K223479 · B.Braun Medical, Inc. · FIP · Aug 16, 2023 · Gastroenterology, Urology
Device Facts
Record ID
K223479
Device Name
AQUAbase nX
Applicant
B.Braun Medical, Inc.
Product Code
FIP · Gastroenterology, Urology
Decision Date
Aug 16, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5665
Device Class
Class 2
Indications for Use
The AQUAbase nX is intended for use with hemodialysis systems to remove organic and inorganic substances and microbial contaminants from the water used for treating hemodialysis patients. The AQUAbase nX is to be used at dialysis clinics or hospitals The AQUAbase nX is a reverse osmosis unit intended to be a component in a complete water purification system, and is not a complete water treatment system. This reverse osmosis unit must be preceded by pre-treatment devices. Whether a particular device is included in an individual water treatment system will be dictated by local conditions. The reverse osmosis unit may need to be followed by post-treatment devices as well. The AQUAbase nX is designed to meet current AAMI/ISO and Federal (U.S.) standards.
Device Story
AQUAbase nX is a single-stage reverse osmosis (RO) system for hemodialysis water purification. Input: raw water; Process: high-pressure cross-filtration through semipermeable polyamide thin-film composite spiral-wound membranes; Output: purified permeate. Used in dialysis clinics/hospitals; operated by B. Braun certified technicians and trained personnel. Features touchscreen for monitoring production parameters, conductivity-based water conversion control, and alarms for poor water quality/pressure. HT models include integrated heaters for heat disinfection; standard models support chemical disinfection. System design emphasizes low dead-space stainless steel piping to reduce biofilm risk. Output affects clinical decision-making by ensuring water quality meets AAMI/ISO standards for dialysate preparation, protecting patients from contaminants.
Clinical Evidence
Bench testing only. No clinical data. Testing included biocompatibility (cytotoxicity, hemolysis, chemical characterization), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and performance testing per ISO 23500-3. Additional validation included chemical/heat disinfection efficacy and bacterial/endotoxin count verification.
Technological Characteristics
Single-stage RO system; polyamide thin-film composite spiral-wound membranes. Stainless steel piping with low dead-space design. Integrated touchscreen control. Conductivity-based water conversion control. Optional integrated heater for heat disinfection. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2. Complies with ISO 23500-3. Modular cabinet design.
Indications for Use
Indicated for use in dialysis clinics or hospitals to purify water for hemodialysis by removing organic/inorganic substances and microbial contaminants. Must be used as a component within a larger water treatment system requiring pre-treatment.
Regulatory Classification
Identification
A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.
Special Controls
*Classification.* Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the words "Convert to PDF" in a clear, sans-serif font. The text is horizontally aligned and appears to be a call to action or a title indicating a conversion process. The background is a light, neutral color, providing good contrast for the text.
Image /page/0/Picture/2 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
August 16, 2023
B. Braun Medical Inc. Rushtin Chaklader Regulatory Affairs Specialist 901 Marcon Blvd Allentown, Pennsylvania 18109
Re: K223479
> Trade/Device Name: AQUAbase nX, AQUAbase nX HT Regulation Number: 21 CFR 876.5665 Regulation Name: Water Purification System for Hemodialysis Regulatory Class: II Product Code: FIP Dated: July 14, 2023 Received: July 17, 2023
#### Dear Rushtin Chaklader:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpm/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part
{1}------------------------------------------------
801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic provisions (Sections (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Gema Gonzalez, MS Acting Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
# K223479
Device Name AQUAbase nX
Indications for Use (Describe)
The AQUAbase nX is intended for use with hemodialysis systems to remove organic and inorganic substances and microbial contaminants from the water used for treating hemodialysis patients. The AQUAbase nX is to be used at dialysis clinics or hospitals
The AQUAbase nX is a reverse osmosis unit intended to be a component in a complete water purification system, and is not a complete water treatment system. This reverse osmosis unit must be preceded by pre-treatment devices. Whether a particular device is included in an individual water treatment system will be dictated by local conditions. The reverse osmosis unit may need to be followed by post-treatment devices as well.
The AQUAbase nX is designed to meet current AAMI/ISO and Federal (U.S.) standards.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
complying with 21 CFR 807.92
# AQUAbase nX
#### I. APPLICANT AND SUBMITTER
#### Applicant
B. Braun Avitum AG Schwarzenberger Weg 73-79 34212 Melsungen, Germany Establishment Registration Number: 3007007791
#### Submitter
B. Braun Medical Inc. 901 Marcon Boulevard Allentown, PA 18109-9341 610-266-0500
#### Contact
Rushtin Chaklader Regulatory Affairs Specialist Phone: (610) 596-2789 E-mail: rushtin.chaklader@bbraunusa.com
Date prepared: 11/08/2022
#### II. DEVICE
| Name of device: | AQUAbase nX |
|----------------------|--------------------------------------------------------------|
| Classification name: | Water Purification System for Hemodialysis (21 CFR 876.5665) |
| Regulatory class: | II |
| Product code: | FIP |
| Product code name: | Subsystem, Water Purification |
| FDA Review Panel: | Gastroenterology/Urology |
{4}------------------------------------------------
#### III. PREDICATE DEVICE
Aquaboss (Eco)RO Dia I + II (HT) 510(k) number: K124059
### IV. DEVICE DESCRIPTION
The AQUAbase nX is a single stage reverse osmosis system. A graphical touchscreen allows access and monitoring of all operating parameters at any time. Customized parameters make a high water vield possible, even under poor raw water conditions. Raw water consumption is based solely on the end user's permeate needs. The touchscreen makes it possible for the user to monitor all production parameters as well as design every system function, including disinfection mode, individually and reproducibly.
#### Operating principle
The AQUAbase nX works on the reverse osmosis principle. Reverse osmosis describes the process of pressure-operated cross-filtration. Water flows at high pressure tangentially over a semipermeable membrane. As is the case with normal filtration, the system is cleaned by allowing one component (water) of the mixture to be separated to pass through the membrane with almost no hindrance, while other components (dissolved and undissolved water contents) are held back to a greater or lesser extent and leave the filtration unit in the concentrate flow. This is a purely physical separation process in the molecular range which does not change the components being separated either chemically, biologically or thermally.
| Single pass / Single staged Chemical disinfection | Flow rate of dialysis water (permeate)<br>in l/h (gph) @ 10°C (50°F) | |
|---------------------------------------------------|----------------------------------------------------------------------|------------|
| | L/h @ 10°C | gph @ 50°F |
| AQUAbase® nX 300 | 300 L/h | 79 gph |
| AQUAbase® nX 600 | 600 L/h | 158 gph |
| AQUAbase® nX 900 | 900 L/h | 237 gph |
| | | |
| Single pass / Single staged Heat disinfection | | |
| AQUAbase® nX HT 250 | 250 L/h | 66 gph |
| AQUAbase® nX HT 500 | 500 L/h | 132 gph |
| AQUAbase® nX HT 750 | 750 L/h | 198 gph |
Following models are available:
## Device characteristics
The system has the following design features:
- Disconnection from mains/free water intake .
- User-friendly touchscreen control .
{5}------------------------------------------------
- Password protection of configurable device data ●
- Compact design
- Low power consumption
- Single pipe construction
- o The single pipe construction ensures easy servicing of the membrane. The membrane pipe is made of stainless steel.
- AQUAbase nX HT hot water-disinfectable, full-fit elements ●
- o All hot-cleanable reverse osmosis systems in the AQUAbase nX HT series come with special full-fit reverse osmosis membrane elements whose external, high precision-manufactured, knurled polypropylene texture makes them especially suitable for use in microbially sensitive water treatment systems.
- Low dead-space, stainless steel piping ●
- The entire system is designed to achieve the lowest possible dead-space. High flow rate and the resulting shear forces also significantly reduce the risk of biofilm growth on the pipe walls.
# V. INTENDED USE AND INDICATIONS FOR USE
Intended Use:
Device that is intended for use with a hemodialysis system to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate.
#### Indications for Use:
The AQUAbase nX is intended for use with hemodialysis systems to remove organic and inorganic substances and microbial contaminants from the water used for treating hemodialysis patients.
The AQUAbase nX is to be used at dialysis clinics or hospitals
The AQUAbase nX is a reverse osmosis unit intended to be a component in a complete water purification system, and is not a complete water treatment system.
This reverse osmosis unit must be preceded by pre-treatment devices. Whether a particular
device is included in an individual water treatment system will be dictated by local conditions.
The reverse osmosis unit may need to be followed by post-treatment devices as well.
The AQUAbase nX is designed to meet current AAMI/ISO and Federal (U.S.) standards.
{6}------------------------------------------------
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Comparing the subject device to the predicate device, some similarities and differences are noted in the design employed to accomplish the same intended use:
Both systems utilize reverse osmosis for purification of water to be used for Hemodialysis. Product water complies with the international standard ISO 23500-3, as recognized by FDA.
The subject and predicate devices are based on the following same technological elements:
- Both utilize a polyamide, thin film composite, and spiral wound membrane. ●
- Both are modular system families.
- Similar sensors and same general setup of the cabinet ●
- Both shall be installed and commissioned by B. Braun certified technicians and shall . be operated by trained personnel.
- . Both systems shall undergo annual technical inspections
- Optional heat disinfection available for both ●
- Use of substantially equivalent water contact materials in both systems ●
- . Chemical disinfection using peracetic acid disinfection products or heat disinfection possible for both
- Both display poor water quality and pressure alarms
- Both monitor water quality on digital displays
The following differences exist between the subject and predicate devices:
- Differences in indications for use, based on the scope of the predicate device ● submission (system including pre-treatment, reverse osmosis and post-treatment). The subject device is compared to the reverse osmosis subset of the predicate device only.
- User interface of the subject device and predicate device is different. ●
- . The predicate device offers a back washing option ("Eco") which is not offered in the subject device.
- Difference in water conversion control (volume controlled in the predicate device and conductivity controlled in the subject device)
- Systems differ in output specification. The subject device provides output range for smaller clinics whereas the predicate device provides output range for bigger clinics.
- . Predicate device is available as an option as double staged system. Subject device is available as single staged system.
- Optional heat disinfection is available for both. Heater option for optional heat ● disinfection is integrated in the subject device. The predicate device must be linked to a separate inline hot cleaning system.
{7}------------------------------------------------
- . Remote alarm monitoring system available in the predicate device. Acoustic alarms available in the subject device.
- Difference in the supervision of the controller function (via watchdog in the predicate . device and observer system in the subject device).
- . Emergency operation mode available in the predicate device for the optional double staged system. Not required in the subject device being a single staged system.
# VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
#### Biocompatibility testinq
The biocompatibility evaluation for the AQUAbase nX was conducted in accordance with the international standard ISO 10993-1 "Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process", as recognized by FDA.
The battery of testing included the following tests:
- cytotoxicity ●
- hemolysis ●
- chemical characterization (leach test) ●
The AQUAbase nX has indirect blood contact with the patient and contact duration is less than 24 hours for one single treatment and long term for cumulative use.
## Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the AQUAbase nX. The system complies with the IEC 60601-1 standard for electrical safety and with IEC 60601-1-2 standard for EMC.
## Software verification and validation testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" as well as the draft FDA guidance recently issued.
{8}------------------------------------------------
Additional performance bench testing conducted on the device include:
- performance testing according to the international standard ISO 23500-3, as . recognized by FDA
- . chemical and heat disinfection testing
- . bacterial count and endotoxins
# VIII. CONCLUSIONS
The conclusions drawn from the nonclinical bench tests demonstrate that the AQUAbase nX is as safe, as effective and performs as well as than the legally marketed device identified in section III.
Testing results indicate that the device complies with all requirements stated in applicable recognized consensus standards for water purification systems for hemodialysis.
The device has equivalent technical features and intended use as the predicate device and the differences between the two do not introduce new issues of safety and effectiveness.
The AQUAbase nX is therefore considered substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.