Spine Planning (2.0), Elements Spine Planning, Elements Planning Spine
K223553 · Brainlab AG · OLO · Aug 2, 2023 · Neurology
Device Facts
Record ID
K223553
Device Name
Spine Planning (2.0), Elements Spine Planning, Elements Planning Spine
Applicant
Brainlab AG
Product Code
OLO · Neurology
Decision Date
Aug 2, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Vertebra labeling
Convolutional Network (CNN)
—
—
—
—
—
—
Spinal parameter measurement
Convolutional Network (CNN)
—
—
—
—
—
—
Pedicle screw placement proposal
Convolutional Network (CNN)
—
Similar to predicate algorithm
—
—
—
—
Indications for Use
Spine Planning is intended for pre- and intraoperative planning of open and minimally invasive spinal procedures. It displays digital patient images (CT, Cone Beam CT, MR, X-ray) and allows measurement and planning of spinal implants like screws and rods.
Device Story
Spine Planning 2.0 is a software tool for pre- and intraoperative planning of spinal surgeries. It accepts 2D X-ray images and 3D datasets (CT, MR, Cone Beam CT) as input. The software uses CNN-based AI/ML algorithms to detect anatomical landmarks on 2D images for vertebrae labeling and to detect vertebrae on Digitally Reconstructed Radiographs (DRR) for atlas registration. It provides automated proposals for spinal parameters, measurements, and screw/rod placement. Used in office or operating room settings by clinicians, the software allows manual review, adjustment, and approval of all automated proposals. The output is displayed to the surgeon to assist in surgical planning and implant positioning. By providing automated anatomical labeling and implant trajectory suggestions, the device aims to improve planning accuracy and efficiency for spinal procedures.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including software verification (unit/integration tests), AI/ML landmark detection assessment (quantifying object detection and vertebra level assignment), screw proposal algorithm evaluation (comparing new algorithm to predicate), and summative usability testing. Usability testing confirmed the device can be used by the intended user group without critical use-related problems.
Technological Characteristics
Software-based planning tool. Utilizes CNN-based AI/ML algorithms for landmark detection and vertebra labeling. Supports CT, MR, Cone Beam CT, and X-ray modalities. Features include automated vertebra labeling, spinal parameter measurement, and screw/rod placement proposals. Operates on standard clinical computing hardware. Software verification covers SOUP and cybersecurity.
Indications for Use
Indicated for pre- and intraoperative planning of open and minimally invasive spinal procedures in patients requiring spinal implant placement (screws/rods) using CT, Cone Beam CT, MR, or X-ray imaging.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
August 2, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Brainlab AG Sadwini Suresh OM Consultant Olof-Palme-Str.9 Munich, BY 81829 Germany
Re: K223553
Trade/Device Name: Spine Planning 2.0 Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO, LLZ Dated: June 30, 2023 Received: June 30, 2023
Dear Sadwini Suresh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali-S
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223553
Device Name Spine Planning 2.0
Indications for Use (Describe)
Spine Planning is intended for pre- and intraoperative planning of open and minimally invasive spinal procedures. It displays digital patient images (CT, Cone Beam CT, MR, X-ray) and allows measurement and planning of spinal implants like screws and rods.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "BRAINLAB" in pink, with a stylized pink symbol to the left of the word. The symbol appears to be three curved lines stacked on top of each other. The word "BRAINLAB" is written in all capital letters and is in a sans-serif font. The overall image has a clean and modern look.
# K223553 510(k) Summary
June 30, 2023
| General Information | |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Brainlab AG; Olof-Palme Str.9; 81829, Munich, Germany |
| Establishment Registration | 8043933 |
| Trade Names | Spine Planning 2.0<br>• Spine Planning<br>• Elements Spine Planning<br>• Elements Planning Spine |
| Classification Name | Orthopedic Stereotaxic Instrument |
| Product Code | OLO |
| Regulation Number | 882.4560 |
| Regulatory Class | II |
| Panel | Orthopedic |
| Predicate Device | K212245<br>Spine and Trauma Navigation System, Spine & Trauma 3D<br>Navigation, Instrument Selection, Fluoro 3D, Registration Software<br>Fluoro 3D, Registration Software Paired Point, Registration Software<br>Spine Surface Matching, Spine Planning, Elements Screw Planning<br>Spine, Elements Spine Screw Planning |
| Contact Information | |
| Primary Contact | Sadwini Suresh<br>QM Consultant<br>Regulatory Affairs<br>Phone: +49 89 99 15 68 0<br>Email: regulatory.affairs@brainlab.com |
| Alternate Contact | Chiara Cunico<br>Senior Manager Regulatory Affairs<br>Phone: +49 89 99 15 68 0<br>Email: chiara.cunico@brainlab.com |
#### INDICATIONS FOR USE 1
Spine Planning is intended for pre- and intraoperative planning of open and minimally invasive spinal procedures. It displays digital patient images (CT, Cone Beam CT, MR, X-ray) and allows measurement and planning of spinal implants like screws and rods.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Brainlab logo. The logo consists of a stylized pink symbol on the left, resembling a brain or a network, followed by the word "BRAINLAB" in pink, block letters. The overall design is clean and modern, suggesting a focus on technology and innovation in the field of neuroscience or medical technology.
#### DEVICE DESCRIPTION 2
The Spine Planning software allows the user to plan spinal surgery pre-operatively or intra-operatively. The software is able to display 2D X-Ray images and 3D datasets (e.g. CT or MR scans). The software consists of features for automated labelling of vertebrae and proposals for screw and rod implants, proposals for measurement of spinal parameters.
The device can be used in combination with spinal navigation software during surgery, where preplanned or intra-operatively created information can be displayed, or solely as a pre-operative tool to prepare the surgery.
AI/ML algorithms are used in Spine Planning for
- . Detection of landmarks on 2D images for vertebrae labeling and measurement and
- . Vertebra detection on Digitally Reconstructed Radiograph (DRR) images of 3D datasets for atlas reqistration (labeling of the vertebra).
The AI/ML algorithm is a Convolutional Network (CNN) developed using a Supervised Learning approach. The algorithm was developed using a controlled internal process that defines from the inspection of input data to the training and verification of the algorithm.
#### SUBSTANTIAL EQUIVALENCE ന
The Subject Device has similar intended use and technological features as the predicate devices. An overview of the similarities and differences can be found in the tables below:
| Device | Name |
|----------------------------|--------------------|
| Subject Device | Spine Planning 2.0 |
| Predicate device (K212245) | Spine Planning 1.0 |
| Topic/ Feature | Predicate Device (Spine<br>Planning 1.0 (K212245)) | Subject Device (Spine<br>Planning 2.0) | Comment |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | Spine Planning is intended for pre- and intraoperative planning of open and minimal invasive spinal procedures. It displays digital bio imaging and allows measurement and planning of spinal implants like screws and rods. | Spine Planning is intended for pre- and intraoperative planning of open and minimally invasive spinal procedures. It displays digital patient images (CT, Cone Beam CT, MR, X-ray) and allows measurement and | No change in the indications. Image modalities that are supported have been included. |
| | | planning of spinal implants<br>like screws and rods. | |
| Intended use<br>environment | The planning part of the use case<br>shall be done in an office or in the<br>operating room. | The device can be used in an<br>office / team room or in the<br>operating room. | Identical |
| Supported Imaging<br>Modalities | - Computed tomography (CT) | - CT<br>- Magnetic resonance (MR)<br>- Cone Beam CT (XT)<br>- X-ray | Additional modalities<br>effect measurement<br>or screw planning<br>features |
| Vertebra Labeling | Automatic detection of thoracic<br>and lumbar vertebrae. User<br>approval of calculated automatic<br>labeling. No functionality to<br>change the calculated labeling. | Automatic detection of<br>cervical, thoracic, lumbar,<br>and sacral vertebrae.<br>Ability to adjust automatic<br>labeling for correction if<br>needed. Automatic labeling<br>can be discarded and<br>performed manually. User<br>approval of labeling. | New features:<br>- Addition of cervical<br>and sacral spinal<br>regions in subject<br>device.<br>- Ability to adjust<br>and correct labeling.<br>- Possible to do<br>performed manual<br>labeling.<br>Functionality tested<br>during verification<br>testing. |
| Anatomical<br>Variations | N/A | Adjustment of anatomical<br>chain to account for<br>anatomic variations.<br>Depending on patient<br>anatomy, it is possible to<br>exclude T12 and/or L5 and<br>include T13 and/or L6.<br>To be added manually by<br>user. | New feature. |
| Spinal Parameter<br>Proposals | N/A | Able to add spinal<br>parameters on X-rays based<br>on AI/ML generated<br>landmarks. To be manually<br>reviewed and approved. | New feature. |
| | | | |
| Manual<br>Measurements | N/A | Where AI/ML landmarks are<br>unavailable all spinal<br>parameters can be added<br>manually, reviewed, and<br>approved. | New feature. |
| Automatic Screw<br>Proposals | Automatic screw proposals<br>provided for thoracic and lumbar<br>pedicle screws. | Manual placement and<br>adjustment of angles and<br>open angles on X-rays. | Similar function to<br>predicate. New<br>screw proposal<br>algorithm with<br>additional<br>functionality used in<br>subject device. |
| Screw Planning | Automatic screw proposals<br>provided for cervical (lateral<br>mass or pedicle), thoracic<br>(pedicle), lumbar (pedicle),<br>sacral (pedicle and alar iliac),<br>and iliac regions. | Adjustment of screw length,<br>diameter, and position. | Adjustment of<br>screws, creation of<br>manual screws and<br>screw approval is<br>similar to predicate. |
| | Adjustment of screw length,<br>diameter, and position. | Creation of manual screws. | New features.<br>Possible to change<br>type of screw<br>between lateral<br>mass and pedicle<br>screw types<br>(cervical region) or<br>pedicle and alar iliac<br>screw types (Sacral<br>region) |
| | Creation of manual screws. | Addition of automatic screw<br>proposals. | |
| | Addition of automatic screw<br>proposals. | Change of screw type<br>(cervical and sacral regions<br>only) | |
| | User approval of all screws<br>required. | User approval of all screws<br>required. | |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "BRAINLAB" in all capital letters. To the left of the word is a pink abstract symbol. The word and symbol are in a bright pink color.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Brainlab. The logo consists of a stylized pink caduceus symbol on the left, followed by the word "BRAINLAB" in pink, block letters. The caduceus symbol is a medical emblem featuring a staff with two snakes coiled around it.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the Brainlab logo. The logo consists of a stylized pink symbol on the left, resembling a medical symbol or abstract figure. To the right of the symbol, the word "BRAINLAB" is written in capital letters, also in pink. The overall design is clean and modern.
#### 4 PERFORMANCE DATA
## Software Verification:
Software verification was performed, verifying the software requirements through integration tests, and unit tests. Incremental test strateqies have been set up after verification of the first release candidate for changes with limited scope. In this case, an impact analysis of the modifications is performed and tests to be performed are identified and planned correspondingly. That means, not all tests have to be performed but only a subset, as some of the previous tests are not affected by the change and remain therefore valid.
Software verification verifies all specifications, including SOUP items and cybersecurity.
For newly added components, integration tests were carried out, in addition to the individual component verification.
## Al/ML Detected X-Ray Landmarks Assessment:
This test was conducted to assess the AI/ML detected landmarks on X-rays which are used by the software for automatic measurement proposals. This was done by quantifying the object detection, quality of vertebra level assignment, the quality of landmark predictions, and the performance of the observer view direction for 2D X-rays from the Universal Atlas Transfer Performer 6.0.
### Screw Proposal Alqorithm Evaluation:
The purpose of this testing was to compare the newly proposed screw proposal algorithm to the predicate and back-up algorithms. Thoracic and lumbar pedicle screw proposals generated by the new algorithm were found to be similar to thoracic and lumbar pedicle screw proposals generated by the predicate algorithm.
### Usability Evaluation:
Summative usability testing was planned and performed in order to validate that the Spine Planning 2.0 can be used by the intended user group. The summative usability testing identified no critical use related problems.
#### 5 CONCLUSION
The performed verification and validation activities established that the set requirements were met and that the device performs as intended.
The Subject Devices' comparison with the Predicate Device establishes that they have similar functionality, intended use and technological characteristics. Therefore, we consider that the Subject Device can be considered substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.