Antimicrobial susceptibility of Streptococcus pneumoniae
Discriminant Analysis
EA > 90%, CA > 90%
98.3% EA, 99.7% CA
—
—
Re-analysis of 289 clinical and challenge isolates from K111893
—
Antimicrobial susceptibility of Streptococcus pyogenes
Discriminant Analysis
EA > 90%, CA > 90%
99.7% EA, 100.0% CA
—
—
Re-analysis of 308 clinical and challenge isolates from K111893
—
Antimicrobial susceptibility of Streptococcus spp. β-Hemolytic Group
Discriminant Analysis
EA > 90%, CA > 90%
97.7% EA, 96.0% CA
—
—
Re-analysis of 530 clinical and challenge isolates from K111893 and additional challenge set
—
Indications for Use
VITEK® 2 Streptococcus Tetracycline is designed for antimicrobial susceptibility testing of Streptococcus species and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Streptococcus Tetracycline is a quantitative test. Tetracycline has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections: Streptococcus pneumoniae Streptococcus pyogenes* *The VITEK® 2 Streptococcus Susceptibility Card also reports the susceptibility of the following additional organisms as listed on the FDA Susceptibility Test Interpretative Criteria website (STIC): Streptococcus spp. B-Hemolytic Group (other than S. pyogenes). The VITEK® 2 Streptococcus Susceptibility Card is intended for use with the VITEK® 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Streptococcus pneumoniae, beta-hemolytic Streptococcus, and Viridans Streptococcus to antimicrobial agents when used as instructed.
Device Story
VITEK 2 AST-Streptococcus Tetracycline card; miniaturized, automated doubling dilution technique for MIC determination. Input: bacterial isolate diluted in saline. Operation: card contains 64 wells with premeasured antimicrobial concentrations; system fills, seals, and incubates card; optical scanner measures light attenuation every 15 minutes to monitor growth. Output: MIC value and interpretive category (S, I, R). Used in clinical laboratories; operated by technicians. Results assist clinicians in selecting appropriate antimicrobial therapy for Streptococcus infections. This submission updates device labeling to reflect current FDA-recognized breakpoints.
Clinical Evidence
Bench testing only. Re-analysis of 1127 clinical/challenge isolates (auto-dilution) and 163 challenge isolates (manual dilution) from original submission (K111893), plus 63 additional challenge isolates. Primary endpoints: Essential Agreement (EA) and Category Agreement (CA) against CLSI broth microdilution reference. Results: Combined EA 98.3% (S. pneumoniae), 99.7% (S. pyogenes), 97.7% (S. spp. β-Hemolytic). CA 99.7%, 100%, and 96.0% respectively. Minor discrepancies noted; limitation added for S. dysgalactiae equisimilis.
Technological Characteristics
Miniaturized 64-well AST card; automated growth-based detection via light transmittance optics. Energy source: visible light scanner. Connectivity: VITEK 2/VITEK 2 Compact systems. Software: Discriminant Analysis algorithm. Sterilization: Not specified.
Indications for Use
Indicated for antimicrobial susceptibility testing of Streptococcus pneumoniae, Streptococcus pyogenes, and other beta-hemolytic Streptococcus species and Viridans Streptococcus in clinical laboratories using VITEK 2 or VITEK 2 Compact systems.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K223481
B Applicant
bioMérieux, Inc
C Proprietary and Established Names
VITEK 2 Streptococcus Tetracycline (≤0.25 - ≥16 μg/mL)
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| LON | Class II | 21 CFR 866.1645 - Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System | MI - Microbiology |
## II Submission/Device Overview:
### A Purpose for Submission:
To update the VITEK 2 AST-Streptococcus Tetracycline device labeling to include updated FDA-recognized breakpoints for Streptococcus pneumoniae, Streptococcus pyogenes (Group A Beta-Hemolytic streptococci) and Streptococcus spp. β-Hemolytic Group as published in the FDA STIC website.
### B Measurand:
Tetracycline ≤0.25 – ≥16 μg/mL
### C Type of Test:
Automated quantitative or qualitative antimicrobial susceptibility test for tetracycline
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
VITEK 2 Streptococcus Tetracycline is designed for antimicrobial susceptibility testing of *Streptococcus* species and is intended for use with the VITEK 2 and VITEK 2 COMPACT Systems as a laboratory aid in the determination of *in vitro* susceptibility to antimicrobial agents. VITEK 2 Streptococcus Tetracycline is a quantitative test. Tetracycline has been shown to be active against the microorganisms listed below, according to the FDA label for this antimicrobial.
Active *in vitro* and in clinical infections:
*Streptococcus pneumoniae*
*Streptococcus pyogenes*
*The VITEK 2 *Streptococcus* Susceptibility Card also reports the susceptibility of the following additional organisms as listed on the FDA Susceptibility Test Interpretative Criteria website (STIC): *Streptococcus* spp. β-Hemolytic Group (other than *S. pyogenes*).
The VITEK 2 *Streptococcus* Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an *in vitro* test to determine the susceptibility of *Streptococcus pneumoniae*, beta-hemolytic *Streptococcus*, and Viridans *Streptococcus* to antimicrobial agents when used as instructed.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
Limitation:
Perform an alternative method of testing prior to reporting of results for the following antibiotic/organism combination(s):
Tetracycline (te01n): *Streptococcus dysgalactiae equisimilis* when the VITEK 2 MIC is 2 μg/mL
D Special Instrument Requirements:
VITEK 2 and VITEK 2 Compact Systems
IV Device/System Characteristics:
A Device Description:
The VITEK 2 AST card is a miniaturized, abbreviated and automated version of the doubling dilution technique for determining the minimum inhibitory concentration (MIC). Each VITEK 2 AST card contains 64 wells. A control well(s) which contain only nutrient medium is resident on all cards. The remaining wells contain premeasured portions of antimicrobials combined with the nutrient media. The isolate to be tested is diluted to a standardized concentration with 0.45% to
K223481 - Page 2 of 12
{2}
0.50% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 System will automatically (or allow operator to manually) dilute the bacterial suspension to prepare an inoculum for susceptibility cards. Then, the VITEK 2 will fill, seal and place the card into the incubator/reader. The VITEK 2 Compact has a manual filling, sealing, and loading operation. The VITEK 2 Systems monitor the growth of each well in the card over a defined period of time. The analysis program determines when a well demonstrates growth based on attenuation of light measured by an optical scanner. This data is used to determine the minimum inhibitory concentration or "MIC" values for the antimicrobial agent. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antimicrobial contained on the card.
VITEK 2 Streptococcus Tetracycline has the following concentrations in the card: 0.125, 0.5, 1 and 4 µg/mL (equivalent standard method concentration by efficacy in µg/mL). The MIC result range for the VITEK 2 Streptococcus Tetracycline is ≤0.25 - ≥16 µg/mL.
## B Principle of Operation:
The VITEK 2 and VITEK 2 Compact Systems utilize automated growth-based detection using attenuation of light measured by an optical scanner. The optics in the systems use visible light to directly measure organism growth within each of the 64 micro-wells. Transmittance optics is based on an initial light reading of a well before significant growth has begun. Every 15 minutes throughout the incubation cycle (defined period of time based on the VITEK 2 card), light transmittance readings of each well determine organism growth by the amount of light that is prevented from passing through the well. At the completion of the incubation period, the MIC values and their associated interpretive category results for each antimicrobial on the test card are displayed in an automatically generated report.
## V Substantial Equivalence Information:
A Predicate Device Name(s):
VITEK 2 Streptococcus Tetracycline
B Predicate 510(k) Number(s):
K111893
C Comparison with Predicate(s):
| Device & Predicate Device(s): | Device: K223481 | Predicate: K111893 |
| --- | --- | --- |
| Device Trade Name | VITEK 2 Streptococcus Tetracycline (≤0.25 - ≥16 μg/mL) | VITEK 2 Streptococcus Tetracycline |
| General Device Characteristic Similarities | | |
| Intended Use | VITEK 2 Streptococcus Tetracycline is designed for antimicrobial susceptibility testing of Streptococcus species | VITEK 2 Streptococcus Tetracycline is designed for antimicrobial susceptibility testing of Streptococcus species. |
K223481 - Page 3 of 12
{3}
K223481 - Page 4 of 12
| | and is intended for use with the VITEK 2 and VITEK 2 COMPACT Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK 2 Streptococcus Tetracycline is a quantitative test. Tetracycline has been shown to be active against the microorganisms listed below, according to the FDA label for this antimicrobial.
**Active in vitro** and in clinical infections:
*Streptococcus pneumoniae*
*Streptococcus pyogenes**
*The VITEK® 2 Streptococcus Susceptibility Card also reports the susceptibility of the following additional organisms aslisted on the FDA Susceptibility Test Interpretative Criteria website (STIC): *Streptococcus* spp. β-Hemolytic Group (other than *S. pyogenes*).
The VITEK 2 Streptococcus Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an *in vitro* test to determine the susceptibility of *Streptococcus pneumoniae*, beta-hemolytic Streptococcus, and Viridans Streptococcus to antimicrobial agents when used as instructed. | VITEK 2 Streptococcus Tetracycline is a quantitative test intended for use with the VITEK 2 and the VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. Tetracycline has an antimicrobial activity against the microorganisms listed below, according to the FDA label for this antimicrobial.
**Active in vitro** and in clinical infections:
*Streptococcus pneumoniae*
*Streptococcus pyogenes*
Viridans group streptococci
The VITEK 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK 2 System for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, *Staphylococcus* spp., *Enterococcus* spp., *Streptococcus* spp. and clinically significant yeast |
| --- | --- | --- |
| Test Methodology | Automated quantitative antimicrobial susceptibility test for use with the VITEK 2 and VITEK 2 Compact Systems to determine the in vitro susceptibility of microorganisms | Same |
| Inoculum | Saline suspension of organism | Same |
| Test Card | Streptococcus (AST-ST) Susceptibility Card | Same |
| Analysis Algorithms | Discriminant Analysis | Same |
| Instrument | VITEK 2 and VITEK 2 Compact Systems | Same |
{4}
| Antimicrobial Agent | Tetracycline | Same |
| --- | --- | --- |
| General Device Characteristic Differences | | |
| Indicated Organisms | Tetracycline has been shown to active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial.
**Active in vitro** and in clinical infections:
*Streptococcus pneumoniae*
*Streptococcus pyogenes**
*The VITEK 2 *Streptococcus* Susceptibility Card also reports the susceptibility of the following additional organisms a slisted on the FDA Susceptibility Test Interpretative Criteria website (STIC):
*Streptococcus* spp. β-Hemolytic Group (other than *S. pyogenes*) | Tetracycline has been shown to active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial.
**Active in vitro** and in clinical infections:
*Streptococcus pneumoniae*
*Streptococcus pyogenes*
Viridans group streptococci |
| Breakpoints | *Streptococcus pneumoniae*
S≤1, I 2, R≥4
*Streptococcus* spp. β-Hemolytic Group (including *S. pyogenes*)
S≤1, I 4, R≥8 | *Streptococcus* spp.
S≤4, I 8, R≥16 |
VI Standards/Guidance Documents Referenced:
1. MacLowry, J.D. and Marsh, H.H., Semi-automatic Microtechnique for Serial Dilution Antibiotic Sensitivity Testing in the Clinical laboratory, Journal of Laboratory Clinical Medicine, 72:685-687, 1968.
2. Gerlach, E.H., Microdilution 1: A Comparative Study, p. 63-76. Current Techniques for Antibiotic Susceptibility Testing. A. Balows (ed.), Charles C. Thomas, Springfield, IL, 1974.
3. Barry, A.L., The Antimicrobic Susceptibility Test, Principles and Practices, Lea and Febiger, Philadelphia, PA, 1976.
4. CLSI M7-A5, "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard - Fifth Edition", (January 2000)
5. CLSI M07, "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard – 11th Edition", (January 2018)
6. CLSI M100, "Performance Standards for Antimicrobial Susceptibility Testing; Thirty Second Edition", Vol. 32, No. 1 (February 2022)
K223481 - Page 5 of 12
{5}
7. FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA (Issued August 28, 2009)
## VII Performance Characteristics (if/when applicable):
### A Analytical Performance:
1. **Precision/Reproducibility:**
Refer to K111893. Reproducibility testing for the VITEK 2 Streptococcus Tetracycline was performed in support of clearance of K111893 and was determined acceptable.
2. **Linearity:**
No applicable
3. **Analytical Specificity/Interference:**
Not applicable.
4. **Assay Reportable Range:**
Not applicable.
5. **Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):**
Quality Control (QC) testing, Inoculum and Growth failure for the VITEK 2 Streptococcus Tetracycline was performed in support of clearance of K111893 and was determined acceptable. QC testing specific to the changes in breakpoints was not performed in the current submission.
The QC organism recommended by both the FDA and CLSI, namely *Streptococcus pneumoniae* ATCC 49619, was tested against tetracycline during the clinical study a minimum 20 times/site by the automatic and the manual dilution. The QC organism was tested by the VITEK 2 AST card and the reference broth microdilution method. The QC results for both the reference and test panels appeared to be acceptable.
At the time of the original submission (K111893), the QC range for *S. pneumoniae* ATCC 49619 was $0.12 - 0.5\mu \mathrm{g / mL}$. The recommended CLSI QC range for this strain changed to include lower concentrations with a range of $0.06 - 0.5\mu \mathrm{g / mL}$. However, the VITEK 2 MIC reporting range is $\leq 0.25 - \geq 16\mu \mathrm{g / mL}$. The VITEK 2 systems do not provide results lower than $0.25\mu \mathrm{g / mL}$. Therefore, QC results for *S. pneumoniae* ATCC 49619 may be off scale. An MIC value of $\leq 0.25\mu \mathrm{g / mL}$ indicated that the quality control test results were acceptable (Table 1).
K223481 - Page 6 of 12
{6}
Table 1. Quality Control Results for VITEK 2 (Auto-Dilution and Manual Dilution Methods) and VITEK 2 Compact (Manual Dilution Method)
| Organism | VITEK 2 Result Range^{1} | BMD Result Range (μg/mL) | VITEK 2 Auto-Dilution | BMD | VITEK 2 Manual Dilution | BMD | VITEK 2 Compact Manual Dilution | BMD |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| S. pneumoniae ATCC 49619
Expected Result: 0.06 - 0.5 μg/mL | | ≤0.016 | | | | | | |
| | | 0.03 | | | | | | |
| | | 0.06 | | 1 | | 1 | | 0 |
| | | 0.12 | | 165 | | 167 | | 19 |
| | ≤0.25 | 0.25 | 222 | 56 | 223 | 55 | 68 | 49 |
| | 0.5 | 0.5 | 0 | 0 | 0 | 0 | 0 | 0 |
| | 1 | 1 | | | | | | |
| | 2 | 2 | | | | | | |
| | 4 | 4 | | | | | | |
| | 8 | 8 | | | | | | |
| | ≥16 | 16 | | | | | | |
| | | ≥32 | | | | | | |
¹VITEK 2 Card range is ≤0.25 - ≥16. The lowest dilution of the VITEK 2 Streptococcus Tetracycline MIC range is ≤0.25 μg/mL. Obtaining this value was considered an indicator that the quality control test results were acceptable.
bioMérieux included the following footnote under the Quality Control table in the device labeling:
“Does not include the full CLSI/FDA recommended dilution range for QC testing with this organism”.
The QC data is acceptable for the purpose of the re-evaluation of the performance of Streptococcus spp. when the new breakpoints are applied.
6. Detection Limit:
Not applicable.
7. Assay Cut-Off:
Not applicable.
B Comparison Studies:
1. Method Comparison with Predicate Device:
The VITEK 2 Streptococcus Tetracycline (≤0.25 - ≥16 μg/mL) which was originally cleared in K11893, included indications for Streptococcus pneumoniae, Streptococcus pyogenes (Group A Beta-Hemolytic streptococci) and Streptococcus spp. (Viridans Group) tested with the VITEK 2 auto dilution and the VITEK 2 manual dilution and compared to the CLSI broth microdilution reference method.
K223481 - Page 7 of 12
{7}
The performance of the VITEK 2 AST-ST Tetracycline card with Streptococcus pnuemoniae, Streptococcus pyogenes (Group A Beta-hemolytic streptococci) and Streptococcus spp. (Beta-Hemolytic Group) using revised interpretive criteria currently recognized by FDA was evaluated in the current submission using data obtained in support of K111893. Streptococcus spp. (Viridans Group) is no longer recognized by FDA STIC webpage and these data were removed from analysis.
Since there was no change in the design or the dilution range of the VITEK 2 AST-ST card, the performance evaluation was achieved via re-analysis of the MIC data of the original 510(k) submission (K111893). An additional 63 beta-hemolytic Streptococcus spp. (not including S. pyogenes) challenge organisms were tested and performance was assessed in this submission to determine performance for this indication. For this review, the current Tetracycline interpretive criteria are applied to Streptococcus pneumoniae and Streptococcus spp. $\beta$ -Hemolytic Group according to the FDA STIC webpage.
A total of 1127 (clinical and challenge) Streptococcus spp. isolates were tested using the VITEK 2 with automatic dilution. The isolates included 964 clinical isolates and 163 challenge isolates.
Results of the re-analyzed 1127 Streptococcus spp. clinical and challenge isolates (including 289 Streptococcus pneumoniae, 308 Streptococcus pyogenes (Group A Beta-Hemolytic streptococci) and 530 Streptococcus spp. Beta-Hemolytic Group (other than S. pyogenes) are summarized in Table 2.
The previously-collected data for the 163 challenge isolates tested at a single site with the VITEK 2 and the VITEK 2 Compact Systems using the manual dilution was also analyzed using the current breakpoints for Streptococcus pneumoniae, S. pyogenes and Streptococcus spp. Beta-Hemolytic Group (other than S. pyogenes). The results of the analysis are outlined in Table 3.
Table 2. Re-analysis of the Performance of Streptococcus spp. for the Auto Dilution and the VITEK 2 AST-ST Tetracycline
| | Total | #EA | %EA | Eval EA Total | Eval EA # | Eval EA % | #CA | %CA | #R | #S | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| S. pneumoniae ≤1 (S), 2 (I), ≥4 (R) | | | | | | | | | | | | | |
| Clinical | 239 | 234 | 97.9 | 6 | 1 | 16.7 | 238 | 99.6 | 66 | 173 | 1 | 0 | 0 |
| Challenge | 50 | 50 | 100.0 | 0 | 0 | 0.0 | 50 | 100.0 | 17 | 33 | 0 | 0 | 0 |
| Combined | 289 | 284 | 98.3 | 6 | 1 | 16.7 | 288 | 99.7 | 83 | 206 | 1 | 0 | 0 |
| S. pyogenes (Group A Beta-Hemolytic Streptococcus spp.) ≤2 (S), 4 (I), ≥8 (R) | | | | | | | | | | | | | |
| Clinical | 258 | 257 | 99.6 | 3 | 2 | 66.7 | 258 | 100.0 | 22 | 236 | 0 | 0 | 0 |
| Challenge | 50 | 50 | 100.0 | 1 | 1 | 100.0 | 50 | 100.0 | 6 | 44 | 0 | 0 | 0 |
| Combined | 308 | 307 | 99.7 | 4 | 3 | 75.0 | 308 | 100.0 | 28 | 280 | 0 | 0 | 0 |
| Streptococcus spp. (Beta-Hemolytic Group other than S. pyogenes)a,b ≤2 (S), 4 (I), ≥8 (R) | | | | | | | | | | | | | |
| Clinical | 467 | 455 | 97.4 | 44 | 32 | 72.7 | 446 | 95.5 | 295 | 156 | 18 | 1 | 2 |
| Challenge# | 63 | 63 | 100 | 3 | 3 | 100 | 63 | 100 | 53 | 10 | 0 | 0 | 0 |
| Combined | 530 | 518 | 97.7 | 47 | 35 | 74.5 | 509 | 96.0 | 348 | 166 | 18 | 1 | 2 |
a Including the following tested species: S. agalactiae (338), S.canis (4), S. dys dysgalactiae (1), S.dys.equisimilis (185) and S.equi zooepidemicus (2).
b The interpretive criteria are applied to Streptococcus spp. Beta-Hemolytic Group according to the FDA STIC webpage.
K223481 - Page 8 of 12
{8}
${}^{c}$ Additional testing was conducted to assess performance for Beta-Hemolytic Streptococcus spp. (not including S. pyogenes) against a challenge set of organisms.
EA - Essential Agreement
CA - Category Agreement
EVAL - Evaluable isolates
R - Resistant
min - minor discrepancies
maj - major discrepancies
vmj - very major discrepancies
S - Susceptible
When applying the new breakpoints of $\leq 1$ (S), $2(\mathrm{I}), \geq 4(\mathrm{R})$ for Streptococcus pneumoniae, the performance of the combined isolates tested with the VITEK 2 and automatic dilution is acceptable with $98.3\%$ EA and $99.7\%$ CA. There were no very major errors and no major errors. One (1) minor error out of 289 isolates $(0.3\%)$ tested was observed (Table 2).
When applying the new breakpoints of $\leq 2$ (S), $4(\mathrm{I}), \geq 8(\mathrm{R})$ for Streptococcus pyogenes, the performance of the combined isolates tested with the VITEK 2 and automatic dilution is acceptable with $99.7\%$ EA and $100.0\%$ CA. There were no very major errors, no major errors and no minor errors observed (Table 2).
When applying the new breakpoints of $\leq 2$ (S), $4(\mathrm{I}), \geq 8(\mathrm{R})$ for Streptococcus spp. (Beta-Hemolytic Group other than S. pyogenes), the performance of the combined isolates tested with the VITEK 2 and automatic dilution is acceptable with $97.7\%$ EA and $96.0\%$ CA. There were two very major errors out of 348 isolates $(0.6\%)$ and one major error out of 166 isolates $(0.6\%)$ . Eighteen (18) minor errors out of 530 isolates $(3.4\%)$ tested were observed (Table 2). The two very major errors $(3.4\%$ of 58 isolates), one major error $(0.9\%$ of 113 isolates) and 16 of the minor errors $(8.6\%$ of 185 isolates) were due to S. dysgalactiae subsp. equisimilis; therefore, the following limitation was added to the package insert to address these errors generated with the automatic dilution option:
Perform an alternative method of testing prior to reporting of results for the following antibiotic/organism combination(s):
Tetracycline (te01n): Streptococcus dysgalactiae equisimilis when the VITEK 2 MIC is 2 $\mu \mathrm{g} / \mathrm{mL}$
Table 3. Re-analysis of the Performance of Streptococcus spp. for the Manual Dilution with the VITEK 2 and the VITEK 2 Compact Systems AST-ST Tetracycline
| VITEK 2 Systems | Total | #EA | %EA | Eval Total | No Eval EA | Eval EA % | #CA | %CA | #R | #S | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| S. pneumoniae ≤1 (S), 2 (I), ≥4 (R) | | | | | | | | | | | | | |
| VITEK 2 | 50 | 49 | 98.0 | 1 | 0 | 0.0 | 50 | 100.0 | 17 | 33 | 0 | 0 | 0 |
| VITEK 2 Compact | 50 | 48 | 96.0 | 12 | 10 | 83.3 | 50 | 100.0 | 17 | 33 | 0 | 0 | 0 |
| S. pyogenes (Group A Beta-Hemolytic Streptococcus spp.) ≤2 (S), 4 (I), ≥8 (R) | | | | | | | | | | | | | |
| VITEK 2 | 50 | 50 | 100.0 | 1 | 1 | 100.0 | 50 | 100.0 | 6 | 44 | 0 | 0 | 0 |
| VITEK 2 Compact | 50 | 50 | 100.0 | 1 | 1 | 100.0 | 50 | 100.0 | 6 | 44 | 0 | 0 | 0 |
| Streptococcus spp. (Beta-Hemolytic Group other than S. pyogenes)a,b ≤2 (S), 4 (I), ≥8 (R) | | | | | | | | | | | | | |
| VITEK 2c | 63 | 63 | 100 | 3 | 3 | 100 | 63 | 100.0 | 53 | 10 | 0 | 0 | 0 |
| VITEK 2 Compactc | 63 | 63 | 100 | 4 | 4 | 100 | 62 | 98.4 | 53 | 10 | 1 | 0 | 0 |
a Including the following tested species: S. agalactiae (56), S.canis (1), S. dys. dysgalactiae (1), S.dys.equisimilis (4) and S.equi zooepidemicus (1).
K223481 - Page 9 of 12
{9}
b The interpretive criteria are applied to Streptococcus spp. Beta-Hemolytic Group according to the FDA STIC webpage.
c Additional testing was conducted to assess performance for Beta-Hemolytic Streptococcus spp. (not including S. pyogenes) against a challenge set of organisms.
| EA – Essential Agreement | min – minor discrepancies |
| --- | --- |
| CA – Category Agreement | maj – major discrepancies |
| EVAL – Evaluable isolates | vmj – very major discrepancies |
| R – Resistant | S – Susceptible |
When applying the new breakpoints of ≤1 (S), 2 (I), ≥4 (R), the performance of Streptococcus pneumoniae challenge isolates tested with the manual dilution and the VITEK 2 and the VITEK 2 Compact systems is acceptable with % EA and % CA >90%. There were no very major errors, no major errors and no minor errors observed with both the VITEK 2 and the VITEK 2 Compact out of 50 challenge isolates tested (Table 3).
When applying the new breakpoints of ≤2 (S), 4 (I), ≥8 (R), the performance of Streptococcus pyogenes challenge isolates tested with the manual dilution and the VITEK 2 and the VITEK 2 Compact systems is acceptable with % EA and % CA >90%. There were no very major errors, no major errors and no minor errors observed with both the VITEK 2 and the VITEK 2 Compact out of 50 challenge isolates tested (Table 3).
When applying the new breakpoints of ≤2 (S), 4 (I), ≥8 (R) for Streptococcus spp. (Beta-Hemolytic Group other than S. pyogenes) challenge isolates tested with the manual dilution and the VITEK 2 and VITEK 2 Compact systems is acceptable with % EA and % CA >90%. There were no very major errors and no major errors observed. One (1) minor error out of 63 isolates (1.6%) tested was observed for the VITEK 2 Compact system (Table 3).
As required under 511A(2) (2) (B) of the Federal Food, Drug and Cosmetic Act, the following statement was included under Precautions Section in the device labeling:
Per the FDA-Recognized Susceptibility Test Interpretive Criteria website, the safety and efficacy of antimicrobial drugs, for which antimicrobial susceptibility is tested by this AST device, may or may not have been established in adequate and well-controlled clinical trials for treating clinical infections due to microorganisms outside of those found in the indications and usage in the drug label. The clinical significance of susceptibility information in those instances is unknown. The approved labeling for specific antimicrobial drugs provides the uses for which the antimicrobial drug is approved.
## Trending
A trending re-analysis was performed using the combined (challenge and clinical) data obtained in K111893 from the VITEK 2 auto-dilution method for all Streptococcus spp. to align with current trending analysis.
This trending calculation takes into account the MIC values that are determined to be one or more doubling dilutions lower or higher than the reference method irrespective of whether the device MIC values are on-scale or not. Results that are not clearly at least one dilution lower, at least one dilution higher or in exact agreement with the CLSI reference method are not considered in the trending analysis.
K223481 - Page 10 of 12
{10}
When the difference between the percentage of isolates with higher vs. lower readings was > 30% and for which the confidence interval was determined to be statistically significant were considered to show evidence of trending. Trending that showed higher or lower MIC values compared to the reference is addressed in the labeling.
No trending was observed with *Streptococcus pneumoniae*, *S. pyogenes* (Group A Beta-Hemolytic streptococci) and *Streptococcus* spp. (Beta-Hemolytic Group other than *S. pyogenes*) (Table 4).
Table 4. Trending Re-analysis for *Streptococcus* spp. with Tetracycline and the VITEK 2 Automated Dilution
| Organism | Total Evaluable for Trending | ≥ 1 Dilution lower No. (%) | Exact No. (%) | ≥ 1 Dilution Higher No. (%) | Percent Difference (CI) | Trending Noted |
| --- | --- | --- | --- | --- | --- | --- |
| Streptococcus pneumoniae | 12 | 4 (33.3) | 1 (8.3) | 7 (58.3) | 25.0% | No |
| Streptococcus pyogenes | 6 | 3 (50.0) | 0 (0.0) | 3 (50.0) | 0% | No |
| Streptococcus spp. Beta-Hemolytic Group (other than S. pyogenes) | 50 | 27 (54.0) | 7 (14.0) | 16 (32.0) | -22.0% | No |
2. Matrix Comparison:
Not applicable.
C Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable.
D Clinical Cut-Off:
Not applicable.
E Expected Values/Reference Range:
The FDA recognized susceptibility interpretative for Tetracycline are listed in Table 5.
K223481 - Page 11 of 12
{11}
Table 5. FDA Recognized Interpretative Criteria for Tetracycline
| Pathogen | Minimum Inhibitory Concentrations (μg/mL)1 | | |
| --- | --- | --- | --- |
| | S | I | R |
| S. pneumoniae | ≤1 | 2 | ≥4 |
| Streptococcus spp. β-Hemolytic Group2 | ≤2 | 4 | ≥8 |
S = Susceptible; I = Intermediate; R = Resistant
1 FDA-Recognized Antimicrobial Susceptibility Test Interpretive Criteria Website https://www.fda.gov/Drugs/DevelopmentApprovalProcess/DevelopmentResources/ucm410971.htm
2Includes S. pyogenes (Group A β-Hemolytic streptococci) and other β-hemolytic streptococci other than S. pyogenes.
## VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device when evaluated with the current FDA-recognized Tetracycline breakpoints.
## IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
To support the implementation of changes to FDA-recognized susceptibility test interpretive criteria (i.e., breakpoints), this submission included a breakpoint change protocol that was reviewed and accepted by FDA. This protocol addresses future revisions to device labeling in response to breakpoint changes that are recognized on the FDA STIC webpage (https://www.fda.gov/Drugs/DevelopmentApprovalProcess/DevelopmentResources/ucm410971.htm). The protocol outlined the specific procedures and acceptance criteria that bioMérieux intends to use to evaluate the VITEK 2 System with Tetracycline when revised breakpoints for Tetracycline are published on the FDA STIC webpage. The breakpoint change protocol included with the submission indicated that if specific criteria are met, bioMérieux will update the Tetracycline device label to include (1) the new breakpoints, (2) an updated performance section after re-evaluation of data in this premarket notification with the new breakpoints, and (3) any new limitations as determined by their evaluation.
K223481 - Page 12 of 12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.