← Product Code [OLO](/productcode/OLO) · K223511

# Excella Navigation Instruments (K223511)

_Innovasis, Inc. · OLO · May 2, 2023 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K223511

## Device Facts

- **Applicant:** Innovasis, Inc.
- **Product Code:** [OLO](/productcode/OLO.md)
- **Decision Date:** May 2, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The Excella® Navigation Instruments are intended to be used during the preparation and placement of Excella II®, Excella 3® & Excella MIS® screws during spinal surgery to assist the surgeon in locating anatomical structures in either open or percutaneous spinal procedures. The Excella Navigation Instruments are specifically designed for use with the Medtronic® StealthStation® System S8 and associated tracking arrays, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra can be identified relative to an acquired image (CT, MR, 2D/3D fluoroscopic image reconstruction) and/or an image-based model of the anatomy.

## Device Story

Excella Navigation Instruments are surgical tools used for preparation and insertion of spinal screws; compatible with Medtronic StealthStation System S8. Instruments feature tracking arrays allowing real-time visualization of tool position relative to patient anatomy; input data includes pre-operative or intra-operative images (CT, MR, 2D/3D fluoroscopy). System transforms spatial tracking data into visual guidance on surgical display; assists surgeons in navigating anatomical structures during open or percutaneous spinal procedures. Output provides visual feedback to surgeon for precise screw placement; enhances accuracy in spinal instrumentation. Used in OR by surgeons; benefits include improved surgical precision and reduced risk of malpositioning.

## Clinical Evidence

Bench testing only. Testing included navigation accuracy analysis per ASTM F2554-18, navigation simulated use, CAD model evaluation, implant/instrument mating verification, and functional testing of the spine tools package. Results confirmed all measurements were within specifications and comparable to the predicate device.

## Technological Characteristics

Orthopedic stereotaxic instruments designed for integration with Medtronic StealthStation S8 tracking arrays. Materials are biocompatible and equivalent to predicate. Connectivity is achieved via physical tracking arrays compatible with optical/electromagnetic navigation systems. Testing performed per ASTM F2554-18 for positional accuracy.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Medtronic Navigated CD Horizon Solera Screwdrivers and Taps ([K140454](/device/K140454.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

May 2, 2023

Innovasis, Inc. Michael Thomas Regulatory Affairs Manager 614 E 3900 S Salt Lake City, Utah 84107

Re: K223511

Trade/Device Name: Excella Navigation Instruments Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: [NOTE: Use date of most recent supplement] Received: May 5, 2023

Dear Michael Thomas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali-S

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

Submission Number (if known)

K223511

Device Name

Excella Navigation Instruments

#### Indications for Use (Describe)

The Excella® Navigation Instruments are intended to be used during the preparation and placement of Excella II®, Excella 3® & Excella MIS® screws during spinal surgery to assist the surgeon in locating anatomical structures in either open or percutaneous spinal procedures. The Excella Navigation Instruments are specifically designed for use with the Medtronic® StealthStation® System S8 and associated tracking arrays, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra can be identified relative to an acquired image (CT, MR, 2D/3D fluoroscopic image reconstruction) and/or an image-based model of the anatomy.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

Prepared on: 2022-11-18

### Contact Details

| Applicant Name              | Innovasis, Inc.                                    |
|-----------------------------|----------------------------------------------------|
| Applicant Address           | 614 E 3900 S Salt Lake City UT 84107 United States |
| Applicant Contact Telephone | 8012612236                                         |
| Applicant Contact           | Mr. Michael Thomas                                 |
| Applicant Contact Email     | mthomas@innovasis.com                              |

Device Name 21 CFR 807.92(a)(2)

| Device Trade Name   | Excella Navigation Instruments    |
|---------------------|-----------------------------------|
| Common Name         | Stereotaxic instrument            |
| Classification Name | Orthopedic Stereotaxic Instrument |
| Regulation Number   | 882.4560                          |
| Product Code        | OLO                               |

Legally Marketed Predicate Devices 21 CFR 807.92(a)(3)

| Predicate #                                              | K140454                                                     |
|----------------------------------------------------------|-------------------------------------------------------------|
| Predicate Trade Name (Primary Predicate is listed first) | Medtronic Navigated CD Horizon Solera Screwdrivers and Taps |
| Product Code                                             | OLO                                                         |

Device Description Summary 21 CFR 807.92(a)(4)The Innovasis® Excella® Navigation Instruments used for the preparation and insertion of Excella III , Excella 3° & Excella MIS® screws, in either open or percutaneous procedures. These instruments are with the Medtronic® StealthStation® System or as standalone, unnavigated instruments.

## Intended Use/Indications for Use

The Excella® Navigation Instruments are intended to be used during the preparation and placement of Excella 3° & Excella MIS® screws during spinal surgery to assist the surgeon in locating anatomical structures in either open or percutaneous spinal procedures. The Excella Navigation Instruments are specifically designed for use with the Medtronice StealthStation® System S8 and associated tracking arrays, which is indical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra can be identified relative to an acquired image (CT, MR, 2D/3D fluoroscopic image reconstruction) and/or an image-based model of the anatomy.

## Indications for Use Comparison

Indications for use are substantially equivalent. The subject Innovasis indications are a subset of the predications (i.e., specifically spinal procedures).

### Technological Comparison

The subject device has the same technological characteristics as the as the predicate device.

- Technology is substantially equivalent.

- Design is substantially equivalent.

#### K223511, Page 1 of 2

## 21 CFR 807.92(a)(5)

21 CFR 807.92(a)(6)

# 21 CFR 807.92(a)(5)

{4}------------------------------------------------

- Size is substantially equivalent.

- Indications for use and intended use are substantially equivalent.

- Materials (biocompatibility profile) are substantially equivalent.

#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

Testing was conducted in accordance with "ASTM F2554-18 Standard Practice for Measurement of Positional Accuracy of Computer Assisted Systems". The data shows all measurements were within specifications and comparable to the predicate product.

Navigation Accuracy Analysis: Confirmed navigated instrument accuracy.

Navigation Simulated Use: Confirmed navigation system functionality under expected use conditions.

CAD Model Evaluation: Verified that the CAD models are accurately reflected in the application software.

lmplant/Instrument Mating Conditions: Verified that the instruments can be assembled with the appropriate devices according to their intended use.

Spine Tools Package Functional Testing: Verified that the Spine Tools package has met the required interface needs of the spine application software.

---

**Source:** [https://fda-staging.innolitics.com/device/K223511](https://fda-staging.innolitics.com/device/K223511)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
