Uretero1 Single-Use Digital Flexible Ureteroscope (standard deflection), PN UROSD101; Uretero1 Single-Use Digital Flexible Ureteroscope (reverse deflection), PN URORD101; Vision1 Imaging Console and Display System, PN VIS101
Applicant
STERIS Corporation
Product Code
FGB · Gastroenterology, Urology
Decision Date
Feb 17, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The single use Uretero1™ Ureteroscope System is intended to visualize organs, cavities and canals in the urinary tract (urethra, bladder, ureter, calyces and renal papillae) via transurethral or percutaneous access routes. It can also be used in conjunction with endoscopic accessories to perform various diagnostic and therapeutic procedures in the urinary tract.
Device Story
Uretero1 Ureteroscope System consists of a single-use, flexible digital ureteroscope and a Vision1 Imaging Console and Display System (CCU). Ureteroscope features articulation lever for tip steering, working channel for instruments/irrigation, and camera/illumination at distal tip. CCU processes video from ureteroscope for display on integrated touchscreen or external monitor via HDMI. System used in clinical settings by physicians for urinary tract visualization and procedures. Ureteroscope has 4-hour usage limit with on-screen countdown notifications. CCU provides GUI for image management, photo/video capture to encrypted USB, and printer connectivity. No ePHI storage; no network connectivity. System aids clinical decision-making by providing real-time visualization of urinary anatomy and diagnostic/therapeutic access.
Clinical Evidence
Bench testing only. Performance verified via functional and regression testing (IEC 62304), optical performance (ISO 8600-3/5, ISO 12233, ISO 9039, ISO 15739), biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1), and packaging integrity (ASTM F1886, F2096, F88).
Technological Characteristics
Single-use flexible ureteroscope; CCU utilizes NXP i.MX8MQ processor and LATTICE ECP5 FPGA. Software built on Linux (YOCTO) with QT5 GUI and g-streamer video pipeline. Connectivity: HDMI, USB. Sterilization: Ethylene Oxide (SAL 10^-6). Materials include various medical-grade adhesives (Loctite series, Dymax).
Indications for Use
Indicated for patients undergoing urological endoscopic procedures requiring visualization of the urinary tract (urethra, bladder, ureter, calyces, renal papillae) and performance of diagnostic or therapeutic procedures.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 17, 2023
STERIS Corporation Carroll L. Martin Regulatory Affairs Director 5976 Heislev Road Mentor, OH 44060
Re: K223466
> Trade/Device Name: Uretero1TM Ureteroscope System Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FGB Dated: January 13, 2023 Received: January 19, 2023
Dear Carroll L. Martin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mark J. Antonino -S
Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K223466
Device Name Uretero1TM Ureteroscope System
## Indications for Use (Describe)
The single use Uretero1™ Ureteroscope System is intended to visualize organs, cavities and canals in the urinary tract (urethra, bladder, ureter, calyces and renal papillae) via transurethral or percutaneous access routes. It can also be used in conjunction with endoscopic accessories to perform various diagnostic and therapeutic procedures in the urinary tract.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for STERIS. The word STERIS is in all capital letters and in a bold, sans-serif font. Below the word STERIS is a graphic of several horizontal, wavy lines in a light blue color. The registered trademark symbol is to the right of the word STERIS.
K223466
Page 1 of 11
# 510(k) Summary for the Uretero1™ Ureteroscope System
STERIS Corporation 5960 Heisley Road Mentor, OH 44060
| Contact: | Carroll Martin |
|----------|----------------------------------|
| | Regulatory Affairs Director |
| | Tel: 440-358-6259 |
| | Email: Carroll_Martin@steris.com |
February 6, 2023 Submission Date:
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
{4}------------------------------------------------
#### 1. Device Name
| Trade Name: | Uretero1TM Ureteroscope System |
|--------------------|--------------------------------|
| Device Class: | Class II |
| Regulation Name: | Endoscope and Accessories |
| Common/usual Name: | Ureteroscope |
| Regulation Number: | 21 CFR 876.1500 |
| Product Code: | FGB |
#### 2. Predicate Device
Uretero1 Ureteroscope System, K211347 Class II (21 CFR 876.1500), product code FGB
#### Device Description 3.
The Uretero1™ Ureteroscope System contains an all-in-one camera control unit (CCU) for image processing and display. The CCU operates at 110 V - 230V ~ 50 - 60 Hz and has a simple graphical user interface that prompts the clinician through actions to get a live view via the Ureteroscope. After unpackaging the sterile, single-use ureteroscope, the user is required to plug in the single-use ureteroscope device via a 16-pin male connector inserted into the CCU to obtain a live image. There is a process of authentication, in which the CCU recognizes the firmware of the inserted product, and the image from the ureteroscope is displayed on live view. The Uretero1TM CCU software platform is based on i.MX8MQ processor from NXP and LATTICE ECP5 FPGA. The ECP5 provides the camera interface plus ISP functionality. The CCU software is based on Linux board support package provided by NXP. The board support package and Linux distribution are configured and built using YOCTO. The main functionality of the system is to capture the video from an external camera and display it to an external display. Graphical user interface is developed using QT5. Video pipeline is implemented through g-streamer.
The ureteroscope is disposable and can be utilized for up to 4 hours. The ureteroscope possesses a 4-hour countdown timer and after 4 hours of usage, the ureteroscope will no longer produce a live image and a use notification message will be displayed on the screen informing the user that the ureteroscope usage time has expired. The ureteroscope has 3 use notification messages. One occurs at 60 minutes left on the timer and appears on the side menu bar of the display screen on the CCU. A second occurs with 30 minutes left on the timer and also appears on the side menu bar. The final notification message appears with 5 minutes of time remaining on the ureteroscope and it is in the middle of the screen and requires user interaction to acknowledge the use notification message by pressing OK. When the duration of time has elapsed, the ureteroscope will no longer produce a live image and the user will be directed to a screen that states the ureteroscope no longer has any usable time remaining and to insert new ureteroscope.
Once the flexible shaft is inserted into the patient, the distal tip is steered via the articulation lever on the ureteroscope handle. The flexible shaft allows for passive secondary deflection while accessing the patient's urinary tract. The flexible shaft contains a working channel to allow surgical accessories and procedural solutions to be delivered through the distal tip to the surgical field. The handle button allows the user to take photos, record video and zoom
{5}------------------------------------------------
in and out. In addition to the articulation lever and the handle button, the handle contains an accessory access port and a connection port for procedural solutions. An irrigation system (irrigation tubing and an irrigation source) can be attached to one port and the biopsy port cover can be attached to the other port. In this way, the user can use endoscopic instruments through the biopsy port cover and irrigate at the same time. The Uretero1TM Single-Use Digital Flexible Ureteroscope must be used in conjunction with the Vision1™ Imaging Console and Display System (CCU).
In addition to image processing, the CCU supports video connectivity in two additional ways. The user is allowed to mirror image from the CCU to an external display which produces the same image the user sees on the CCU screen on any other compatible display system that accepts HDMI input. In addition, the CCU can accept an HDMI input signal from an external HDMI output and the CCU can act as a dummy display panel for the external HDMI imaging source. There are two USB data ports to provide data export (photos and video taken during procedure) or allow connection of selected accessories (printer). No data, including Electronic Protected Health Information (ePHI) data is stored on this device (e.g., last name, SSN, DOB, MRN) and there is no ethernet, Wi-Fi or Blue Tooth capability.
The ureteroscope will be offered in two different models: UreterolTM Ureteroscope (Standard Deflection) and UreterolTM Ureteroscope (Reverse Deflection). To save procedural photo or video, a proprietary system encrypted USB drive is provided to Customers with the purchase of the CCU. The Vision1™ Imaging Console and Display System communicate with the encrypted USB drive, the Uretero1TM Single-Use Digital Flexible Ureteroscope, and the external printer.
The Uretero1 Ureteroscope System comes packaged with single-use, disposable accessory port cover. The accessory port cover is adjustable and is intended to resist the backflow of fluid around an instrument inserted through the working channel of the ureteroscope. It is designed to facilitate the passage of instruments during a procedure. It has a cap feature to allow instruments to be inserted and secured. A clicking mechanism is used to facilitate the open, semi-closed and closed position of the valve. It also has a side arm y-port with a luer cap to allow for irrigation or other procedural solutions, if needed. This accessory attaches to the instrument port of the Uretero1 Ureteroscope via a distal luer connection.
#### 4. Indications for Use
The single use Uretero1™ Ureteroscope System is intended to be used to visualize organs, cavities and canals in the urinary tract (urethra, bladder, ureter, calyces and renal papillae) via transurethral or percutaneous access routes. It can also be used in conjunction with endoscopic accessories to perform various diagnostic and therapeutic procedures in the urinary tract.
{6}------------------------------------------------
#### Comparison of Technological Characteristics with the Predicate Device 5.
A comparison of technical characteristics between the proposed and predicate devices is summarized in Table 1.
| Features | Uretero1™ Ureteroscope System<br>K211347 | Uretero1™ Ureteroscope<br>System (Modified Device) | Comparison |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The single use Uretero1™<br>Ureteroscope System is intended<br>to be used to visualize organs,<br>cavities and canals in the urinary<br>tract (urethra, bladder,<br>ureter, calyces and renal papillae)<br>via transurethral or percutaneous<br>access routes. It can also be used<br>in conjunction with endoscopic<br>accessories to perform various<br>diagnostic and therapeutic<br>procedures in the urinary tract. | The single use Uretero1™<br>Ureteroscope System is<br>intended to be used to visualize<br>organs, cavities and canals in<br>the urinary tract (urethra,<br>bladder,<br>ureter, calyces and renal<br>papillae) via transurethral or<br>percutaneous access routes. It<br>can also be used in conjunction<br>with endoscopic accessories to<br>perform various diagnostic and<br>therapeutic procedures in the<br>urinary tract. | Identical |
| Ureteroscope<br>Construction /<br>Components | Handle<br>Articulation Lever<br>Handle Button<br>Accessory Port<br>Irrigation Port<br>Connector Cable<br>Connector Cable Plug<br>Flexible Shaft<br>Articulation Section<br>Distal Tip (camera, illumination<br>optics, and the working channel,<br>video signal cables, articulation<br>wires and light fiber)<br>Instrument Channel<br>Light (illumination fiber)<br>Camera | Handle<br>Articulation Lever<br>Handle Button<br>Accessory Port<br>Irrigation Port<br>Connector Cable<br>Connector Cable Plug<br>Flexible Shaft<br>Articulation Section<br>Distal Tip (camera, illumination<br>optics, and the working channel,<br>video signal cables, articulation<br>wires and light fiber)<br>Instrument Channel<br>Light (illumination fiber)<br>Camera | Identical |
| Monitor Components | Frame<br>Touchscreen with Antiglare<br>USB Ports<br>Scope Connector<br>Power Button<br>HDMI In<br>HDMI Out<br>USB Type B Port<br>AD/DC Power Cable | Frame<br>Touchscreen with Antiglare<br>USB Ports<br>Scope Connector<br>Power Button<br>HDMI In<br>HDMI Out<br>USB Type B Port<br>AD/DC Power Cable | Identical |
| Sterile/Non-sterile | Sterile | Sterile | Identical |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide | Identical |
| Sterilization<br>Assurance Level | 10-6 | 10-6 | Identical |
| Usage | Single use, disposable | Single use, disposable | Identical |
| Features | Uretero1TM Ureteroscope System<br>K211347 | Uretero1TM Ureteroscope<br>System (Modified Device) | Comparison |
| Page 3 of 1 | | | |
| Functionality | Irrigation capability through<br>irrigation port<br>Ability to use endoscopic tools<br>through instrument port<br>Ability to record images (picture<br>and video)<br>Ability to record to external media<br>(USB flash)<br>Audio (notifications to the user) | Irrigation capability through<br>irrigation port<br>Ability to use endoscopic tools<br>through instrument port<br>Ability to record images (picture<br>and video)<br>Ability to record to external<br>media (USB flash)<br>Audio (notifications to the user) | Identical |
| Ureteroscope<br>Working Distance | 2mm – 50mm | 2mm – 50mm | Identical |
| Dimensions | Tip Diameter: 6.6F (2.2mm)<br>Outer Diameter: 9.5F (3.2mm)<br>Instrument Channel internal<br>diameter: 3.6F (1.2mm) | Tip Diameter: 6.6F (2.2mm)<br>Outer Diameter: 9.5F (3.2mm)<br>Instrument Channel internal<br>diameter: 3.6F (1.2mm) | Identical |
| Ureteroscope Tip<br>Deflection | 270° in both directions | 270° in both directions | Identical |
| Target Population | Patients undergoing a urological<br>endoscopic procedure | Patients undergoing a urological<br>endoscopic procedure | Identical |
| Energy<br>Used/Delivered | None | None | Identical |
| Accessories Provided | Accessory port cover | Accessory port cover | Identical |
| Method of<br>Application | Manual | Manual | Identical |
| Software System | The Uretero1TM CCU software<br>platform is based on i.MX8MQ<br>processor from NXP and LATTICE<br>ECP5 FPGA. The ECP5 provides<br>the camera interface plus ISP<br>functionality. The CCU software is<br>based on Linux board support<br>package provided by NXP. The<br>board support package and Linux<br>distribution are configured and built<br>using YOCTO. The main<br>functionality of the system is to<br>capture the video from an external<br>camera and display it to an external<br>display. Graphical user interface is<br>developed using QT5. Video<br>pipeline is implemented through g-<br>streamer. | The Uretero1TM CCU software<br>platform is based on i.MX8MQ<br>processor from NXP and<br>LATTICE ECP5 FPGA. The<br>ECP5 provides the camera<br>interface plus ISP functionality.<br>The CCU software is based on<br>Linux board support package<br>provided by NXP. The board<br>support package and Linux<br>distribution are configured and<br>built using YOCTO. The main<br>functionality of the system is to<br>capture the video from an<br>external camera and display it to<br>an external display. Graphical<br>user interface is developed using<br>QT5. Video pipeline is<br>implemented through a streamer | Identical |
| Features | Uretero1™<br>Ureteroscope System | Uretero1™<br>Ureteroscope System<br>(Modified Device) | Comparison |
| | K211347 | | |
| Software Update:<br>Allow unlimited<br>addresses for scope<br>detection. | Has finite number of<br>addresses for scope<br>detection. | Has unlimited addresses<br>for scope detection. | Similar. No impact on<br>safety, effectiveness or how<br>the device is used. This<br>software change is a<br>convenience feature for the<br>user. |
| Software Update:<br>Synchronization check on<br>incoming video frame. | Synchronization check<br>not performed on<br>incoming video. | Synchronization check<br>performed on incoming<br>video. | Similar. No impact on<br>safety, effectiveness or how<br>the device is used by the<br>user. Change is intended to<br>increase manufacturability. |
| Software Update:<br>Modification in Image<br>Signal Processing (ISP) | Image Signal Processing<br>(ISP) utilizes an external<br>clock source as a root<br>clock. | Image Signal Processing<br>(ISP) utilizes an internal<br>clock source as a root<br>clock | Similar. No impact on<br>safety, effectiveness or how<br>the device is used by the<br>user. Change is intended to<br>increase manufacturability. |
| Software Update:<br>Modification in behavior<br>of HDMI output to<br>external monitors | Use of an incompatible<br>monitor causes a "stutter"<br>effect. | Use of an incompatible<br>monitor results in no<br>video from the<br>ureteroscope. | Different. No impact on<br>safety, effectiveness or how<br>the device is used by the<br>user. The Uretero1 System<br>allows a user to connect a<br>monitor other than the<br>Vision1. The change is for<br>the convenience of the user. |
| Software Update:<br>Turn off scope LED drive<br>when the scope is not<br>connected. | LED drive remains on<br>when scope is<br>disconnected. | LED drive is turned off<br>when the scope is<br>disconnected. | Different. No impact on<br>safety, effectiveness or how<br>the device is used by the<br>user. This change is to<br>extend the life of the LED<br>driver chip. |
| Software Update:<br>Update initialization for<br>Hardware Chip | Error checks performed<br>only at boot up | Additional error checks<br>performed at boot up and<br>screen transitions. | Similar. No impact on<br>safety, effectiveness or how<br>the device is used by the<br>user. This change improves<br>the boot up performance of<br>the system. |
| Software Update:<br>Remove dormant<br>software class objects<br>from memory | Dormant software class<br>objects remain in<br>memory. | Dormant software class<br>objects removed from<br>memory. | Different. No impact on<br>safety, effectiveness or how<br>the device is used by the<br>user. This change improves<br>memory usage of the<br>system. |
| Software Update:<br>Update the Uretero1<br>System application to<br>eliminate the digital<br>artifact | Sensor registers are not<br>monitored once they are<br>set. | Sensor registers are<br>constantly monitored<br>once they are set. | Different. No impact on<br>safety, effectiveness or how<br>the device is used by the<br>user. This change returns the<br>device to its original<br>specifications. |
| Features | Uretero1TM<br>Ureteroscope System<br>K211347 | Uretero1TM<br>Ureteroscope System<br>(Modified Device) | Comparison |
| Software Update:<br>Provide the user the ability<br>to disable the scope buttons<br>in the Settings menu of the<br>console. | Scope button is always<br>enabled. | Option given to the user<br>to disable the scope<br>button. | Different. This change does<br>not impact safety or<br>effectiveness. This change is<br>for the convenience of the<br>user. |
| Software Update:<br>Implement on-screen<br>annotation of Left vs. Right<br>kidney. | No ability to annotate<br>left and right kidney. | Ability for user to<br>annotate left and right<br>kidney. | Different. This change does<br>not impact safety or<br>effectiveness. This change is<br>to aid the user |
| Software Update:<br>Change email address<br>listed in HELP menu from<br>vision1@steris.com to<br>urology@steris.com. | Email address is<br>vision1@steris.com | Email address is<br>urology@steris.com | Different. This change does<br>not impact safety,<br>effectiveness or how the<br>device is used. The change is<br>for the convenience of the<br>user |
| Software Update:<br>Remove the red static text<br>"Insert Thumb drive to<br>take photos or videos" in<br>the patient data box. | Red static text remains<br>regardless of whether<br>or not the thumb drive<br>has been inserted. | The red static text is<br>removed. | Different. This change does<br>not impact safety,<br>effectiveness or how the<br>device is used. The change is<br>for the convenience of the<br>user. |
| Software Update:<br>Make the pop-up message<br>"Photo or video not saved,<br>insert USB drive"<br>disappear after 5 seconds,<br>without requiring user<br>intervention. | User intervention is<br>required to remove the<br>pop-up message. | User intervention is not<br>required to remove the<br>pop-up message. | Different. This change does<br>not impact safety,<br>effectiveness or how the<br>device is used. The change is<br>for the convenience of the<br>user. |
| Software Update:<br>Grey out all languages<br>except English. | All languages appear<br>available. | Only English appears<br>available. | Different. This change does<br>not impact safety,<br>effectiveness or how the<br>device is used. The change is<br>for the convenience of the<br>user. |
| Software Update:<br>Refactor device power<br>board software code and<br>GUI software code. | Unoptimized code<br>structure and memory<br>usage. | Improved code structure<br>and memory usage. | Different. This change does<br>not impact safety,<br>effectiveness or how the<br>device is used. |
| Features | Uretero1TM<br>Ureteroscope System<br>K211347 | Uretero1TM<br>Ureteroscope System<br>(Modified Device) | Comparison |
| Software Update:<br>Lock the HDMI output<br>resolution to 1080p/60. | HDMI output capable<br>of 720p and 1080p | HDMI output capable of<br>1080p/60 | Different. This change does<br>not impact safety,<br>effectiveness or how the<br>device is used. This change is<br>to ensure monitor<br>compatibility. |
| Software Update:<br>Provide a selection button<br>on the monitor that allows<br>the user to print a snapshot<br>of the live view during a<br>procedure to a printer. | No printer functionality<br>available. | Printer functionality<br>available.…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.