ClosureFast™ Endovenous Radiofrequency Ablation Catheter

K223488 · Medtronic, Inc. · GEI · Mar 23, 2023 · General, Plastic Surgery

Device Facts

Record IDK223488
Device NameClosureFast™ Endovenous Radiofrequency Ablation Catheter
ApplicantMedtronic, Inc.
Product CodeGEI · General, Plastic Surgery
Decision DateMar 23, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ClosureFast catheter is intended for endovascular coagulation of blood vessels in patients with superficial vein reflux.

Device Story

ClosureFast™ Endovenous Radiofrequency Ablation (RFA) Catheter is a sterile, single-use, disposable device; used for endovascular coagulation of blood vessels in patients with superficial vein reflux. Device operates by delivering thermal energy to treatment site via radiofrequency heating of catheter heating element (coil). Designed for use with ClosureRFG™ Radiofrequency Generator. Features 6F profile, 8 cm heating element, and integrated connection cable. Includes single-use limiter in handle PCBA to establish 2-hour treatment window after first energy cycle. Physician operates device in clinical setting; output is thermal energy for vein ablation. Benefits include minimally invasive treatment of superficial vein reflux.

Clinical Evidence

Bench testing only. Performance data included mechanical, dimensional, tensile strength, electrical, environmental, simulated use, packaging verification, software verification/validation, usability, benchtop validation, sterilization validation, and biocompatibility testing. All tests met acceptance criteria.

Technological Characteristics

6F profile catheter; 8 cm length, 2.00 mm diameter heating element; integrated connection cable; PCBA-based single-use limiter; compatible with ClosureRFG™ Radiofrequency Generator; sterile, single-use, disposable.

Indications for Use

Indicated for endovascular coagulation of blood vessels in patients with superficial vein reflux.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. March 23, 2023 Medtronic, Inc. Aditi Dave Senior Regulatory Affairs Specialist 2300 Berkshire Lane North Plymouth, Minnesota 55441 Re: K223488 Trade/Device Name: ClosureFast™ Endovenous Radiofrequency Ablation (RFA) Catheter Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: December 22, 2022 Received: December 23, 2022 Dear Aditi Dave: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2023.03.23 10:40:48 -04'00' Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. 510(k) Number (if known) K223488 Device Name ClosureFast™ Endovenous Radiofrequency Ablation (RFA) Catheter Indications for Use (Describe) The ClosureFast catheter is intended for endovascular coagulation of blood vessels in patients with superficial vein reflux. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the text 'K223488' in a simple, sans-serif font. The text is presented in a horizontal arrangement, with the letter 'K' followed by the numbers '223488'. The text is black and the background is white. Image /page/3/Picture/1 description: The image shows the word "Medtronic" in a blue, sans-serif font. The letters are bold and evenly spaced. The word appears to be a logo or brand name. The background is plain white. Page 1 of 4 Medtronic, Inc. Peripheral Vascular Health 2300 Berkshire Lane North Plymouth, MN 55441 # Traditional 510(k) Summary ClosureFast Catheter | 510(k) Summary | This summary of 510(k) safety and effectiveness information is<br>being submitted in accordance with the requirements of 21 CFR<br>807.92. | |--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------| | Applicant/ Submitter | Medtronic, Inc.<br>2300 Berkshire Lane North<br>Plymouth, MN 55441 | | Contact Person | Aditi Dave<br>Senior Regulatory Affairs Specialist<br>Tel: 763-398-7029<br>Email : aditi.b.dave@medtronic.com | | Secondary Contact Person | Laura Moen-Ftacek<br>Regulatory Affairs Director<br>Tel : 763-398-7359<br>Email : laura.moen-ftacek@medtronic.com | | Date Prepared | March 21, 2023 | | Device Trade Name | ClosureFast™ Endovenous Radiofrequency Ablation<br>Catheter | | Device Common Name | Electrosurgical Device | | Classification Name | Electrosurgical, Cutting & Coagulation & Accessories | | Regulation Number | 21 CFR 878.4400 | | Classification | Class II | | Classification Panel | General & Plastic Surgery | | Product Code | GEI | {4}------------------------------------------------ K223488 Image /page/4/Picture/1 description: The image shows the word "Medtronic" in a bold, sans-serif font. The word is a dark blue color. The letters are closely spaced together, creating a solid block of text. Page 2 of 4 Medtronic, Inc. Peripheral Vascular Health 2300 Berkshire Lane North Plymouth, MN 55441 | Predicate Device | ClosureFast™ Radiofrequency Catheter,<br>Models CF7-7-60, CF7-7-100 and CF7-3-60 | | |----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------| | Predicate 510(k) Number | K061373 and K111887 | | | Predicate Regulation<br>Number | 21 CFR 878.4400 | | | Indications for Use | The ClosureFast catheter is intended for endovascular<br>coagulation of blood vessels in patients with superficial vein<br>reflux. | | | Device Description | The ClosureFast™ Radio Frequency Ablation catheter is a sterile,<br>single use, disposable device with an integrated connection cable.<br>The catheter has a 6F profile with a heating element (coil) 8 cm in<br>length and 2.00 mm in diameter. It is available in two working<br>lengths, 60 cm and 100 cm. A single-use limiter feature is added<br>in the Printed Circuit Board Assembly (PCBA) housed inside the<br>handle of the device. It establishes an effective treatment window<br>of 2 hours after the first energy cycle. The catheter's function is<br>to provide thermal energy to the desired treatment site via<br>radiofrequency heating of the catheter heating element. The<br>ClosureFast catheter is designed to be used with the<br>ClosureRFG™ Radiofrequency Generator. | | | Comparison of<br>Technological<br>Characteristics: | The proposed device has the same characteristics as the predicate<br>devices, with the exception of the following design modifications:<br>Heating element dimensions (length/diameter) Catheter Index Mark spacing Addition of bridge tube Addition of Single Use Limiter Instructions for use updates Table 1 below outlines the comparison in technological<br>characteristics between the predicate and proposed device that are<br>not identical but are considered substantially equivalent. | | | Parameters | Proposed<br>Device | Predicate Device | | Introducer Sheath<br>Compatibility | 6F | 7F | | Heating Element<br>Length | 8 cm | 7 cm 3 cm | | Heating Element<br>Diameter | 2.00 mm | 2.3 mm | | Catheter<br>Markings | 7.5 cm | 6.5 cm 2.5 cm | | Bridge Tube | Yes, new<br>component | N/A | | Single Use Limiter | Yes, new<br>component | N/A | | Instructions for Use<br>(IFU)<br>(specifications) | Paper IFU (8.5"<br>x 11" saddle<br>stitch) and eIFU<br>leaflet (3.5'' x<br>3.5'' folded) | Paper IFU (4" x 9.94"<br>Map fold) | {5}------------------------------------------------ ### K223488 ## Table 1: Comparison between proposed and predicate device ## Performance data To demonstrate substantial equivalence of the proposed device to the predicate device, the following design verification and validation testing was performed: - Mechanical ● - Dimensional - Mechanical / Tensile Strength ● - Electrical ● - Environmental - Simulated Use - Packaging Verification ● - Software Verification and Validation - Usability - Benchtop Validation - Sterilization Validation - Biocompatibility Test results demonstrate that the proposed ClosureFast catheter meets acceptance criteria and performance requirements and is acceptable for its intended use. {6}------------------------------------------------ K223488 Page 4of 4 Medtronic, Inc. Peripheral Vascular Health 2300 Berkshire Lane North Plymouth, MN 55441 ## Summary of Substantial Equivalence The proposed device has the following equivalencies to the predicate device: - Intended Use ● - Indications for Use - Principle of Operation - Fundamental scientific technology - Materials of construction - Compatibility to ClosureRFGTM Radiofrequency Generator . - Packaging materials - Sterilization method The proposed device modifications do not impact the overall safety and effectiveness of the device. There are no new risks identified or significantly modified existing risks are identified due to these device modifications. - Conclusion Based on the same intended use, technological characteristics, safety and performance testing included in the submission, Medtronic concludes the proposed ClosureFast Catheter to be substantially equivalent to the predicate ClosureFast Catheters.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...