications for use. The difference in technological characteristics and intended use do not raise new or different questions of safety and effectiveness for the subject device.
### NON-CLINICAL PERFORMANCE TESTS
Results of tests performed on the subject Plato® 17 Microcatheter demonstrates it met the testing acceptance criteria, performs as well as the predicate device, and/or meets requirements of relevant standards and FDA guidance documents.
#### Biocompatibility
The biocompatibility evaluation for the subject device, Plato® 17 Microcatheter, was performed in accordance with ISO 10993-1: 2018. The previously conducted testing from the predicate device submission, K210601, was used to support the biocompatibility of the subject device and the following
{7}------------------------------------------------
Scientia Vascular, Inc. Traditional 510(k) Plato® 17 Microcatheter
additional testing was also conducted:
| Table 2: Summary of Subject Device Biocompatibility Testing Performed. | | |
|------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|-----------------------|
| Name of Test | Test Summary | Conclusion of Testing |
| Cytotoxicity: MEM<br>Elution | Cell culture was observed for<br>cytotoxic reactivity. | Non-cytotoxic. |
| Direct Contact and<br>Extract method<br>Hemolysis Test | The difference between the<br>hemolytic indexes of the subject<br>device and the negative control was<br>evaluated. | Non-hemolytic. |
| Partial<br>Thromboplastin<br>Time (PTT) Test | The clotting time was observed for<br>both the subject device and the<br>predicate. | Non-activator. |
| Platelet Leukocyte Counts | The concentration of platelets and<br>leukocytes were determined on an<br>automated hematology analyzer. | Non-activator. |
#### Functional Testing
Functional testing was performed in accordance with the following standards:
- ISO 80369-7:2021 - Small bore connectors for liquids and gases in healthcare applications -Connectors for intravascular or hypodermic applications,
- . ISO 80369-20:2015 - Small bore connectors for liquids and gases in healthcare applications -Common test methods,
- . ISO 10555-1:2013 (corrected 2014) - Intravascular catheters - Sterile and single-use catheters -Part 1: General requirements,
as well as the FDA Guidance Documents:
- Premarket Notifications [510(k)] Submissions for Short-Term and Long-Term Intravascular ● Catheters (March 1995),
- . Non-Clinical Engineering Tests and Recommended Labeling for Intravascular Stents and Associated Delivery Systems (April 2010),
- Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 2019),
- Intravascular Catheters, Wires, and Delivery Systems with Lubricious Coatings Labeling . Considerations (October 2019).
The design of the subject device is unchanged when compared to the predicate device, K210601. Therefore, the functional testing from K210601 for the predicate device is utilized and adopted for the subject device. Table 3 Summarizes the additional functional verification testing performed as part of this submission.
{8}------------------------------------------------
| Test | Attribute and Acceptance Criteria | Results |
|----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Accessory<br>Verification | The Peel-Away Introducer must facilitate introduction of<br>catheter into a guide catheter. | PASS |
| Therapeutic Device<br>Compatibility and<br>Simulated Use | The Plato® 17 Microcatheter can successfully deliver<br>representative therapeutic devices, including stent, in a<br>tortuous neurovascular model.<br>The Plato® 17 Microcatheter was evaluated for delivery to<br>target location and functionality post therapeutic device<br>delivery. | PASS |
| Embolization<br>Material<br>Compatibility | Visual inspection, lumen check, liquid leakage, static burst,<br>tensile and force to withdraw testing was conducted for the<br>Plato® 17 Microcatheter using representative embolization<br>materials in a tortuous neurovascular model. | PASS |
Table 3: Summary of Subject Device Functional Testing.
Additionally, the sterilization process of the subject device is identical to the predicate device K210601. The shelf-life testing, packaging integrity, and sterilization including EO, ECH residuals and bacterial endotoxin levels for the subject device are supported by the testing conducted for the predicate device.
#### CONCLUSION
The Plato® 17 Microcatheter has similar intended use and technological characteristics when compared to the predicate device. The identified differences in the indications for use statement and accessory introducer diameter do not raise new questions of safety or effectiveness. Risk evaluation along with testing, functional and biological, was completed for the subject device. The testing and risk evaluation demonstrate that the subject device is substantially equivalent to the predicate. The subject device and the predicate device have similar intended use, the same fundamental technological characteristics, and similar functional performance, as demonstrated through testing.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.