The Cardioblate™ Gemini™ Surgical Ablation Device is intended to ablate cardiac tissue during cardiac surgery using radiofrequency energy. The system is indicated for use, under direct or endoscopic visualization, in surgical procedures, including minimally invasive surgical procedures.
Device Story
Handheld, single-use, bipolar radiofrequency (RF) ablation device; used in cardiac surgery (open or closed chest) to create linear transmural lesions in atria. Device features dual linear electrodes (6.3 cm) and saline irrigation system to cool tissue during energy delivery. Physician operates device; flexible neck (0-180 degrees) allows anatomical access. Squeezing handle locks parallel jaws to ensure tissue contact; trigger actuation initiates RF energy and saline flow. External generator (Cardioblate 68000 or Valleylab FT10) delivers medium frequency (434 KHz) alternating current. Energy creates resistive heating; tissue temperature >50°C causes irreversible cell death, blocking abnormal electrical conduction pathways. Benefits include precise lesion creation to treat cardiac arrhythmias.
Clinical Evidence
Bench testing only. No clinical data presented. Verification testing and updated risk management (DFMECA) confirmed that modifications to IFU and directions for use do not impact safety or effectiveness.
Technological Characteristics
Bipolar RF ablation device; dual linear electrodes (6.3 cm); saline irrigation system. Energy source: external generator (434 KHz ± 10%). Flexible neck (0-180°). Materials: patient-contacting materials unchanged from predicates. Sterilization: ethylene oxide. Single-use, disposable. No software-based algorithm described; operation is controlled by external generator impedance monitoring.
Indications for Use
Indicated for patients undergoing cardiac surgery requiring ablation of cardiac tissue, including minimally invasive procedures, under direct or endoscopic visualization.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 22, 2022
Medtronic Inc. Rajitha Nair Sr. Regulatory Affairs Specialist 7611 Northland Drive Minneapolis, Minnesota 55448
Re: K223508
Trade/Device Name: Cardioblate Gemini-s 49260 Surgical Ablation Device; Cardioblate Gemini-s 49351 Surgical Ablation Device Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: OCL Dated: November 21, 2022 Received: November 22, 2022
Dear Rajitha Nair:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Aneesh S. Deoras -S
Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K223508
#### Device Name
Cardioblate Gemini-s 49260 Surgical Ablation Device; Cardioblate Gemini-s 49351 Surgical Ablation Device
Indications for Use (Describe)
The Cardioblate™ Gemini™ Surgical Ablation Device is intended to ablate cardiac tissue during cardiac surgery using radiofrequency energy. The system is indicated for use, under direct or endoscopic visualization, in surgical procedures, including minimally invasive surgical procedures.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="text-align:left;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|--------------------------------------------------------------------------------------------------------------------------------|
| <span style="text-align:left;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### K223508 510(k) Summary Complying with 21 CFR 807.92
510(k) Summary
Submitter
| Date Prepared: | November 21, 2022 |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | Medtronic, Inc.<br>Medtronic Perfusion Systems<br>7611 Northland Drive<br>Minneapolis, MN 55428<br>Establishment Registration Number: 2184009 |
| Contact Person: | Rajitha Nair<br>Senior Regulatory Affairs Specialist<br>Medtronic Perfusion Systems<br>Phone: 609-760-1907<br>Email: rajitha.nair@medtronic.com |
| Alternate Contact: | Megan Schlichting<br>Regulatory Affairs Manager<br>Medtronic Perfusion Systems<br>Phone: 763-526-3087 |
#### Device
| Trade Name | Common<br>Name | Classification<br>Name | Classification | Classification<br>Panel | Regulation | Product<br>Code |
|-----------------------------------------------------------------|----------------------------|----------------------------------------------------------------------------|----------------|-------------------------|--------------------|-----------------|
| Cardioblate<br>Gemini-s 49260<br>Surgical<br>Ablation<br>Device | Electrosurgic<br>al device | Electrosurgical<br>cutting and<br>coagulation<br>device and<br>accessories | Class II | Cardiovascular | 21 CFR<br>878.4400 | OCL |
| Cardioblate<br>Gemini-s 49351<br>Surgical<br>Ablation<br>Device | Electrosurgic<br>al device | Electrosurgical<br>cutting and<br>coagulation<br>device and<br>accessories | Class II | Cardiovascular | 21 CFR<br>878.4400 | OCL |
Email: megan.schlichting@medtronic.com
#### Predicate Device
| Trade Name | 510(k) | Clearance Date |
|-----------------------------------------------------|---------|----------------|
| Cardioblate Gemini-s 49260 Surgical Ablation Device | K121767 | 7/13/2012 |
| Cardioblate Gemini-s 49351 Surgical Ablation Device | K200514 | 6/3/2020 |
No reference devices were used in this submission.
{4}------------------------------------------------
### K223508 510(k) Summary Complying with 21 CFR 807.92
#### Device Description
The Cardioblate Gemini surgical ablation device (Model 49351), figure 1 below, is a handheld, single-use, bipolar, radio-frequency (RF) ablation device for use in cardiac surgery. It has a saline irrigation system to deliver fluid at the contact point between the tissue and electrode to cool the tissue during RF energy delivery. This device is intended for intermittent operation
Model 49260 of the Cardioblate Gemini surgical ablation device is intended for use with Cardioblate 68000 Generator (figure 2), whereas Model 49351 is intended for use with the Valleylab™ FT10 Energy(figure 3), whereas model 49260 is intended for use with Cardioblate 68000 Generator(figure 3). The devices are provided sterile and nonpyrogenic, are disposable, and are for single use only. They are sterilized using ethylene oxide.
The Cardioblate Gemini-s is a bipolar electrode device that transmits radio frequency energy, generated by an external generator, between the two electrodes. The device has dual linear electrodes, 6.3 cm in length, with integral fluid delivery to the electrodes. The Gemini-s device is port accessible, intended for open or closed chest procedures to ablate tissue through resistive heating due to radiofrequency energy passing through tissue. The device is designed to reproduce the Cox-Maze lesion pattern by creating linear transmural lesions in both atria of the heart during cardiac procedures.
The Gemini-s surgical ablation device has a flexible neck design that provides the physician with the ability to access various anatomies utilizing a neck curve through the full range from 0 to 180-degree configuration.
Once the electrodes are correctly positioned, the parallel jaws lock by squeezing the handle. The electrodes have full contact with the tissue and the locking mechanism must be fully engaged to ensure accurate transmurality readings. Simultaneous with the actuation of the trigger and jaw closure is the opening of the (normally closed) fluid path to start the flow of saline. The device automatically shuts off the saline when the jaws are open and when the device is not in use.
Image /page/4/Picture/7 description: The image shows a Medtronic surgical tool. The tool has a gray and blue handle with a trigger and a long, black, flexible shaft. At the end of the shaft are two small, curved metal prongs.
Figure 1: Cardioblate Gemini-s (Model 49260, 49351)
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a portable medical device, possibly a defibrillator or similar diagnostic tool. It features a screen that is flipped open, revealing a user interface with buttons and possibly a display area. The device is primarily white with blue accents, and it has a handle for easy carrying.
## Figure 2 Cardioblate 68000 Generator Medtronic VLFT10GEN Generator
| Acres problem of | |
|------------------|--|
| | |
| | |
| DON . | |
Image /page/5/Figure/4 description: This image is a title for a figure. The title says "Figure 3: Medtronic VLFT10GEN Generator". The title is written in a bold, sans-serif font. The title is centered on the image.
### Principles of Operation
The device applies endocardial radiofrequency energy. The energy platform monitors tissue impedance and adjusts the amount of power delivered to the tissue between the active electrodes to create transmural lesions in the intracardiac tissue. Irrigated Radiofrequency Frequency (RF) ablation uses medium frequency alternating current to generate heat, which destroys abnormal electrical conduction pathways while cooling the cardiac tissue. Radiofrequency (RF) ablation is a procedure where medium frequency (434 KHz ± 10%) energy is delivered into cardiac tissue. The radiofrequency energy is delivered from a generator, through the irrigated electrode (patient contacting end of handheld device) and into the tissue. When the energy encounters the higher impedance of cardiac tissue, heat is produced, causing cells to die. As tissue temperature rises above 113° F (50° C), protein is permanently damaged and cell membranes fuse causing irreversible cardiac cell death. Once the cells are no longer viable, they are unable to propagate electrical signals. This type of electrical energy has been used in medical procedures for decades.
#### Indications for Use
The Cardioblate® Gemini® Surgical Ablation Device is intended to ablate cardiac tissue during cardiac surgery using radiofrequency energy. The system is indicated for use, under direct or endoscopic visualization, in surgical procedures, including minimally invasive surgical procedures.
#### Comparison of Technological Characteristics with the Predicate Device
When compared to the predicate devices, the Cardioblate Gemini Surgical Ablation Devices presented in this submission have the same:
- intended use/indications for use
- contraindications
- operating principle
- energy type
{6}------------------------------------------------
#### K223508
510(k) Summary Complying with 21 CFR 807.92
- mechanism of action
- . radiofrequency generator and delivery system
- . patient contacting materials
- shelf life
- sterilization process
- dimensions
- performance
- irrigation
- Electrical
- packaging configuration and materials
When compared to the predicate devices, the Cardioblate Gemini Surgical Ablation Devices presented in this submission have the following differences:
- . IFU updates (new warning and updated directions for use)
#### Performance Data
The modifications described in this Special 510(k) Notification were verified through design verification testing. The Design Failure Modes and Effects Analysis (DFMECA) file was updated along with the Risk Management File to reflect the current status and risk evaluations at the device and system level. Risk Assessment concluded that no new biocompatibility testing was needed for the proposed Cardioblate Gemini Surgical Ablation Devices because there are no changes to:
- Intended use
- Technological characteristics
- Operating principle
- Design features
- Device and packaging materials
- Sterilization requirements
- Shelf life
#### Conclusions
The Predicate devices are legally marketed under K121767 (49260) and K200514 (49351), have the same intended use, and the same technological and performance characteristics as the subject devices. The proposed change to the IFU adds a new warning to align with the DFMECA. Additionally, directions for use have been updated to provide additional clarity and minimize use error. These changes do not arise from different technological characteristics of the devices or raise different questions of safety and effectiveness.
In summary, this submission demonstrates that the Cardioblate Gemini Ablation Devices are substantially equivalent2 to the legally marketed predicate devices.
<sup>+</sup> The term 'substantially equivalent' as used herein is intended to be a determination of substantial equivalency under the Federal Food, Drug, and Cosmetic Act and relates to the fact the product can be marketed without premarket approval or reclassification. Such a determination is not intended to have any bearing whatsoever on the resolution of patent infringement suits or other patent matters." (Federal Register, Vol. 42, No. 163, Aug. 23, 1977, page 42525 and 42529).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.