K223541 · Jkh Health Co., Ltd. · IRP · Oct 27, 2023 · Physical Medicine
Device Facts
Record ID
K223541
Device Name
Cold/Hot Compression
Applicant
Jkh Health Co., Ltd.
Product Code
IRP · Physical Medicine
Decision Date
Oct 27, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Cold/Hot Compression combines cold, heat, contrast, and compression therapy. It is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated. It is intended to treat post traumatic and post-surgical medical and/or surgical conditions for which localized thermal therapy (hot or cold) are indicated. Cold/Hot Compression also provide DVT therapy. It is intended to reduce the risk of the formation of deep venous thrombosis (DVT) by aiding blood flow back to the heart via lower extremity limb compression. Cold/Hot Compression is intended to be used by, or on the order of, licensed health care professionals in rehabilitation facilities, outpatient clinics, athletic training settings, and home settings.
Device Story
Prescription device combining cold/hot thermal therapy and air compression; treats post-surgical/acute injuries and provides DVT prophylaxis. System comprises main control unit and optional DVT/thermal wraps; wraps applied over clothing to prevent skin irritation. Operates via thermoelectric cooling/heating and pneumatic compression; user-adjustable settings for temperature, pressure, and cycle duration via touch screen. Used in rehabilitation, clinics, athletic settings, or home by professionals or patients under prescription. Provides localized thermal therapy and limb compression to aid blood flow and reduce edema/pain. Battery-powered with 3-year shelf life; stores usage data.
Clinical Evidence
No clinical data; bench testing only. Verification included electrical safety (ANSI AAMI ES60601-1), electromagnetic compatibility (IEC 60601-1-2), and performance testing (UI, cooling/heating, pressure, alarms, cycle times, leakage/burst).
Indicated for patients requiring cold, heat, contrast, or compression therapy for post-surgical/acute injuries (edema, swelling, pain) or post-traumatic/post-surgical conditions requiring localized thermal therapy. Also indicated for DVT prophylaxis via lower extremity compression. Intended for use by or on order of licensed healthcare professionals in clinical or home settings.
Regulatory Classification
Identification
A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 27, 2023
JKH Health Co., Ltd. Bill Quanqin Dai, Ph.D. Official Correspondent 14271 Jeffrev Rd. #246 Irvine, California 92620
Re: K223541
Trade/Device Name: Cold/Hot Compression Regulation Number: 21 CFR 890.5650 Regulation Name: Powered Inflatable Tube Massager Regulatory Class: Class II Product Code: IRP, ILO, JOW Dated: September 26, 2023 Received: September 28, 2023
Dear Dr. Dai:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Heather L. Dean -S
> Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Rehabilitation Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223541
Device Name Cold/Hot Compression
#### Indications for Use (Describe)
Cold/Hot Compression combines cold, heat, contrast, and compression therapy. It is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated. It is intended to treat post traumatic and post-surgical medical and/or surgical conditions for which localized thermal therapy (hot or cold) are indicated.
Cold/Hot Compression also provide DVT therapy. It is intended to reduce the risk of the formation of deep venous thrombosis (DVT) by aiding blood flow back to the heart via lower extremity limb compression.
ColdHot Compression is intended to be used by, or on the order of, licensed health care professionals in rehabilitation facilities, outpatient clinics, athletic training settings, and home settings.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(k) Summary
This 510(k) Summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
### 1. Submitter's Information
Submitter: JKH Health Co., Ltd. Address: 4-5F, Building 12, Hengmingzhu Industrial Park, Xinqiao Tongfuyu Industrial Area, Shajing, Baoan, Shenzhen, China Contact Person: Pu Jiang Tel: +86-755-27926589 Fax: +86-755-29970323 Email: sales@JKHhealth.com Date of Preparation: October 24, 2022
## 2. Subject Device
Trade/Device Name: Cold/Hot Compression Regulation Description: Name: Powered Inflatable Tube Massager; Hot Or Cold, Water Circulating Pack; Compressible Limb Sleeve Regulation Medical Specialty: Physical Medicine, Cardiovascular Review Panel: Physical Medicine, Cardiovascular Product Code: IRP, ILO, JOW Regulation Number: 21 CFR 890.5650, 21 CFR 890.5720, 21 CFR 870.5800 Device Class: II Use: Prescription
## 3. Predicate device
Predicate Device: Therm-X 510(k) Number: K193550 Clearance Date: February 28, 2020 Submitter: Zenith Technical Innovations, LLC
Reference Device: NanoTherm and VascuTherm Systems 510(k) Number: K061866 Clearance Date: August 22, 2006 Submitter: ThermoTek, Inc.
## 4. Description of Subject Device
The subject device is a prescriptive device, which combines cold/hot therapy and air compression. It is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated. It is intended to treat post traumatic and post-surgical medical and/or surgical conditions for which localized thermal therapy (hot or cold) are indicated. It is intended to reduce the risk of the formation of deep venous thrombosis (DVT) by aiding blood flow back to the heart via lower extremity limb compression.
The subject device comes with the thermal-therapy wraps and the DVT wraps. The thermal-therapy wrap is nylon fabric on one side and velvet cloth on the other side; the DVT wrap is velvet on both sides. To avoid any potential adverse skin reactions such as redness, irritation, and cold/hot injury, the sock/clothing should be worn by the patient prior to use.
{5}------------------------------------------------
The subject device includes a main device and an optional DVT device, and has a limited shelf life of 3 years, based on the charge retention characteristics of the device's battery. Please find the detailed storage condition and maintenance in the user manual.
## 5. Indications for Use
Cold/Hot Compression combines cold, heat, contrast, and compression therapy. It is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated. It is intended to treat post traumatic and post-surgical medical and/or surgical conditions for which localized thermal therapy (hot or cold) are indicated.
Cold/Hot Compression also provide DVT therapy. It is intended to reduce the risk of the formation of deep venous thrombosis (DVT) by aiding blood flow back to the heart via lower extremity limb compression.
Cold/Hot Compression is intended to be used by, or on the order of, licensed health care professionals in rehabilitation facilities, outpatient clinics, athletic training settings, and home settings.
#### 6. Summary of Substantial Equivalence
The comparison Table 1 summarizes the detailed comparison between the subject device and the predicate device, indicating the technical characteristics, specifications, and intended use of the subject device are substantially equivalent to those of the predicate device.
| | Subject Device | Predicate Device | Reference Device | Equivalence |
|--------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K223541 | K193550 | K061866 | N/A |
| Submitter | JKH Health Co., Ltd. | Zenith Technical Innovations,<br>LLC | ThermoTek, Inc. | N/A |
| Device Name/Model | Cold/Hot Compression | Therm-X | VascuTherm | N/A |
| Indications for Use | Cold/Hot Compression<br>combines cold, heat, contrast,<br>and compression therapy. It is<br>intended to treat post-surgical<br>and acute injuries to reduce<br>edema, swelling, and pain for<br>which cold and compression<br>are indicated. It is intended to<br>treat post traumatic and post-<br>surgical medical and/or<br>surgical conditions for which<br>localized thermal therapy (hot<br>or cold) are indicated.<br>Cold/Hot Compression also<br>provide DVT therapy. It is<br>intended to reduce the risk of<br>the formation of deep venous<br>thrombosis (DVT) by aiding<br>blood flow back to the heart via<br>lower extremity limb<br>compression.<br>Cold/Hot Compression is<br>intended to be used by, or on | Therm-X (Therm-X Home and<br>Therm-X AT) combines cold,<br>heat, contrast, and compression<br>therapy. Therm-X is intended<br>to treat post-surgical and acute<br>injuries to reduce edema,<br>swelling, and pain for which<br>cold and compression are<br>indicated. It is intended to treat<br>post traumatic and post-<br>surgical medical and/or<br>surgical conditions for which<br>localized thermal therapy (hot<br>or cold) are indicated.<br>Therm-X Home systems also<br>provide DVT therapy. Therm-<br>X Home systems with DVT<br>therapy are intended to reduce<br>the risk of the formation of<br>deep venous thrombosis (DVT)<br>by aiding blood flow back to<br>the heart via lower extremity<br>limb compression. | Treatment of disorders<br>associated with vascular or<br>lymphatic insufficiency such as<br>Chronic Venous Insufficiency<br>(CVI), venous stasis ulcers,<br>post-mastectomy edema and<br>chronic lymphedema.<br>Reduction of edema associated<br>with soft tissue injuries such as<br>bums, postoperative edema,<br>and ligament sprains.<br>Localized thermal therapy (hot<br>or cold) for post traumatic and<br>post surgical medical and/or<br>surgical conditions.<br>Decrease the risk of deep<br>venous thrombosis (DVT).<br>Aids the blood flow back to the<br>heart. | Identical |
| | the order of, licensed health<br>care professionals in<br>rehabilitation facilities,<br>outpatient clinics, athletic<br>training settings, and home<br>settings. | Therm-X (Therm-X Home and<br>Therm-X AT) is intended to be<br>used by, or on the order of,<br>licensed health care<br>professionals in rehabilitation<br>facilities, outpatient clinics,<br>athletic training settings, and<br>home settings. | Treat and assist healing of<br>cutaneous ulceration (wounds),<br>reduce would healing time,<br>enhance arterial circulation<br>(blood flow), reduce<br>compartmental pressures,<br>reduce edema (swelling),<br>reduce the need for<br>anticoagulant (blood thinning)<br>medications. | |
| Intended Users | Health Care Professionals and<br>lay users (under prescription) | Health Care Professionals and<br>lay users (under prescription) | Health Care Professionals and<br>lay users (under prescription) | Identical |
| Prescription or OTC | Prescription | Prescription | Prescription | Identical |
| Power Source(s) | 100-240 VAC, 50/60 Hz | 100-240 VAC, 50/60 Hz | 100-240 VAC, 50/60 Hz | Identical |
| Compliance with<br>Voluntary Standards? | Yes | Yes | Yes | Identical |
| Electrical Safety<br>Mechanical Safety<br>Chemical Safety<br>Thermal Safety<br>Radiation Safety? | Yes | Yes | Yes | Identical |
| Cold Therapy | Default: 50°F<br>Custom: 41°F-55°F<br>Default, continuous: 50°F<br>Custom, continuous: 41°F - 50°F | Default: 34°F, 45°F, 55°F<br>Custom: 34°F - 55°F<br>Default, continuous: 40°F, 45°F,<br>50°F<br>Custom, continuous: 40°F - 50°F | Default: 49°F<br>Custom: 43°F-49°F<br>Default, continuous: 49°F<br>Custom, continuous: 43°F - 49°F | Equivalent Note 1 |
| Heat Therapy | Default: 105°F<br>Custom: 105°F-109°F<br>Default, continuous: 105°F<br>Custom, continuous: 105°F -<br>109°F | Default: 105°F, 107°F, 110°F<br>Custom: 105°F -110°F<br>Default, continuous: 105°F, 107°F<br>Custom, continuous: 105°F -<br>107°F | Default: 105°F<br>Custom: 100°F-105°F<br>Default, continuous: 105°F<br>Custom, continuous: 100°F -<br>105°F | Equivalent Note 2 |
| Edema Pressure<br>Levels | Compression Therapy Alternate<br>mode: 15 -75 mm Hg<br>Compression Therapy Continuous<br>mode: 15 -75 mm Hg | Available in four levels:<br>Lite (5 mm Hg)<br>Low (20 mm Hg)<br>Medium (45 mm Hg)<br>High (70 mm Hg)<br>For continuous treatment,<br>available in three levels:<br>Low (20 mm Hg)<br>Medium (45 mm Hg)<br>High (70 mm Hg) | Compression Therapy 3 Settings:<br>Low (15 mm Hg)<br>Medium (30 mm Hg)<br>High (50 mm Hg) | Equivalent Note 3 |
| Static or Intermittent<br>Pressure | Both | Both | Intermittent Pressure available | Identical |
| DVT Only | Available | Available for Therm-X Home<br>Model | Available | Identical |
| DVT Pressure | Calf: 55 mmHg<br>Foot: 130 mmHg | Calf: 50-70 mmHg<br>Foot: 90 - 130 mmHg | Calf: 45 mmHg<br>Foot: 100 mmHg | Equivalent Note 4 |
| Cycle Length (for<br>Heat, Cold, And<br>Compression) | Default: 30 minutes<br>Custom: 1 - 60 minutes | Default: 10 or 20 minutes<br>Custom: 3 - 40 minutes<br>Continuous: 10-40 minutes<br>active, 30-60 minutes rest | Default: 90 minutes | Equivalent Note 5 |
| Contrast Therapy | Heat: 105°F<br>Cold: 49°F | Heat: 105°F<br>Cold: 38°F | Heat: 105°F<br>Cold: 49°F | Equivalent Note 6 |
| Cycle Length (for<br>Contrast Therapy) | Heat: 10 minutes<br>Cold: 20 minutes | Heat: 3-10 minutes<br>Cold: 3-10 minutes<br>Total treatment: 6-60 minutes | Heat: 10 minutes<br>Cold: 20 minutes | Equivalent Note 7 |
| DVT Cycle Length | No specified time interval. DVT<br>can be stopped at any time by the<br>user. | No specified time interval. DVT<br>can be stopped at any time by the<br>user. | No specified time interval. DVT<br>can be stopped at any time by the<br>user. | Identical |
| Edema Compression<br>and DVT Compression<br>at the same time | Available<br>Edema Compression must be<br>combined with cold, heat, or<br>contrast therapy | Available<br>Edema Compression (Lite, Low,<br>Medium, High) must be | Available<br>Edema Compression (Lite, Low,<br>Medium, High) must be | Identical |
| | | combined with cold, heat, or<br>contrast therapy | combined with cold, heat, or<br>contrast therapy | |
| Inflation and Deflation | Inflation: Up to 120 seconds<br>Deflation: Up to 30 seconds | Inflation: Up to 120 seconds<br>Deflation: Up to 30 seconds | Inflation: Up to 120 seconds<br>Deflation: Up to 30 seconds | Identical |
| Store Cycle Usage<br>Data | Available | Available | Available | Identical |
| Chilling Mechanism | Thermoelectric | Thermoelectric | Thermoelectric | Identical |
| Heating Mechanism | Thermoelectric | Thermoelectric | Thermoelectric | Identical |
| Reservoir Fluid<br>Capacity | 350 mL | 650 mL | 300mL | Equivalent Note 8 |
| Recommended<br>Coolant | 90% Distilled Water,<br>10% Isopropyl Alcohol | 90% Distilled Water,<br>10% Isopropyl Alcohol | 90% Distilled Water,<br>10% Isopropyl Alcohol | Identical |
| User Interface | Touch Screen…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.