K223523 · Ge Medical Systems,Llc (Ge Healthcare) · LNH · May 30, 2023 · Radiology
Device Facts
Record ID
K223523
Device Name
Sonic DL
Applicant
Ge Medical Systems,Llc (Ge Healthcare)
Product Code
LNH · Radiology
Decision Date
May 30, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
AI/ML, Pediatric
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Cardiac functional measurements
Deep learning based image reconstruction algorithm
—
Inter-method variability was smaller than inter-observer intra-method variability for conventional ASSET Cine images.
—
—
Reader evaluation study: 107 image series from 57 subjects (46 patients, 11 healthy volunteers) across 7 sites.
3 (radiologists)
Cardiac image quality
Deep learning based image reconstruction algorithm
—
Rated as being of diagnostic quality with significant reduction in scan time.
—
—
Blinded image quality assessment study: 127 image sets including short axis, long axis, and aortic valve views.
3 (radiologists)
Indications for Use
Sonic DL is a Deep Learning based image reconstruction technique that is available for use on GE Healthcare 1.5T and 3.0T MR systems. Sonic DL reconstructs MR images from highly under-sampled data, and thereby enables highly accelerated acquisitions. Sonic DL is intended for cardiac imaging, and for patients of all ages.
Device Story
Sonic DL is a deep learning-based software reconstruction feature for GE Healthcare 1.5T and 3.0T MR systems. It processes highly under-sampled raw MR data from cardiac cine exams to reconstruct diagnostic-quality images. By enabling acceleration factors up to 12x, it allows for single-heartbeat (1 R-R) acquisition, facilitating both breath-hold and free-breathing imaging. The feature is integrated into the MR system software and activated via a license key. Radiologists use the reconstructed images to perform quantitative cardiac measurements, such as left ventricular volumes, ejection fraction, and cardiac output. The primary clinical benefit is a significant reduction in scan time compared to conventional parallel imaging (e.g., ASSET), improving patient comfort and throughput while maintaining diagnostic image quality.
Clinical Evidence
Two reader studies evaluated 107 image series from 57 subjects (46 patients, 11 healthy volunteers, age range 12 weeks to 82 years). Study 1 (n=57) compared quantitative measurements (LV volumes, ejection fraction, cardiac output) between Sonic DL and conventional ASSET Cine; inter-method variability was smaller than inter-observer intra-method variability. Study 2 (n=127 image sets) performed blinded image quality assessments, confirming diagnostic quality. Additional testing on 19 healthy volunteers and 10 subjects confirmed 12x acceleration and 1 R-R free-breathing capability, with functional measurements showing close agreement to breath-hold ASSET Cine. Bench testing included MRXCAT and digital phantom simulations using PSNR, RMSE, SSIM, and MAE metrics.
Technological Characteristics
Software-based deep learning image reconstruction feature for 1.5T and 3.0T MR systems. Operates on under-sampled k-space data to reconstruct cardiac cine images. Integrated into existing MR system software; activated via software option key. Utilizes deep learning architecture for image reconstruction to enable high acceleration factors (up to 12x).
Indications for Use
Indicated for cardiac imaging in patients of all ages, including those with structural heart disease, ischemic cardiomyopathy, and non-ischemic cardiomyopathy.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 30, 2023
GE Medical Systems,LLC (GE Healthcare) % Glen Sabin Regulatory Affairs Director 3200 N Grandview Blvd. WAUKESHA WI 53188
Re: K223523
Trade/Device Name: Sonic DL Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH Dated: April 28, 2023 Received: April 28, 2023
Dear Glen Sabin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
D.R.X.
Daniel M. Krainak, Ph.D. Assistant Director Magnetic Resonance and Nuclear Medicine Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K223523
Device Name Sonic DL
#### Indications for Use (Describe)
Sonic DL is a Deep Learning based image reconstruction technique that is available for use on GE Healthcare 1.5T and 3.0T MR systems. Sonic DL reconstructs MR images from highly under-sampled data, and thereby enables highly accelerated acquisitions. Sonic DL is intended for cardiac imaging, and for patients of all ages.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the GE HealthCare logo. On the left is the GE monogram in purple. To the right of the monogram is the text "GE HealthCare" in purple.
# 510(k) Summary
K223523
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | 24 May 2023 |
|-----------------------|-------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Medical Systems, LLC<br>3200 N. Grandview Blvd.<br>Waukesha, WI 53188 |
| Primary Contact: | Glen Sabin<br>Regulatory Affairs Director<br>Phone: 262 894-4968<br>Email: Glen.Sabin@GE.com |
| Secondary Contact: | Sandra Westphal<br>Senior Regulatory Affairs Leader<br>Phone: 262 720-8872<br>Email: Sandra.Westphal@GE.com |
| Device Trade Name: | Sonic DL |
| Common / Usual Name: | MR System |
| Classification Name: | Magnetic Resonance Diagnostic Device |
| Regulation Number: | 21 CFR 892.1000 |
| Primary Product Code: | LNH |
| Predicate Device: | |
| 510(k) Number: | K213668 |
| Device Name: | SIGNA Hero |
| Manufacturer: | GE Medical Systems, LLC |
### Device Description:
Sonic DL is a new software feature intended for use with GE Healthcare MR systems. It consists of a deep learning based reconstruction algorithm that is applied to data from MR cardiac cine exams obtained using a highly accelerated acquisition technique.
Sonic DL is an optional feature that is integrated into the MR system software and activated through a purchasable software option key.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the GE HealthCare logo. The logo consists of a circular emblem on the left and the text "GE HealthCare" on the right. The emblem is purple and features a stylized design. The text is also purple and is written in a sans-serif font.
#### Indications for Use:
The Indications for Use statement for the proposed device is provided below:
Sonic DL is a Deep Learning based image reconstruction technique that is available for use on GE Healthcare 1.5T and 3.0T MR systems. Sonic DL reconstructs MR images from highly under-sampled data, and thereby enables highly accelerated acquisitions. Sonic DL is intended for cardiac imaging, and for patients of all ages.
### Comparison of Technological Characteristics:
The proposed Sonic DL software feature that is the subject of this 510(k) is similar to the 2D FIESTA Cine feature with ASSET acceleration that is included in the predicate SIGNA Hero. K213668. The predicate device uses breath holds and a conventional parallel imaging acceleration technique (ASSET) that typically achieves acceleration factors between 2 and 4, and acquires each slice in 6 to 10 heartbeats. Sonic DL can achieve acceleration factors as high as 12, and enables single heartbeat (1 R-R) imaging with either breath holds or free breathing.
#### Summary of Nonclinical Testing:
The performance of the Sonic DL reconstruction algorithm has been evaluated using simulated data generated from an MRXCAT phantom, and a digital phantom. Model accuracy metrics such as Peak-Signal-to-Noise (PSNR), Root-Mean-Square Error (RMSE), Structural Similarity Index Measure (SSIM), and Mean Absolute Error (MAE) were used to compare simulated Sonic DL images with different levels of acceleration and numbers of phases to the fully sampled images. Additionally, line profiles across the heart at different phases were used to evaluate both inplane and temporal sharpness.
The nonclinical testing demonstrated that Sonic DL is capable of reconstructing Cine images from highly under sampled data that are similar to the fully sampled Cine images in terms of image quality and temporal sharpness.
### Summary of Clinical Testing:
Two reader evaluation studies were performed using Sonic DL images acquired from clinical sites and from healthy volunteers at GE Healthcare facilities. In one study, three radiologists were asked to determine various quantitative measurements such as left ventricular volumes, ejection fraction, and cardiac output using both conventional Cine and Sonic DL image sets as summarized below:
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for GE HealthCare. The logo consists of a circular emblem on the left and the text "GE HealthCare" on the right. The emblem features a stylized, swirling design within a circle, and the text is in a clean, sans-serif font, all in a matching purple color.
| Number of image series evaluated: | 107 |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Number of unique subjects: | 57 |
| Patients: | 46 |
| Healthy volunteers: | 11 |
| Sites contributing data: | 7 |
| GE Healthcare facilities: | 2 |
| External collaborators (clinical sites): | 5 |
| Gender of subjects: | |
| Male: | 35 |
| Female: | 22 |
| Age range of subjects: | 12 weeks to 82 years old |
| Pathology: | Subjects from clinical sites included examples of various cardiac pathology, including structural heart disease, ischemic cardiomyopathy, and non-ischemic cardiomyopathy. |
| Equipment Used: | GE HealthCare 1.5T and 3.0T MR Systems |
| Acquisition Parameters: | |
| Field of View: | 28 - 46 cm |
| In-plane resolution: | 1.3 – 2.2 mm<br>(both x and y directions) |
| Acquired Temporal Resolution (frame rate) | 23.7 – 70.2 ms |
The results showed that the inter-method variability (coefficient of variability comparing functional measurements taken with Sonic DL images versus measurements using the conventional ASSET Cine images) was smaller than the inter-observer intra-method variability for the conventional ASSET Cine images for all parameters, indicating that Sonic DL is suitable for performing functional cardiac measurements.
In a second study, three radiologists were asked to perform blinded image quality assessments of both conventional 2D FIESTA Cine images with ASSET acceleration, and Sonic DL images obtained with different number of R-R intervals and different acceleration factors. This study involved 127 image sets which included a subset of the subjects described above, and included different cardiac views such as short axis, long axis, and aortic valve images. The results of this study showed that on average the Sonic DL images were rated as being of diagnostic quality while providing a significant reduction in scan time compared to the conventional ASSET Cine images.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the GE HealthCare logo. The logo consists of a circular emblem on the left and the text "GE HealthCare" on the right. The emblem is a stylized version of the letters "GE" inside a circle, and the text is in a sans-serif font.
In addition to the two reader studies described above, other tests were performed. Similar to the nonclinical testing using the digital phantom, model accuracy and temporal sharpness evaluations were conducted using in vivo cardiac cine images obtained from 19 health volunteers.
Finally, clinical testing has been performed to demonstrate that Sonic DL is capable of achieving a 12 times acceleration factor and obtaining free-breathing images in a single heartbeat (1 R-R). Functional measurements using Sonic DL 1 R-R free breathing images from 10 subjects were compared to functional measurements using the conventional ASSET Cine breath hold images, and showed close agreement. Additionally, the image quality of 13 Sonic DL 1 R-R free breathing cases was evaluated by a U.S. board certified radiologist, and scored higher than the corresponding conventional free breathing Cine images from the same subjects.
The results of the clinical testing confirmed that the Sonic DL feature provides images that are of diagnostic quality, and that allow functional measurements that are within typical inter reader variability to the same measurements taken with conventional images. Additionally, the Sonic DL feature provided significantly shorter scan times that the conventional Cine imaging.
#### Conclusions Drawn from Performance Testing:
The nonclinical and clinical testing demonstrated that Sonic DL satisfies the product claims of reducing scan times while preserving image quality, enabling single heartbeat functional imaging, and enabling rapid free-breathing functional imaging.
The proposed Sonic DL software feature has been developed under GE Healthcare's quality system and is at least as safe and effective as the 2D FIESTA Cine feature with ASSET acceleration that is used in the legally marketed predicate device. The performance testing did not identify any new hazards, adverse effects, safety concerns, or performance concerns that are significantly different from those associated with cardiac cine MR imaging in general.
Therefore, GE Healthcare believes that Sonic DL is substantially equivalent to the predicate device and is safe and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.